Data Sheet
TekConnect®TCA-1MEG Buffer Amplifier
System Expands the Functionality of Tektronix
High-performance Oscilloscopes
The TCA-1MEG high-impedance buffer amplifier system extends the
capabilities of Tektronix high-performance oscilloscopes, making them ideal
for a variety of general-purpose measurements. The TCA-1MEG amplifier
system provides a 1 MΩpath that is easily removed and replaced with a
wide array of TekConnect probes, amplifiers, and adapters.
This amplifier system delivers versatility to Tektronix high-performance
oscilloscopes through the addition of input coupling (selectable), bandwidth
limit (selectable), and a 1 MΩinput that provides access to a wide array of
measurement solutions. These solutions include general-purpose passive
probes, high-voltage probes (passive single-ended and active differential),
microvolt differential probes, and current probes.
With this tool, Tektronix high-performance oscilloscopes may now perform
measurements such as primary and secondary power supply voltage levels,
currents, and elevated voltages, and assess the power requirements of
the device-under-test.
For today’s high-speed measurement solutions, having the 50 Ωand1MΩ
terminations on the same input path compromises signal fidelity. By making
the 1 MΩpath replaceable, the TCA-1MEG high-impedance buffer amplifier
system eliminates a permanent degradation of the high-speed signal path.
This also allows users to easily and quickly configure each channel for the
input characteristics required for their measurements. By exchanging the
TCA-1MEG for a high-performance 50 Ωpath or other high-speed probing
solution, maximum signal fidelity is maintained at the oscilloscope’s input.
TekConnect®Interface Delivers Superior Signal
Fidelity, Unparalleled Versatility, and Ease-of-use
The TekConnect interface ensures superior signal fidelity with useful
bandpass up to 18 GHz at the oscilloscope input, while offering unparalleled
versatility with the world’s widest array of accessory signal acquisition
solutions for high-performance, real-time oscilloscopes. This interface
delivers a more robust oscilloscope interface for next-generation products
with multi-GHz analog bandwidths, overcoming many of the inherent
bandwidth limitations of BNC-based interfaces. The TekConnect interface
preserves a low-voltage standing wave ratio (VSWR) 50 Ωenvironment as
well as a reliable electrical connection. A convenient, one-button release
and locking mechanism provides quick, easy installation and removal of
probes, amplifiers, and adapters.
Characteristics
Model Specifications
Characteristic Description
Bandwidth (-3 dB) DC to ≥500 MHz
(host instrument bandwidth >1.5 GHz)
Probe Tip Bandwidth
(with P6139A at -3 dB)
DC to ≥500 MHz
(host instrument bandwidth >1.5 GHz)
BW Limit Full, 100 MHz ±25%, 20 MHz ±25%
DC Gain Accuracy ±3% (TCA-1MEG with or without P6139A probe)*1
Propagation Delay
(Input-to-Output)
<2 ns
Input Impedance 1MΩ/10pF(at DC)
Maximum Input Voltage
(Derated with Frequency)
150 VRMS CAT I
100 VRMS CAT II
Above 200 kHz derate at 20 dB per decade; limit to
13 Vpk AC at 3 MHz and above
*1Probe calibration required in TekConnect host instrument.
Typical
Characteristic Description
Displayed System Input
Offset
±0.2 div
Rise Time (calculated
small signal tr= 0.4/F3dB
)
≤0.8 ns (minimum rise time)
Linear Dynamic Range Thelesserof±5divisionsorthedynamicrangeofthe
TekConnect host instrument
Linearity ±0.2%
Nominal
Inputs – 1 (TEKPROBE™ BNC 1 MΩ).
Input Coupling – DC, AC or GND.
Physical Characteristics
Dimensions mm in.
Length 180.08 4.255
Width 31.5 1.240
Height 46.1 1.815
Safety Certifications U.S. NRTL Listing – UL3111-1.
UL3111-2-032.
Canadian Certification – CAN/CSA C22.2 No.1010.1.
CAN/CSA C22.2 No.1010.2.032.
European Union Compliance – EN61010-1/A2.
EN61010-2-032.
Other – IEC61010-1/A2.
IEC61010-2-032.
GPIB Programmable – Through Host Instrument Commands.
Power Requirements – TekConnect.
Adapter Model Compatibility – Refer to TDS7000 Series TekConnect Adapters
Compatibility Table.
Warranty – One year.
2www.tektronix.com