Agilent U3606A
Multimeter|DC Power Supply
Data Sheet
The real two instruments in one box
Features
•Two independent instruments
in one box
•Nine basic measurements
as well as 4-wire milliohm
measurement and eight math
functions
•Measurement speed: up to
37 readings/s
•Low error rate: up to 0.025%
DCV accuracy
•30-W dual-range power supply
with remote sensing
•Excellent load regulation: up to
0.01%+3 mV
•OVP and OCP load protection
•Auto scan and ramp, and
built-in 4.8 kHz square-wave
generator
•USB-TMC488.2 and GPIB
connectivity
•Kensington lock slot security
Pick both
Digital multimeter (DMM) and power
supply? Simultaneous and inde-
pendent? Small and cost-effective?
Get all this and more in the Agilent
U3606A multimeter|DC power supply.
This convenient new hybrid combines
a 5½-digit DMM and 30-W dual-range
supply in a single unit. Operating
simultaneously and independently,
the instruments provide efficient,
affordable testing while saving space
on the bench or in a rack.
The 5.5-digit DMM
The 5.5-digit DMM includes nine
essential multimeter capabilities as
well as 4-wire milliohm measurement
and eight built-in math functions. The
DMM also provides a fast measure-
ment speed of up to 37 readings/s,
and a low error rate of up to 0.025%
DCV accuracy.
Physical security and seamless
system integration
Instruments may be at risk of theft or
misplacement when left unattended
on the bench. With the hybrid multi-
meter’s rear Kensington lock slot, you
can secure your instrument and be
assured that it is where you expect it
to be for your continued testing the
next day. The rackmountable U3606A
also enables seamless integration
into your system via popular GPIB
and USB-TMC488.2 interfaces,
programmable with standard SCPI
commands.
Get to know the U3606As
replacement - the U3606B
www.agilent.com/find/dmm-power
2
The 30-W DC power supply
The 30-W DC power supply provides
a dual-range output of 30 V/1 A
and 8V/3 A, with an excellent load
regulation of up to 0.01%+3 mV. The
power supply adds overvoltage and
overcurrent load protection (OVP and
OCP), a built-in square-wave genera-
tor, and auto scan and ramp for multi-
level DC bias testing. Remote sensing
capability further ensures accurate
supply of power at load end.
Square-wave generator
Square-wave output is a unique
function for many applications, such
as pulse-width modulation (PWM)
output, adjustable voltage control,
and synchronous clock (baud rate
generator). You can also use this
function to check and calibrate flow-
meter displays, counters, tachom-
eters, oscilloscopes, frequency
converters, frequency transmitters,
and other frequency input devices.
The U3606As square-wave output
provides selectable frequencies up to
4.8 kHz with variable duty cycles and
amplitudes.
Sweep functions
Sweep functions in the U3606A are
auto ramp and scan outputs for low-
speed multilevel DC bias tests, such
as margin tests, power cycling tests
and relay control. Both functions are
conveniently configurable from the
front panel to sweep up to 100 steps
for scan and 10,000 steps for ramp,
programmable to 105% of full scale.
Auto scan output
Auto ramp output
Amplitude (V/A)
Amplitude (V/A)
Amplitude
end position
Amplitude
end position
Scan dwelling time
Total dwelling time
Total dwelling time
Time (t)
Time (t)
Number
of steps
Number of steps
3
Take a closer look
VFD display
Source
terminals
Local/Shift key
Source operation keys
Power
On/Off
button
Multimeter
operation keys
Kensington
lock slot
GPIB (IEEE-488)
connector
USB 2.0 Type B
connector
Rear output terminals
for remote sense
Autorange
and manual
range keys
Multimeter
terminals
Figure 1. Front panel of the U3606A.
Figure 2. Rear panel of the U3606A.
