Keysight Technologies

Oscilloscopes




2


See www.keysight.com/find/4000X-series for more details.


Consider the InfiniiVision 4000 X-series
Industry’s first 12.1-inch capacitive touch display
InfiniiScan Zone touch trigger capability
200 MHz - 1.5 GHz DSO and MSO models
1,000,000 wfms/sec
Fully upgradable 5 instruments in 1
20 MHz dual-channel WaveGen with arbitrary waveform
    
Analog channels 2 or 4 2 or 4 2 or 4 2 or 4
Digital channels
16 (MSO models or upgrade) 16 (MSO models or upgrade) 16 (MSO models or upgrade)
8 (MSO models or upgrade)
Bandwidth (upgradable) 1, 2.5, 4, 6 GHz
200, 350, 500 MHz, 1, 1.5 GHz
100, 200, 350, 500 MHz, 1 GHz
70, 100, 200 MHz
Max sampling rate 20 GSa/s 5 GSa/s 4 GSa/s (≤ 500 MHz)
5 GSa/s (1 GHz)
2 GSa/s
Max memory depth 4 Mpts 4 Mpts 2 Mpts (standard)
4 Mpts (option)
100 kpts (standard)
1 Mpts (option)
Max waveform update rate > 450,000 waveforms/sec > 1,000,000 waveforms/sec > 1,000,000 waveforms/sec > 50,000 waveforms/sec
Display 12.1 inches, capacitive,
multi-touch, gesture enabled
12.1 inches, capacitive 8.5 inches 8.5 inches
InfiniiScan Zone trigger Standard Standard No No
Voice control Standard No No No
WaveGen 20-MHz funtion/
arbitrary waveform generator
Dual-channel AWG (option) Dual-channel AWG (option) Single-channel AWG
(option)
Single-channel funtion
(option)
Integrated digital voltmeter Option Option Option Option
Integrated hardware counter 5 digits (standard),
10 digits + totalizer (option)
5 digits (standard) 5 digits (standard) 5 digits (standard)
Search and navigate Standard, lister supported Standard Standard Standard
Segment memory Standard Standard Option Option
Mask/limit test Option Option Option Option
Serial protocol analysis options I2C/SPI, UART, CAN/LIN,
FlexRay, I2S, MIL-STD1553,
ARINC429, USB2.0
I2C/SPI, UART, CAN/LIN,
FlexRay, I2S, MIL-STD1553,
ARINC429, USB 2.0
I2C/SPI, UART, CAN/LIN,
FlexRay, I2S, MIL-STD1553,
ARINC429
I2C/SPI, UART, CAN/LIN
Advanced analysis options Power analysis, USB 2.0
signal quality test, HDTV
analysis, FPGA
Power analysis, USB 2.0
signal quality test, HDTV
analysis, FPGA
Power analysis, HDTV
analysis
No
Color grade Standard No No No
Histogram Standard No No No
FFT Standard enhanced FFT Standard Standard Standard
Jitter analysis with clock
recovery
Option No No No
Realtime eye diagram Option No No No
Advanced math Standard, display four
funtions simultaneously
Standard, display one
funtion
Option, display one funtion No
Connectivity Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard support
Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard support
Standard USB 2.0 (LAN/
video/GPIB option), USB
keyboard support
Standard USB 2.0 (LAN/
video/GPIB option), USB
keyboard support

3000 X-Series – oscilloscopes redefined
Key features
See more:
One million waveforms per second update rate
MegaZoom IV smart memory technology
Large 8.5-inch WVGA display
Optional segmented memory
Do more:
Industry’s first 5-instruments-in-1 (oscilloscope,
digital channels, built-in 20 MHz function/arbitrary
waveform generator with modulation, integrated digital
voltmeter and protocol analyzer)
Get more:
Investment protection with Industry’s only fully-
upgradable oscilloscope, including bandwidth to
1 GHz
Industry’s leading application solutions
The InfiniiVision 3000 X-Series redefined oscilloscopes. It
sees the most signal detail, does more functions than any
other oscilloscope, and gives you maximum investment
protection.
The 3000 X-Series’s innovation starts with the industry’s
only 5-instruments-in-1 integration. The industry-leading
one million waveforms per second update rate is 20 times
faster than the competition to display the most signal
detail. The 3000 X-Series provides maximum investment
protection with fully-upgradable 5-instruments-in-1; even
bandwidth is upgradable. Our breakthrough technology
delivers more scope for the same budget.


InfiniiVision 3000 X-Series with MegaZoom IV smart memory technology

4

The best signal visibility starts with the largest display.
The InfiniiVision 3000 X-Series comes with a large 8.5-inch
WVGA display so you can view analog, digital and serial
signals easily on the screen.

If you can’t see the problem, it is hard to troubleshoot it.
With Keysight’s MegaZoom IV smart memory technology,
the 3000 X-Series updates waveforms up to 1 million times
per second, which gives you the highest probability of cap-
turing random and infrequent events that you would miss
on an oscilloscope with a lower waveform update rate.

With up to 4 Mpts of MegaZoom IV deep memory, you can
capture long, non-repeating signals while maintaining a
high sample rate, then quickly zoom in on areas of interest.
The InfiniiVision 3000 X-Series optimizes your deep-
memory oscilloscope measurements by using MegaZoom
IV technology to make the most effective trade-offs in
sample rate, memory depth and waveform update rate
automatically. Although many people think deeper memory
is always better, usually deep memory means making
tradeoffs.
Oscilloscopes with deep memory require additional wave-
form processing time to acquire deep memory waveforms,
which means waveform update rates will be reduced sig-
nificantly. For this reason, most other oscilloscopes have
manual memory-depth selections, and the typical default
memory depth setting is usually relatively shallow (10 to
100 kpts). If you want to use deep memory in these other
oscilloscopes, you must manually turn it on and deal with
the update rate tradeoff.

