MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 1
Features
2” x 4” x 1.3” Package
For 1U Applications
Class I and Class II versions
175W w/air, 120W convection cooled
Universal Input 90-264 VAC
Average Efficiency meets Level V
Requirements
Approved to CSA/EN/IEC/UL60601-1, 3
rd
Edition
2 x MOPP Isolation
12V fan output
Efficiency 90% typical at Full Load
3 Year Warranty
RoHS Compliant
Description
The MINT1175 Series models provide the highest power density for a power supply in a 2” x 4” size. Ideal for use in
Home Healthcare and Portable Medical Device Applications and more. With input & output monitoring alarms plus
isolated 12V/0.4A fan, it is one the most feature rich supplies and can easily fit in 1U chassis. Exceeding the medical
device strict requirements, the MINT1175 has the lowest earth leakage current with less than 200µA, satisfies 2MOPP
reinforced isolation requirements of IEC60601-1 3rd edition, and can be purchased as Class I or II configurations.
Model Selection
Model
Output Current
Ripple &
Total
OVP***
Number
Volts
Convection
Fan Output
Noise**
Regulation
Threshold***
MINT1175A1206K01 12V 14.6A 10.0A 12Vdc/0.4A 0.5%RMS,
1.2% pk-pk ±3% 14.0 ± 1.1V
MINT1175A1506K01 15V 11.7A 8.0A 12Vdc/0.4A
0.5%RMS,
1.0% pk-pk ±3% 18.5 ± 1.5V
MINT1175A2406K01 24V 7.3A 5.0A 12Vdc/0.4A
0.5%RMS,
1.2% pk-pk ±3% 28.0 ± 2.5V
MINT1175A4806K01 48V 3.6A 2.5 A 12Vdc/0.4A
0.5%RMS,
1.0% pk-pk ±3% 55.0 ± 4.0V
MINT1175A5606K01 56V 3.1A 2.1A 12Vdc/0.4A
0.5%RMS,
1.0% pk-pk ±3% 59.0 ± 1.0V
Notes: * Total convection power is 120 Watts.
** Measured with noise probe directly across output terminals, and load terminated with 0.1µF ceramic and 10µF low ESR capacitors.
*** Output adjustment on 56V model will not exceed 56.2 volts.
General Specifications
AC Input
100-240Vac, ±10%, 47-63Hz, 1
120–370Vdc
Turn On Time
Less than 2 sec. @115Vac (inversely
proportional to input voltage and thermistor
temperature)
Input Current
115Vac: 2A, 230Vac: 1A
Hold-up Time
16mS, typical (120W output load)
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 2
General Specifications
(continued)
Inrush Current
264Vac, cold start: will not exceed 50A
Signals
AC Power Fail, DC OK, Inhibit
Input Fuses
F1, F2: 4A, 250VAC fuses provided on all
models
Overload
Protection
Hiccup Mode
Earth Leakage
Current
<200µA@264Vac, 60Hz, NC; <400µA SFC
Short Circuit
Protection
Provided - no damage will occur if the output is
shorted.
Efficiency
90% typical at 115Vac
Overvoltage
Protection
OVP firing reduces output voltage to <50% of
nominal in <50ms. See chart for trip range.
Output Power
175W continuous, with 200 lfm airflow, 120W
convection cooled – See chart for specific
voltage model ratings.
Switching
Frequency
PFC: Fixed, 65kHz.
Main Converter: Variable 35-200kHz, 65-70kHz
at full load.
Transient
Response
50% load step. i/
t: <0.2A/µS
Max Volt Deviation = 3%
Isolation
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 1500Vac
Ripple and Noise
See chart
Operating
Temperature
-10°C to +70°C
Start Up at -40°C, full load
Output Voltage
See chart
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
Voltage
Adjustability
+/-
5% from nominal
Storage
Temperature
-40°C to +85°C
Minimum Load
Not required
Altitude
Operating: -500 to 10,000 ft.