Ventilation fan
Chassis
ground lug
AC power
connector
AC line fuse
4
Digital multimeter specifications
Specification assumptions:
•Specifications stated are after 60-minutes of warm-up and for 5½-digit resolution
•One-year calibration cycle, with calibration temperature of 18 °C to 20 °C
•Operating temperature: 18 °C to 28 °C (64.4 °F to 82.4 °F)
•Accuracy is expressed as ±(% of reading + % of range)
•Temperature coefficient: Add [0.1 x (the specified accuracy) / °C] for 0 °C to 18 °C and 28 °C to 55 °C
•Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for 30 °C to 55 °C
DC specifications
Function Range1
Test current
or burden
voltage
Accuracy ±(% of reading + % of range)
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
24 hours2
23 ºC ± 1 ºC
90 days
23 ºC ± 5 ºC
1 year
23 ºC ± 5 ºC
DC voltage 100.000 mV
1.00000 V
10.0000 V
100.000 V
1000.00 V
0.012 + 0.008
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.015 + 0.008
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.025 + 0.008
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.0015 + 0.0005
0.0010 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
DC current310.0000 mA
100.000 mA
1.00000 A
3.0000 A
< 0.2 V
< 0.2 V
< 0.3 V
< 0.7 V
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.15 + 0.007
0.15 + 0.007
0.0060 + 0.0005
0.0060 + 0.0005
0.0100 + 0.0005
0.0150 + 0.0010
Resistance4100.000 
1000.00 
10.0000 k
100.000 k
1.00000 M
10.0000 M
100.000 M
0.83 mA
0.83 mA
100 A
10 A
900 nA
205 nA
205 nA || 10 M
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.05 + 0.008
0.05 + 0.005
0.05 + 0.005
0.05 + 0.005
0.06 + 0.005
0.25 + 0.005
2.00 + 0.005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0150 + 0.0005
0.1500 + 0.0005
Continuity 1.0000 k 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.0050 + 0.0005
Diode51.0000 V 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.0050 + 0.0005
Capacitance61.000 nF
10.00 nF
100.00 nF
1.000 F
10.00 F
100.0 F
1000 F
10000 F
0.75 A
0.75 A
8.3 A
83 A
83 A
83 A
0.83 mA
0.83 mA
2.0 + 0.8
1.0 + 0.5
1.0 + 0.5
1.0 + 0.5
1.0 + 0.5
1.0 + 0.5
1.0 + 0.5
2.0 + 0.5
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
0.02 + 0.001
1. 20% over-range on all ranges, except for 1000 Vdc range.
2. Relative to calibration standards.
3. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset current im-
mediately after a high-current measurement, ensure that the U3606A has cooled down.
4. Specifications stated are for 2-wire resistance measurements using Null math operation. Without Null, add a 0.2 Ω error. To eliminate noise interfer-
ence which may be induced by the test leads, a shielded test cable is recommended for resistances above 100 kΩ.
5. Specifications stated are for the voltage measured at the input terminals only. The test current (1 mA) is typical. Variation in the current source will
create some variation in the voltage dropped across a diode junction.
6. Specifications stated are for open test lead measurements with film capacitor or better, using Null math operation.
5
Digital multimeter specifications (continued)
Low-resistance specifications
Range Test current
Accuracy ±(% of reading + % of range)1
1 year (23 ºC ± 5 ºC)
100 m 1.0000 A 0.25 + 0.05
1000 m 0.1000 A 0.25 + 0.03
10  0.1000 A 0.25 + 0.03
1. Four-wire measurement method is used. Test current is sent from the FORCE terminals and resistance is measured at the SENSE terminals.
AC specifications
Function Range1
Frequency
range
Accuracy ±(% of reading + % of range)
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
24 hours2
23 ºC ± 1 ºC
90 days
23 ºC ± 5 ºC
1 year
23 ºC ± 5 ºC
True rms AC
voltage3
100.000 mV
to 750.00 V4
20 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 30 kHz
30 kHz to 100 kHz7
0.60 + 0.1
0.16 + 0.1
0.80 + 0.1
3.00 + 0.2
0.60 + 0.1
0.16 + 0.1
0.80 + 0.1
3.00 + 0.2
1.00 + 0.15
0.20 + 0.1
1.00 + 0.16
3.00 + 0.28, 9
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.05 + 0.02
True rms AC
current
10.0000 mA
to 3.0000 A10
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz11
0.80 + 0.1
0.40 + 0.1
2.00 + 0.2
0.80 + 0.1
0.40 + 0.1
2.00 + 0.2
1.50 + 0.1
0.50 + 0.1
2.00 + 0.2
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
1. 20% over-range on all ranges, except for 750 Vac range.
