183
QUICK FIND
the smart approach to instrumentation
™
®
Temperature & Voltage Measurement
tel: 440-439-4091 fax: 440-439-4093 sales@iotech.com www.iotech.com
DataShuttle
™
Specifications
Specifications
General
Power Requirements: +5 to +9 VDC @ 300 mA; AC
adaptor included
Operating Temperature: 0˚ to 50˚C
Resolution* Speed
20 bits** 45/55 Hz†
18 bits 45/55 Hz†
16 bits 200 Hz
14 bits 550 Hz
12 bits 1000 Hz
11 bits 1500 Hz
Dimensions: 190 mm W x 140 mm D x 50 mm H
(7.5” x 5.5” x 2.0”)
Voltage Measurements
(Applies to General Purpose version, Thermocouple
version, and when the range resistor is removed,
the RTD version)
8 differential inputs with screw terminal signal con-
nections
Ranges:
6 ranges, individually software selectable: -5 mV
to +50 mV, -25 mV to +25 mV, -50 mV to +500 mV,
-250 mV to +250 mV, -1V to +10V, -5V to +5V
Input Protection: 50V continuous, 150V for one
second
Input Resistance: >109 Ohm (Ω)
Input Current: <3 nA
Accuracy: 0.05% of reading, +0.01% of range (15˚ to
35˚C)
Temperature Coefficient: ±(40 ppm + 1 µV), 0˚ to
15˚C, and 35˚ to 50˚C
Common Mode Rejection Ratio (CMRR): >80 dB,
±5V and 10V ranges, >100 dB on all other ranges
Normal Mode Rejection Ratio (NMRR): 50 or 60 Hz,
40 dB min @ 18-bit resolution
Thermocouple Measurements
Thermocouple Types: E, J, K, T, B, R, S, C, D, G, N
Thermocouple Model
Isothermal Block insures accurate readings even
under thermally dynamic conditions
Accuracy: 0.3˚C (with environment @ 15˚ to 35˚C)
Resolution: 0.01˚C
General Purpose and RTD Model
(with range resistor removed)
Accuracy: 0.8˚C
Resolution: 0.01˚C
Isolated 5B Model
Accuracy: 1.0˚C
RTD Measurements
8 channels of 2- or 3-wire measurements
RTD Types: 100 Ohm Platinum
(can also accommodate 50 and 200 Ohm)
115C Model 750C Model
Range -200˚ to +115˚C -200˚ to +750˚C
Resolution 0.005˚C 0.01˚C
Accuracy 0.8˚C 1.0˚C
(with environment @ 15˚ to 35˚C)
Isolated Measurements
(Applies to Isolated 5B model only)
8 sockets are provided for user-installation of up to 8
isolated 5B modules
Each channel has screw-terminals for signal con-
nections
Each channel has a cold-junction sensor for thermo-
couple measurements
Isolation: Channel-to-Channel or Channel-to-
Common, 500V max
Accuracy: Dependent on module
Analog Output Modules: Up to 2 channels can be
configured for isolated analog output when the
analog output version is purchased (DS-16-8-5B-AO)
* Resolution refers to full-scale readings; up to 10 times more resolution is provided at the lower 10% span
of each range
** 20 bits of resolution are possible in the 18 bit resolution mode for inputs <10% of full scale
† 45 Hz when in 50 Hz low-noise mode, 55 Hz when in 60 Hz low-noise mode
Digital I/O
8 I/O lines, individually programmable as input or
output
Each I/O line can be TTL-compatible, or open-
collector
Optional modules can be installed on each channel,
providing high-isolation and high-voltage switch-
ing and sensing on a per-channel basis
Input Voltage Range: 0 to +7V max
Input Voltage Logic Levels: TTL-compatible
Output Voltage Range: 0 to 30V max (with user-
supplied pull-up resistors)
Output Current Sink: 50 mA/channel continuous,
400 mA max per 8 channels
Speed: Up to 1500 Hz (maximum is the same as the
selected analog input measurement speed)
Analog Output (Models with –AO)
Two independent 12-bit programmable channels
Voltage Output Ranges (software selectable):
0 to +10V, 0 to +5V, 0 to +2V, -5V to +5V,
-2.5V to +2.5V, -1V to +1V
Voltage Output Speed: Up to 1500 Hz (maximum is
equal to maximum analog input speed)
Current Output Range: 4 to 20 mA (5 to 15V com-
pliance)
Speed: Up to 1500 Hz (maximum is the same as the
selected analog input measurement speed)
Accuracy: ±0.024% of full scale
Output Protection: -12V to +15V in volts mode, 0 to
+15V in current mode