Non-operating: -500 to 40,000 ft.
Total Regulation
+/- 3%. See chart
Relative Humidity
5% to 95%, non-condensing
Vibration
Operating: 0.003g
²
/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g²/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Shock
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
Dimensions
W: 2.0” x L: 4.0” x H: 1.3”
Safety Standards
EN/CSA/UL/IEC 60601-1, 3
rd
Edition
Weight
210g
MTBF
331,000 hours, 25°C Ambient, 110Vac input
Auxiliary Signals
AC Power Fail:
During normal operation, stays HIGH. Signal
goes LOW with 5mS warning before loss of
DC output from AC failure.
DC OK:
Open collector logic signal goes and stays HIGH
100mS to 500mS after main output reaches
regulation.
Inhibit:
Connect to inhibit pin to output common to
inhibit the DC output.
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 3
EMI/EMC Compliance
Conducted Emissions
EN55011/22 Class B, FCC Part 15, Subpart B, Class B
Radiated Emissions
EN55011/22 Class A; FCC Part 15, Subpart A, Class A w/6db margin
Static Discharge Immunity
EN61000-4-2, Criteria A, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity
EN61000-4-3, 3V/m. Criteria A
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz, Criteria A
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
Conducted RF Immunity
EN61000-4-6, 3Vrms, Criteria A
Power Frequency Magnetic Field Immunity
EN61000-4-8, 3A/m, Criteria A
Voltage Dip Immunity
EN61000-4-11, 0% Vin, 0.5cycle; 40% Vin, 5 cycles; 70% Vin, 25 cycles;
Criteria A
Line Harmonic Emissions
EN61000-3-2, Class A, B, C, & D
Flicker Test
EN61000-3-3, Complies (dmax<6%)
Mechanical Drawing
When installing the power supply into the end-product, a non-conductive Insulator should be placed between the unit and
any conductive metal chassis or mounting platform.
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 4
J300–
Pin 1
Connector Information
Input Connector
J100
Ground
J101
DC Output Connector
J300
Fan Output Connector
J500
Signal Connector
J600
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
.187 x .032 FASTON TAB
PIN 1) +Vout PIN 4) -Vout
PIN 2) +Vout PIN 5) -Vout
PIN 3) +Vout PIN 6) -Vout
PIN 1) +12Vfan
PIN 2) RTN
PIN 1) Inhibit
PIN 2) PF/DC OK
PIN 3, 4) Common
Mating Connector:
Molex 09-50-3031
Pins= 08-52-0072
Mating Connector:
Molex 01-90020009
Mating Connector:
AMP 640250-6
Pins = 640252-1
Mating Connector:
Molex 1375820-2
Pins = 1375819
Mating Connector:
Molex 1375820-4
Pins = 1375819
Fan Output – J500
J500 provides a 12V@0.4mA output to support a system cooling fan. The fan output is always available when AC input is
present, so it also can be used for a 12V standby output is so desired.