2. Relative to calibration standards.
3. Specifications stated are for input signals greater than 5% of range.
4. Available ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V.
5. For 750 V range accuracy is specified for input less than 200 Vrms.
6. For 100 mV range, the 10 kHz to 30 kHz accuracy is 1.5 + 0.3.
7. Additional error 0.003% of full scale per kHz to be added when signal input changes less than 10% of range.
8. For 100 mV range, the 30 kHz to 100 kHz accuracy is 5 + 0.3.
9. For 750 V range accuracy is specified for input less than 300 Vrms.
10. Available ranges: 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A.
11. For 1 A and 3 A ranges, the accuracy is specified for frequencies less than 5 kHz.
6
Digital multimeter specifications (continued)
Frequency specifications
Function Input range
Frequency range:
minimum input
frequency = 1 Hz
Accuracy
±(% of reading + % of range)
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC1 year (23 ºC ± 5 ºC)
Frequency Voltage:
100 mV to 750 V1
< 2 Hz
< 20 Hz
20 Hz to 100 kHz
100 kHz to 300 kHz
0.18 + 0.003
0.04 + 0.003
0.02 + 0.003
0.02 + 0.003
0.005
0.005
0.005
0.005
Current:
10 mA to 3 A
< 2 Hz
< 20 Hz
20 Hz to 10 kHz
0.18 + 0.003
0.04 + 0.003
0.02 + 0.003
0.005
0.005
0.005
Input range2
Minimum frequency sensitivity for voltage measurement (RMS sine wave)
20 Hz to 100 kHz 100 kHz to 300 kHz 300 kHz to 1 MHz
100 mV
1.0 V
10 V
100 V
750 V
50 mV
100 mV
1 V
10 V
100 V
50 mV
120 mV
1.2 V
12 V
0.5 V
0.5 V
Input range
Minimum frequency sensitivity for current measurement (RMS sine wave)
20 Hz to 100 kHz
10 mA
100 mA
1 A
3 A
1 mA
10 mA
100 mA
300 mA
1. For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V input signal.
2. Maximum input for the specified accuracy = 10 x range or 1000 Vdc.
All frequency counters are susceptible to errors when measuring low-voltage, low-frequency signals. Shielding inputs from
external noise pickup is critical for minimizing measurement errors.
Duty cycle and pulse width specifications
Function Range Resolution Accuracy at full scale
Duty cycle 100.000%10.001% 0.3% + 0.2% per kHz
Pulse width 199.999 ms2
1999.99 ms2
0.001 ms
0.01 ms
Duty cycle/frequency
Duty cycle/frequency
1. The range is from {10 µs × frequency × 100%} ~ {[1 – (10 µs × frequency)] × 100%}. For example, a 1 kHz signal can be measured from
1% ~ 99%.
2. The positive or negative pulse width must be greater than 10 µs. The range of the pulse width is determined by the frequency of the signal.
7
Digital multimeter specifications (continued)
Measurement speeds (typical)
Function Rate
Reading speed1
(readings/second)
Reading speed
over USB2
(readings/second)
Reading speed
over GPIB3
(readings/second)
DC voltage
(10 V)
Slow (5½ digits)
Fast (4½ digits)
15
70
17
31
17
32
DC current
(1 A)
Slow (5½ digits)
Fast (4½ digits)
15
70
17
37
17
36
AC voltage
(10 V at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
15
70
17
31
17
32
AC current
(1 A at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
15
70
16
37
17
37
Resistance
(100 k)
Slow (5½ digits)
Fast (4½ digits)
15
70
17
27
17
31
Capacitance
(10 F)
Slow/Fast (3½ digits) 5 4.4 4.6
Frequency
(voltage path at
10 V, 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
9
9
2.7
2.7
2.7
2.7
Frequency
(current path at
10 V, 1 kHz)
Slow (5½ digit)
Fast (4½ digit)
9
9
2.7
2.7
2.7
2.7
1. Reading rate of the A/D converter.
2. Number of measurements per second that can be read through USB using SCPI “READ?” command.
3. Number of measurements per second that can be read through GPIB using SCPI “READ?” command.
Supplementary specifications
DC voltage
Measurement method Sigma Delta A-to-D converter
Maximum input voltage 1000 Vdc on all ranges
Input impedance 10 M ± 2% range (typical) in parallel with capacitance < 120 pF
Input protection 1000 Vrms on all ranges, < 0.3 A short circuit
Response time Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
DC current
Measurement method Sigma Delta A-to-D converter
Maximum input current 10 mA to 3.0 A DC
Burden voltage and shunt resistance •< 0.2 V, 10  for 10 mA range
•< 0.2 V, 1  for 100 mA range
•< 0.3 V, 0.1  for 1 A range
•< 0.7 V, 0.01  for 3 A range
Input protection Protected with 3.15 A/500 V, FF fuse
Response time Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
8
Digital multimeter specifications (continued)
Supplementary specifications
AC voltage
Measurement method AC coupled true rms
Maximum input voltage 750 Vrms/1200 Vpeak/3 x 107 V-Hz of product
Input impedance 1 M ± 2% range (typical) in parallel with capacitance < 120 pF
Input protection 750 Vrms on all ranges
Crest factor For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
Peak input 300% of range. Limited by maximum input.
