Digital Multimeter Series
Series
Pocket DMM
Series of 4.5-digit Handheld Multimeters
Series of 3.5-digit Handheld Multimeters
Series of 3.5-digit Handheld Multimeters
of 3.5-digit Pocket Digital Multimeter
Bulletin DMM-E
0.020% Maximum Measurement Accuracy
0.09% Accuracy RMS Measurement Low-end Model
1
Glossary
Integral Action Time
Measurement Accuracy
Root Mean Square Value
Mean Value
Form Factor
Frequency Characteristic
Input Impedance
Decibel
CE Mark
Electromagnetic Compatibility (EMC)
Safety Standards
Crest Factor
Peak-to-Peak (P-P) value
Measurement categories (CAT)
In order to ensure the safety of the user, IEC 60664 defines the ranges of use of
measuring instruments by classifying power levels into measurement categories
II through IV and O (None, other). This is because the excessive impulse or surge
levels induced in a power line vary depending on the location of measurement
(category). Categories
with higher numerals
designate locations that
include larger surge
voltages. Instruments that
are designed for category
III can thus withstand
higher surge voltages than
instruments designed for
category II.
Internal wiring
Outlet
Service
drop
Table 1. RMS Value, Average Value, Waveform Factor and Crest Factor
for a Typical Periodic Waveform
Item Waveform RMS Mean value Waveform
factor Crest
factor
Figure 1. RMS and Mean Values of Sine Wave Figure 2. RMS of Distorted Waves
Sine wave
Half
rectification
wave
Full
rectification
wave
Triangular
wave
Square
wave
E (RMS value)
RMS value
Emean (mean value)
Ep (Amplitude)
(energy)
Mean value
(surface area)
Calibration of RMS value by
mean value rectification
P-P value
E (RMS value)
Instantaneous value and spectrum
Emean (mean value)
Ep (crest value)
RMS of each spectrum
RMS value
Crest factor (CF)
CF = Crest value
RMS value
Waveform factor = RMS value
mean value
DC component Fundamental wave
component Harmonic
component
Digital multimeters (DMMs) employ an A/D converter with a dual-integration
system, which determines the measurement value by converting the input
voltage into time using an integration AD converter. The interval to perform
an integral action periodically is referred to as the integralaction time.
Refers to a characteristic that shows variations in input, measurement, or
response with frequency. When measuring alternating current signals, a
measured signal does not have a simple frequency, but often includes
various frequencies ranging from lower frequencies to higher harmonics.
To measure such signals more accurately, it is preferable to use a
measurement device that has a broader frequency characteristic range.
To prevent the measured object from being influenced during voltage
measurement, you should use a measurement device with an extremely
high input impedance.
A unit used for describing the change in electrical signal amplitude or noise
level, or transmission systems in wired devices, etc. This parameter is also
used to represent the level differences in voltage, current or related values,
but is generally restricted to cases characterized by the relationship: (I
1
/I
2
)
2
=
(V
1
/V
2
)
2
= P
1
/P
2
. In the abbreviation "dB," "d" (deci) denotes 1/10, and "B"
(Bell) denotes logarithm.
The products of Yokogawa Meters & Instruments Corporation are subjected
to design and evaluation testing to ensure compliance with the safety and
EMC standards in accordance with the directives issued by the EC.
The parameters EMI and EMS are referred to as electromagnetic
compatibility as they relate to compatibility within an electromagnetic
environment.
These standards lay out safety requirements that are to be met by a product
with the objective of the preservation of human life and property. The
applicable international standard is IEC 61010, and while a product must
conform to this standard, there are also domestic standards laid out by
individual countries. With these safety regulations, the range of use of a
measurement device is specified by categorization in measurement categories
I through IV to ensure the safety of the user. The designations "CAT II,
1000 V" or "CAT III, 600 V" at the input terminals of a measurement device,
for example, indicates the applicable category and the maximum voltage for
the device in terms of safety.
