SERVICE
If the instrument fails to operate, check battery, test leads, etc and
replace as necessary. If the instrument still malfunctions, please call
the phone number listed below:
Service Division
AMPROBE INSTRUMENT
Miami, Florida 33150
Tel: 800-327-5060
outside the U.S.A. the local Amprobe representative will assist you.
Visit our Web-site HTTP://www.AMPROBE.COM
|AMPROBE®
DIVISION OF CORE INDUSTRIES INC.
Miami, Florida 33150
Tel: 800-327-5060
USER MANUAL
MODEL: ACDC-620T
|AMPROBE®
Miramar, Florida 33025
Tel: 800-327-5060
AUNIQUE AC / DC TRUE RMS
CLAMP-ON MULTIMETER
UL
®
C
LISTED
Manual P/N: 978760
8/97
z
1
LIMITED WARRANTY
Congratulations! You are now the owner of an AMPROBE®instrument. It has been
quality crafted according to quality standards and contains quality components and
workmanship. This instrument has been inspected for proper operation of all its
functions. It has been tested by qualified factory technicians according to the long-
established standards of AMPROBE®.
Your A M P R O B E® instrument has a limited warranty against defective materials
and/or workmanship for two years from the date of purchase provided that, in the
opinion of the factory, the instrument has not been tampered with or taken apart.
Should your instrument fail due to defective materials, and/or workmanship
during the two-year period, please have your dated bill of sale which must
identify the instrument model number and serial number and call the number
listed below:
Service Division
AMPROBE INSTRUMENT
Miami, Florida 33150
Tel: 800-327-5060
2
CONTENTS
PÁGINA
LIMITED WARRANTY 1
SAFETYINFORMATION 4
AUNIQUE CLAMP-ON MULTIMETER 6
INTRODUCTION 6
USING THE METER SAFELY 8
LCD DISPLAYILLUSTRATION 10
GETTING ACQUAINTED WITH YOUR METER 12
ALIGNMENTMARKS 12
ROTARYSWITCH 13
INPUTTERMINAL 14
PUSH BOTTONS 15
PUSH-BOTTONS OPERATIONS 16
POWER ON OPTIONS 18
SPECIALFUNCTIONS INSTRUCTIONS 19
Dynamic Recording 19
Data Hold 22
Zero (Relative)
Analog Bargraph 23
Auto Power Off and Sleep Mode 24
Disable Auto Power Off 25
Demostrate Annunciator 25
Continuity Function For Ohms Measurement 26
1 ms Peak Hold 27
CONTENTS continued
PAGE
HOW TO OPERATE
AC CURRENTMEASUREMENT 29
DISTRIBUTION TRANSFORMER MEASUREMENT 29
AC MOTOR CURRENTMEASUREMENT 31
AC VOLTAGE MEASUREMENT 33
RESSISTENCE / CONTINUITYMEASUREMENT 35
TEMPERATURE MEASUREMENT 39
CAPACITANCE MEASUREMENT 43
ESPECIFICACIONES GENERALES 45
GENERALSPECIFICATIONS 47
ACCESSORIES AND REPLACEMENT PARTS 48
CURENT HARMONICS THEORY 51
TRUE RMS MEASUREMENT 53
WAVE FORM COMPARISON 54
REFRIGERATION 56
MAINTENANCE 61
SERVICE
BATERYREPLACEMENT 61
CLEANING 64
3
4
SAFETYINFORMATION
cSAFETYINFORMATION:
To ensure that you use meter safely, follow the safety guidelines listed below.
This meter is for indoor use, altitude up to 2000m.
Avoid working alone. Take precautions when working around moving parts.
Use the meter only as specified in this manual. Otherwise, the protection pro
vided by the meter may be impaired.
Never measure current while the test leads arer inserted into the input termi
nals.
Do not use the metr if it looks dameged.
Inspect the leads for damage insulation or exposed metal. Check test lead
continuity. Replace damaged leads
Disconnect the power and discharge all high-voltage capacitors before testing.
Use caution when working above 60V DC or 30V AC RMS. Such voltages
pose a pose a shock hazard
When making measurements, keep your fingers behind the finger guards on
the probe.
Set the proper function and renge before attaching the metr to circuit. To avoid
damaging the meter disconnect the test leads from test points before changing
functions.
Read this operation manual completly before using the meter and follow all
safety instructions.
The meter is safety-certified in compliance with UL3111-1, C22.2 NO.1010.1-
92 and EN61010(IEC 1010-1, 1010-2031,IEC 1010-2-32) Instalation Category
ll 1000V or instalation Category lll 600V. In order to mantain its insulation
properties, please be sure to use ULListed Category ll 1000V or Category lll
600V probes.
insrtallation category (CAT) ll is an environment with smaller trasient overvolt-
age than Installation Category lll.
CE requirement: Under the influence of R.F field according to standard, the
supplied test leads will pick up induced noice. To have better shielding tests
are required in order to conform to CE:
1. IEC 801-2: ESD (electro-static discharge) test.
2. IEC-801-3: RFI (Radio frequency Interference) test.
Condition : 27 ~ 500MHz, signal intensity is 3 volts per meter.
3. IEC 801-4: EFT(electro-fast transient) test.
4. EN 55011: EMI (electromagnetic interference) test.
Do not allow the temperature sensor to contact a surface which is energized
above 30 V RMS or 60V DC, such voltages pose a shock hazard.
5
6
AUNIQUE CLAMP-ON MULTIMETER
cINTRODUCTION
Measuring current accurately is difficult job in today's industrial plants and com-
mercial buildings.An increasing number of personal computers, adjustable
speed motor drivers, and other types of electronic equipment in short pulses and
are reffered to as non-linear loads.
non-linear loads draw high peak current, causing harmonics in the load current.
this may result in unexplained circuit breaker tripping, or dangerous overheating
of neutral conductors and transformers. Currents containing harmonics can only
be accurately measured with a true rms or clamp meter. this CLAMP-ON MULTI-
METERis shown in Figure1.
