FEATURES
8 universal analog input channels
8 digital I/Os (individually configured as
inputs or outputs)
2 relay outputs (NO, NC and common)
RS485 interface
(Modbus master or slave)
24Vdc output to power up to eight
4 to 20mA transmitters
Ethernet interface for LAN and
internet use
USB-device interface for configuring,
monitoring and download
USB-host interface for logged data
retrieval through a USB flash drive
Up to 32 configurable alarms
Up to 128 virtual channels
Detachable display (optional)
Up to 100 channels can be logged at a
configurable rate
Data downloading through the
configuration software or through the
freely distributed download DLL
4 input channel types: analog, digital,
remoted and virtual
16GB SD card interface
DATA LOGGERS
8 to 16 Channel Logger
Models DL-1080 & DL-1081
Versatile, powerful and cost
effective data loggers handling
analog, digital and other
types of variables with high
resolution and speed
Special
product
features
R
S
4
8
5
C
O
M
P
A
T
I
B
L
E
D
C
I
N
P
U
T
S
D
I
G
I
T
A
L
I
/
O
U
N
I
V
E
R
S
A
L
E
T
H
E
R
N
E
T
P
R
O
G
R
A
M
M
A
B
L
E
Inputs rated CAT II
NEW!
IP
Rated
20
SPECIFICATIONS
MODELS DL-1080 & DL-1081
ELECTRICAL
Input Type Measuring Range: Accuracy:
J-184º to 1832ºF (-120 to 1000ºC)
± 0.2% (F.R.) ± 1ºC
K-202º to 2501.6ºF (-130 to 1372ºC)
T-202º to 752ºF (-130 to 400ºC)
E-202º to 1436ºF (-130 a 780ºC)
N-202º to 2372ºF (-130 to 1300ºC)
R68º to 3214.4ºF (20 to 1768ºC)
± 0.2% (F.R.) ± 3ºC
S68º to 3214.4ºF (20 to 1768ºC)
B212º to 3308°F (100 to 1820ºC)
Pt100 -328º to 1562ºF (-200 to 850ºC) ± 0.15% (F.R.)
Pt1000 -328º to 1562ºF (-200 to 850ºC)
Linear 0 to 20mA
User Programmable ± 0.15% (F.R.)*
Linear 4 to 20mA
Linear 0 to 20mV
Linear 0 to 50mV
Linear 0 to 60mV
Linear -20 to 20mV
Linear 0 to 5V
Linear 0 to 10V
Digital Inputs Logic level “0”: from 0 to 0.8Vdc
Logic level “1”: from 3 to 30Vdc
Internal Memory 2MB
Excitation Current Pt100s: 360μA; ; Pt1000s: 320μA
Maximum Pt100/Pt1000
Compensated Cable
Resistance
40Ω
Maximum Input Voltage 30Vdc
Input Current @ 30Vdc
(typical) 3mA
Digital Outputs Maximum output voltage: 30Vdc
Maximum output current: 200mA
Maximum relay current: 3A @ 250Vac; 3A @ 30Vdc
Storage Rate from 1ms to 24 hours
Maximum Channel Logged 100
Supported Modbus
Commands Read Coil Status (01h)
Read Holding Registers (03h)
Write Single Coil (05h)
Write Single Register (06h)
Write Multiple Registers (0Fh)
Power Supply 100 to 240Vac, 50/60Hz. 20VA (max)
ENVIRONMENTAL
Operating Temperature 32° to 122°F (0 to 50°C)
Relative Humidity 80% up to 85°F (30°C)
For temperatures higher than 85°F (30°C),
decrease 3% per °C
Altitude < 2000m
Protection IP20
F.R. = Full Range = Span
*The full scale refers to the input of the sensor signal and not the range of configured indication.
Technical Assistance (800) 343-1391 www.aemc.com
CATALOG NO. DESCRIPTION
2134.61 Data Logger Model DL-1080 (8 Analog to 8 Digital Channel, no LCD)
2134.62 Data Logger Model DL-1081 (8 Analog to 8 Digital Channel, LCD)
PRODUCT INCLUDES
DL-1080 & DL-1081 software
User manual and drivers
also available on USB stick
Model DL-1081
Color Functional Displays
"FAVORITES" Screen
"CONFIGURATION" Screen
"CHART" Screen Software Screens
"ALARMS" Screen
Shows a six-position grid where you can
assign a channel to be displayed real-time
in each position.
Data logger and display parameters can
be changed and viewed on this screen.
Displays a plot of the favorite
channels values.
A 32-position grid where each numbered
position is related to its equivalent alarm or
event. When an alarm is active, its number