The MegaZoom IV smart memory technology combines the
capabilities of an oscilloscope, digital channels, protocol
analyzer, WaveGen built-in function generator and DVM
in a compact form factor. Fourth generation MegaZoom
technology enables the industry’s fastest waveform update
rate with responsive deep memory.
Fast update rate of the 3000 X-Series displaying the rare
metastable signal


4M deep memory captured a long time span without losing
the details

5


The 3000 X-Series is the first instrument in its class to
offer an integrated and upgradable digital channels. Digital
content is everywhere in today’s designs and traditional 2
and 4 channel oscilloscopes do not always provide enough
channels for the job at hand.
With an additional 16 integrated digital channels, you
now have up to 20 channels of time-correlated triggering,
acquisition and viewing on the same instrument. Buy a 2
or 4 channel DSO and at anytime, and upgrade it yourself
to an MSO with a license to turn on those integrated
16 digital channels.

The InfiniiVision 3000 X-Series features up to 4 Mpts of
memory with Keysight’s patented MegaZoom IV technol-
ogy that is always enabled and always responsive provid-
ing the industry’s fastest update rate at up to 1 million
waveforms per second, with no compromise if you turn on
measurements, digital channels or protocol decodes.
In addition, the 3000 X-Series offers 35 (50 with N2820A
Series high-sensitivity AC/DC current probe) automated
measurements, 10 parametric triggers, serial protocol
triggers, as well as waveform math functions including FFT.




An industry first, the 3000 X-Series offers an integrated
built-in 20 MHz function/arbitrary waveform generator,
available with modulation support (DSOX3WAVEGEN). The
integrated function generator provides stimulus output of
sine, square, ramp, pulse, DC, Sinc (x), exponential rise/
fall, cardiac, Gaussian Pulse and noise waveforms to your
device under test. The modulation feature supports AM,
FM, and FSK modulations with modulation shapes of sine,
square, and ramp. With AWG functionality, you can store
the waveforms from analog channels or reference memory
to the arbitrary memory and output from WaveGen. Easily
create/edit the waveform using built-in editor or by using
Keysight’s free Benchlink Waveform Builder Basic:
www.keysight.com/find/33503.

6




Embedded serial triggering and analysis (I2C, SPI)
Computer serial triggering and analysis
(RS232/422/485/UART)
Automotive and industrial serial triggering and analysis
(CAN,LIN)
FlexRay automotive triggering and analysis
Audio serial triggering and analysis (I2S)
Aerospace and defense serial triggering and analysis
(MIL-STD 1553 and ARINC 429)
Keysight’s InfiniiVision Series oscilloscopes are the
industry’s first scopes to use hardware-based serial
protocol decoding. Other vendors scopes use software
post-processing techniques to decode serial packets
frames. With these software techniques, waveform and
decode-update rates tend to be slow (sometimes
seconds per update). That’s especially true when using
deep memory, which is often required to capture multiple
packetized serial bus signals. Faster decoding with
hardware-based technology enhances scope usability,
and more importantly, the probability of capturing
infrequent serial communication errors.
After capturing a long record of serial bus communication
using the InfiniiVision scope’s MegaZoom IV deep
memory, you can easily perform a search operation
based on specific criteria, and then quickly navigate to
bytes/frames of serial data that satisfy that search criteria.
Sometimes it may be necessary to correlate data
from one serial bus to another. Keysight’s InfiniiVision
3000 X-Series oscilloscope can decode and list two
serial buses simultaneously using hardware-based
decoding. as well as display the captured data in a
time interleaved “Lister” display.

An industry first, the 3000 X-Series offers an integrated
3-digit voltmeter (DVM) and 5-digit frequency counter
inside the oscilloscopes (DSOXDVM). The voltmeter
operates through the same probes as the oscilloscope
channels, however, the measurements are de-coupled
from the oscilloscope triggering system so that both the
DVM and triggered oscilloscope measurements can be
made with the same connection. The voltmeter results
are always displayed, keeping these quick characteriza-
tion measurements at your fingertips.

7

oscilloscope

Project needs change, but traditional oscilloscopes are
fixed – you get what you pay for at the time of purchase.
With the 3000 X-Series, your investment is protected. If
you need more bandwidth (up to 1 GHz), digital channels,
WaveGen, DVM or measurement applications in the future,
you can easily add them all after the fact.
Add at the time of your purchase or upgrade later:
Bandwidth
Digital channels (MSO)
WaveGen built-in 20 MHz function/arbitrary
waveform generator
Integrated digital voltmeter
Measurement applications
Serial protocol analysis
Power measurement analysis
HDTV video triggering and analysis
Advanced math analysis
Mask/limit testing
Segmented memory
Educators’ lab kit

Whether performing pass/fail tests to specified standards
in manufacturing or testing for infrequent signal anomalies
in R&D debug, the mask test option can be a valuable
productivity tool. The 3000 X-Series features hardware-
based mask testing and can perform up to 270,000 tests
per second.
Multiple test criteria can be selected including the ability
to run tests for a specific number of acquisitions, time, or
until detection of a failure. Pass/fail masks can be auto-
matically created based on an input reference waveform
along with user-specified tolerance bands, or can be cre-
ated on a PC and then imported via a USB memory stick.
See pages 27 and 28 for more detailed information on
available upgrades.
Mask test evaluated more than 27 million waveforms in just
over two minutes
Limit testing made easy with the “automask” feature

8

When capturing low-duty cycle pulses or data bursts, you
can use segmented memory acquisition to optimize acqui-
sition memory. Segmented memory acquisition not only
lets you selectively capture and store important segments
of signals without capturing unimportant signal idle/
deadtime, but it also allows you to run post-capture inter-
segment analysis such as segment play back, waveform
measurements, and waveform overlay. Segmented memory
acquisition is ideal for applications including packetized
serial buses, pulsed laser, radar bursts and high-energy
physics experiments. Up to 1000 segments can be captured
on the 3000 X-Series models with a minimum re-arm time
under 1 µs. Segmented memory works simultaneously with
serial bus decodes as well.

