6
dc1710aaf
DEMO MANUAL DC1710A-A
quick start proceDure
measurement equipment anD setup
source and load match to reduce reflections, which
may degrade measurement accuracy.
5. A high dynamic range spectrum analyzer, such as the
Rohde & Schwarz FSEM30, should be used for linearity
measurement.
6. Use narrow resolution bandwidth (RBW) and engage
video averaging on the spectrum analyzer to lower the
displayed average noise level (DANL) in order to improve
sensitivity and to increase dynamic range. However, the
trade off is increased sweep time.
7. Spectrum analyzers can produce significant internal
distortion products if they are overdriven. Generally,
spectrum analyzers are designed to operate at their best
with about –30dBm at their input filter or preselector.
Sufficient spectrum analyzer input attenuation should be
used to avoid saturating the instrument, but too much
attenuation reduces sensitivity and dynamic range.
8. Before taking measurements, the system performance
should be evaluated to ensure that:
a. Clean input signals can be produced. The two-tone
signals’ OIP3 should be at least 15dB better than the
DUT’s IIP3.
b. The spectrum analyzer’s internal distortion is mini-
mized.
c. The spectrum analyzer has enough dynamic range
and sensitivity. The measurement system’s IIP3
should be at least 15dB better than the DUT’s OIP3.
d. The system is accurately calibrated for power and
frequency.
A SPECIAL NOTE ABOUT RF TERMINATION
The LTC5590 consists of high linearity passive double-
balanced mixer cores and IF buffer amplifiers. Due to the
bidirectional nature of all passive mixers the LO±IF mixing
products, also referred to as pseudo-image spurs, are
always present at the RF input, typically at a level 12dB
below the RF input signal. Mismatched impedances at the
pseudo-image spur frequencies, such as when filters are
used for SSB NF measurements, can significantly impact
the linearity and noise figure measurements. To avoid
interference from the pseudo-image spurs, terminate the
RF input port with an isolator, diplexer, or attenuator. In the
recommended measurement setups presented in Figure 6
and Figure 7, the 6dB attenuator pad at the demonstration
circuit’s RF input serves this purpose.
Demonstration circuit 1710A-A is easy to set up to evaluate
the performance of the LTC5590. Refer to Figure 5, Figure 6,
and Figure 7 for proper equipment connections. The fol-
lowing procedures describe performing measurements on
Mixer Channel A. The measurement procedures for Mixer
Channel B are identical.
NOTE. Care should be taken to never exceed absolute
maximum input ratings. Make all connections with RF
and DC power off.
RETURN LOSS MEASUREMENTS
1. Configure the Network Analyzer for return loss meas-
urement, set appropriate frequency range, and set the
test signal to –3dBm.
2. Calibrate the Network Analyzer.
3. Connect all test equipment as shown in Figure 5 with
the signal generator and the DC power supply turned off.
4. Increase the DC power supply voltage to 3.3V, and verify
that the total current consumption is close to the figure
listed in the Typical Demonstration Circuit Performance
Summary. The supply voltage should be confirmed at
the demo board VCC, VCCIF and GND terminals to ac-
count for lead ohmic losses.