This meter has many functions which are shown below:
TRUE RMSmeasurement for non-linear and traditional loads.
1ms Peak. Hold feature to capture glitch or in-rush current.
Differentiation reading for Dual temperature measurement.
Start capacitor measurement.
Dynamic Recording helps to record the variation of test.
Hand Guard for prevention of accidental contact with conductors.
Carryng case with shoulder strap
Data Hold to freeze displayed digital value.
Relative(zero) function
Auto and Manual Ranging
Figure 1. AUnique Clamp-On Multimeter
7
Start capacitor
measurement
Ohm measurement and
k-type temperature
Voltage measurement
Current measurement
1ms Peak Hold for
glitch capture
Different Temperatures
measurement
Hand guard design for
more safety.
Dynamic Recording
helps to record the
variation of tests. And
one touch DATA
HOLD.
Rotary switch for easy
operation.
Dual display to indicate the
of and °C of temperature
simultaneously.
Relative (ZERO) mode for
deviation measurements.
AUTO/MANUAL
SELECT
USING THE METER SAFELY
cWARNING
Read " SAFETYINFORMATION" before using the meter..
cNOTE
Some typical tests are provided in this manual. These tests are designed to help
you understand how to use the Meter. Consult original manufacturer service
manualfor the test procedures that apply to your particular piece of equipment.
Your Clamp-on multimeter is a hand-held, bateryy operated instrument for test-
ing and troubleshooting electrical and electronic systems. If the meter is dam-
aged or something is missing, contact the place of purchase immediately.
AWARNING identifies conditions and actions that pose hazard(s) to the user; a
CAUTION identifies and actions that may damage the Meter.InternTIONl electri-
cal symbols used are explained inTable 1.
Tabla 1. Símbolos Eléctricos Internacionales
8
AC- Alternating Current
DC- Direct Current
AC and DC-Alternating and Direct Current
Ground
Double Insulation
See Explanation In The Manual
w
y
y
v
G
G
x
LCD DISPLAYILLUSTRATION
1) - Negative polarity Annunciator
2) @OFF Auto Power Off Enabled Annunciator
3) q Low Battery Annunciator
4) DC Direct Current or Volytage Annunciator
5) AC Alterning Current or Volytage Annunciator
6) AUTO AUTOrange Mode Annunciator
7) DH Data hold Annunciator
8) DH MAX Peak hold Annunciator
9) MAX AVG MIN Dynamic recording mode, presasent reading
10) MAX Maximum reading
11) MIN Minimum reading
12) AVG Average reading
13) n)))Continuity function annunciator
14) T1-T2 Delta Temperature (withDKTA-620 adapter)
15) T1 T2 T1 or T2 temperature measurement
16) °C, °F Unit of temperature
17) V Unit of Voltage measurement
18) A Unit of Curent measurement
19) kW Unit of Resistance (ohm) measurement
20) _ Zero (Delta) mode annunciator
21) 8.8.8.8 Digital display for degree °F, A, V and
22) 8000 8000 Analog bar-graph annunciator with sacle indicator.
-IIIIIIIIIl4
23) µFUnit of Capacitor measurement
24) - 8.8.8.8 Digital display for degree C
10
11
Figure 2. LCD Display
12
Figur3 3. Alignment Marks
In order to meet the meeter accuracy spacifications when making a current meas-
urement, the conductor must be inside the jaws and centered within the indicated
marks as much as possible (see Figure 3).
Getting Acquainted WIth Your Meter
cALIGNMENT MARKS
Conductor
Mark
MARK
MARK
Rotary Switch
To turn the meter on and select a function, turn the rotary switch
(Figure 4) to a switch setting. The whole display lights for one second.
Then the meter is ready for use. (if you press and hold down any push
button while turning the meter from OFF to ON, the display remains [it
until the push button is released.)
1. OFF: Power off position.
2. A: AC or DC Current. Default is AC current.
El valor predetermina.
do es corriente alterna (CA).
3. V: AC or DC voltage. Default is AC voltage.
4. Temp : Temperature/Differential Temperature, Ohms,
and Continuity. The continuity buzzer sounds when
test value that is displayed is below 100 counts.
5. µF: Capacitance measurement.
13
5 µF
4 Temp.
3 V
2 A
1OFF
Figure 4. Rotary Switch
y
y
INPUT TERMINAL
WARNING
To avoid damaging the meter, do not exceed input limit shown below
Table 1:
Table 1. Input limit specificationTable 1.
The meter has two input terminals (Figure 5) that are protected
against overloads to the limits shown in the specifications.
1. Terminal común para todas las mediciones excepto la de corriente.
2. Voltios, Ohmios, Prueba de Capacidad y mediciones de
Temperatura.
14
ROTATORYSWITVH FUNCTION
AC 400 ~ 1000V(CATII)
AC 400 ~ 600V(CATIII)
DC 400 ~ 1000V(CATII)
DC 400 ~ 600V(CATIII)
DC 400 ~ 1000 A
AC 400 ~ 1000 A
OHM ()
Capacitance (µF)
Temperature
INPUTTRMINAL
V - - µF & COM
Clamp jaw
V - - µF & COM
INPUTLIMIT
CATII
1000V de CA/1000V DC
CATIIl
600V
1000ARMS
600V RMS
21
Figure 5. Input Terminal
Figure 6. Push bottons
15
1. Press to lselect
DC.AC,DC+AC
Press and hold for
more than 1 second
to toggle, between
PEAK" and DC.
Press to toggle 0
and Temperature
measurements.
5. Press to toggle Continuity
ON/OFF for Ohm measurement.
Press to select T1-T2-T2 for tem
perature measurement.
3. Press and hold for more
than 1 second to set
Dynamic Recording then
press to step through
MAX,MIN, AVG and present
readings..
6. In manual range press to step
up 1 range at a time. Press
and hold for 1 sec to select
Auto range.