With DMMs, the measurement accuracy is generally expressed as: ±__% of
reading + __digits. ("Reading" refers to the reading value, and is abbreviated
as "rdg"; "digits" refers to the number displayed in the smallest decimal
place, and is abbreviated as "dgt.") This expresses the range of values that
a DMM may measure or represent for a given actual value.
The value most directly related to the energy of a given waveform. Refers to
the square root of a value found by averaging the squares of instantaneous
values of a waveform over a single cycle. (See Table 1,Figures 1 and 2.)
Refers to the average of the sum of instantaneous values, determined for a
current half-wave. It is equivalent to calculating the surface area of a
waveform.
Ratio of RMS value with respect to average value.
Form factor = RMS value/mean value (See Figures 1 and 2.)
Ratio of maximum value to RMS value.
Crest factor = maximum value/RMS value(See Figures 1 and 2.)
Refers to the distance between the smallest and largest amplitudes in a
waveform (see Figure 1).
Measurement category
Appliances,
portable equipments, etc.
RemarksDescription
Switchboard,
circuit breaker, etc.
Overhead wire,
cable systems, etc.
Other circuits that are not directly connect to MEAINS.
For measurement performed on circuits directly connected to
the low-voltage installation.
For measurement performed in the building installation.
For measurement performed at the source of the low-voltage
installation.
O (None, other)
2
Digital Multimeter Selection Guide
External View
Model
Type
Max. V alue
Dual
Bar Graph
Back-lit
AC RMS
Voltage
AC + DC
Current
AC + DC A
Resistance
Continuity Check
Diode Test
Frequency
Temperature
Capacitance
Filter
Communication
Data Memory
Max./Min. V alue Memory
Relative Value Computation
Logarithm Computation
Data Hold
Auto Hold
Peak Hold
Overvoltage Input Warning
PC Connection
Auto Power Off
TY710
TY720
TY520
TY530
73201
73202
73203
73204
73101
50000
4300
6000
4300
Handheld
Pocket-
sized
Display Measurement Items Additional Functions
: Also functions as excessive current input warning.
3
Maximum Measurement Accuracy
Closed Case Calibration
0.020% rdg + 2 dgt (DC voltage)
True RMS measurement
In addition to the above, the sub-display can display the reference value for
differential calculation, memory storage numbers for measured data,
minimum/maximum/average value recording times, and standard resistance
during decibel calculation.
Loaded with Measurement Functions
Duty ratio (%) measurement
AC+DC measurement
Measures RMS of a waveform in which ripple waveforms are superimposed
on a direct current.
Supports waveforms of 1 ms or greater. You can capture instantaneous crest
values not possible with ordinary maximum measurement functions.
Can display the measured values as the values relative to a reference value
(defined by the REL key; even after data hold) or as the percentages of the
reference value.
Percentage calculation: (Measured value – reference value) / (reference
value), expressed as percentage.
Computes the logarithm of an alternating current, and uses it together with
the relative value computation to display the relative value. You can select
the standard resistance according to the application, such as audio or
communication circuit signal measurement.
50,000-count, 51-segment bar graph display
Backlight provided as standard for when working in dark places.
Simultaneous display of frequency and voltage, frequency and
duty ratio or decibels and voltage on the dual display.
Displays the duty ratio of a pulse waveform:
(High level period/1 cycle of waveform) x 100 = (t2/t1) x 100 [%]
Automatically hold the data measured when the test leads are disconnected
from the measured object, thus freeing both hands for performing reliable
measurement.
Allows recording of minimum, maximum and average values along with their
respective times (time passed since the start of measurement)
Peak hold function (TY720, for DC V/A measurement)
Relative and percentage value computation
Auto hold
Minimum/maximum/average display
The optional communication package*1 sold separately (Model 92015)
allows you to connect to a PC for transmitting large amounts of data that
cannot be saved in the DMM internal memory.
You can transmit the saved data from the internal memory to a PC and
process it using application software or spreadsheet software (Excel*2) for
data management.
For details of the application software, refer to page 7.