Response time Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
Overload ranging Will select higher range if peak input overload is detected during auto range. Overload is
reported in manual ranging.
AC current
Measurement method AC coupled true rms
Maximum input current 10 mA to 3.0 A DC or AC rms
Burden voltage and shunt resistance •< 0.2 V, 10  for 10 mA range
•< 0.2 V, 1  for 100 mA range
•< 0.3 V, 0.1  for 1 A range
•< 0.7 V, 0.01  for 3 A range
Input protection Protected with 3.15 A/500 V, FF fuse
Crest factor For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
Peak input 300% of range. Limited by maximum input.
Response time Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
Resistance
Measurement method Two-wire, open-circuit voltage limited to < 5 V
Open circuit voltage < +5.0 Vdc
Input protection 1000 Vrms on all ranges, < 0.3 A short circuit
Response time Approximately 0.15 seconds for 1 M and ranges below 1 M
Low-resistance
Measurement method Four-wire; test current is sent from the FORCE terminals and resistance measured at the
SENSE terminals.
Input protection •FORCE terminals: 3.15 A/250 V
•SENSE terminals: 1000 Vrms on all ranges, < 0.3 A short circuit
Continuity
Measurement method 0.83 mA ± 0.2% constant current source
Open circuit voltage < +5.0 Vdc
Audible tone Continuous beeping when reading is less than the threshold resistance of 10  at 1.0 k
range
Input protection 1000 Vrms on all ranges, < 0.3 A short circuit
9
Digital multimeter specifications (continued)
Supplementary specifications
Diode
Measurement method 0.83 mA ± 0.2% constant current source
Open circuit voltage < +5.0 Vdc
Audible tone •Continuous beeping when level is below +50 mV DC
•Single tone for normal forward-biased diode or semiconductor junction where
0.3 V ≤ reading ≤ 0.8 V
Input protection 1000 Vrms on all ranges, < 0.3 A short circuit
Capacitance
Measurement method Computed from constant current source charge time, typical 0.2 V to 1.4 Vac signal level
Maximum voltage at full scale •For 1 nF to 10 F range: < 1.5 V
•For 100 F to 10000 F: 0.33 V
Input protection 1000 Vrms on all ranges, < 0.3 A short circuit
Response time Approximately 1 s for 100 F and ranges below 100 F
Charge and discharge voltage 5 Vpp (approximately from +3 V to –2 V)
Frequency
Measurement method Reciprocal counting technique
Signal level 0.2 V to 1.4 V
Input protection •Voltage path: 1000 Vrms on all ranges, < 0.3 A short circuit
•Current path: Protected with 3.15 A/500 V, FF fuse
Maximum display counts (excluding frequency)
5½ digits 120,000
4½ digits 12,000
Noise rejection
Common mode rejection ratio (CMRR)
for 1 k unbalanced in LO lead
•DC: 140 dB
•AC: 70 dB
Normal mode rejection ratio (NMRR) •60 Hz ± 0.1%:
5½ digits: 65 dB, 4½ digits: 0 dB
•50 Hz ± 0.1%:
5½ digits: 55 dB, 4½ digits: 0 dB
10
Power supply specifications
Specification assumptions:
•Specifications stated are after 60-minutes of warm-up with no load
•Operating temperature: 18 °C to 28 °C (64.4 °F to 82.4 °F)
•Accuracy is expressed as ±(% of output + offset) at 23 °C ± 5 °C
•Temperature coefficient: Add [0.1 x (the specified accuracy) / °C] for 0 °C to 18 °C and 28 °C to 55 °C
•Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for 30 °C to 55 °C
DC power supply specifications
Output ratings •Range S1: 0 V to 30 V, 0 A to 1 A
•Range S2: 0 V to 8 V, 0 A to 3 A
Programming accuracy
1 year (@ 23 ºC ± 5 ºC), ±(% of output + offset)
•0.05% + 5 mV
•0.15% + 3 mA
Readback accuracy
1 year over GPIB and USB or front panel with respect to
actual output (@ 23 ºC ± 5 ºC), ±(% of output + offset)
•0.05% + 5 mV
•0.15% + 3 mA
Ripple and noise
With outputs ungrounded, or with either output
terminal grounded, 20 Hz to 1 MHz3
•< 2 mVrms; < 30 mVpp
•< 1 mArms
Front terminal load regulation
±(% of output + offset)
•< 3 mV1
•< 0.03% + 0.3 mA
Rear terminal load regulation
±(% of output + offset)
•< 0.01% + 3 mV
•< 0.03% + 0.3 mA
Line regulation Voltage: 3 mV
Current: 1.5 mA
Programming resolution 1 mV, 0.1 mA
Readback resolution 1 mV, 0.1 mA
Front panel resolution 1 mV, 0.1 mA
Transient response time Less than 450 ms for output to recover to within 50 mV following a
change in output current from full load to half load or vice versa
Command processing time Average time for output voltage to begin to change after receipt of