4. Press to toggle the relative
(ZERO) mode ON/OFF.
2. Press to toggle
"DH" ON/OFF.
16
Push-button Operations
The operation of the push-buttons are outlined below. When a buton is pushed,
an annunciator lights, and the unit beeps. turning the rotary switch to another
switch setting resets all push buttons to thei default states. The pushbuttons are
shown in (page15).
1. Shift / Peak :
This push-button is used for selecting the measurement of either
an Alternating or Direct source, Ohms or Temperature, or for selecting the
PEAK hold function.
Press this button momentarily to toggle DC and AC test.
or between and Temperature.
To select PEAK hold, press and hold this button until the display shows " DH
MAX" and indicates the PEAK hold mode.
2. Hold : Hold Q: DATAHOLD or Refresh Data Hold
0The data HOLD function allows operator to freeze the
displayed digital value while the analog bargraph
continues to display the present reading.
Press this button momentarily to toggle DH on or off.
The display shows "DH" to indicate the hold function.
If you select " Refresh Data Hold " by Power-ON
Options, the reading is updated to the display
automatically when the reading changes. The beeper
sounds a tone to remind user, that an update has
occurred. Press this button momentarily to toggle DH on or off.
. MAX * MIN 0: Dynamic Recording
3. MAX MIN : Dynamic recording
To enter or exit dynamic recording mode, press and hold this button to toggle
recording mmode,press and ghold this buton to toggle recording mode on or off.
Records maximum, minimum, and calculates tru avarage.
Press this button momentarily to cycle through maximum, minimum, average
and present (MAX AVGMIN) readings.
17
4. Zero :
Push this button momentarily to zero the residual current. Note: Allow the
meter to stabilize before zeroing the display.
The " " swill also be displayed.
5. ))) / T m: continuity, Delta Temperature
In the position, press this button momenterily to toggle ")))"
continuity ON/OFF. The continuity buzzer sound when test value is below 100
counts (10,0 on auto range). pushing this button for more than 1 secondwill
exit the continuity function and return to the auto-ranging ohm measurement.
Press to restart 1ms PEAKhold test afte entering PEAK mode.
in the temperature test, push this button momentarily to toggle between T1 T2
temperature and T1 - T2 diferential temperatures test. Note: T1, T2 or T1 - T2
selected on the DKTA-620acessory adapter (optional)..
6. AUTO / RANGE :
In auto-range press this button momentarily to select manual range and turn
off the "AUTO" annunciator.
In manual range, press this button momentarily to step up 1 range at one time,
press. this buttonfor more than 1 second to enter auto-range.
in auto-range, the "AUTO" annunciator is lit an the meter will select an
appropiate renge for measurement being made. If a reading is greater than max-
imum available renge, "OL" (overload) is displayed on the screen. The meter
selects a lower range when reading is less thn about 9% of fill scale.
POWER-ON OPTIONS
cSELECTING POWER-ON OPTIONS
Some options can only be selected when you turn the meter on. These power-on
options are listed in Table 2.
To select power-on press and hold the appriopriate pushbutton while turning the
rotary switch to any ON position. Power-on optons remain selected until the
meter is turned off.
18
PUSH
BUTTON OPTIONDESCRIPTION
Turns off all beeper functions.
)))
T
DH
MAXMIN
Demonstrate Annunciators
To demonstrate the annunciators. to demonstrate all annunciators.
momentarily press any button to exit the demonstration mod.
Disable auto-power off
In general, the auto-power off function turns the meter off if neither
rotary switch nor push button is activated for 15 minutes. You can dis-
able auto-power off function by this option. When auto-power off is dis-
abled the meter will stay in on continuously. Auto-power off is auto dis-
able in Dynamic Recording.
Enable "Refresh data Hold".
Table 2. Power-ON Options
SPECIALFUNCTIONS INSTRUCTIONS
This clamp-on multimeter provides the operator with various functions including:
cDynamic Recording
cData Hold
cZero (Relative)
cAnalog bargraph
cAuto Power Off and Sleep Mode
cDisable Auto Power Off
cDemonstrate Annunciator of Display
cContinuity Function For Ohms Measurement
c1ms Peak hold
cDYNAMIC RECORDING
The dynamic recording mode can be used to catch intermittent and turn on or off
surges, verify performance while you are away, or take readings while you are
operating the equipment under test and can not watch the meter.
the average reading is useful for smoothing out unstable or changing inputs, esti-
mating the the percent of time a circuit is operational, or verifying circuit perform-
ance
The operational procedures are described below::
1. Press and hold the "MAXMIN" push button to toggle recording mode on or
off. the dynamic recording ode is indicated when the MAX AVG MINannunci
ator turns on. The present valve is stored to memories of maximum, minimum
and average.
2. Press this button momentarily to cycle through maximum, average and present
readings. The MAX, MIN, AVGannunciator turns on respectively to indicate
what value is being displayed. See Figure 7.
19
3. The beeper sounds when a new maximun or minimum value is recorded.
4. If an overload is recorded the averaging function is stopped. An average value
becomes "OL" (overloaded.
5. In dynamic recording the auto powe off feature is disabled and the"@OFF"
turns off.
6. By selecting dynamic recording in the auto range, the meter will record the value
of MAX, MIN or AVG for different ranges.
7. The record speed of dynamic reording is about 100 milli-seconds (0.1 second).
8. The average value is the true average of all measured values taken since the
recording mode was entered.
20
Figure 7. Display of Dynamic Recording
21
1. Press for more than 1 second to
enter Dynamic Recording
cDATAHOLD
The data HOLD function allos operator to freeze the displayed digital value while
the aanlog bargraph displays present readings. Press "DH" Hold button to enter the
data Hold mode, and the "DH" annunciator is displayed. Press the button again to
exit. The present reading is now shown.
Figura 8. Operación de Retención de Datos (Data Hold)
cZERO (RELATIVE)
The ZERO (relative) function subtracts a stored value from the present measure-
ment and displays the result..