AC Power Failure/DC OK and Inhibit Signals – J600
Timing Sequence
J300–Pin 1
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 5
Isolation Specifications
Parameter Conditions/Description Min Nom Max Units
Insulation Safety Rating
Input/Ground
Input/Output
Output/Ground
1 MOPP
2 MOPP
Operational
Electric Strength Test Voltage
Input/Ground
Input/Output
Output/Ground
1800
4000
1500 - -
Vac
Vac
Vac
Input Specifications
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter Conditions/Description Min Nom Max Units
Input Voltage 90 115/230 264 Vac
Turn-On Input Voltage Ramping up 82.7 Vac
Turn-Off Input Voltage Ramping down 67.0 Vac
Input Frequency 47 50/60 63 Hz
Inrush Current Limitation 264Vac, cold start - - 50 A
Power Factor Vi
nom
,
Io
nom
0.9 - -
Efficiency
Vi
nom
,
Io
nom
MINT1175A1206K01
MINT1175A1506K01
MINT1175A2406K01
MINT1175A4806K01
MINT1175A5606K01
- 90% - %
Output Specifications
Parameter Conditions/Description Min Nom Max Units
Output Voltage Setpoint Accuracy
MINT1175A1206K01 Vi
nom
,
, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom
Output Voltage Setpoint Accuracy
MINT1175A1506K01 Vi
nom
,
, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom
Output Voltage Setpoint Accuracy
MINT1175A2406K01 Vi
nom
,
, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom
Output Voltage Setpoint Accuracy
MINT1175A4806K01 Vi
nom
,
, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom
Output Voltage Setpoint Accuracy
MINT1175A5606K01 Vi
nom
,
, Io1 @ ADC, TC = 25 °C -3 - 3 % Vo nom
Output Current V1
Output Current V2 MINT1175A1206K01 0
0
10.0
-
14.6
0.4
ADC
ADC
Output Current V1
Output Current V2 MINT1175A1506K01 0
0
8.0
-
11.7
0.4
ADC
ADC
Output Current V1
Output Current V2 MINT1175A2406K01 0
0
5.0
-
7.3
0.4
ADC
ADC
Output Current V1
Output Current V2 MINT1175A4806K01 0
0
2.5
-
3.6
0.4
ADC
ADC
Output Current V1
Output Current V2 MINT1175A5606K01 0
0
2.1
-
3.1
0.4
ADC
ADC
Static Line Regulation V1 Vi min-Vi max, Vi
nom
,
0-100% Io nom -1 - 1 % Vo nom
Static Load Regulation V1
(Droop Characteristic) Vi min-Vi max, Vi
nom
,
0-100% Io nom -3 - 3 % Vo nom
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 6
Output Specifications
(continued)
Parameter Conditions/Description Min Nom Max Units
Hold-Up Time Starting at Vi = 230 VAC, Po nom - 16 - ms
Dynamic Load Regulation Load change =50%, di/dt = 0.2A/µS
voltage deviation 3% 0 3 % Vo nom
Start-Up Time Vi
nom
,
Io nom 0 - 2 s
Protection
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter Conditions/Description Min
Nom Max Units
Input Fuse Not user accessible
Input Transient Protection 2KV(CM) and 1KV(DM) surge 2 KV (CM)
Output
No-load and short circuit proof Hiccup or Latch
short circuit proof Hiccup or Latch
overload (latch style) Hiccup
Overvoltage Protection Latch style Latch
Over temperature Protection Automatic power shutdown at T
C
=155°C
Characteristic Curves
Output vs. Temperature
120W convection cooled and 175W continuous with 200 LFM airflow. Derate output power to 50% at 70°C.
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 7
Efficiency vs. Loading
The high efficiency is achieved by using LLC technology, PFC topology minimizing switching losses. Synchronous SCHOTTKY or ultra-fast diode is
used as rectifier in MINT1175 family because of high output voltage level.
Noise & Ripple
To verify that the output ripple and noise does not exceed the level specified in the product specification. Measured using a scope probe socket with
0.1µF ceramic and a 10µF electrolytic capacitor connected in parallel across it, BW limit with 20MHz.
MINT1175
175W Single Output
Medical Series
www.slpower.com MINT1175 REV 1.4, 5-AUG -15 8
Output Transient Response
50% load step within the regulation limits of minimum and maximum load, di/dt<0.2A/µSec. Recovery time not specified as there is no laps in regulation
with a 50% Load Step. Maximum voltage deviation is 3%, This test is performed on the MAIN OUTPUT ONLY.
Output Overload Characteristic
Supply shall protect itself against Overload conditions. The Power Supply shall recover from Overload Conditions without operator intervention.
Output Overshoot
When supply is turned on, off or when the load is stepped 100%, the Power Supply Overshoot/Undershoot shall not exceed minimum or maximum of
output voltage regulation.
Data Sheet © 2012 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.