oscilloscope

When working with switching power supplies and power
devices, the DSOX3PWR power measurements application
provides a full suite of power measurements and analysis
that runs in the oscilloscope. Measurements include:
Current harmonics
Ef ciency
Inrush current
Modulat ion
Power qualit y
Sw itching response
Transient respons e
Turn on/turn off
Outpu t ripple
Power Supply Rejection Ratio (PSRR)
Slew rate
Also included at no additional charge is a license for
the U1881A PC-based power analysis software package
which provides additional offline measurements and report
generation. U1881A additional measurements include:
Safe operating area (SOA)/SOA mask editor
Dynamic on resistance (Rds)
On/offline analysis
Capture 1000 very infrequent glitches over 100 seconds
using segmented memory, then run inter-segment
measurement and overlay analysis on the 1000 segments.
An example screen for power quality analysis

9

oscilloscope

analysis.
Whether debugging consumer electronics with HDTV or
characterizing a design, the DSOX3VID measurement appli-
cation provides support for a variety of HDTV standards
including:
4 8 0p/6 0
567p/50
7 20 p/5 0
72 0p/ 60
1080i/50
1080i/60
108 0p/24
108 0p/25
1080p/30
1080p/50
1080p/60
Generic (custom bi-level and tri-level sync video standards)

In addition to the standard waveform math functions (add,
subtract, multiply, integrate, differentiate, square root,
FFT), the optional DSOX3ADVMATH application provides
additional advanced waveform transforms, filters, and
visualization tools including:
Transforms
Ax + B
Sq uare (x 2)
Absolute value (|x|)
Common logarithm (log)
Natural logarithm (ln)
E xponent ial (eX)
Base 10 exponential (10X)
Filters
Low pass filter (4th order Bessel-Thompson filter
with selectable -3dB frequency)
High pass filter (single-pole high pass filter with
selectable -3dB frequency)
Visualizations Tools
Magnif y
Measurement tr end
Chart logic bus timing
Chart logic bus state




Keysight’s Infiniium Offline PC-based analysis oscilloscope
software allows you to do additional signal viewing,
analysis and documentation tasks away from your scope.
Capture waveforms on your scope, save to a file, and
recall the waveforms into Infiniium Offline. The application
supports a variety of popular waveform formats from
multiple oscilloscope vendors and includes the
following features:


Use familiar scope controls to quickly navigate and zoom
in to any event of interest.
Navigate and view
Navigate in time or between bookmarks
Up to eight waveforms simultaneously
One, two, or four grids
Measurements
More than 50 automated measurements
View up to 20 simultaneously
User-customizable result window
Analyze
20 math operators including FFT and filters
Up to four independent/cascaded math functions
Measurement histogram
Documentation
Up to 100 bookmarks
Markers with dynamic delta value updates when moved
One step save/load setup and all waveforms
Analysis upgrades (optional)
Protocol decode
Jitter analysis
Serial data analysis
For more information, visit
www.keysight.com/find/N8900A


This PC-based software package connects to the scope
via USB or ethernet connection and uses the Keysight
I/O libraries to communicate. It provides advanced FFT
frequency domain analysis at a cost-effective price as well
as spectrum and spectrogram analysis with an intuitive
user interface that RF engineers are familiar with.
For more information visit
www.keysight.com/find/ASV_InfiniiVision
Waterfall view for ASV spectrogram measurement



Store up to two waveforms in the scope’s non-volatile
reference waveform memory locations. Compare these
reference waveforms with live waveforms, and perform
post analysis and measurements on stored data. You can
also store waveforms on a removable USB memory device
in *.h5 format and recall them back into scope’s reference
waveform memory later. Save and/or transfer waveforms
to a PC as XY data pairs in a comma-separated values
format (*.csv) or store bitmap images and transfer them
to a PC for documentation purposes in a variety of image
formats including: 8-bit bitmaps (*.bmp), 24-bit bitmaps
(*.bmp), and PNG 24-bit images (*.png).

Operate the scope in the language most familiar to you.
The graphical user interface, built-in help system, and front
panel overlays are available in 13 languages. Choose from:
English, Japanese, simplified Chinese, traditional Chinese,
Korean, German, French, Spanish, Russian, Portuguese and
Italian. During operation, access the built-in help system
just by pressing and holding any button.

Get the most out of your 3000 X-Series scope by using the
right probes and accessories for your application. Keysight
offers a complete family of innovative probes and acces-
sories for the InfiniiVision 3000 X-Series oscilloscopes
including the innovative N2820A Series high-sensitivity
current probes for ultra-low current measurements. For
the most up-to-date and complete information about
Keysight’s probes and accessories, visit our Web site at
www.keysight.com/find/scope_probes.
Also available is the N2744A T2A (Tektronix TekProbe®
interface to Keysight AutoProbe) probe interface adapter.
This adapter allows users of Tektronix TekProbe active
probes to connect directly to the InfiniiVision 3000 X-Series
AutoProbe interface BNC input.



Quickly display any active signals and automatically set
the vertical, horizontal and trigger controls for optimal
viewing with the press of the autoscale button. (This
feature can be disabled or enabled for the education
environment via a USB thumb drive file with a SCPI
remote command.)



Through improved quality processes and rigorous test-
ing, the Keysight InfiniiVision X-Series oscilloscopes are
now able to perform at specification for two years without
yearly calibration thereby reducing cost of ownership
to you.


The secure erase feature comes standard with all 3000
X-Series models. At the press of a button, internal non-vol-
atile memory is cleared of all setup, reference waveforms,
and user preferences, ensuring the highest level of security
in compliance with National Industrial Security Program
Operation Manual (NISPOM) chapter 8 requirements.

The traditional VNC connection through your favorite PC
browser lets you:
Operate the 3000 X-Series remotely
Save/recall data and setup files
Get screen image
Get instrument status
In addition to the traditional VNC connection, the 3000
X-Series supports remote oscilloscope control from any
html5-enabled browser on your tablet devices. The virtual
front panel looks and acts like the real front panel on the
oscilloscope with the same associated keys and knobs.