1) Press the ZERO button momentarily to set the relative mode. This sets the dis
play to zero and stores the present reading as a reference value. The " "
annunciator will also be displayed. Press this button again to exit the relative
mode.
2) The ZERO (relative) mode can be set in both the autorange or manual range
mode. The relative mode can't be set when an overload has occurred.
3. When the Dc Current measurement mode is entered, the display will indicate a
non-zero DC Curent (positive or neative) value due to residual magnetism of the
jaw and of the internal sensor. You can use the Zero (relative) function to Zero-
Adjust the display.
22
cANALOGBATRGRAPH
The analog bargraph display provides a 12-segment analog reading representa
tion. The unit of the bargraph is 100 counts/bar.
Figure 9. relative(Zero) Operation.
23
Figura 10. Analog Bar graph
cAUTO POWER OFFAND SLEEPMODE
There are two modes for power saving.
1 ) The instrument will enter the "sleep" mode within 15 minutes, unless:
1-1. Any push buttons have been pressed
1-2. The rotary switch has been changed to another function
1-3. The unit has been set to Dynamic recording mode
1-4. The unit has been set to I ms PEAK hold mode.
1-5. The auto power off has been disabled with powerup option
2) In the sleep mode, the LCD will display a blinking
"@OFF . . . "annunciator.
2-1. To wake-up sleep mode, press any push button for 0.5 sec or rotate rotary
switch.
2-2. Without wake-up, after 15 minutes, the meter will automatically shut off com
pletely.
3) You must turn the rotary switch to the OFF position, then turn it back to a func
tion to activate the meter after an auto power off.
24
25
cDISABLE AUTO POWEROFF
When the meter is to be used for long periods of time you may want to disable the
auto power off. Once the auto power off function is disabled, the meter will stay on
continuously. The meter is shut off by turning the rotary switch to the OFF position.
To activate this function, press and hold the "HOLD/MAX MIN" button and turn the
rotary switch from the OFF position to the desired function. When all annunciators
are displayed, press any button momentarily to exit demonstrate mode, and the
"@OFF" annunciator will be off.
cDEMONSTRATE ANNUNCIATOR
To demonstrate the annunciators, press "HOLD/MAX MIN" button and turn on the
meter simultaneously. All annunciators will be displayed. Press any button to exit
demonstrate mode. Auto power off will be disabled.
Figure 11. Sleep Mode
@OFF ❚ ❚
cCONTINUITY FUNCTION FOR OHMS MEASUREMENT
To enable the continuity function, set the meter to the range.
Press )))button momentarily to toggle the CONTINUITYfunction ON/OFF.
The continuity range is 0-400.0 and the beeper will sound if the resistance is less
than 10.0 . If another range is selected, the unit will beep if the value displayed is
less than 100 counts. Momentarily pushing this button again will toggle the beeper
and annunciator on or off.
Figure 12. Demostrate Annunciator.
26
c1ms Peak Hold
You can use this Meter to analyze components such as power distribution trans-
formers and power factor correction capacitors. The additional features allow the
measurement of the half-cycle peak current by using the I ms peak hold feature.
This allows the determination of the crest factor:
Crest factor = Peak value/True rms value
1 ) Press PEAK button for more than 1 second to toggle 1 ms peak hold mode
ON/OFF.
2) Press PEAK button momentarily to select PEAK+ measurement after entering
the peak mode. The display shows "DH MAX" to indicate the PEAK +. See
Figure 16.
27
Figure 14. Continuity Operation
IIIIIIIIIl4IIIIIIIIIl4
3) If the reading is " OL", then you can push RANGE button momentarily to change
measuring range and restart the PEAK+ measurement after setting the peak
mode.
4) Press )))button to re-set the I ms peak hold again after setting peak mode.
Note: Acrest factor of 1.4 indicates a sinusoidal waveform.
Figure 16. 1ms Peak Hold Display.
Press and Hold
1 second
Press Peak
and 1 second
Hold
V
V
V
DH MAX
I
IIIl
I
28
HOW TO OPERATE
cAC CURRENT MEASUREMENT
WARNING: MAKE CERTAIN THATALLTESTLEADS ARE DISCONNECTED
FROM THE METER TERMINALS.
1 ) Set the rotary switch to " A".
2) Open the meter jaws and clamp around a single conductor. The most accurate
reading will be obtained by keeping the conductor aligned with the centering
marks on the jaws. Make sure that the jaws are fully closed.
3) Read the display.
29
Figure 17. Measuring AC Current.
CORRECT
INCORRECT
30
cDISTRIBUTION TRANSFORMER MEASUREMENT
You can measure current, phase imbalance between phases, and true RMS neutral
current. True RMS measurement yields the effective value.
1) Set the rotary switch to " A".
2) Clamp around a phase wire of the transformer. Be sure the jaws are completely
closed or measurement will not be accurate.
3) Observe the display for true RMS current.
4) Repeat your measurement for each phase to determine balance. Imbalanced
phases and/or harmonics can cause neutral currents.
5) Observe the display for true RMS current reading. If the phases are balance, any
significant current flow on the neutral may indicate the presence of harmonic cur
rents.
6) Press the HOLD/MAXMINbutton to freeze the digital display.
7) Press and hold the SHIFTbutton (>I sec) to enter the PEAK mode (DH MAX dis
played). Measure the half cycle PEAK current. Divide first reading into the sec
ond reading to determine crest factor. Acrest factor other than 1.4 is an indication
of harmonic current.
8) Press and hold the SHIFT button (>11 sec) to exit the PEAK mode.
9) Press and hold the HOLD/MAXMIN (> 1 sec) to enter dynamic recording
mode. Momentarily press HOLD/MAXMIN button to review recorded maxi
mum, minimum, and average values.
10) Press and hold the HOLD/MAXMIN button (> 1 sec) to exit recording.