Real-time measurement
SAVE-mode memory
A mode for manually saving any data
Logging-mode memory
A mode for automatically saving data at a specified interval
Logging interval: 1 second to 30 minutes
Two memory modes
Decibel calculation
* Selectable standard resistance values:
4/8/16/32/50/75/93/110/125/135/150/200/250/300/500/600/800/900/1000/1200
*1 Communication cable and application software are included.
*2 Excel is a registered trademark of Microsoft Corporation in the United States.
*3 The communication cable employs an infrared system, so the device is
electrically insolated.
Full Display Functions
Safe Design
Display: V AC and V DC measurements
Sub-display
Full Support for Data Management
Model
TY710
TY720
SAVE-mode memory*
100
Logging-mode memory*
1000
10000
* Saved data can be checked on the display.
Memory capacity
Series Most Reliable Handheld Digital Multimeters
TY720
TY710
Model
4.5
digits
4.5
digits
50
,
000
count
50
,
000
count
Terminal
shutters
Terminal
shutters
USB
USB
USB
USB
0.020%
(DCV)
0.020%
(DCV)
AC50mV
(TY720)
AC50mV
(TY720)
LPF
(TY720)
LPF
(TY720)
RMS
RMS Shutters prevent erroneous insertion of test leads into current
measurementterminals (terminal shutters)
Conforms to EN61010-1 safety standard
The current terminals have terminal shutters that prevent erroneous setting
of the measurement function and leadwire connections resulting from
operational errors. The terminal shutters open and close according to the
function switch position.
Conforms to measurement category 1000 V AC/DC, CAT and 600 V
AC/DC, CAT Ⅳ
The TY series, simply performing special operations via front panel allows
for quick and reliable adjustment. In addition, the series allows for one-
touch adjustment of AC voltage- and AC current-to-frequency
characteristics. The user calibration function leads to improved operation
efficiency and cost reduction.
• External standard instrument required for calibration.
User calibration function
4
TY700 General Specifications
Performance
Model and Specification Code
Measurement Functions
:DC voltage, AC voltage, DCV+ACV, DC current, AC current, DCA+ACA, resistance, frequency, temperature, capacitance,
duty cycle, decibel calculation, continuity check, diode test, low-power resistance (TY720 only)
For AC voltage/current, RMS/MEAN detection can be switched (TY720 only).
For AC voltage/current, the low-pass filter can be turned on/off (TY720 only).
Additional Functions :Data hold/auto hold/peak hold (TY720 only), range hold, maximum/minimum/average values
resistance, capacitance zero, relative and percentage value calculation, manual-mode memory, logging-mode memory,
auto power off, backlight (white LED)
Display :5-digit LCD: 7-segment
Digital display: Main display; [50,000] counts
Sub-display ; [50,000] counts
Bar graph display: 51-segment
Polarity indicator: “-” appears automatically when the polarity is negative
Overrange indicator: “OL”
Low-battery indicator: ” appears at or below the minimum operating voltage.
