1 Pt100 KN 1508 P The KN Series Ceramic Wire Wound PRTDs are suitable for general applications requiring temperature stability. Applications: Industrial resistance thermometers, especially in chemical, power generation plants and analytical equipment. Construction: A platinum coil is sealed inside a high purity aluminum oxide ceramic body. Lead wires are shear force resistant and assure proper connection to extension leads and cables. Types Self Heating Dimensions in mm Product 1Pt100 KN 1508 P Tolerance Order No. L W0.3 32.206.968 32.206.969 32.206.970 +2 15- 0 W0.15 W0.1 D 0.80.1 0.150.01 10.00.5 Response time Air: V=3m/s Water: V= 0.4m/s t 0.9 t 0.9 t 0.5 t 0.5 l d 0.14 0 .2 0 .3 3.0 9.0 Technical Specification 100 Ohm @ 0 C Measuring current: Temperature range: W0.3 (Class B) = -196 to +660 C Tolerance class: W0.15 (Class A) = -196 to +600 C* (ST exceeds IEC 60751: -100 to +450 C) W0.1 (Class 1/3 B) = -100 to +350 C *For a limited time: 250 hours Temperature coefficient: Tc = 3850 ppm/K Leads: Platinum-gold alloy Insulation resistance after assembly: > 100 MOhm @ 25 C Temperature stability: Also available: 1 mA - According to IEC 60751:2008 - Other standards and narrower tolerances are available on request Excellent long-term stability - Different temperature coefficients (3916 ppm/K - old JIS) - Extension leads Status 07/2014. Technical modifications reserved. Nominal resistance: The measuring point is located at 8 mm from the end of the sensor body Sensor Technology reserves the right to make changes without notice in the specifications of this product Heraeus Sensor Technology USA 1901 Route 130 North Brunswick, NJ 08902 Phone 732-940-4400 Fax 732-940-4445 Email info.hst-us@heraeus.com www.hst-us.com Status: HSTUSA 01/15 Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: Heraeus: 32206212 32206214 32206213