31
Figure 18. Measuring AC Curreft
32
cAC MOTOR CURRENT MEASUREMENT
You can measure starting (inrush) current , running current, and current imbal-
ance in AC Motor circuits. Inrush current is typically 6-8 times the value of run-
ning current, depending on the motor type..
1. St the rotary switch to "A".
2. Press and hold the PEAK button(>1sec) to enter 1ms PEAK hold mode.
3. Clamp around a motor phase conductor. Be Sure the clamp jaws are com
pletely closed, or measurement will nor be accurate..
4. press ZERO to set the display to zero.
5. Turn the mmotor on.When the motor gets to the desired speed, obbserve
the display for inrush current reading.
6. If the reading is "OL" , you can push the RANGE button momentarily to
change measuring range. Turn off the motor.
7. repeat your measurament from step 2 through 6 for each phase. Avoltage
imbalance or a shorted motor winding may cause imbalance current..
33
Figure 19. Measuring AC Motor Curent
34
cAC VOLTAGE MEASUREMENT
1. Set the rotary switch to "V".
2. Insert the black test lead to "COM" terminal and red test lead to
"V- - µF" terminal.
3. touch the probes to the test points and read the displayed AC Voltage.
35
Figure 20. Measuring Voltage.
36
cRESISTANCE / CONTINUITYMEASUREMENT
CAUTION: Make sure that power is removed and all
capacitors have been discharged before measuring.
1) Set the rotary switch to "TEMP. ". OLis displayed.
2) Insert the black test lead to "COM" terminal and red test lead to
" V - - µF" terminal.
3) Short the test leads together and momentarily press the ZERO button to subtract
test lead resistance from measurement.
4) Touch the test leads to the circuit (Fuse Cartridge or other) and read resistance
value in the display.
5) Press )))button momentarily to enter continuity function if desired.
6) Repeat steps 3 and 4. The beeper sounds if continuity reading is less
than 10.0 Ω..
7) OL (overload) is displayed if the resistance across the input terminals is greater
than the full-scale rating on the range setting of the instrument. Be sure that the
contact between the probes and the circuit is clean. Dirt, oil, paint, rust or other for-
eign matter can seriously effect resistance measurements.
37
Figura 21. Measuring Resistance and Continuity.
FUSE
CARTRIDGE
38
cTEMPERATURE (K-TYPE) MEASUREMENTS
The ACDC-620TLCD contains a dual display for
Temperature Measurement. The larger digits display oF (Fahrenheit), and the small
digits display oC (Celsius).
Both readings are displayed simultaneously.
CAUTION: Do not allow the temperature sensor to contact a surface which is ener-
gized above 30 V RMS or 60 V DC, such voltages pose a shock hazard.
To perform single input temperature measurement:
1) Set the rotary switch to "TEMP. ",
2) Insert the model TAC-DMM banana to K-type thermocouple adapter into the
meter. The "+" end of adapter must be inserted into"V - - µF"terminal.
3) Insert the model TPK-56 Type K thermocouple into the TAC-DMM.
4) Secure the beaded end of the K thermocouple to the surface you wish to
measure.
5) Press SHIFT button momentarily to enter temperature measurement mode.
6) Read the displayed temperature.
39
Figure 22. Surface Temperature Measurement.
40
Press
Shift
To perform dual input and/or differential temperature measurement:
1) Set the rotary switch to "TEMP. ".
2) Insert the model DKTA-620 Dual Type K thermocouple adapter into the meter.
The "+" end of the adapter must be inserted into "V - - µF" terminal.
3) 1 nsert two of the model TPK-56 Type K thermocouples into the DIKTA-620.
4) Secure the beaded end of the K thermocouples to the surfaces you wish to
measure.
5) Press SHIFTbutton momentarily to enter temperature measurement mode.
The "T1 T2' annunciator will be displayed.
6) Since the displayed temperature will be based on the position of the selector
switch located on the DKTA-620 adapter, select either T1 or T2 for the desired
measurement.
7) Read the display.
To measure differential temperature:
Note: T1 temperature must be greater than T2 for
differential measurement to be correct.
1 ) Set the DKTA-620 selector switch to the "T1-T2" position.
2) Push the Tbutton on the ACDC-620Tmomentarily to enter the Differential
Temperature Mode. The "TI-T2" and "A" annunciators will be displayed.
3) Read the displayed differential temperature.
4) For information on basic refrigeration theory refer to page 56.
41
Figure 23. Measuring Temperature on Refrigeration System.
Contact where
you want to measuring
PRESS
T
42
cCAPACITANCE MEASUREMENTS
In many instances a motor can not be started due to a failed start/run capacitor. To
test the capacitor:
: Discharge the capacitor before testing.
1 ) Set the rotary switch to "µF" position.
2) Insert the black test lead to "COM" terminal and red test lead to "V - - µF"
terminal.
3) Remove at least one leg of the capacitor from circuit board or device.
4) Touch the test leads across the capacitor and read the display.
Note: If you are measuring a polarity sensitive electrolytic capacitor, touch the red
test lead to the positive end of the capacitor, and the black test lead to the negative
end of the capacitor.
43
Figure 24. CAPACITANCEMEASUREMENTS.
44
Negro
(-)
Rojo
(+)
GENERALSPECIFICATIONS
Display:
Display: Fully annunciated 4-digit liquid crystal display (LCD) with maximum reading
of 4,000 count. Dual display in Temperature mode. 12 segments analog bar graph.
Automatic polarity indication.
Functions: DCV, ACV, DCA, ACA, OHM, Capacitance and Temperature.
Measuring rate:
3.3 times per second for V, A, Ohm and Temperature tests.
0.5 - 7.6 seconds per time for Capacitance test.
Low battery indicator:
The "'q" appears when the battery voltage drops below 7V
(approx.).
Operating temperature:
0 * C to 50 * C (32 * F to 122 * F), 0 - 80% R.H.
Storage temperature:
-20'C to 60 * C (-4'F to 140 - F), 0 - 80% R. H. with
BATTERYREMOVED.
