ACDC
Specification
15 Watts
•Ultra Compact Size
•Medical Approvals
•Single Outputs from 3.3 to 48 V
•PCB Mount, Open Frame & Chassis Mount
•Encapsulated PCB & Chassis Mount
•Class II
•Peak Load Capability
EML15 Series
Input
Input Voltage • 85-264 VAC (120-370 VDC)
Input Frequency • 47-63 Hz
Input Current • 0.3 A rms at 230 VAC
Inrush Current • 20 A at 115 VAC, 40 A at 230 VAC,
Earth Leakage Current • Class II construction no earth
Input Protection • Internal 2 A fuse
Output
Output Voltage • See tables
Output Voltage Trim • ±5% (not encapsulated units)
Initial Set Accuracy • ±1%
Minimum Load • No minimum load required
Start Up Delay • 2 s max
Start Up Rise Time • 14 ms max
Hold Up Time • 12 ms min at full load & 115 VAC
Line Regulation • ±0.5% max
Load Regulation • ±1% max
Transient Response • 4% max deviation, recovery to within
1% in 500 µs for a 25% load change
Ripple & Noise • 3.3-5 V versions: 50 mV pk-pk,
12-15 V versions: 120 mV pk-pk,
24-48 V versions: 200 mV pk-pk,
20 MHz BW
Overvoltage Protection • 115-135% Vnom, recycle input to reset
Overload Protection • 120-150%
Short Circuit Protection • Trip and restart (Hiccup mode)
Temperature • 0.05%/°C
Coefficient
xppower.com
General
Efficiency • See tables
Isolation • 4000 VAC Input to Output
Protection Level • Primary to Secondary 2 MOPP
Switching Frequency • 70 kHz typical
Power Density • 4.8 W/In3(PCB Mount version)
MTBF • >400 kHrs per MIL-HDBK-217F
Environmental
Operating Temperature • -20 °C to +70 °C, derate linearly from
100% at +50 °C to 50% at +70 °C
Cooling • Convection-cooled
Operating Humidity • 95% RH, non-condensing
Storage Temperature • -40 °C to +85 °C
Operating Altitude • 3048 m
Vibration • IEC68-2-6, 2 g, 10 Hz to 500 kHz,
10 mins/cycle, 60 mins each cycle
Shock • IEC68-2-27, 30 g, 11 ms half sine, 3 times
in each of 6 axes
EMC & Safety
Emissions • EN55022, level B conducted & radiated
Harmonic Currents • EN61000-3-2, EN61000-3-3
ESD Immunity • EN61000-4-2, level 3 Perf Criteria A
Radiated Immunity • EN61000-4-3, 10 V/m 80% mod
Perf Criteria A
EFT/Burst • EN61000-4-4, level 3 Perf Criteria A
Surge • EN61000-4-5, level 3 Perf Criteria A
Conducted Immunity • EN61000-4-6, 10 Vrms Perf Criteria A
Magnetic Fields • EN61000-4-8, 10 A/m Perf Criteria A
Dips & Interruptions • EN60601-1-2, 70% UTfor 500 ms,
40%UTfor 100 ms, <5% UTfor 10 ms,
<5% UTfor 5000 ms,
Perf Criteria A, A, A, B (see Note 4)
Safety Approvals • EN60601-1, AN51/AAIMT ES60601-1.
IEC60601-1