ALM65 Series AC-DC Power Supplies 65 Watts * Medical & IT Safety Approvals * Energy Efficiency Level VI * 4th Edition Medical EMC * IP32 Environmental Rating * Class I and Class II Versions * <0.21 W Standby Power * 0 C to 60 C Operation * Low Earth Leakage Current * 3 Year Warranty Dimensions: The ALM65 series of medical external power supplies is fully approved to international medical safety standards. It has been designed with very high efficiency and low standby power, enabling it to meet the latest environmental legislation. The unit has a fully sealed enclosure complying with IP32 and a smooth surface finish making it easier to wipe down in a clinical setting. ALM65: 4.94 x 2.19 x 1.32" (125.5 x 55.5 x 33.5 mm) Models & Ratings Output Power Output Voltage Output Current 12.0 V 15.0 V 19.0 V 24.0 V 48.0 V 65 W 5.40 4.30 3.40 2.70 1.35 A A A A A Total Regulation Efficiency(1) 5% 88% 88% 88% 88% 88% Model Number(2,3,4) ALM65US12 ALM65US15 ALM65US19 ALM65US24 ALM65US48 Notes 1. Typical average of efficiencies measured at 25%, 50%, 75% and 100% load and 115VAC input. 2. For class II versions, add suffix 'C2-8' to the end of the part number e.g. ALM65US24C2-8. 3. For optional input connector retention clip add suffix `-A' to the model number, e.g. ALM65US24-A (not available for C2 versions) 4. 5.5 x 2.1 mm output connector add suffix B1 to the part number, e.g. ALM65US24-B1, ALM65US24C2-8B1, ALM65US24-AB1. Mechanical Details Class I IEC 320-C14 4.941 (125.5) 1.319 (33.5) Class II IEC 320-C8 1.512 (38.4) 2.185 (55.5) 0.177 (4.50) 35.433 0.8 (90020) o 0.685 (17.4) 1.969 0.394 (5010) 0.374 0.012 (9.500.30) 1.319 (33.5) 0.039 (1.0) RETURN ( - ) OUTPUT ( + ) 4.488 (114.0) 1.732 (44.0) Notes 1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except 5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8 for C8 versions. output cable length. 2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8 for C8 versions. 3. Weight: 0.68 lbs (308 g) approx. 4. For European mains lead order part EU-MAINS-IEC for C14 versions, or EU-MAINS-8 for C8 versions. 1 www.xppower.com ALM65 Series AC-DC Power Supplies Input Units Notes & Conditions 264 VAC Derate linearly from 100% load at 90 VAC to 80% load at 80 VAC 63 60/120 Hz A A 250 0.21 A W Max. 48 2 Units VDC % 2 40 0.5 5 s ms ms % % 4 % Ripple and Noise 1.5 % pk-pk Overshoot Overload Protection Overvoltage Protection Short Circuit Protection Thermal Protection Temperature Coefficient Patient Leakage Current 10 175 170 % % % 100 %/C A 264 VAC, 60 Hz Maximum +60 Units C Notes & Conditions Derate from 100% load at 40 C to 50% load at 60 C Characteristic Minimum Input Voltage 80 Input Frequency Input Current Inrush Current Power Factor Earth Leakage Current No Load Input Power Input Protection 47 Typical Maximum 1.2/0.6 Measured at 115/230 VAC 115/230 VAC, cold start at 25 C EN61000-3-2 Class A 264 VAC, 60 Hz T3.15A/250 VAC internal fuse in both line Output Characteristic Output Voltage Initial Set Accuracy Minimum Load Start Up Delay Start Up Rise Time Hold Up Time Line Regulation Total Regulation Min. 12 Typ. 20 Transient Response 115 145 Trip and restart (hiccup), auto resetting Measured internally, auto resetting 0.05 Notes & Conditions See Models and Ratings table At 50% load No minimum load required Full load and 230 VAC Maximum