Special product features GER L C ER OG & L T DS U P TO RE C A MP ME OR Model CL601 Designed for plant and field service applications and for use in harsh environments UT E B TO N ' O PE N ' O 6 C 00 AA N DATA LOGGERS Simple Logger(R) II TRMS Clamp-On Current R AT I O SPECIFICATIONS MODEL ELECTRICAL CL601 Channels One Input Split CT - AC Current Measurement Range 0 to 600Aac Resolution 0.1A Accuracy (50/60Hz) 0 to 5A: unspecified 5 to 50A: (1% of Reading + 1A) 50 to 400A: (1% of Reading + 0.5A) 400 to 600A: (3% of Reading + 1A) 64 samples/cycle Sample Rate Storage Rate IP 40 Rated Programmable from 8 every second to 1 every day Storage Modes Recording Length Memory Start/Stop, FIFO, Extended Recording Mode (XRMTM)* and Alarm 15 minutes to 8 weeks, programmable using DataView(R) 240,000 measurements (512KB) - The recorded data is stored in non-volatile memory and will be retained even if the battery is low or removed. Communication Power Source USB 2.0 optically isolated 2 x 1.5V AA-cell Alkaline batteries (included) Battery Life 100 hours to >45 days (dependent on sample rate and recording length) MECHANICAL Dimensions 9.25 x 4.0 x 1.63" (235 x 102 x 41mm) Max Conductor Size Weight (with battery) Case 1 conductor - O 1.42" (36mm) 2 conductor - O 1.00" (25mm) ea 17.1 oz (485g) UL94-V0 Vibration IEC 68-2-6 (1.5mm, 10 to 55Hz) Shock IEC 68-2-27 (30G) Drop IEC 68-2-32 (1m) FEATURES * * * * * * * * * * * 0 to 600Arms True RMS measurements Self contained, no exposed connections Overload indication Optically isolated USB 2.0 communication cable included One button operation Alarm function 5 LED indicators quickly and clearly display logger status Powered by standard Alkaline batteries Includes FREE DataView(R) software for data storage, real-time display, analysis and report generation EN 61010-1; 300V CAT IV, 600V CAT III APPLICATIONS *Extended Recording Mode (XRMTM) This unique recording mode provides the opportunity to continuously record over long periods of time by reducing the stored samples of the oldest data and maintaining matching resolution for the newest data. Each time the memory fills up using XRMTM every other of the oldest stored samples is discarded making room for newer samples. This process continues until the recording is manually stopped. * * * * * Machine load monitoring HVAC troubleshooting Load profiling Electrical troubleshooting Start-Stop time stamping CATALOG NO. DESCRIPTION 2126.01 Simple Logger(R) II Model CL601 (1-Channel, TRMS Clamp-On, 600Aac, DataView(R) software) Technical Assistance (800) 343-1391 DATAVIEW(R) Software for Data Loggers Data Analysis and Reporting Software for Data Loggers Typical DataView(R) Functional Displays Configure all data logger functions of the Simple Logger(R) II Models Quick and simple configuration of all functions and settings from one dialog box. * D isplay and analyze real-time data on your PC * C onfigure all data logger functions and parameters from your PC including sample rate, recording length, channel configuration and more * Create and store a complete library of configurations that can be uploaded to the logger as needed * Zoom in and out and pan through sections of the graph to analyze the data * Download, display and analyze recorded data * Display waveforms, trend graphs, harmonics (AC models) and text summaries * Create custom views and reports * Print reports using standard or custom templates you design * Free software upgrades are available on our website www.aemc.com Configure all alarm functions with straightforward selections. Real-time display of all active inputs on computer through DataView(R) software. Real-time view of trend, waveform and status screens. Technical Assistance (800) 343-1391