MIT415
Insulation Tester
DESCRIPTION
The new Megger MIT415 insulation and continuity tester
has been designed for electrical testing by utilities,
industrial, commercial and domestic electricians. The wide
range of test voltages makes the MIT415 series ideal for the
engineer or technician involved with combined applications
of both communications circuits and power circuits such as
would be found in elevator service/repair and routine main-
tenance. The low test voltage insulation ranges also makes
the MIT415 ideal for ESD testing.
The MIT415 replaces the well-established BMM2000ESD
insulation tester, gives greater functionality with simplified
operation, greater application range and an ergonomically
designed case.
INSULATION TESTING
nTest voltages -10 V, 25 V, 50 V, 100 V, 250 V and 500 V
nTest Lock -Locks insulation test on continuous.
nTest voltage display -The actual test voltage is displayed
on the smaller digital readout, with the insulation result
on the larger digital display.
nAnalogue arc -The display also features an analogue arc
to replicate the response of a moving coil display.
nPI and DAR - Polarisation Index (PI) and Dielectric
Absorption Ratio (DAR) functions
n50 G-Insulation testing from 1 Gto 50 G
nSilicone leads -High quality flexible silicone test leads
are comfortable to use and prevent measurement er rors
on higher Granges.
nTest inhibit -prevents testing if system voltages in excess
of 50 V are detected when making insulation tests.
nInsulation buzzer -The buzzer can operate if the
insulation resistance is above a preset limit, set via the
Setup menu.
CONTINUITY TESTING
nAuto-test -Auto test on circuit contact enables real two
handed operation without the need to press the test
button.
n200 mA or 20 mA -Either 200 mA or 20 mA continuity
test currents are available. 20 mA test cur rent will
considerably increase batterylife.
nLead null -Lead resistance compensation (NULL)
operates up to 10 of resistance.
nBuzzer -ON-OFF selected by simple push button.
nBuzzer limit -Continuity buzzer limit alarm provides
adjustment of the maximum resistance at which the
continuity buzzer sounds. This is adjustable from 1 to
20 in 5 steps.
nResistance - krange extends continuity measurement
to 1 M.
n10 V and 100 V ESD testing
n25 V & 50 V signal cable insulation testing
n250 V & 500 V Insulation testing up to
50 G(500 V)
nContinuity testing at 200 mA or 20 mA
from 0.01 to 999 k
nDC and TRMS AC voltage measurement
nLarge backlit display
nCombined analogue and dual digital
readout
nWeatherproof to IP54
nSupplied with AA alkaline cells
MIT415
Insulation Tester
DISPLAY
The display offers a combination of analogue arc and a dual
digital readout:
Analogue arc:
nFull display width analogue arc.
nPatented arc display shows essential charge and discharge
characteristics not visible on a digital numeric only
display.
nSingle pointer “needle” response is similar to a moving
coil meter.
nLogarithmic display for better low insulation value
measurements.
Dual digital display:
nLarge main digital readout for good visibility of all main
measurement results
nSecond digital display for additional data such as:
Insulation test voltage.
Insulation leakage current.
Supply frequency (when measuring volts).
Test mode eg. PI, DAR or TI (Timed mode).
OTHER FEATURES
W
We
ea
at
th
hp
pr
ro
oo
of
f -
-Every tester is sealed to IP54, providing a
weatherproof case to reduce the chances of water ingress,
including the battery and fuse compar tment.
T
To
ou
ug
gh
h h
ho
ou
us
si
in
ng
g- Rubber over moulding combines the
tough shock absorbing outer protection with excellent grip,
on a strong modified ABS housing, providing an almost
indestructible case.
B
Ba
at
tt
te
er
ri
ie
es
s- Battery requirements are 5 AA batteries of
either standard Alkaline or Nickel Metal Hydride (NiMH)
rechargeable type, providing a minimum of 2200 insulation
tests at 500V.
SAFETY
Designed to be exceptionally safe to use, fast detecting
circuitry prevents damage to the instr uments if accidentally
connected to live circuits or across phases. Specifically, all
instruments:
nMeet the international requirements of IEC1010-2 and
EN61557.
nLive circuit detection inhibits insulation testing on circuits
above 50 V.
nLive circuit detection and test inhibit on continuity
measurements.
nDefault display of live circuit voltage on all ranges.
nDetection and inhibit functions even if the protection
fuse has failed.
nSuitable for use on CAT IV applications and supply
voltages (of up to) 600 V to ground.
