V18 TR60M SERIES 60W MEDICAL SWITCHING ADAPTER Features * Universal Input Range 90~264VAC * Meets EN60601-1 and EN55011 Class B * Continuous Short Circuit Protection * Over Voltage Protection * Meet CEC Level (Output Cable Length1800mm) (TR60M Series meets CEC IV except TR60M05 is Non-CEC Compliant) (TR60M12: Output Cable Length1220mm 16AWG) * Efficiency & Standby Power Meet Level V (Option) (Output Cable Length1800mm) (TR60M12Output Cable Length720mm 16AWG) (TR60M15Output Cable Length1220mm 16AWG) (TR60M18,TR60M19Output Cable Length1500mm 18AWG) * Meets 2 MOPP Ordering information TR60MXX- XX X XX Model No. DC Plug Type OVP E: With OVP DC Cable Length and Type 01: 720mm 02: 1220mm 03: 1800mm 11: 720mm with Ferrite Core 12:1220mm with Ferrite Core 13:1800mm with Ferrite Core *18AWG/UL1185 OUTPUT MODEL VOLTAGE RIPPLE & VOLTAGE LINE LOAD NOISE ACCURACY REGULATION REGULATION NOTE 2 NOTE 1 NOTE 3 NOTE 4 OUTPUT % EFF. (Typ.) CURRENT NOTE 5 TR60M05 5V 6A 50 mV 4% 1% 6% 75% TR60M12 12 V 5A 120 mV 2% 1% 5% 85% TR60M15 15 V 4A 150 mV 2% 1% 3% 85% TR60M18 18 V 3.33 A 180 mV 2% 1% 2% 86% TR60M19 19 V 3.15 A 190 mV 2% 1% 2% 86% TR60M24 24 V 2.5 A 240 mV 2% 1% 2% 87% TR60M36 36 V 1.66 A 360 mV 2% 1% 2% 87% TR60M48 48 V 1.25 A 480 mV 2% 1% 2% 87% Specifications are subject to change without notice. 1 V18 Specifications TR60M Series Derating Curve Voltage .............................................................................. 90~264Vac Frequency ......................................................................... 47 to 63Hz Inrush Current ........................... Cold Start @25 80A max. @240Vac Conducted EMI ................................................ CISPR/FCC Class B Leakage Current .............................................................. 0.1mA max. OUTPUT SPECIFICATIONS: Hold-up Time ......................................................... 8ms typ. @115Vac Short Circuit Protection ...................................................... Continuous Over Voltage Protection ....................................................................... Yes 100% Rated Load(%) INPUT SPECIFICATIONS: 75% 50% 20LFM(Natural Convection) 25% 0% 0 10 20 30 40 Ambient Temperature(oC) 50 60 Temperature Coefficient ............................................................0.05%/ GENERAL SPECIFICATIONS: NOTE: Isolation ................................................... Input to output =5,656VDC Operating Temperature ................................... 0 ~ 60 (see derating curve) Storage Temperature ........................................................... -20 ~ 85 Humidity .................................................. 93% RH max. Non condensing Cooling ................................................................ Natural Convection Switching Frequency ................................................ 100KHz Typical 1. Voltage accuracy at 60% full load. 2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for Ripple & Noise measurement @20MHz BW. 3. Line regulation is measured from 100Vac to 240Vac full load. 4. Load regulation is measured from 60% to 100% full load and from 60% to 20% full load (60% +/- 40% full load). MTBF ......... MIL-HDBK-217F, GB, at 25/115VAC .......... 200Khrs min. Altitude ................................................................................. 3000m Dimensions ........... . 5.197x2.283x1.201 inches (132.00x58.00x30.50mm) Weight ........................ ....................................... 345g(0.76 Pounds) 5. Typical efficiency at 230VAC and full load at 25. 6. "Various TR Series adapters are PSE certified. PSE certification alone is not sufficient for importation into Japan. A valid PSE mark must contain the name of the importer as shown in the example below. SAFETY AND EMC: If PSE mark is required, the name of the registered importer must be Emission and lmmunity (Ed. 4.0) ........ N55011, EN60601-1-2, EN61000-3-2 supplied to Cincon on order placement. Product labels will not contain EN61000-3-3 Safety (Ed. 3.1) ... Class II, IEC60601-1:2005+A1:2012, EN60601-1:2006/A1:2013 UL ANSI/AAMI ES60601-1:2005 PSE mark if importer name is not supplied. Consult factory or local representative for details". Mechanical Specification All Dimensions are in inches(mm) Tolerance:Inches:X.XXX0.02 Millimeters:X.XX0.5 5.197[132.00] 70.866[1800.00]1.969[50.00] 2.283[58.00] LED DC Plug type: V+ ) VDC Plug :Right Angle(5.52.1) L12mm 18AWG1800mm 1.201[30.50] IEC320/C8 Typical at 25C, nominal line and 75% load, unless otherwise Specified 2