Built-in USB host (one front, one back) and USB device
ports make PC connectivity easy. Operate the scope from
your PC and save/recall stored waveforms as well as
set-up files via LAN. The optional LAN/VGA module gives
you network connectivity and complete LXI class C support
as well as the ability to connect to an external monitor. An
optional GPIB module is also available. Only one module
may be used at a time.
34840B BenchVue lets you visualize the 3000 X-Series and
multiple measurements simultaneously. Save time with
the ability to export measurement data to Excel, Word and
MATLAB in three clicks. Monitor and control your 3000
X-Series with a mobile device from anywhere. Learn more at
www.keysight.com/find/BenchVue



When capturing long complex waveforms using the
scope’s deep acquisition memory, manually scrolling
through stored waveform data to find specific events
of interest can be slow and cumbersome. But with the
InfiniiVision 3000 X-Series scope’s automatic search &
navigation capability, you can easily set up specific search
criteria and then quickly navigate to “found and marked”
events using the scope’s front panel forward and back
navigation keys. Available search criteria include: edges,
pulse width (time-qualified), rise/fall times (time-qualified),
runt pulses (time- and level-qualified), and serial.

With today’s more complex signals, it is also often neces-
sary to trigger on complex signal conditions in order to
synchronize the scope’s acquisition on specific events
of interest. Keysight’s InfiniiVision 3000 X-Series scopes
can trigger on the following conditions: edge, pulse width
(time-qualified), pattern, rise/rall time, Nth edge burst,
runt, setup & hold, video, USB 2.0 full/low speed, Serial1,
and Serial2.


The oscilloscope captured
1 ms of a digital data stream.
It was able to find, mark,
and then quickly navigate to
20 “runt” pulses using the
search and navigation feature
Examples of advanced triggers


Teach your students what an oscilloscope is and how to
perform basic measurements with the Educator’s Oscil-
loscope Training Kit (DSOXEDK). It includes training tools
created specifically for electrical engineering and physics
undergraduate students and professors. It contains an
array of built-in training signals, a comprehensive oscil-
loscope lab guide and tutorial written specifically for the
undergraduate student and an oscilloscope fundamentals
PowerPoint slide set for professors and lab assistants. For
more information, refer to: www.keysight.com/find/EDK.

The 3000 X-Series comes with a one-time 30-day all-
optional-features trial license. You can choose to start
the 30-day trial at any time. In addition you can redeem
individual optional feature 30-day trial licenses at any
time by visiting www.keysight.com/find/30daytrial. This
enables you to receive in effect 60 days of trial license of
each optional feature.








Oscilloscope shown actual size















signals

panel




MegaZoom














access to

































3000 X-Series specification overview
3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Bandwidth* (– 3 dB) 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Calculated rise time
(10-90%)” ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps
Input channels DSOX 2 4 4 2 4 2 4 2 4
MSOX 2 + 16 4 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16
* For example, if you choose 500 MHz, 4+16 channels, the model number will be MSOX3054A.
Choose your bandwidth, number of channels and memory depth.
Tailor your scope with measurement applications to save time and money. After purchase upgrade model
numbers are listed below (values in parentheses are factory-installed option numbers).
Description Model number
Oscilloscope features
Memory DSOX3MEMUP (-040)
Segmented memory DSOX3SGM (-SGM)
MSO upgrade DSOX3MSO
MSO upgrade for 1 GHz models DSOXPERFMSO
Serial protocols
Embedded serial triggering and analysis (I²C, SPI) DSOX3EMBD (LSS)
Computer serial triggering and analysis (RS232/UART) DSOX3COMP (-232)
Automotive serial triggering and analysis (CAN/LIN) DSOX3AUTO (-AMS)
FlexRay serial triggering and analysis DSOX3FLEX (-FLX)
Audio serial triggering and analysis (I²S) DSOX3AUDIO (-SND)
MIL-STD 1553 and ARINC 429 serial triggering and analysis DSOX3AERO (-AER)
Measurement applications
WaveGen 20 MHz arbitary/function generator DSOX3WAVEGEN (-001)
Integrated 3-digit digital voltmeter (DVM) DSOXDVM (-DVM)
Power analysis application DSOX3PWR (-PWR)
Mask limit testing DSOX3MASK (-LMT)
Enhanced video/TV application package DSOX3VID (-VID)
Advanced math DSOX3ADVMATH (-MAT)
Productivity tools
Education and training kit DSOXEDK (-EDK)
Infiniium Offline oscilloscope analysis software N8900A
Keysight spectrum visualizer 64997A
Vector signal analyzer software 89601B (version 15 and higher)
Benchlink waveform builder pro and basic 33503A


Choose your probes - For complete list of compatible probes, see Keysight document 5968-8153EN
and visit 
Probes 3000 X-Series
N2862B Passive probe 150 MHz 10:1 attenuation 1 per channel included 100 MHz models
N2863B Passive probe 300 MHz, 10:1 attenuation 1 per channel included 200 MHz models
N2890A Passive probe 500 MHz, 10:1 attenuation 1 per channel included 350/500 MHz/1 GHz models
N6450-60001 16 digital channel MSO cable 1 per scope included on all MSO models and
MSO upgrades
N2889A Passive probe 350 MHz 10:1/1:1 switchable attenuation Optional
10076B Passive probe 250 MHz 100:1 attenuation Optional
N2771B Passive probe 50 MHz 1000:1 attenuation Optional
N2795A Single-ended active probe 1 GHz ±8 V with AutoProbe Optional
N2750A InfiniiMode differential probe 1.5-GHz 700-fF 200-kΩ with AutoProbe Optional
N2790A Differential active probe 100 MHz ±1.4 kV with AutoProbe Optional
N2791A Differential active probe 25 MHz ±700 V Optional
N2792A Differential active probe 200 MHz ±20 V Optional
N2793A Differential active probe 800 MHz ±15 V Optional
1146A AC/DC Current probe 100 kHz 100 A Optional
1147B AC/DC Current probe 50 MHz 15 A with AutoProbe Optional
N2893A AC/DC Current probe 100 MHz 15 A with AutoProbe Optional
N2820A 2-channel high-sensitivity current probe 50 uA - 5 A with AutoProbe Optional
N2821A 1-channel high-sensitivity current probe 50 uA - 5 A with AutoProbe Optional
Step 4. Add the final touches.
Recommended accessories 3000 X-Series
LAN/VGA connection module DSOXLAN
GPIB connection module DSOXGPIB
Rack mount kit N6456A
Soft carrying case and front panel cover N6457A
Hard transit case for 2000 and 3000 X-Series CaseCruzer 3F1112-1510J
(available from http://www.casecruzer.com/)
Hard copy manual N6459A
Front panel cover only N2747A