MIT415
IInsulation Tester
600 V CAT IV
All MIT415 instruments are designed to meet the safety
requirements for use on CAT IV 600 V.
APPLICATIONS
(A) Electrical installations testing:
The MIT415 includes all the features required for
electricians and engineers working in a range of industries.
Available features are selected to make testing easy and fast
in a range of situations. Typical industries include:
Electrical supply companies
Large and small scale electrical installation
Periodic inspection and testing
Cable testing
(B) Combined power and communications
applications:
Typically power circuits are tested at 250V or 500V, whereas
communications circuits for signalling are tested at lower
voltages such as 25V or 50V. Both systems can be tested
with the MIT415. Typical applications would include:
Elevator service engineers
Street lighting (pedestrian controls)
Machinery and safety interlock commissioning/
service
Heating, ventilation and air conditioning systems
Robotic power and controls
(C) Electrostatic Discharge testing
The 10V and 100V insulation test voltage ranges are ideal for
ESD testing, including servicing of equipment and routine
maintenance of ESD conductive flooring, bench mats and
grounding systems etc. Typical industries include:
Electronic manufacturing
Electronic servicing and repair
Calibration houses
SPECIFICATION
All quoted accuracies are at +20°C.
Insulation
Nominal test voltages
MIT415 10 V, 25 V, 50 V, 100 V, 250 V, 500 V
Insulation resistance range (see note 1)
10 V 1 G
25 V 2 G
50 V 5 G
100 V 10 G
250 V 20 G
500 V 50 G
Range Full Scale Accuracy
All ranges ±2% ±2 digits up to 100 M.
Then:
500 V ±3% ±2 digits ±0.4% per G
250 V ±3% ±2 digits ±0.8% per G
100 V ±3% ±2 digits ±2.0% per G
50 V ±3% ±2 digits ±4.0% per G
25 V ±3% ±2 digits ±2.0% per 100 M
10 V ±3% ±2 digits ±2.0% per 100 M
Analogue range: 1 Gfull scale
Resolution: 0.01 M
Short circuit cur rent: 2 mA +0% -50%
Open circuit terminal voltage: -0% + 20% ±1 V
Test cur rent on load:
1 mA at min. pass value of insulation specified in BS7671, HD384
and IEC364, 2 mA max.
EN61557 Operating range: 0,10 Mto 1,00 G
Leakage current range 10 µA 2000 µA
Leakage current: 10% ±3 digits
Voltage display: 3% ±3 digits ±0.5% of rated
voltage
Polarisation Index (PI): 10 min / 1 minute ratio
Dielectric Absorption Ratio (DAR): 60 sec / 30 sec ratio
Timer: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 minute
countdown in seconds
Notes:
(1) All ranges measure from 0,00 Mupwards.
(2) Above specifications only apply when high quality silicone leads
are being used.
Continuity
Measurement: 0,01 to 99,9 (0 to 100 on
analogue scale)
Accuracy: ±2% ± 2 digits (0 to 100 )
Open circuit voltage: 5 V ± 1 V
Test cur rent: 200 mA (±1 mA) (0.01 to 9.99 )
20 mA (±1 mA ) (10.0 to 99.9 )
Zero offset at probe tips: 0,10 typical
Insulation range
10 V Maximum reading = >1.00 G
25 V Maximum reading = >2.00 G
50 V Maximum reading = >5.00 G
100 V Maximum reading = >10.0 G
250 V Maximum reading = >20 G
500 V Maximum reading = >50 G
Leakage current display 3
INS test voltage display 3
Continuity
200 mA (0.01 to 99.9 ) 3
Variable Current Limit. 200 mA / 20 mA 3
Fast buzzer- Selectable threshold 3
K range to 999 K 3
Voltage
TRMS & DC Measurement to 600 V 3
Default voltmeter on other ranges 3
Measurement ranges 3
Frequency Hz- 15 to 450 Hz 3
Safety
Live circuit warning at 25 V 3
Testing inhibit at 50 V 3
IEC61010 compliant 3
Installation Category IV 600 V to ground 3
Features
Backlight 3
Battery condition display 3
Insulation Timed - PI – DAR Tests 3
Test button plus lock button 3
Auto power down 3
Remote switch probe (SP5) 3
Calibration certificate with product 3
MIT415
Insulation Tester
UK
Archcliffe Road Dover
CT17 9EN England
T +44 (0) 1304 502101
F +44 (0) 1304 207342
UKsales@megger.com
UNITED STATES
4271 Bronze Way
Dallas TX 75237-1088 USA
T 800 723 2861 (USA only)
T +1 214 333 3201
F +1 214 331 7399
USsales@megger.com
OTHER TECHNICAL SALES OFFICES
Norristown USA, Sydney AUSTRALIA,
Toronto CANADA, Trappes FRANCE,
Kingdom of BAHRAIN, Mumbai INDIA,
Johannesburg SOUTH AFRICA and
Conjure THAILAND.