Temperature coefficient:
0.12 % / *C(fr0m O'C to 18 *C or 28 *C to 50 'C),
0.067 % / 'F(from 32 * F to 64.4 * F or 82.4 * F to 122 * F), or
otherwise specified.
Power supply: Single standard NEDA1604, JIS006P,IEC6F22 carbonzinc or alka-
line type 9V battery.
45
MAX. Jaw Opening:
To Accommodate Circuit Cables 2" ( 50.8 mm ) diameter.
Dimension:
32 (H) x 64 (W) x 260 (L) mm
1.26"(H) x 2.52"(W) x 10.24"(L)
Weight: 840 grams with battery included. (1.85 lbs with battery included.)
Standard Accessories: Test leads (pair), Manual, Battery and Carrying case.
Optional Accessories: K-type bead probe and Dual input adopter.
Safety: Designed and manufactured to conform to UL3111-1, C22.2 NO. 1010. 1 -
92 and EN61 010 (IEC1 010-1, IEC1010-2-031, fEC-1010-2-032) Installation
Category (Overvoltage Category) 111 OOOV or Installation Category III 60OV,
Pollution Degree 2 environment.
Note: Meter has been submitted for approval to above standard at the time of print-
ing of this manual. Product will be marked accordingly upon approval.
46
47
Accessories and ReplacementsParts
Safety Test Leads(included)
Carryimg Case(included)
Instruction Manual(included)
9 Volt Alkaline Battery(included)
Dual input thermocouple adapter(opcional)
Single input thermocouple(opcional)
Type K bead probe(opcional)
DTL-3000
CC-ACDC
978760
MN-1604
DKTA-620
TAC-DMM
TPK-56
Amprobe P/N Description
ELECTRICALSPECIFICATIONS
Accuracy is giveN as ± % of reading + the number of least significant digits at
23oC ± 5oC, with relative hunidity Less than 80% R.H.
Range Resolution Accuracy Overload
Protection
400V 0.1V
1000v 1V
±(1% rdg + 3 dgt) 1000V
AC RMS
Range Resolution Accuracy Overload
45Hz ~ 400 Hz Protection
400V 0.1V
1000v 1V
±(1,5% rdg + 3 rdg) 1000V
AC RMS
c DC VOLTAJE
c AC Voltage
(TRUERMS: Froml 10%al 100%of range)
Input Impedance: 10M// less than 100pF.
crest Factor: <3:1
Input Impedance: 10M.
Range Resolution Accuracy Overload
Protection
400V 0.1V
1000v 1V
±(1% rdg + 3 rdg) 1000 Voltios CA
Eficaces (RMS)
c VOLTAGE ( 1ms PEAKHOLD)
SPECIFIED ACCURACY+/- 40 Digits for Changes >1ms in duration .
Input Impedance: 10M.
48
49
Range Resolution Accuracy
400A 0.1A
1000A 1A
±(1,5%rdg+3rdg)
±(2% rdg+ 5rdg)
cDC CURRENT
Range Resolution Accuracy
400A 0.1A
1000A 1A
45Hz~65Hz 65Hz~2kHz
±(2% rdg+5dgt) ±(3%rdg+5 dgt)
±(3%rdg+5 dgt)
±(2.5% rdg + 5dgt) ±(3% de la
cAC CURRENT
(TRUE RMS: From 10% al 100% of range)
Crest Factor: <3:1
Range Resolution Accuracy
400A 0.1A
1000A 1A
±(2%rdg+ 43dgt)
±(2%rdg+43dgt)
cCORRIENTE (RETENCIÓN DE PICO DE 1ms)
Precisión especificada ± 40 dígitos para cambios con duración mayor de 1ms.
Range Resolution Accuracy Maximum Overload
Tets Voltage Protection
4000.1
10001±(1%rdg+ 3dgt)
3.3V
1.25V 600V(RMS)
cVOLTAJE de CC
In contituity mode, built-in buzzer sounds when resistence is les than 10.0 .
RANG Resolution Accuracy *V Maximum
-40°C ~1372°C 1°C
-40°F ~ 2502°F 1°F
±(0,5%rdg+3°C)
±(0,5%rdg+6°F)
30 V RMS
or 60 V DC
cK-TYPETEMPERATUREMEASUREMENTRANGE
Gama Resolución Precisión *V Máximo
-50°C< T<100°C 1°C
-58°F< T<180°F 1°F ±(0,5%rdg+3°F)
±(0,5%rdg+6°F) 30 V RMS
or 60 V DC
cTEMPERATURADIFERENCIALTIPO K
RANGE Resolution Accuracy Overload Protection
400µF 0.1µF
400µF1µF
±(3%rdg+4dgt)
±(3.5%rdg+4dgt),
>2mF, NO Spec
600 V
RMS
cCAPACITANCE
Note: Do not alow the temperature sensor to contact a surfece which is energized above
30 V RMS or 60 V DC, such voltages pose a shock hazard.
The T1 input must be greater than -T2 input for T1-T2 measuring.
the accuracy does not include the tolerance of thermocouple probe.
MODELTPK-56, BEADTYPE-K THERMOCOUPLE (optional)
Temperature Rating (wire): continuous 204°C, 260°C (intermittent)
Temperature measuring Range (bead): -50oC to +800°C (MAX) Accuracy.
±2.2°C or ±0.75% of reading from 0°C to 800°C (whichever is greater)
±2.2°C or ±2% of reading from 0°C to -50oC (whichever is greater)
Caution: The bead end can be touched to surfacve temperatures above 2040C
for very short durations. However, wire insulation or wire may be damage if wire
subjected to the same temperature.
With film capacitor or better, use ZERO to cancel residual.