deviation, recovering to less than 1% within 500 s for 25% step load 20 MHz bandwidth, measured with 20 MHz Bandwidth and 10 F electrolytic in parallel with 0.1 F ceramic capacitor. At turn on / turn off Recycle mains to reset Environmental Characteristic Operating Temperature Cooling Operating Humidity Storage Temperature Operating Altitude Shock Vibration Minimum 0 Natural convection 5 -40 Typical 90 %RH +80 C 5000 m IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes IEC68-2-6, 10-500 Hz, 2 g 10 mins/sweep, 60 mins for each of 3 axes Non-condensing Derating Curve Output Power (%) 100% 80% 60% 40% 20% 0 20 Ambient Temperature (C) www.xppower.com 40 60 2 ALM65 Series AC-DC Power Supplies General Characteristic Minimum Efficiency Typical Maximum Notes & Conditions Units % 88 Switching Frequency Power Density Mean Time Between Failure Weight See Models and Ratings table and curves. 4000 VAC 2 x MOPP 1500 VAC 1 x MOPP (Class I versions only) 500 VAC Class I versions only 65 KHz kHz PWM 4.55 W/in3 Isolation: Input to Output Input to Ground Output to Ground >300 kHrs 0.68 (308) lb (g) MIL-HDBK-217F at 25 C GB Efficiency Curves ALM65US12 100% 90% 80% Efficiency 70% 60% 50% 115V 40% 230V 30% 20% 10% 0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% Load ALM65US24 100% 90% 80% Efficiency 70% 60% 50% 115V 40% 230V 30% 20% 10% 0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100% Load EMC: Emissions Phenomenon Emissions Harmonic Current Voltage Flicker Standard EN55022 EN61000-3-2 EN61000-3-3 Test Level Level B Class A www.xppower.com Notes & Conditions Conducted & Radiated 3 ALM65 Series AC-DC Power Supplies EMC: Immunity Phenomenon ESD Radiated EFT/Burst Surge Conducted Magnetic Fields Test Level 3 10 V/m 3 Installation Class 3 10 V 4 Dip: 30% 500 ms Dip: 60% 200 ms Dip: 80% 5000 ms Dip: 100% 5000 ms Dip: 30% 25 AC Cycles Dip: 60% 5 AC Cycles Dip: 100% 0.5 AC Cycles Int.: >95% 5000 ms Standard EN61000-4-2 EN61000-4-3 EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-8 EN61000-4-11 Dips and Interruptions EN60601-1-2 Criteria A A A A A A A/B A/B B B A A A B Notes & Conditions 15 kV air/8 kV contact 80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies High Line/Low Line High Line/Low Line 230 VAC 100% load, 100 VAC 75% 230 VAC 100% load, 100 VAC 20% At 8 angles Safety Approvals Safety Agency Safety Standard UL Notes & Conditions UL60950-1 2nd Edition TUV EN60950-1 CB IEC60950-1 CE LVD UL ANSI/AAMI ES 60601-1 CSA CSA C22.2 No. 60601 TUV EN60601-1 Information Technology CB IEC60601-1 Others CCC, PSE, KC & RCM Medical May require additional importer information Mechanical Details Class I IEC 320-C14 4.941 (125.5) 1.319 (33.5) Class II IEC 320-C8 35.433 0.8 (90020) 1.512 (38.4) 2.185 (55.5) 0.177 (4.50) o 0.685 (17.4) Optional Input Connector Retention (-A) 1.40 (35.7) 1.969 0.394 (5010) 0.374 0.012 (9.500.30) 1.319 (33.5) 0.039 (1.0) RETURN ( - ) OUTPUT ( + ) 4.488 (114.0) 1.732 (44.0) Notes 1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except 5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8 for C8 versions. output cable length. 2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8 for C8 versions. 3. Weight: 0.68 lbs (308 g) approx. 4. For European mains lead order part EU-MAINS-IEC for C14 versions, or EU-MAINS-8 for C8 versions. www.xppower.com 17 Sept 19