Registered to ISO 9001:2000 Cert. no. Q 09290
Registered to ISO 14001-1996 Cert. no. EMS 61597
MIT415_DS_en_V01
www.megger.com
Megger is a registered trademark
ORDERING INFORMATION
Item (Qty) Order No.
Insulation and continuity tester MIT415-EN
Included accessories
Red/Black silicone test leads with probes and clips
SP5 remote switch probe
Rubber boot with stand
Owners information CD
Batteries
Hard carry case
Item (Qty) Order No.
Optional accessories
Replacement lead set 6220-813
SP5 remote switch probe 6220-812
Rubber boot with stand 6231-802
Hard case 5410-420
MIT415
Insulation Tester
Lead resistance zeroing: Up to 9.00
Buzzer: Variable limit 1 , 2 , 5 ,
10 , 20
Resistance
Measurement: 0.01 kto 1000 k(0 to 1 M
on analogue scale)
Accuracy: ±3% ±2 digits
Open circuit voltage: 5 V ±1 V
Short circuit cur rent: 20 µA ±5 µA
Voltage range
0 to 600 V d.c. ±2% ±2 digits
10 mV to 600 V TRMS sinusoidal (40 to 400 Hz) ±2% ±1 digit
0 to 1000 V on analogue scale
Unspecified input level 0 - 10 mV (40 to 400 Hz)
For non-sinusoidal waveforms additional specification
apply:
±3% ±2 digits 101 mV to 600 V TRMS and ±8% ±2 digits 10 mV to
100 mV TRMS
Default Voltmeter: Operates at >25 V a.c. or d.c.
on any range except OFF
Frequency: 40-450 Hz (40 Hz - 99.9 Hz)
±0.5% ±1 digit (100 Hz to
450 Hz)
Power Supply:
5 x 1,5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP) Alkaline
NiMH rechargeable cells may be used.
Batter y life: 2200 insulation tests with duty cycle of 5 sec ON /55
sec OFF @ 1000 V into 1 M
Dimensions
Instrument: 220 x 92 x 50 mm (8.66 in. x 3.63 in. x 1.97 in.)
Instrument + case: 456 x 178 x 89 mm (18 in. x 7 in. x 3.5 in.)
Weight
Instrument only: 590 gms, 775 gms with boot (20.73 oz.
27.22 oz.)
Instrument plus case: 1.75kg (3.86 lb)
Fuse
Use only a 500 mA (FF) 1000 V 32 x 6 mm ceramic fuse of high
breaking capacity HBC 50 kA minimum. Glass fuses MUST NOT be
fitted.
Safety Protection
The instruments meet IEC 61010-1 to 600 V phase to ear th,
Category IV. Refer to safety warnings supplied.
E.M.C.
In accordance with IEC 61326-1
Temperature effects
Temperature coefficient: <0,1% per °C up to 1 G
<0,1% per °C above 1 G
Environmental
Operating range: -10 to +55 °C
Operating humidity: 90% RH at 40 °C max.
Storage temperature range: -25 ºC to +70 °C
Calibration Temperature: +20 °C
Maximum altitude: 2000 m
Dust and water protection:
IP54 Protected against dust and splashing water