Temperature coefficient: 0.15 x specified accuracy / °C(from 18 o C or 28 o C to 50 o C) or 0.084 x
specified accuracy/ oF(from 32 o F to 64.4 oF or 82.4 oF to 122 oF)
50
CURRENT HARMONICS THEORY
True-RMS current is very important because it directly relates to the amount of
heat dissipated in wiring, transformers, and loads. Most clamp-on meters
already in the field measure average current, not true RMS current, even if
this average value is displayed on a scale calibrated in RMS. These average-
sensing meters are accurate only for sinusoidal signals.
All current signals are distorted in some way. The most common is harmonic
distortion caused by non-linear loads such as office machines, medical equip-
ment, personal computers, or speed controls for motors. Harmonic distortion
causes significant currents at frequencies that are odd multiples of the power
line frequency. Harmonic current can cause a substantial load on the neutral
wires of wye-connected power distribution systems.
In most countries, 50Hz or 60Hz power distribution systems include 3-phase
delta primary - wye secondary transformers. The secondary generally pro-
vides 120V AC from phase to neutral, and 208V AC from phase to phase.
Historically, balancing the loads on each phase was a big headache for the
electrical system designer.
Typically, the vector addition of the phase currents in the transformers' neutral
wire is zero or quite low in a wellbalanced system. Typical devices that pres-
ent linear loading include incandescent lighting and small motors. The result is
essentially a sine wave current in each phase and a low neutral current at a
frequency of 5OHz: or 60Hz.
Devices such as TV sets, fluorescent lighting, video machines, and microwave
ovens are now commonly drawing power line current for only a fraction of
each cycle so that they cause non-linear loading and subsequent
51
non-linear current. This generates odd harmonics of the 5OHz or 60Hz line
frequency. Therefore, the current waveform from the transformer could contain
not only a 60Hz component, but also a 180Hz component, a 30OHz compo-
nent, etc.
The vector addition in a properly balanced power distribution system feeding
non-linear loads may still be quite low. However, the vector addition does not
cancel all the harmonic currents. The odd multiples of the 3rd harmonic (called
the "TRIPLENS") are added together in the neutral. These harmonics can cre-
ate an RMS current in the transformers neutral wire that is 130( of the total
RMS current measured in any individual phase. For example, phase currents
of 80 amperes may cause 104 amperes of harmonic current in the neutral, the
most common harmonic being the 3rd. The electrical designer must consider
the following issues when designing a power distribution system that will con-
tain harmonic current.
1. The AC neutral wires must be of sufficient gauge to
allow for harmonic current.
2. The distribution transformer must have additional cooling to continue opera
tion at its rated capacity. This is because the harmonic current in the sec
ondary neutral wire is circulating in the deltaconnected primary winding.
This circulating harmonic current heats up the transformer.
3. Phase current harmonics are reflected to the primary winding and they
continue back towards the power source, This can cause distortion of the
voltage wave so that any power factor correction capacitors on the line can
be easily overloaded.
We can use this Meter to analyze components such as power distribution
transformers and power factor correction capacitors. An additional feature
allows the measurement of half-cycle peak current by using the 1 ms peak
hold feature. This allows the ability to determine crest factor:
Crest factor + Peak value/True rms value
NOTE: If Crest factor exceeds 1.1 harmonic distortion is present
52
53
cTRUERMSMEASUREMENT
The meter measures the TRUERMS value of AC voltages and currents. In physical
terms, the RMS (Root-Mean-Square) value of a waveform is the equivalent DC val-
ue that causes the same amount of heat to be dissipated in a resistor. TRUE RMS
measurement greatly simplifies the analysis of complex AC signals. Since the RMS
value is the DC equivalent of the original waveform, it provides a reliable basis for
comparing dissimilar waveforms.
By contrast, many meters use average-responding AC
converters rather than TRUE RMS converters. The scale
factor in these meters are adjusted so that they display the
RMS value for a harmonic-free sine wave. If a signal is not
sinusoidal, average-responding meters do not display
correct RMS readings.
For a free video on Amprobe's Harmonalyzer, HA-2000
harmonic/waveform analyzer, please contact Amprobe
directly.
cWAVEFORMCOMPARISON
Ta b l e3. Illustrates the relationship between AC and DC components for common
waveforms, and compares readings for TRUE RMS meters and average-responding
meters. For example, consider the first waveform, a 1414V (zero-to-peak) sine wave.
Both this Clamp-on meter and RMS-calibrated average-responding meters display
the correct RMS reading of100.0V (the DC component equals 0). However, consider
the 200V (peak-to-peak) square wave, both types of meter correctly measure the DC
component (oV). The clamp meter correctly measures the AC component (100.0V).
The average-responding meter measures 111.1V, which amounts to an 11% error.
The conversion factors in Table 3 show the differences between average sensing
instruments and true RMS instrument measurements.
54
Table 3. WAVEFORMCOMPARISONCHART
55
* RMS CALIS THE DISPLAYED VALUEFOR AVERAGE RESPONDING METERS
THATARE CALIBRATED RMS FOR SINE WAVES.
56
REFRIGERATIN THEORY
Thermocouples are widely used in the process control and refrigeration industry. A
review of basic refrigeration theory will help you to apply the meter in refrigeration
applications.
In any refrigeration application the goal is to transfer heat from one place to another.
The transfer is made by pumping a refrigerant from one area to another area whose
temperature is different.
The refrigerant has a low boiling point and a high latent heat of vaporization. That is,
a great deal of heat is required to convert the refrigerant from a liquid to a gas at its
boiling temperature. The resulting gas is at the same temperature, but it contains
more heat energy than the liquid did.
At atmospheric pressure, common refrigerant gasses such as Freon and Ammonia
boil far below water freezing point, and absorb heat from their surroundings in the
process. Likewise, as gasses condense, they release heat back into their environ-
ment.
We can control the boiling point by raising or lowering the pressure, forcing the gas
to dissipate heat or absorb it. If we raise the pressure enough by compressing, it will
become liquid. In a refrigeration system there is a compressor and a system of
valves to change the pressure of a contained gas. The gas absorbs or dissipates
heat in heat exchangers called evaporators or condensers, depending on how they
are used. Figure 25 shows the basic parts of a refrigerator.
