AC-DC
Specification
175 Watts
•Medical Safety Approval
•High Efficiency Resonant Topology
•12 V Fan Output
•Remote Sense
•5V Standby Option
•Remote On/Off & Power Good Signal Options
•3 Year Warranty
CLC175 Medical Series
Input
Input Voltage • 85-264 VAC (120-370 VDC), derate output
power <90 VAC by 10%
Input Frequency • 47-63 Hz
Input Current • 3.0 A typical at 115 VAC, full load
1.5 A typical at 230 VAC, full load
Inrush Current • 40 A max at 230 VAC, cold start 25 ºC
Power Factor • EN61000-3-2, class A
Earth Leakage Current • 230 µA maximum at 264 VAC at 60 Hz
90/185 µA 115 V 60 Hz/230 V 50 Hz typical
Input Protection • Internal T5 A 250 V fuse in both line and
neutral
Output
Output Voltage • 12-48 V (see tables)
Output Voltage Trim • ±5%
Initial Set Accuracy • ±1% V1, ±5% V2, ±3% V3
Minimum Load • 0.1A required for V2 regulation when used
Start Up Delay • 1.5 s typical, see longform datasheet for
more details
Start Up Rise Time • 5 ms typical
Hold Up Time • 16 ms min at 115 VAC
Drift • ±0.2% after 20 min warm up
Line Regulation • ±0.5% max
Load Regulation • ±1% V1, ±5% V2 & V3 max
Over/Undershoot • 5% typical
Transient Response • 4% max. deviation, recovery to within
1% in 500 µs for a 50-75-50% load
change
Ripple & Noise • 1% pk-pk V1, others 2%,
20 MHz bandwidth
Overvoltage Protection • 115-140% Vnom, recycle input to reset
Overload Protection • 110-150% V1 only
Short Circuit Protection • Continuous trip and restart (Hiccup mode)
Temperature • 0.05%/˚C
Coefficient
Remote Sense • Compensates for 0.5 V total voltage drop
Remote On/Off • Uncommitted isolated optocoupler diode,
(Inhibit/Enable) powered diode inhibits V1 & V2
(Option -A)
xppower.com
General
Efficiency • 87% typical
Isolation • 4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
Switching Frequency • 70 kHz typical
Signals (Option -A) • Power OK - open collector,
Remote On/Off, 5 V Standby
MTBF • 390 kHrs per MIL-HDBK-217F
at 25 ºC, GB
Power Density • 9.0 W/in3
Environmental
Operating Temperature • 0 ˚C to +70 ˚C derate linearly from +50 ˚C
at 2.5%/˚C to 50% load at +70 ˚C. See
derating curve and longform datasheet for
thermal information.
Cooling • Forced-cooled >10 CFM or use ‘-TF’ option
Operating Humidity • 95% RH, non-condensing
Storage Temperature • -40 ˚C to +85 ˚C
Operating Altitude • 3000 m
Shock • 30 g pk, half sine, 6 axes
Vibration • 2 g rms, 5 Hz to 500 Hz, 3 axes
EMC & Safety
Low Voltage PSU EMC • EN61204-3, high severity level as below
Emissions • EN55022 level B conducted
EN55022 level A radiated
Harmonic Currents • EN61000-3-2, class A
Voltage Flicker • EN61000-3-3
Radiated Immunity • EN61000-4-3, level 3 Perf Criteria A
EFT/Burst • EN61000-4-4, level 3 Perf Criteria A
Surge • EN61000-4-5, installation class 3
Perf Criteria A
Conducted Immunity • EN61000-4-6, level 3 Perf Criteria A
Dips & Interruptions • EN61000-4-11, 30% 10 ms,
60% 100 ms, 100% 5000 ms
Perf Criteria A, B, B
Safety Approvals • IEC60601 CB report, UL60601,
TUV EN60601