(R) MIHW2000 SERIES DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety FEATURES Industrial Standard DIP-24 Package Ultra-Wide 4:1 Input Voltage Range Fully Regulated Output Voltage I/O Isolation 4000VAC with Reinforced Insulation, rated for 1000Vrms Working Voltage Low Leakage Current < 2A Operating Ambient Temp. Range -40 to +85 Under-Voltage, Overload and Short Circuit Protection Conducted EMI EN 55011/22 Class A & FCC Level A Approved Medical EMC Standard with 4th Edition of EMI EN 55011 and EMS EN 60601-1-2 Approved Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking PRODUCT OVERVIEW The MINMAX MIHW2000 series is a range of high performance DC-DC converter modules with a reinforced insulation system .The I/O isolation voltage is specified for 4000VAC with reinforced insulation, which rated for 1000Vrms working voltage.The product comes in a small DIP-24 package. All models features an ultra-wide input voltage range including 36-160VDC for railroad applications. Full SMD design with exclusive use of ceramic capacitors guarantees a high reliability with calculated MTBF of >1 million hours. These high isolation DC-DC converters are the perfect solution solution for many demanding applications in industrial and railroad systems, in medical instrumentation, everywhere where a certified supplementary or reinforced insulation system is required to comply with specific industrial or medical safety standards. Model Selection Guide Model Number MIHW2022 MIHW2023 MIHW2026 MIHW2027 MIHW2032 MIHW2033 MIHW2036 MIHW2037 MIHW2042 MIHW2043 MIHW2046 MIHW2047 Input Voltage (Range) Output Voltage VDC VDC 5 12 12 15 5 12 12 15 5 12 12 15 24 (9 ~ 40) 48 (18 ~ 80) 110 (36 ~ 160) Output Current Input Current Max. Min. @Max. Load @No Load Reflected Ripple Current mA 600 250 125 100 600 250 125 100 600 250 125 100 mA 90 37.5 18.8 15 90 37.5 18.8 15 90 37.5 18.8 15 mA(typ.) 160 151 151 151 80 75 75 75 35 33 33 33 mA(typ.) mA (typ.) 20 15 10 8 5 3 Max. capacitive Load F 1000 470 220# 220# 1000 470 220# 220# 1000 470 220# 220# @Max. Load # For each output E-mail:sales@minmax.com.tw 2018/10/31 REV:12 Tel:886-6-2923150 Page 1 of 5 Efficiency (typ.) % 78 83 83 83 78 83 83 83 78 83 83 83 (R) MIHW2000 SERIES DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety Input Specifications Parameter Model 24V Input Models 48V Input Models 110V Input Models 24V Input Models 48V Input Models 110V Input Models 24V Input Models 48V Input Models 110V Input Models Input Surge Voltage (1 sec. max.) Start-Up Threshold Voltage Under Voltage Shutdown Short Circuit Input Power Input Filter Conducted EMI Min. -0.7 -0.7 -0.7 8 13 26 --------- Typ. Max. Unit --50 --100 --180 8.5 9 15 17 VDC 30 34 --8.5 --16 --32 --2000 mW Internal Pi Type Compliance to EN 55011/22,class A and FCC part 15,class A All Models Output Specifications Parameter Conditions / Model Output Voltage Setting Accuracy Min. Typ. Max. Unit --- --- 1.0 %Vnom. Output Voltage Balance Dual Output, Balanced Loads --- 0.5 2.0 % Line Regulation Vin=Min. to Max. @Full Load --- 0.3 0.5 % Load Regulation Io=25% to 100% Ripple & Noise 0-20 MHz Bandwidth Transient Recovery Time --- 0.5 1.0 % 5V Output Models --- 75 100 mV P-P Other Output Models --- 100 150 mV P-P --- 150 500 sec --- 3 6 % --- 0.02 0.05 %/ 120 150 --- % 25% Load Step Change Transient Response Deviation Temperature Coefficient Over Load Protection Foldback Short Circuit Protection Continuous, Automatic Recovery Isolation, Safety Standards Parameter I/O Isolation Voltage Conditions Min. Typ. Max. Unit 60 Seconds Reinforced insulation, rated for 1000Vrms working voltage 4000 --- --- VACrms Leakage Current 240VAC, 60Hz --- --- 2 A I/O Isolation Resistance 500 VDC 10 --- --- G I/O Isolation Capacitance 100kHz, 1V --- 7 13 pF UL/cUL 60950-1, CSA C22.2 No. 60950-1 Safety Standards ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1 IEC/EN 60950-1, IEC/EN 60601-1 3rd Edition 1xMOPP & 2xMOOP UL/cUL 60950-1 recognition (UL certificate), IEC/EN 60950-1 (CB-report) Safety Approvals ANSI/AAMI ES60601-1 1xMOPP & 2xMOOP recognition (UL certificate), IEC/EN 60601-1 3rd Edition (CB-report) General Specifications Parameter Conditions Switching Frequency MTBF(calculated) MIL-HDBK-217F@25, Ground Benign E-mail:sales@minmax.com.tw 2018/10/31 REV:12 Tel:886-6-2923150 Page 2 of 5 Min. Typ. Max. --- 150 --- Unit kHz 1,000,000 --- --- Hours (R) MIHW2000 SERIES DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety Environmental Specifications Min. Max. Unit Operating Ambient Temperature Range (See Power Derating Curve) Parameter -40 +85 Case Temperature --- +100 Storage Temperature Range -50 +125 Humidity (non condensing) --- 95 % rel. H Lead Temperature (1.5mm from case for 10Sec.) --- 260 Power Derating Curve 100 20LFM Output Power (%) 80 400LFM 60 40 20 0 -40 ~ 0 20 40 60 80 100 110 Ambient Temperature C Notes 1 2 3 4 5 6 Specifications typical at Ta=+25, resistive