57
cThe Refrigation Cycle
Inside a refrigerator, a compressed gas in liquid form is released through an expan-
sion valve into the low pressure of an evaporator. Here the liquefied gas evaporates
(boils). The expansion valve controls how much refrigerant boils, and thus, how
much heat ft absorbs. The refrigerant, now in gas form, circulates through the cool-
ing coils carrying the heat away with
FigurE 25. Refrigeration System.
EVAPORATOR
CONDENSER
COMPRESSORS
THERMOSTATICEXPANSION VALVE
Once outside, the gas enters the compressor on the low-pressure side and is
compressed. As it is compressed, the gas temperature rises above the surrounding
air temperature. The hot gas dissipates its heat in the condenser, becoming liquid
once again, before returning to the expansion value.
In large systems the temperature and pressure are closely monitored to ensure long
life and minimum energy usage. In small systems, like a home refrigerator, tempera-
ture is often measured as a troubleshooting aid. Careful temperature measurement
at different points in the system can pinpoint trouble spots. Some of critical tempera-
tures are discussed below.
cImportant Temperature Measurements
Evaporator Temperature
The entire mid-section of the evaporator is at the refrigerant's boiling temperature,
which is related directly to the evaporator pressure. The evaporator temperature
should be lower than the desired cooling temperature.
Evaporator Superheat
By the time the refrigerant reaches the last few turns of the evaporator coils, all the
refrigerant should have become gas with s temperature slightly warmer than the
evaporating temperature. The amount that the gas temperature exceeds its boiling
point is called superheat. Superheat is a sensitive indicator of evaporator efficiency.
The superheat gas temperature is usually taken at the expansion valve's sensing
bulb. Aproper superheat reading, ensures that all liquid has boiled.
58
59
Condenser Temperature
The condenser's entire mid-section should be at the refrigerant's condensing tem-
perature. The condensing temperature is related directly to the condenser's pres-
sure, which is produced by the compressor's head-pressure.
The condenser temperature varies with the system's load, but it is designed to
operate within certain limits. Load factors include the amount of heat absorbed by
the evaporator, air temperature surrounding the condenser (ambient air tempera-
ture), and airflow over the condenser.
Temperature Difference (TD)
The difference between ambient air and condensing temperature is a critical factor.
The greater the temperature difference between hot refrigerant and air, the faster
the heat will dissipate. If TD is too low, the refrigerant will not completely condense.
Condenser Superheat
The temperature of the hot gas discharged from the compressor is higher than the
condensing temperature. In the evaporator, the difference between the gas temper-
ature and condensing temperature is called superheat. However, since the pressure
here is far higher than at the evaporator, so is the temperature. Condenser super-
heat is usually measured as the difference between the first few condenser coils
and the midsection. Abnormal superheat ran indicate troubles such as condenser
overload, lack of refrigerant, and presence of non-condensable gas.
Condenser Sub-cool
Like the evaporator, the condenser is sized so that all the gas will be condensed
before the last few coils, the liquid temperature drops slightly below the condensing
temperature. In the case of an overloaded condenser or one short of refrigerant,
there will be little or no subcooling.
Indication
Efficient
Condenser
Low
Refrigerant
Ambient air
temperature (T2) C o n d e n s i n g
Te m p e r a t u r e
( T 1 )
95°F
35°C
95°F
35°C
95°F
35°C
90°F
32°C
95°F
35°C
102°F
39°C
105°F
41°C
7°F
4°C
170°F
77°C
190°F
88° 88°F
31°C
115°F
46°C 10°F
-12°C
2°F
-17°C
10°F
-12°C
5°F
-15°C
10°F
-12°C
100°F
38°C
95°F
35°C
15°F
-9°C
75°F
24°C
95°F
35°C 120°F
49°C
45°F
7°C
145°F
63°C
10°F
6°C 120°F
49°C
60°F
34°C 225°F
107°C
35°F
19°C 225°F
107°C
150°F
66°C
130°F
54°C
Inefficient
Compressor
Overloaded
Condenser
Non-
Condensable
Gas
This is an example of temperature indications for normal operation and various
malfunctions for typical sir conditioning unit.
125°F
52°C 30°F
17°C
TD
TSuper
Heat
190°F
88°C
120°F
49°C
225°F
107°C
225°F
107°C
170°F
77°C
Vapor
In Liquid
Out Sub-
cool
60
61
MAINTENANCE
WARNING
To avoid electrical shock, do not perform any servicing unless you are qualified to
do so.
cSERVICE
If the instrument fails to operate, check battery, test leads, etc. and replace as nec-
essary. If the instrument still does not operate, double check operating procedure as
described in this instruction manual. When servicing, use only specified replace-
ment parts.
WARNING
To avoid electrical shock or damage to the meter, do not get water inside the case.
Remove the test leads and any input signals before opening the case.
0 BATTERYREPLACEMENT
The meter is powered by a single 9V battery, with
NEDA1604, S006P,IEC6F22 carbon-zinc alkaline, or
similar battery. Replace battery if the low battery sign
(~)is displayed and flashing. Use the following procedure
to replace the baftery:
1 . Remove the meter from the circuit and turn the rotary
switch to the OFF position.
2. Disconnect the test leads from the instrument.
3 Loosen the screw on the battery cover.
4. Pull the cover up slightly and slide the battery cover off (see Figure 26 and
Figure 27).
5. Replace the defective battery.
6. Reverse the procedure of opening cover to close the
battery cover.
62
Figure 26. step 1 of Battery replacement.
Pull up slightly
63
Figure 27. Step 2 of Battery Replacement.
Pull and move to right.
64
c CLEANING
To clean the instrument, use s soft cloth dampened in a solution of mild detergent
and water. Do not spery cleaner directly onto the instrument, since it may leak
intothe cabinet and cause damage.
Do not use chemicals containing benzine, benzene, touene, xylene, acetone or sim-
ilar solvents.