Measuring Rate :6 times/sec (Frequency: 1 time/sec, Capacitance: max. 0.03 times/sec (50mF), Resistance: 4 times/sec)
Bar graph display: 15 times/sec
Operating Temp. and Humidity
:-20 to 55°C; 80% RH or less (no condensation) 40 to 55°C: 70% RH or less
Storage Temp. and Humidity
:-40 to 70°C; 70% RH or less (no condensation)
Temperature Coefficient
:
Add the accuracy 0.05/°C to the basic accuracy at a temperature within -20 to 18°C and 28 to 55°C. For continuous
measurements, add 1 digit/°C for DC voltage (DCV) and DC current (DCA). (Add 3 digits/°C for 50mV, 5A, and 10A ranges)
Power Supply :Four AA (R6) dry cells
Battery Life :Approx. 120 hours (for continuous DC voltage measurement with alkaline cells)
Withstanding Voltage :6.88kV for 5 seconds (between input terminals and casing)
Dimensions :Approx. 90(W) x 192(H) x 49(D) mm
Weight :Approx. 560g (including batteries)
Compliance with Standards
:Safety EN61010-1, EN61010-2-030, EN61010-031, 1000V CAT#, 600V CAT$, pollution level 2, indoor, 2000m max. above sea level
UL 61010-1, CAN/CSA-C22.2 No. 61010-1,
UL 61010-031, CAN/CSA-C22.2 No. 61010-031
EMC: EN61326-1 Class B, EN55011 Class B Group 1, EN61326-2-2
Standard Accessories :AA (R6) dry cells: 4, Test lead set (98015): 1, Fuse (installed) 440mA/1000V and 10A/1000V, Instruction manual: 1
+
--
Name Model
TY710
Digital Multimeter
TY720
Optional Accessories
Name Model Specification
DMM communication package 92015 USB communication adapter + USB
communication cable + Application software
Test leads 98073 1000V CAT#, 600V CAT$ Red/black (1 set)
Test leads with Alligator Clip 99014 1000V CAT#, 600V CAT$ Red/black (1 set)
Fuse 99015 440 mA/1000V (1 piece/1 unit)
99016 10 A/1000 V (1 piece/1 unit)
TC-K temperature probe 90050 -50 to 600°C (For liquids)
90051 -50 to 600°C (For liquids)
90055 -20 to 250°C (For surfaces)
90056 -20 to 500°C (For surfaces)
Current clamp probe 96001 For 400A, AC Output: 10mV/A, AC
Carrying case 93029
Hard type (Houses the DMM, the test leads and communication cable)
Test conditions: Temperature and humidity = 23 ± 5°C, 80% RH or less; Accuracy = ± (% rdg + dgt).
Note: A response time is the time required for achieving the accuracy specified for the corresponding range.
AC Current Measurement [RMS] (~A) RMS detection crest factor: 3
Range Resolution
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Voltage Drop Maximum Input
Current
500µA 0.01µA<0.11mV/µA
5000µA 0.1µA 1.5+20 1+20
50mA 0.001mA 1+20 0.75+20 1+30 <4mV/mA
500mA
*
3
0.01mA
5A 0.0001A 1.5+20 1+20
<0.1V/A 10A
10A 0.001A 1.5+20 1+20 2+30 fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: 1 second or less
*3: Maximum testing current at 500mA of range is 440mA.
10 - 20Hz 20Hz - 1kHz 1k - 5kHz
440mA
fuse-protected
AC Voltage Measurement [MEAN] (~A)
Mean-value detection and RMS-value calibration
(sinusoidal wave)
Range Resolution
Accuracy TY720 Voltage Drop Maximum Input
Current
500µA0.01µA<0.11mV/µA
5000µA0.1µA2+20 1.5+20 2+30
50mA 0.001mA <4mV/mA
500mA
*
3
0.01mA
5A 0.0001A 3+20 2+20 4+30 <0.1V/A 10A
10A 0.001A
fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: Approx. second or less
*3: Maximum testing current at 500mA of range is 440mA.
10 - 20Hz 20 - 500Hz 500Hz - 1kHz
440mA
fuse-protected
15+40*
2
1000V DC
1000V rms AC
Range Resolution Accuracy Input Resistance Maximum Input Voltage
TY710,TY720
50mV 0.001mV 0.05+10
500mV 0.01mV 0.02+2 Approx. 100M
2400mV 0.1mV
5V 0.0001V 0.025+5
50V 0.001V 10M
500V 0.01V 0.03+2
1000V 0.1V
NMRR: 80dB or greater for 50/60Hz ± 0.1%At 50mV of range, 70dB or greater for 50/60Hz ± 0.1%
CMRR: 100dB or greater for 50/60Hz(Rs=1k) Response time: 0.3 seconds or less
DC Voltage Measurement( V)
Range Resolution
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Voltage Drop Maximum Input
Current
500µA0.01µA <0.11mV/µA
5000µA 0.1µA2+10 1.5+10
50mA 0.001mA 1.5+10 1+10 1.5+10 <4mV/mA
500mA
*
3
0.01mA
5A 0.0001A 2+10 1.5+10
<0.1V/A 10A
10A 0.001A 2+10 1.5+10 3+10
fuse-protected
Shown above is the accuracy at 5 to 100% of range (10 to 100% for 10A range). Response time: Approx. 2 seconds *3: Maximum testing current for 500mA of range is 440mA.
