Specifications
Fluke 66 Fluke 68/68IS*
Temperature range -32 to 600 °C -32 to 760 °C
(-25 to 1100 °F) (-25 to 1400 °F)
Accuracy -32 to -26 °C (-25 to -15 °F): ±3 °C (±5 °F)
(assumes ambient -26 to -18 °C (-15 to 0 °F): ±2.5 °C (±4 °F)
operating temperature -18 to 23 °C (0 to 73 °F): ±2 °C (±3 °F)
of 23 °C [73 °F]) For targets above 23 °C (73 °F): ±1 % or reading
or ±1 °C (±2 °F), whichever is greater
Repeatability ±0.5 % or ≤±1 °C (±2 °F), whichever is greater
Response time ≤0.5 second (95 % of reading)
Spectral response 8 to 14 µm
Emissivity Digitally adjustable emissivity
(from 0.1 to 1.0 by 0.01)
Ambient operating temp. 0 to 50 °C (32 to 120 °F)
Relative humidity 10 to 90 % RH non-condensing,
at <30 °C (86 °F) ambient
Storage temperature -20 to 60 °C (-13 to 158 °F) without battery
Dimensions 200 x 160 x 55 mm (8 x 6 x 2 in.)
Weight 320 g (11 oz.)
Power 9 V alkaline or NiCd battery
Batter life (alkaline) 20 hours with laser and backlight on 50 %
40 hours with laser and backlight off
Laser sighting (Class II) Offset single point laser point
Typical distance to target 5 m (15 ft.) 8 m (25 ft.)
Distance to Spot (D:S) 30:1 at focus point 50:1 at focus point
MIN, MAX, AVG, DIF
Temperature display•
Data logging 12 points
Probe jack •
Display hold (7 seconds) •
Hi/Low alarm •
LCD backlit •
Temperature display °C or °F selectable
Display resolution 0.1 °C (0.1 °F)
Hard carrying case •
Locking trigger •
Tripod mounting 6.35 mm (.25 in.) 20 UNC threading
Warranty 2 years, conditional
* Approved by Factory Mutual Research for use in hazardous locations.
Optional accessories:
•RTD temperature probe (standard w/68IS)
32 ºC to 260 ºC (-25 ºF to 500 ºF)
•Non-contact thermometer holster
Ordering information
Fluke 66 Infrared Thermometer
Fluke 68 Infrared Thermometer
Fluke 68IS Infrared Thermometer
Included
Use the Fluke 66 within 5 m (15 ft.) of the
intended target. At greater distances, the
measured area will be larger (approximately
the distance divided by 30).
0.9" @
12"
D:S = 30:1
at focal point S
D
1.2" @
36"
2.5" @
60"
Use the Fluke 68 and 68IS within 8 m (25 ft.) of
the intended target. At greater distances, the
measured area will be larger (approximately the
distance divided by 50).
0.8" @
12"
D:S = 50:1
at focal point S
D
0.7" @
36"
1.7" @
60"
*Included w/68IS
Standard Feature (68IS only)
•Factory Mutual Research Intrinsically Safe
approval, including probe (Intrinsically safe,
Class I Division 1 Groups A, B, C, D Class I,
Zone 0, AEx ia IIC, T4 at 50 ºC when used
with 9V alkaline battery.) Groups A, B, C, D
refers to: Acetylane, Hydrogen, Ethylene
and Propane.