load, nominal input voltage and rated output current unless otherwise noted. Transient recovery time is measured to within 1% error band for a step change in output load of 75% to 100%. These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these modules; however, they may not meet all specifications listed. We recommend to protect the converter by a slow blow fuse in the input supply line. Other input and output voltage may be available, please contact factory. Specifications are subject to change without notice. E-mail:sales@minmax.com.tw 2018/10/31 REV:12 Tel:886-6-2923150 Page 3 of 5 (R) MIHW2000 SERIES DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety Package Specifications 3.9 10.5 [0.41] [0.15] Pin Connections 0.2 [0.008] Mechanical Dimensions 0.60 [0.024] 15 23 +Vin +Vin 11 No Pin Common 12 -Vout No Pin 13 +Vout -Vout 15 No Pin +Vout 23 -Vin -Vin 24 -Vin -Vin 13 20.32 [0.80] 31.8 [1.25] 5.08 [0.20] 2.0 [0.08] 2.54 [0.10] All dimensions in mm 2.54 [0.10] Physical Characteristics Case Size : 31.8x20.3x10.5mm (1.25x0.8x0.41 inches) Case Material : Non-Conductive Black Plastic (flammability to UL 94V-0 rated) Pin Material : Copper Alloy with Gold Plate Over Nickel Subplate Weight : 16.2g E-mail:sales@minmax.com.tw 2018/10/31 REV:12 Tel:886-6-2923150 Page 4 of 5 Dual Output 12 Bottom View 24 Single Output 1 20.3 [0.80] 11 15.22 [0.60] 1 Pin Tolerance: X.X0.25 (inches) (X.XX0.01) X.XX0.13 ( X.XXX0.005) Pin diameter 0.6 0.05 (0.0240.002) (R) MIHW2000 SERIES DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety Test Setup Input Reflected-Ripple Current Test Setup Input reflected-ripple current is measured with a inductor Lin (4.7H) and Cin (220F, ESR < 1.0 at 100 KHz) to simulate source impedance. Capacitor Cin, offsets possible battery impedance. Current ripple is measured at the input terminals of the module, measurement bandwidth is 0-500 KHz. To Oscilloscope + + Battery +Vin Lin DC / DC Converter Current Probe Cin +Out -Vin Load -Out Peak-to-Peak Output Noise Measurement Test Use a Cout 0.47F ceramic capacitor. Scope measurement should be made by using a BNC socket, measurement bandwidth is 0-20 MHz. Position the load between 50 mm and 75 mm from the DC-DC Converter. +Vin +Out Copper Strip Single Output DC / DC Converter -Vin Cout -Out Scope +Vin +Out Copper Strip Dual Output DC / DC Converter Com. Copper Strip -Vin -Out Copper Strip Resistive Load Copper Strip Scope Cout Cout Resistive Load Scope Technical Notes Overload Protection To provide protection in a fault (output overload) condition, the unit is equipped with internal current limiting circuitry and can endure current limiting for an unlimited duration. At the point of current-limit inception, the unit shifts from voltage control to current control. The unit operates normally once the output current is brought back into its specified range. Input Source Impedance The power module should be connected to a low ac-impedance input source. Highly inductive source impedances can affect the stability of the power module. In applications where power is supplied over long lines and output loading is high, it may be necessary to use a capacitor on the input to insure startup. By using a good quality low Equivalent Series Resistance (ESR < 1.0 at 100 kHz) capacitor of a 4.7F for the 24V input devices, a 2.2F for the 48V devices and a 1F for the 110V devices, capacitor mounted close to the power module helps ensure stability of the unit. + DC Power Source +Vin + +Out DC / DC Converter Load Cin - -Vin -Out Output Ripple Reduction A good quality low ESR capacitor placed as close as practicable across the load will give the best ripple and noise performance. To reduce output ripple, it is recommended to use 3.3F capacitors at the output. + +Vin - +Out Single Output DC / DC Converter DC Power Source -Vin + Cout +Vin Load -Out - +Out Dual Output DC / DC Com. Converter DC Power Source -Vin -Out Cout Load Load Cout Maximum Capacitive Load The MIHW2000 series has limitation of maximum connected capacitance on the output. The power module may operate in current limiting mode during start-up, affecting the ramp-up and the startup time. Connect capacitors at the point of load for best performance. The maximum capacitance can be found in the data sheet. Thermal Considerations Many conditions affect the thermal performance of the power module, such as orientation, airflow over the module and board spacing. To avoid exceeding the maximum temperature rating of the components inside the power module, the case temperature must be kept below 100. The derating curves are determined from measurements obtained in a test setup. Position of air velocity probe and thermocouple 15mm / 0.6in 50mm / 2in Air Flow DUT 18, Sin Sin Road, An-Ping Industrial District, Tainan 702, Taiwan Tel: 886-6-2923150 Fax: 886-6-2923149 E-mail: sales@minmax.com.tw Minmax Technology Co., Ltd. 2018/10/31 REV:12 Page 5 of 5