digital data when instrument is connected directly to the USB or GPIB
is less than 100 ms
Overvoltage protection (for CC mode) •Accuracy: 0.5% + 0.5 V
•Activation time2: < 2ms
Overcurrent protection (for CV mode) •Accuracy: 0.5% + 0.05 A
•Activation time2: < 2 ms
1. Contacts and leads resistance may contribute an additional error of 6 mV/A (typical).
2. Average time for the detection of OVP or OCP condition.
3. Refer to U3606A User’s and Service Guide (U3606-90013); CV noise effect measurement on page 166.
11
Power supply specifications (continued)
Sweep specifications
Function Range Amplitude1Step Dwelling time
Scan Constant voltage S1 (30 V/1 A)
S2 (8 V/3 A)
0 to 31.500 V
0 to 8.4000 V
1 step to 100 steps 1 s to 99 s
Constant current S1 (30 V/1 A)
S2 (8 V/3 A)
0 to 1.0500 A
0 to 3.1500 A
1 step to 100 steps 1 s to 99 s
Ramp Constant voltage S1 (30 V/1 A)
S2 (8 V/3 A)
0 to 31.500 V
0 to 8.4000 V
1 step to 10,000 steps 300 ms/step (typical)
Constant current S1 (30 V/1 A)
S2 (8 V/3 A)
0 to 1.0500 A
0 to 3.1500 A
1 step to 10,000 steps 300 ms/step (typical)
1. Amplitude start position is fixed at 0 (V or A) by default.
Square-wave output specifications
Parameter Range Resolution Accuracy
Frequency 0.5, 2, 5, 6, 10, 15, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150,
200, 240, 300, 400, 480, 600, 800, 1200, 1600, 2400, 4800 Hz
0.01 Hz 0.005% + 1 count
Duty cycle 0.39% to 99.60% 0.39%10.4%1, 2
Pulse width 1/frequency Range/256 Duty cycle/frequency1, 3
Amplitude S1 (30 V/1 A)
S2 (8 V/3 A)
1 mV
1 mV
0.2 V
0.2 V
1. Specification applies when the positive or negative pulse width is greater than 50 µs.
2. For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is to be added.
Accuracy = 0.4% + ( frequency – 1 ) × 0.1%
100
3. The accuracy of pulse width could also be calculated as [0.4% + (frequency/100 – 1) × 0.1%]/frequency.
Rise/fall time is less than 25 s. Specifications are based on a resistive load.