DC,10 - 20Hz DC,20Hz - 1kHz DC,1k - 5kHz
440mA
fuse-protected
DCA + ACA ( +~) Maximum effective display: 50,000, crest factor: 3
Range Resolution Accuracy
Voltage Drop Maximum Input
TY710,TY720 Current
500µA 0.01µA<0.11mV/µA
5000µA 0.1µA 0.2+5
50mA 0.001mA <4mV/mA
500mA
*
3
0.01mA
5A 0.0001A 0.6+10 <0.1V/A 10A
10A 0.001A 0.6+5 fuse-protected
Response time: 0.3 seconds or less *3: Maximum testing current at 500mA of range is 440mA
440mA
fuse-protected
DC Current Measurement ( A)
Temperature Measurement (TEMP)
Range Resolution
AccuracyTY710,TY720
Input Protection Voltage
-200 - 1372°C 0.1°C1+1.5°C1000V rms
Temperature probe: Type K thermocouple sensor (optional)
Frequency Measurement (Hz)
AC coupling, Maximum effective display: 9999
Range (auto-ranging) Resolution Accuracy TY710,TY720
2.000 - 9.999Hz 0.001Hz
9.00 - 99.99Hz 0.01Hz
90.0 - 999.9Hz 0.1Hz
0.900 - 9.999kHz 0.001kHz
9.00 - 99.99kHz 0.01kHz *
2
*1: At 10 to 100% of input voltage or current range
*2: At 40 to 100% of input voltage or current range
0.02+1*
1
Duty Cycle (%)
Range Resolution
Accuracy TY710,TY720
10 - 90% 1% ± 1%*
1
*1: For input of a square wave with a frequency within 10.00 to
500.0Hz At 40 to 100% of input voltage or current range
Peak Hold Function (PH) TY720 only
Maximum effective display: 5000
Range Accuracy TY720 Response Time
DCV, DCA
±
100 digit >250µs
Capacitance ( ) Maximum effective display:5000
Range Resolution
AccuracyTY710,TY720
Input Protection Voltage
5nF 0.001nF
50nF 0.01nF
500nF 0.1nF
5µF 0.001µF
50µF 0.01µF
500µF 0.1µF2+5
5mF 0.001mF 3+5
50mF 0.01mF
*1: Accuracy after zero calibration
1+5*
1
1000V rms
Diode Test ( )
Range Resolution Accuracy TY710,TY720 Testing Current (Vf = 0.6 V) Open-circuit Voltage Input Protection Voltage
2.4V 0.0001V 1 + 2 Approx. 0.5mA <5V 1000V rms
AC Voltage Measurement [RMS] (~V)
AC coupling, RMS detection, crest factor for 1000V of range: 1.5 ; crest factor for ranges other than 1000V: 3
Input Maximum
Range Resolution Impedance Input Voltage
50mV 0.001mV ———
2+80*
2
0.4+40*
2
5+40*
2
5.5+40*
2
11M<50pF
500mV 0.01mV
5V 0.0001V 1.5+30*
1
2+50*
2
——
50V 0.001V 1+30*
1
1+40*
1
2+70*
2
5+200*
2
500V 0.01V 10M<50pF
1000V 0.1V *
2
*
2
3+30*
2
*
2
*
2
3+30*
2
*1: At 5 to 100% of range *2: At 10 to100% of range CMRR: 80dB or greater for DC to 60Hz(Rs= 1k) Response time: 1 second or less
20Hz -
1kHz
10 -
20Hz
1k -
10kHz
10k -
20kHz
20k -
50kHz
50k -
100kHz
1000V rms AC
1000V DC
0.7+30*
1
0.4+30*
1
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
Input Maximum
Range Resolution Impedance Input Voltage
5V 0.0001V 11M<50pF
50V 0.001V
500V 0.01V
1000V 0.1V *
2
*
