Data Sheet
Battery Analyzers
BA6010 Series
Technical data subject to change
© B&K Precision Corp. 2018 www.bkprecision.com
The BA6010 Series battery analyzers measure
voltage and resistance of modern battery
technologies with high accuracy, resolution, and
speed. Additionally, these instruments provide
auxiliary measurement parameters inductance,
dissipation factor, impedance, quality factor,
reactance, phase angle in degrees, and
capacitance in farads.
The BA6010 Series is suitable for characterizing
battery chemistries that are responsive to a 1 kHz
AC stimulus signal, including lead acid, lithium
and alkaline type batteries used in consumer
products, electric vehicles, power backup,
security, and fire alarm systems. Model BA6011
supports voltage measurements of battery packs
up to 300 V whereas the BA6010 features a 6 V
measurement range ideal for battery cell testing.
The handler and remote interfaces expands the
analyzer’s application to R&D and automated
manufacturing environments.
Features & Benefits
n 4.3 inch color LCD display
n Trace function for graphical display of voltage
and resistance with on-screen cursor
measurements
n 4-wire kelvin test leads with fault monitoring of
drive and sense lines
n Compare and sort using 9 bins with statistical
evaluations
n Δ% mode for quickly determining the percent
difference between batteries
n Pass/Fail indicator with audible tone
n Fast test speed up to 50 measurements per
second to increase manufacturing throughput
n Trigger modes internal, manual, bus and
external
n 100 internal and external storage locations for
setup and screen save
n Handler interface for easy integration with a
component handler or integration with PLC
n Standard RS232, USB (USBTMC and virtual
COM) interfaces
Model Input voltage
range
Basic voltage
accuracy
Voltage
resolution
BA6010 6 V / 60 V 0.05 % 100 µV
BA6011 30 V / 300 V 0.05 % 1 mV
USB RS232
Two user-selected measurements can be displayed
simultaneously, along with stimulus signals
Vm and Im. Unlike comparable battery testers
that only support voltage and resistance
measurements, users can also characterize
additional parameters such as battery capacitance
thus providing additional insight into a battery’s
condition.
Wide range of measurements
Main measurement parameters
Auxiliary measurement parameters
Simplified Randles cell
RS
Cdl
Rct
Battery Analyzers
BA6010 Series
2www.bkprecision.com
Front panel
Rear panel
BNC Trigger
terminal
AC input
terminal
Chassis ground terminal RS232
interface
USB
interface
Handler interfaceFuse line
selector box
Standard RS232, USB (USBTMC and virtual COM) interfaces, handler interface and external BNC trigger
input are useful for production automation.
Bright LCD display Numeric
keypad
USB
host port
Power button
Display and
setup keys
Menu function keys
Function
keypad
Navigation
keys
Input terminals
Pass/Fail Indicator
Large 4.3 inch color LCD screen for easy viewing of configuration and measurements. 4-terminal front panel
connection and quick connect test fixture for high accuracy measurements.
Battery Analyzers
BA6010 Series
3www.bkprecision.com
Flexible operation
On screen monitoring system detects test probe
contact failure and damaged leads for reliable
measurements.
Error: HI drive open
Error: LO drive open
Error: HI sense open
Error: LO sense open
Error: Measure line open
Comparator function
The comparator function evaluates measurements
against a user specified upper and lower limit for
pass/fail (Go/No Go) style testing. Comparative
evaluations can be made using primary, secondary
or both measurements. The front panel PASS /
FAIL indicator will illuminate and a sounder can
be enabled for audible confirmation.
Compare test - below limits
Compare test - above limits
Bin comparator display
Statistical function
Statistical tools menu
The analyzers can perform statistical calculations
on the measurements and display the results
on-screen.
Kelvin clip test fixture
The 4-terminals on the front of the BA6010
Series are used together with the Kelvin clip test
fixture. This system minimizes the influence of
the test lead resistance and improves
measurement accuracy.
Improved measurement accuracy
4-wire kelvin connection
Bins for sorting devices
Binning function
Quickly sort components using up to 9 bins. The
bin results are displayed on-screen with each
cycle. The handler interface includes dedicated
signal pins for each bin, Pass/Fail and end of
measurement. The handler interface is suitable
for integration with device handler systems or
programmable logic controllers (PLC) used in
production automation.
Trace function
Trace display
The trace function samples and plots two
user-selected measurement readings over a
specified time. Enable cursors for viewing plotted
values and time stamp information.
