Amplifi ers for infrared detectors
Low noise amplifi ers for infrared detector
(InSb, InAs, InGaAs, MCT, PbS, PbSe)
www.hamamatsu.com
C4159/C5185 series
C3757-02
1
Accessories
Instruction manual
Power cable (one end with 4-pin connector for connection to amplifi er and the other end unterminated, 2 m) A4372-02
Absolute maximum ratings (Ta=25 °C)
Amplifi ers for photovoltaic detectors (Typ.)
Parameter Value Unit
Operating temperature 0 to +40 °C
Storage temperature -20 to +70 °C
Parameter C4159-01 C4159-04 C4159-05 C4159-06 Unit
Applicable detector InSb (dewar type) *1
(P5968-060, P5968-100)
InSb (dewar type) *1
(P5968-200)
InAs (dewar type)
(P7163)
InAs
(P10090 series) -
Conversion impedance 108, 107, 106
(3 ranges switchable)
2 × 107, 2 × 106, 2 × 105
(3 ranges switchable)
108, 107, 106
(3 ranges switchable)
106, 105, 104
(3 ranges switchable) V/A
Frequency response
(amp only, -3 dB) DC to 100 kHz DC to 45 kHz DC to 15 kHz DC to 100 kHz -
Output impedance 50 50 50 50 Ω
Maximum output voltage
(1 kΩ load) +10 +10 +10 +10 V
Output offset voltage ±5 ±5 ±10 ±5 mV
Equivalent input noise current
(f=1 kHz)
0.15 (108, 107 range)
0.65 (106 range) 0.55 0.15 (108, 107 range)
0.65 (106 range) 6 pA/Hz1/2
Reverse voltage Impossible -
External power supply ±15 V
Current consumption *2+30, -10 Max. +30, -22 Max. mA
Note: Output noise voltage = Equivalent input noise current × Conversion impedance
*1: Custom types for multi-element detectors are also available.
*2: Recommended DC power supply (analog power supply): ±15 V
Current capacity: More than 1.5 times the maximum current consumption
Ripple noise: 5 mVp-p or less
Required power supply specifi cations
· C4159 series: ±15 V (±12.0 to ±17.5 V can also be used)
· C5185 series: ±15 V (±15.0 to ±17.5 V can also be used)
· Power supply voltage: 1.5 times or more of amplifi er's maximum current consumption
· Ripple noise: 5 mVp-p or less
· Analog power supply only