N2820A high-sensitivity high-dynamic range current probe



* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and
2 mV/div sensitivity setting.

3000 X-Series specification overview
3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Bandwidth * (–3 dB) 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Calculated rise time (10 - 90%) ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps
Input channels DSOX 2 4 4 2 4 2 4 2 4
MSOX 2 + 16 4 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16
Maximum sample rate 4 GSa/s half channel, 2 GSa/s all channel 5 GSa/s half ch,
2.5 GSa/s all ch
Maximum memory depth Standard 2 Mpts, optional 4 Mpts and optional segment memory
Display size and type 8.5-inch WVGA display
Waveform update rate > 1 million waveforms per second
Number of active probes
supported
In general, one for 2-channel models and two for 4-channel models.
Contact Keysight for specific configurations.
Vertical system analog channels
Hardware bandwidth limits Approximately 20 MHz (selectable)
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 1.5%
Input sensitivity range 100 MHz ~ 500 MHz models: 1 mV/div to 5 V/div** (1 MΩ and 50 Ohm)
1 GHz models: 1 mV/div to 5 V/div** (1 MΩ), 1 mV/div to 1 V/div (50 Ohm)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With N2862A, N2863A or N2890A 10:1 probe: 300 Vrms
Frequency de-rating (assumes sine wave input): 400 Vpk until 40 kHz. Then de-rates at 20 db/dec
until 6 Vpk
DC vertical accuracy ±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] **
DC vertical gain accuracy* ± 2.0% full scale**
DC vertical offset accuracy ± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation > 100:1 from DC to maximum specified bandwidth of each model
(measured with same V/div and coupling on channels)
Offset range ± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div)
Vertical system digital channels
Digital input channels 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds Threshold per pod
Threshold selections TTL (+1.4 V), 5V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range ± 8.0 V in 10 mV steps
Maximum input voltage ± 40 V peak CAT I; transient overvoltage 800 Vpk
Threshold accuracy* ± (100 mV + 3% of threshold setting)
Maximum input dynamic range ± 10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip
Input capacitance ~8 pF
Vertical resolution 1 bit



* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature.
Horizontal system analog channels
3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Time base range 5 ns/div to
50 s/div
2 ns/div to
50 s/div
1 ns/div to
50 s/div
500 ps/div to
50 s/div
Time base accuracy* 25 ppm ± 5 ppm per year (aging)
Time base delay time range Pre-trigger Greater of 1 screen width or 250 µs
Post-trigger 1 s to 500 s
Channel-to-channel deskew range ± 100 ns
∆ Time accuracy (using cursors) ± (time base acc. x reading) ± (0.0016 x screen width) ± 100 ps
Modes Main, zoom, roll, XY
XY On channels 1 and 2 only. Z Blanking on ext trigger input, 1.4 V threshold.
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: < 0.5 degree.
Horizontal system digital channels
Minimum detectable pulse width 5 ns
Channel-to-channel skew 2 ns (typical); 3 ns (maximum)
Acquisition system
3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Maximum analog channels sample rate
4 GSa/s half channel interleaved, 2 GSa/s all channel
5 GSa/s half
channel inter-
leaved, 2.5 GSa/s
all channel
Maximum analog channels record length 2 Mpts half channel interleaved, 1 Mpts all channel (standard)
4 Mpts half channel interleaved, 2 Mpts all channel (optional with DSOX3MEMUP (-040))
Maximum duration of time captured at
highest sampling rate (all analog channels)
500 us with 4M memory upgrade 400 us with 4M
memory upgrade
Maximum digital channels sample rate 1 GSa/s 1.25 GSa/s
Maximum digital channels record length 1 Mpts (standard - with digital channels only)
2 Mpts (optional with DSOX3MEMUP - with digital channels only)
Modes Normal Default mode
Peak detect Capture glitches as narrow as 250 ps at all time base settings
Averaging Selectable from 2, 4, 8, 16, 64, ... to 65,536
High
resolution
Real time boxcar averaging reduces random noise and effectively increases vertical resolution
12 bits of resolution when ≥ 10 µs/div at 4 GSa/s (5 GSa/s for 1 GHz models)
or ≥ 20-µs/div at 2 GSa/s (2.5 GSa/s for 1 GHz models)
Segmented Segmented memory optimizes available memory for data streams that have long
dead times between activity. Maximum segments = 1000. Re-arm time = 1 µs (minimum
time between trigger events)



Trigger system
Trigger sources Analog channel (1~4), digital channel (D0~D15), line, external, WaveGen (1 or mod) (FM/FSK)
Trigger modes Normal (triggered): requires trigger event for scope to trigger
Auto: triggers automatically in absence of trigger event
Single: triggers only once on a trigger event, press [Single] again for scope to find another
trigger event, or press [Run] to trigger continuously in either Auto or Normal mode
Force: front panel button that forces a trigger
Trigger coupling DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: < 10 Hz (internal); <50 Hz (external)
HF Reject: High frequency reject, cutoff frequency ~ 50 kHz
LF Reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise Reject: Selectable OFF or ON, decreases sensitivity 2x
Trigger holdoff range 40 ns to 10.00 s
Trigger sensitivity
Internal* < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
External* 200 mVpp from DC to 100 MHz
350 mVpp 100 MHz to 200 MHz
Trigger level range
Any channel ± 6 div from center screen
External ± 8 V
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.