Cursor with
measurements
Battery Analyzers
BA6010 Series
4www.bkprecision.com
Specifications
Voltage Measurement (BA6010)
SLOW, MED
Range Maximum Display Value Resolution Accuracy
6 V 6.5000 V 100 µV ±(0.05 % FS)
60 V 65.000 V 1 mV
FAST
Range Maximum Display Value Resolution Accuracy
6 V 6.500 V 1 mV ±(0.1 % FS)
60 V 65.00 V 10 mV
Voltage Measurement (BA6011)
SLOW, MED
Range Maximum Display Value Resolution Accuracy
30 V 35.000 V 1 mV ±(0.05 % FS)
300V 310.00 V 10 mV
FAST
Range Maximum Display Value Resolution Accuracy
30 V 35.00 V 10 mV ±(0.1 % FS)
300 V 310.0 V 100 mV
Model BA6010, BA6011
Measurement
Parameters
Main V, R
Auxiliary L, C, D, Z, X, Q, θd, and θr
Test Frequency 1 kHz ± 0.2 Hz
Display Resolution 5 digits (SLOW & MED), 4 digits (FAST)
Measurement Speed
SLOW, approx. 6.25 measurements/sec
MED, approx. 10 measurements/sec
FAST, approx. 50 measurements/sec
Temperature
Coefficient
Voltage Meas. 0.005 % / ºC
Resistance Meas. 0.05 % / ºC
Triggering Internal, External, Manual, Bus
Delay Time On / Off, 0 ms to 60 s
Averaging 1 to 255 samples
Statistical Calculations
Valid data count, Invalid data count, Mean, Maximum, Minimum,
Standard Deviation, Sample Standard Deviation,
Process Capability Index (Dispersion), Process Capability Index (Deviation)
All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature
range of 20 °C ± 5 °C. Specifications are subject to change without notice.
Battery Analyzers
BA6010 Series
5www.bkprecision.com
v080918
General
Save/
Recall
Instrument Settings
Save / Recall Internal or External Memory: Up to 100
Measurements, Bin Comparator Results, Screenshots
Save External Memory: Up to 100
Remote Interface USBTMC / USB (Virtual COM), RS232,
Display 4.3”, 480 × 272 LCD display
AC Input 110 V ±10 % or 220 V ± 10 %, 47 to 63 Hz
Power Consumption 15 VA Max.
Operating Temperature 0 °C to 40 °C
Storage Temperature -10 °C to 70 °C
Relative Humidity up to 80 %
Dimension (W×H×D) 9.25” x 4.1” x 14.17” (235 x 104 x 360 mm)
Weight 7.9 lbs (3.6 kg)
Three-Year Warranty
Included Accessories User manual (downloadable), power cord, 4-wire kelvin clip test fixture (TLKB1), certificate of calibration & test report
Bin Comparator Function
Limit Setting Mode Tolerance (TOL) or Absolute (ABS) value
Number of Bins 9 sorting bins BIN1-BIN9
Beep Warning OFF, PASS, FAIL
Trace Function
Total Time 1 s - 99999 s
Sampling Interval 1 s - 86400 s
Specifications
Resistance Measurement
SLOW, MED
Range Maximum Display Value Resolution Measurement Current Accuracy
30 mΩ33.000 mΩ1 µΩ100 mA (± 10 %)
±(0.3 % + 0.1 % FS)
300 mΩ330.00 mΩ10 µΩ100 mA (±10 %)
3 Ω3.3000 Ω100 µΩ10 mA (± 10 %)
30 Ω33.000 Ω1 mΩ1 mA (± 10 %)
300 Ω330.00 Ω10 mΩ100 µA (± 10 %)
3 kΩ3.5000 kΩ100 mΩ10 µA (± 10 %)
FAST
Range Maximum Display value Resolution Measurement Current Accuracy
30 mΩ33.00 mΩ10 µΩ100 mA (± 10 %)
±(0.5 % + 0.3 % FS)
300 mΩ330.0 mΩ100 µΩ100 mA (± 10 %)
3 Ω3.300 Ω1 mΩ10 mA (± 10 %)
30 Ω33.00 Ω10 mΩ1 µA (± 10 %)
300 Ω330.0 Ω100 mΩ100 µA (± 10 %)
3 kΩ3.500 kΩ1 Ω10 µA (± 10 %)
Accuracy of Auxiliary Measurement Parameters
L, C, D, Z, X, Q, θd, and θr 5 % typical**
** see user manual for more details