2
10M<50pF
*
2
*
2
*1: At 5 to 100% of range *2: At 10 to 100% of range CMRR: 80dB or greater for DC to 60Hz (Rs = 1k) Response time: Approx. 2 seconds
DC,20Hz
-
1kHz
DC,10
-
20Hz
DC,1k
-
10kHz
DC,10k
-
20kHz
DC,20k
-
50kHz
DC,50k
-
100kHz
1000V rms AC
1000V DC
DCV + ACV ( +~)
AC coupling, RMS detection crest factor for 1000V of range: 1.5 ; crest factor for ranges other than 1000 V: 3
Accuracy (Upper: TY710; Lower: TY720; the display of “—” is not specified)
1.5+10*
1
1+10*
1
2+10*
2
——
1.5+10*
1
0.5+10*
1
1+10*
1
2+10*
2
5+20*
2
Resistance Measurement ()
Range Resolution
Accuracy Maximum Testing Open-circuit Input Protection
Current Voltage Voltage
5000.01<1mA
5k0.0001k<0.25mA
50k0.001k<25µA
500k0.01k<2.5µA
5M0.0001M0.5+2 <1.5µA
50M0.001M1+2 <0.13µA
*1: Accuracy after zero calibration Response time: 1 second or less for 500 to 500k, 5 seconds or less for 5M to 50M
TY710 TY720
<2.5V 1000V rms
0.1+2*
1
0.05+2*
1
Low-power Resistance Measurement (LP-)
Maximum effective display: 5000
Range Resolution
Accuracy Maximum Testing Open-circuit Input Protection
Current Voltage Voltage
5k0.001k<10µA
50k0.01k0.2+3 <1.0µA
500k0.1k<0.6µA
5M0.001M1+3 <0.05µA
TY720
<0.7V 1000V rms
Range Resolution Continuity Beeper TY710, TY720 Testing Current Open-circuit Voltage
Input Protection Voltage
5000.1Buzzer sounds at 100 ± 50 or less. Approx. 0.5mA <5V 1000V rms
Continuity Check ( )
Maximum effective display: 5000
1000V rms AC
1000V DC
AC Voltage Measurement [MEAN] (~V)
AC coupling, Mean-value detection and RMS-value calibration (sinusoidal wave)
Range Resolution
Accuracy Input Maximum
TY720 Impedance Input Voltage
50mV 0.001mV 4+80*
2
1.5+30*
2
5+30*
2
500mV 0.01mV 11M<50pF
5V 0.0001V
50V 0.001V
10M<50pF
500V 0.01V
1000V 0.1V *
2
*
2
*
2
*1: At 5 to 100% of range *2: At 10 to 100% of range CMRR: 80dB or greater for DC to 60Hz (Rs= 1k) Response time: 1 second or less
10 - 20Hz
2+30*
1
20 - 500Hz 500 - 1kHz
1+30*
1
3+30*
1
5
Series
Safe design and supports various maintenance applications.
TY530
TY520
Model
3.5
digits
3.5
digits
6000
count
6000
count
USB
(TY530)
USB
(TY530)
Terminal
shutters
Terminal
shutters 0.09
%
(DCV)
0.09
%
(DCV)
RMS
RMS
Sensor
Sensor LPF
LPF
Maximum Reliability and Safety
Safe Design
Closed Case Calibration
Reliability
High accuracy and safety
Accuracy: 0.09% rdg + 2 dgt (DC voltage)
True RMS measurement
Only TY530 can switch RMS and mean detection.
Conforms to EN61010-1 safety standard.
Shutters prevent erroneous insertion of test leads into current
measurement terminals (terminal shutters).
Conforms to overvoltage category 1000 V AC/DC, CAT and 600 V
AC/DC, CAT IV.