Trigger type selections
Edge Trigger on a rising, falling, alternating or either edge of any source
Edge then edge (B trigger) Arm on a selected edge, wait a specified time, then trigger on a specified count of another
selected edge
Pulse width Trigger on a pulse on a selected channel, whose time duration is less than a value, greater
than a value, or inside a time range
Minimum duration setting: 2 ns (500 MHz, 1 GHz), 4 ns (350 MHz), 6 ns (200 MHz), 10 ns
(100 MHz)
Maximum duration setting: 10 s
Range minimum: 10 ns
Runt Trigger on a position runt pulse that fails to exceed a high level threshold. Trigger on a
negative runt pulse that fails to exceed a low level threshold. Trigger on either polarity runt
pulse based on two threshold settings. Runt triggering can also be time-qualified (< or >) with
a minimum time setting of 2~10 ns and maximum timesetting of 10 s.
Minimum time setting: 2 ns (500 MHz, 1 GHz), 4 ns (350 MHz), 6 ns (200 MHz).
10 ns (100 MHz)
Setup and hold Trigger and clock/data setup and/or hold time violation. Setup time can be set from –7 s to
10 s. Hold time can be set from 0 s to 10 ns.
Rise/fall time
Trigger on rise-time or fall-time edge speed violations (< or >) based on user-selectable threshold.
Select from (< or >) and time settings range between
Minimum: 1 ns (500 MHz, 1 GHz), 2 ns (350 MHz), 3 ns (200 MHz), 5 ns (100 MHz)
Maximum: 10 s
Nth edge burst
Trigger on the Nth (1 to 65535) edge of a pulse burst. Specify idle time (10 ns to 10 s) for framing.
Pattern Trigger when a specified pattern of high, low, and don’t care levels on any combination of
analog, digital , or trigger channels is [entered | exited]. Pattern must have stabilized for a
minimum of 2 ns to qualify as a valid trigger condition.
Minimum duration setting: 2 ns (500 MHz, 1 GHz), 4 ns (350 MHz), 6 ns (200 MHz), 10 ns
(100 MHz)
Maximum duration setting: 10 s
Range minimum: 10 ns
Or Trigger on any selected edge across multiple analog or digital channels
Video Trigger on all lines or individual lines, odd/even or all fields from composite video, or
broadcast standards (NTSC, PAL, SECAM, PAM-M)
Enhanced Video (optional) Trigger on lines and fields of enhanced and HDTV standards (480p/60, 567p/50, 720p/50,
720p/60, 1080p/24, 1080p/25, 1080p/30, 1080p/50, 1080p/60, 1080i/50, 1080i/60).
USB Trigger on start of packet, end of packet, reset complete, enter suspend, or exit suspend.
Support USB low-speed and full-speed.
I2C (optional) Trigger at a start/stop condition or user defined frame with address and/or data values. Also
trigger on missing acknowledge, address with no accq, restart, EEPROM read, and 10-bit write.
SPI (optional) Trigger on SPI (Serial Peripherial Interface) data pattern during a specific framing period.
Supports positive and negative Chip Select framing as well as clock Idle framing and user-
specified number of bits per frame. Supports MOSI and MISO data.
RS-232/422/485/UART (optional) Trigger on Rx or Tx start bit, stop bit or data content or parity error.
I²S (optional) Trigger on 2’s complement data of audio left channel or right channel (=, ≠, <, >, > <, < >,
increasing value, or decreasing value)
CAN (optional) Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start of
frame (SOF) bit (standard). Remote frame ID (RTR), data frame ID (~RTR), remote or data frame
ID, data frame ID and data, error frame, all errors, acknowledge error and overload frame.
LIN (optional) Trigger on LIN (Local Interconnect Network) sync break, sync frame ID, or frame ID and data.
FlexRay (optional)
Trigger on frame ID, frame type (sync, start-up, null, normal), cycle-repetitive, cycle-base, and errors.
MIL-STD 1553 (optional) Trigger on MIL-STD 1553 signals based on word type (Data or Command/Status), Remote
Terminal Address, data, and errors (parity, sync, Manchester encoding).
ARINC 429 (optional) Trigger and decode on ARINC429 data. Trigger on word start/stop, label, label + bits, label
range, error conditions (parity, word, gap, word or gap, all), all bits (eye), all 0 bits, all 1 bits.

22
Waveform measurements
Cursors** Single cursor accuracy:
±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
Dual cursor accuracy:
±[DC vertical gain accuracy + 0.5% full scale]*
Units: Seconds(s), Hz (1/s), Phase (degrees), Ratio (%)
Automatic measurements Measurements continuously updated with statistics. Cursors track last selected
measurement. Select up to four measurements from the list below:
Voltage: peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot,
average- N cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N
cycles, AC RMS- full screen (standard deviation), ratio (RMS1/RMS2)
Time: period, frequency, counter, + width, - width, burst width, duty cycle, rise time, fall
time, delay, phase, X at min Y, X at Max Y
Count: positive pulse count, negative pulse count, rising edge count, falling edge count
Mixed: area- N cycles, area- full screen
Counter Built-in frequency counter:
Source: on any analog or digital channel
Resolution: 5 digits
Maximum frequency: bandwidth of scope
Waveform math
Arithmetic f (g(t))
g(t): { add, subtract, multiply between any 2 channels}
f(t): {FFT(g(t)), differentiate d/dt g(t), integrate ∫ g(t) dt, square root √g(t) }
Enabled between any combination of two channels
Arithmetic DSOX3ADVMath advanced waveform math option adds Ax + B, Square, Absolute,
Common Log, Natural Log, Exponential, Base 10 Exponential, LP Filter, HP Filter, Magnify,
Measurement Trend, Chart Logic Bus (Timing or State).
FFT Up to 64 kpts resolution
Set FFT Window to: Hanning, Flat Top, Rectangular, Blackman-Harris

* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div
sensitivity setting.
Display characteristics
Display 8.5-inch WVGA
Resolution 800 (H) x 480 (V) pixel format (screen area)
Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls.
Format YT, XY, and Roll
Maximum waveform update rate > 1,000,000 wfms/s
Persistence Off, infinite, variable persistence (100 ms - 60 s)
Intensity gradation 64 intensity levels



WaveGen – Built-in function/arbitrary waveform generator (specifications are typical)
WaveGen out Front-panel BNC connector
Waveforms Sine, Square, Ramp, Pulse, DC, Noise, Sine Cardinal (Sinc), Exponential Rise, Exponential Fall,
Cardiac, Gaussian Pulse, and Arbitrary.
Modulation Modulation types: AM, FM, FSK
Carrier waveforms: sine, ramp, sine cardinal, exponential rise, exponential fall, and cardiac.
Modulation source: internal (no external modulation capability)
AM:
Modulation: sine, square, ramp
Modulation frequency: 1 Hz to 20 kHz
Depth: 0% to 100%
FM:
Modulation: sine, square, ramp
Modulation frequency: 1 Hz to 20 kHz
Minimum carrier frequency: 10 Hz
Deviation: 1 Hz to carrier frequency or (2e12 / carrier frequency), whichever is smaller
FSK:
Modulation: 50% duty cycle square wave
FSK rate: 1 Hz to 20 kHz
Hop frequency: 2 x FSK rate to 10 MHz
Sine Frequency range: 0.1 Hz to 20 MHz
Amplitude flatness: ±0.5 dB (relative to 1 kHz)
Harmonic distortion: –40 dBc
Spurious (non harmonics): –40 dBc
Total harmonic distortion: 1%
SNR (50 ohm load, 500 MHz BW) : 40 dB (Vpp > = 0.1 V); 30 dB (Vpp < 0.1V)
Square wave /pulse Frequency range: 0.1 Hz to 10 MHz
Duty cycle: 20 to 80%
Duty cycle resolution: Larger of 1% or 10 ns
Pulse width: 20 ns minimum
Rise/fall time: 18 ns (10 to 90%)
Pulse width resolution: 10 ns or 5 digits, whichever is larger
Overshoot: < 2%
Asymmetry (at 50% DC): ±1% ± 5 ns
Jitter (TIE RMS): 500 ps
Ramp/triangle wave Frequency range: 0.1 Hz to 200 kHz
Linearity: 1%
Variable symmetry: 0 to 100%
Symmetry resolution: 1%
Noise Bandwidth: 20 MHz typical
Sine Cardinal (Sinc) Frequency range: 0.1 Hz to 1.0 MHz
Exponential Rise/Fall Frequency range: 0.1 Hz to 5.0 MHz
Cardiac Frequency range: 0.1 Hz to 200.0 kHz
Gaussian Pulse Frequency range: 0.1 Hz to 5.0 MHz
Arbitrary Waveform length: 1 to 8k points
Amplitude Resolution: 10 bits (including sign bit)***
Repetition Rate: 0.1Hz to 12 MHz
Sample Rate: 100 MSa/s
Filter Bandwidth: 20 MHz

24

* Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 ohms.
** Sinc, Cardiac and Gaussian Pulse: ±1.25 V into Hi-Z; +- 625 mV into 50 ohms
*** Full resolution is not available at output due to internal attenuator stepping.
WaveGen – Built-in function/arbitrary waveform generator (specifications are typical) (Continued)
Frequency Sine wave and ramp accuracy:
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy:
[50+frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude Range:
20 mVpp to 5 Vpp into Hi-Z**
10 mVpp to 2.5 Vpp into 50 ohms**
Resolution: 100 µV or 3 digits, whichever is higher
Accuracy: 2% (frequency = 1 kHz)
DC offset Range:
±2.5 V into Hi-Z**
±1.25 V into 50 ohms**
Resolution: 100 µV or 3 digits, whichever is higher
Accuracy (waveform modes): ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode): ± 1.5% of offset setting ± 3 mV
Trigger output Trigger output available on Trig out BNC
Main Output Impedance : 50 ohms typical
Isolation: not available, main output BNC is grounded
Protection: overload automatically disables output
Digital voltmeter (specifications are typical)
Functions ACrms, DC, DCrms, Frequency
Resolution ACV/DCV: 3 digits Frequency: 5.5 digits
Measuring rate 100 times/ second
Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements.
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds.
Connectivity
Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol.
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices, printers and keyboards
Optional ports GPIB, LAN (10/100Base-T), WVGA video out
Trigger out BNC connector on the rear panel. Supported modes: triggers, mask, and waveform generator
sync pulse

25
General and environmental characteristics
Power line consumption 100 watts
Power voltage range 100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz ± 10% auto ranging
Temperature Operating: 0 to +55 °C
Nonoperating: –30 to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude Operating: up to 4,000 m, Non-operating 15,300 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Vibration Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random
Shock Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random; (operating 30 g,
½ sine. 11 ms duration, 3 shocks/axis along major axis, total of 18 shocks
Dimensions 381 mm (15 in) W x 204 mm (8 in) H x 142 mm (5.6 in) D
Weight Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs)


26
Nonvolatile storage
Reference waveform display 2 internal waveforms or USB thumb drive
Waveform storage Setup, .bmp, .png, .csv, ASCII, XY, reference
waveforms .alb, .bin, lister, mask, HDFS
Max USB flash drive size Supports industry standard flash drives
Set ups without USB flash drive 10 internal setups
Set ups with USB flash drive Limited by size of USB drive