Supplementary specifications
Output programming range
Range Output programming OV/OC OVP/OCP
S1 0 V to 31.500 V/0 A to 1.05 A 31.500 V/1.05 A 33.000 V/1.1 A
S2 0 V to 8.4 V/0 A to 3.15 A 8.4 V/3.15 A 8.8 V/3.3 A
Remote sensing capability
Range S1 (30 V/1 A) Up to a 0.75-volt drop per load lead
Range S2 (8 V/3 A) Up to a 0.5-volt drop per load lead
Temperature coefficient: ± (% of output + offset)/ °C for 0 °C to 18 °C and 28 °C to 55 °C
Voltage 0.005% + 0.5 mV/°C
Current 0.02% + 1 mA/°C
Voltage programming speed (excludes command processing time)
Up 300 ms (full load and no load)
Down 400 ms (full load and no load)
12
General characteristics
Power supply •Universal 100 Vac to 240 Vac ±10%
•AC line frequency 45 Hz to 66 Hz (360 Hz to 440 Hz for 100/120 V operation)
Power consumption 150 VA maximum
Current Input Fuse 3.15 A, 500 V FF fuse (on front panel)
Display Highly visible vacuum-fluorescent display (VFD)
Operating environment •Operating temperature from 0 °C to +55 °C
•Relative humidity up to 80% at 30 °C RH (non-condensing)
•Altitude up to 2000 meters
•Pollution degree 2
•For indoor use only
Storage compliance –40 °C to 70 °C
Safety compliance Certified with:
•IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
•Canada: CAN/CSA-C22.2 No. 61010-1-04
•USA: ANSI/UL 61010-1:2004
EMC compliance Certified with:
•IEC61326-1:2005 / EN61326-1:2006
•CISPR 11:2003 / EN55011:2007 (Group 1 Class A)
•Canada: ICES/NMB-001:2004
•Australia/New Zealand: AS/NZS CISPR11:2004
Shock and vibration Tested to IEC/EN 60068-2
Remote interface •GPIB IEEE-488
•Full Speed USB 2.0 (Type B)
•USBTMC 488.2 Class device
•USB-CDC
Programming language Standard Commands for Programmable Instruments (SCPI)
Measurement category •CAT II, 300 V
•CAT I, 1000 Vdc, 750 Vac rms
•2500 Vpk transient over voltages
Dimensions (H × W × D) 105 × 255 × 329 mm (with rubber bumpers)
87 x 215 x 312 mm (without rubber bumpers)
Weight 3.775 kg (with rubber bumpers)
3.535 kg (without rubber bumpers)
Warranty Three years
Calibration cycle One year
Warm-up time 60 minutes
13
Ordering information
Standard shipped items
•Quick Start Guide
•Product Reference CD
•Agilent IO Library Suite
•Certificate of Calibration
•U8201A Combo Test Lead Kit
•USB 2.0 High-Speed Type-A to
Type-B cable
•AC power cord
Warranty options
•R-51B-001-3C Extended warranty
from one year to three years
•R-51B-001-5C Extended warranty
from one year to five years
U8201A Combo Test Lead Kit U3606A-1CM Rack Mount Kit U8202A Electronic Test Lead Kit
(for DMM function)
34133A Precision Electronic Test Leads
(for DMM function)
34330A Current Shunt (30 A)
(for DMM function)
34136A 40 kV high-voltage probe
(for DMM function)
11059A Kelvin Probe Set and
11062A Kelvin Clip Set (for DMM function)
E3600A-100 Test Lead Kit
(for DC power supply function)
I/O connectivity options
For control via GPIB interface
•82350B/82351A PCI/PCIe
high-performance GPIB
interface card
•82357B USB/GPIB converter
•E5810A LAN/GPIB gateway
•10833D/A/B/C/F/G GPIB cables
•10834A GPIB-to-GPIB adapter
For control via USB interface
•E5813A networked 5-port USB hub
Optional accessories
myAgilent
myAgilent
www.agilent.com/find/myagilent
A personalized view into the information most relevant to you.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe)
is an open standard that extends the AdvancedTCA for general
purpose and semiconductor test. Agilent is a founding member of
the AXIe consortium.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Agilent is a founding member of the
LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation
delivers a rugged, PC-based high-performance measurement and
automation system.
Three-Year Warranty
www.agilent.com/find/ThreeYearWarranty
Beyond product specification, changing the ownership experience.
Agilent is the only test and measurement company that offers
three-year warranty on all instruments, worldwide.
Agilent Assurance Plans
www.agilent.com/find/AssurancePlans
Five years of protection and no budgetary surprises to ensure your
instruments are operating to specifications and you can continually
rely on accurate measurements.
www.agilent.com/quality
Agilent Electronic Measurement Group
DEKRA Certified ISO 9001:2008
Quality Management System
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s measurement expertise and
product breadth, combined with channel partner convenience.
www.agilent.com
www.agilent.com/find/U3606A
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894 4414
Brazil (11) 4197 3600
Mexico 01800 5064 800
United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 375 8100
Europe & Middle East
Belgium 32 (0) 2 404 93 40
Denmark 45 45 80 12 15
Finland 358 (0) 10 855 2100
France 0825 010 700*
*0.125 €/minute
Germany 49 (0) 7031 464 6333
Ireland 1890 924 204
Israel 972-3-9288-504/544
Italy 39 02 92 60 8484
Netherlands 31 (0) 20 547 2111
Spain 34 (91) 631 3300
Sweden 0200-88 22 55
United Kingdom 44 (0) 118 927 6201
For other unlisted countries:
www.agilent.com/find/contactus
(BP-09-27-13)
Product specifications and descriptions in
this document subject to change without
notice.
© Agilent Technologies, Inc. 2011, 2014