Related literature
Publication title Publication type Publication number
Serial Bus Applications for Keysight InfiniiVision 3000 X-Series Oscil-
loscopes
Data sheet 5990-6677EN
Power Measurements for Keysight InfiniiVision 3000 X-Series oscilloscope Data sheet 5990-8869EN
Mask/Waveform Limit Testing For Keysight InfiniiVision Series
Oscilloscopes
Data sheet 5990-3269EN
Included standard with oscilloscope
Factory warranty 3-year warranty (90 days for unserialized accessories such
as passive probes)
Calibration Certificate of calibration, 2-year calibration interval
Standard secure erase
Probes
N2862B Passive probe 150 MHz 10:1 attenuation 1 per channel included 100 MHz models
N2863B Passive probe 300 MHz, 10:1 attenuation 1 per channel included 200 MHz models
N2890A Passive probe 500 MHz, 10:1 attenuation 1 per channel included 350/500 MHz and 1 GHz models
N6450-60001 16 digital channel MSO cable 1 per scope included on all MSO models and
DSOX3MSO (for 500 MHz models and below)
DSOX3PERFMSO (for 1 GHz Models)
Interface and built-in help language support
English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian,
Portuguese, Italian, Thai, and Polish.
Localized power cord
For MET/CAL procedures, click on the Cal Labs solutions link below
http://www.callabsolutions.com/products/Keysight/
These procedures are FREE to customers

27
License Only Bandwidth Upgrade Models
3000 X-Series
DSOX3BW24 100 MHz to 200 MHz, 4 ch, License only
DSOX3BW52 350 MHz to 500 MHz, 2 ch, License only
DSOX3BW54 350 MHz to 500 MHz, 4 ch, License only

Process Description
Ke
ysight sales
partner
Customer
Keysight software
licensing website
USB drive containing
license file
5
Email containing
attached licensed file
Entitlement
certificate Stick-0n
label sheet
Measurement Applications
DSOX3WAVEGEN WaveGen (built -in function generator with AWG)
DSOXDVM Integrated digital voltmeter
DSOXEDK Educator’s kit
DSOX3MASK Mask testing
DSOX3SGM Segmented memory
DSOX3ADVMATH Advanced waveform math
DSOX3VID Enhanced video triggering
DSOX3EMBD Embedded serial triggering and analysis (I2C, SPI)
DSOX3COMP Computer serial triggering and analysis
(RS232/422/485/UART)
DSOX3AUDIO Audio serial triggering and analysis (I2S)
DSOX3AUTO Automotive serial triggering and analysis (CAN,
LIN)
DSOX3FLEX FlexRay serial triggering and analysis
DSOX3AERO Aerospace serial triggering and analysis (MIL-
STD 1553, ARINC 429)
DSOX3PWR Power measurements and analysis
DSOX3MSO MSO upgrade: add 16 digital timing channels
(for 500 MHz and below models)
DSOXPERFMSO MSO upgrade: add 16 digital timing channels
(for 1 GHz models)
Place order for a license only bandwidth upgrade or
measurement application product to a Keysight sales
partner. If multiple bandwidth upgrade steps are needed,
order all the corresponding upgrade products required to
get from current bandwidth to desired bandwidth. In the
case where the new bandwidth requires higher bandwidth
passive probes, they are included with the upgrade. For the
DSOX3BW24, the N2863B 10:1 300 MHz passive probes
(1 per channel) will be sent with the upgrade.
For measurement applications, you will receive a paper
or electronic .pdf Entitlement Certificate. For bandwidth
upgrades only, you will receive a stick-on label document
indicating upgraded bandwidth specification in addition to
a paper Entitlement Certificate.
Use Entitlement Certificate containing instructions and
certificate number needed to generate a license file
for a particular 2000 or 3000 X-Series oscilloscope model
number and serial number unit.
Receive the licensed file and installation instructions via
email.
Copy license file (.lic extension) from email to a USB drive
and follow instructions in email to install the purchased
bandwidth upgrade or measurement application on the
oscilloscope.
For bandwidth upgrades only, attach bandwidth upgraded
stick-on label to front and rear panels of the oscilloscope.
Model number and serial number of the oscilloscope do not
change.
* See page 30 for return-to-Keysight service center upgrade process for these products

1
2
3
4
5
6
28

Return-to-Keysight Bandwidth Upgrade Models
3000 X-Series
DSOX3BW32 100 MHz to 350 MHz, 2 ch, Service center
DSOX3BW34 200 MHz to 350 MHz, 4 ch, Service center
DSOX3BW12 500 MHz to 1 GHz, 2 ch, Service center
DSOX3BW14 500 MHz to 1 GHz, 4 ch, Service center
Process Description
Keysight Business
Center
Keysight sales
partner Keysight service
center
Customer
Ship scope per
instructions
Upgraded scope
shipped back
Place order for a return-to-Keysight Service Center
bandwidth upgrade product to a Keysight sales partner.
Service Center installation, calibration, shipment costs are
in addition to bandwidth upgrade product price. If multiple
upgrade steps are needed, order all the corresponding
upgrade products required to get from current bandwidth
to desired bandwidth. In the case where the new band-
width requires higher bandwidth passive probes, they
are included with the upgrade. For the DSOX3BW32 and
DSOX3BW34, the N2890A 10:1 500 MHz passive probe
(1 per channel) will be sent with the upgrade.
Keysight Business Center will contact you regarding pro-
cess and timing of the Service Center installation. Continue
to use oscilloscope until contacted again later when parts
are available at Service Center.
Ship the oscilloscope per provided instructions to Service
Center.
Service Center ships back upgraded oscilloscope with
stick-on labels applied to front and rear panels indicating
upgraded bandwidth specification. Model number and
serial number of the oscilloscope do not change.
* See page 29 for license-only upgrade process for these products

1
2
3
4
       





myKeysight
www.keysight.com/find/mykeysight

www.axiestandard.org
®


®®

www.lxistandard.org



www.pxisa.org


Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty



Keysight Assurance Plans
www.keysight.com/find/AssurancePlans



www.keysight.com/quality



Keysight Channel Partners
www.keysight.com/find/channelpartners









 
 
 
 

 
China 
Hong Kong 
 
Japan 
 
Malaysia 
 
 
 

 
 
 
 
 
 
 
Italy 
 
 
 
 
 
 



 