If the function is switched to other than current measurement while a test
lead remains inserted in a current measurement terminal, the fuse built into
the DMM can not protect the circuits. The terminal shutters prevent such
accidental errors.
The TY series, simply performing special operations via front panel allows
for quick and reliable adjustment. In addition, the series allows for one-
touch adjustment of AC voltage- and AC current-to-frequency
characteristics. The user calibration function leads to improved operation
efficiency and cost reduction.
• External standard instrument required for calibration.
The DMM can directly read the various sensor output signals (mV DC/AC)
at any scaling. The units can be changed (16 units are available).
Output signal and scaled value are simultaneously displayed.
User calibration function
The optional communication package*
1
sold separately (Model 92015)
allows you to connect to a PC for transmitting large amounts of data that
cannot be saved in the DMM internal memory.
You can transmit the saved data from the internal memory to a PC and
process it using application software or spreadsheet software (Excel*
2
) for
data management.
For details of the application software, refer to page 7.
Real-time measurement
*1 Communication cable and application software are included.
*2 Excel is a registered trademark of Microsoft Corporation in the United States.
*3 The communication cable employs an infrared system, so the device is
electrically insolated.
Data Storage MethodDirect reading of various sensor output signals
Two memory modes (TY530 only)
Selectable from 2 types of memory mode to suit field needs.
SAVE-mode memory
A mode for manually saving any data
Logging-mode memory
A mode for automatically saving data at a specified interval
Model
TY530
SAVE-mode memory
100
Logging-mode memory
1600
Memory capacity
Power
supply
AC/DC clamp-on probe (Model 96095)
Reads maximum 60 A when used with the TY500 series.
Sub-display
Sensor output signal
Main display
Scaled value
Small motor
6
TY500 General Specifications
Performance
Model and Specification Code
Name Model
TY520
Digital Multimeter
TY530
Optional Accessories
Test conditions: Temperature and humidity = 23 ± 5°C, 80% RH or less; Accuracy = ± (% rdg + dgt).
Note: A response time is the time required for achieving the accuracy specified for the corresponding range.
Measurement Functions
:DC Voltage, AC voltage, DC current, AC current, resistance, frequency, temperature, capacitance, continuity check,
diode test For AC voltage/current, RMS/MEAN detection can be switched (TY530 only).
Low-pass filter can be switched on/off
Additional Functions :Data hold/auto hold/range hold, maximum/minimum/average values (TY530 only), resistance, relative and percentage
value calculation, memory function (TY530 only), communication function (TY530 only), logging-mode memory (TY530
only), auto power off, backlight
Display :3.5-digit LCD: 7-segment
Digital display: [6000] counts
Bar graph display: 31-segment
Polarity indicator: “-” appears when the polarity is negative
Overrange indicator: “OL”
Low-battery indicator: ” appears at or below the minimum operating voltage.
Measuring Rate :5 times/sec (Frequency: 1 time/sec, Capacitance: max. 0.14 times/sec (1000µF), Resistance: 2.5 times/sec,
Temperature: 0.7 times/sec), Bar graph display: 25 times/sec (DC voltage, diode test: 5 times/sec)
Operating Temp. and Humidity
:-10 to 55°C; 80% RH or less (no condensation) 40 to 55°C; 70% RH or less
Storage Temp. and Humidity
:-30 to 70°C; 70% RH or less (no condensation)
Temperature Coefficient
:Add the accuracy 0.1/°C to the basic accuracy at a temperature within -10 to 18°C and 28 to 55°C.
Power Supply :Four AA (R6) dry cells
Battery Life :Approx. 300 hours (for continuous DC voltage measurement with alkaline cells)
Withstanding Voltage :6.88kV for 5 seconds (between input terminals and casing)
Dimensions :Approx. 90(W) x 192(H) x 49(D) mm
Weight :Approx. 570g (including batteries)
Compliance with Standards
:Safety EN61010-1, EN61010-2-030, EN61010-031, 1000V CAT#, 600V CAT$, pollution level 2,
2000m max. above sea level
UL 61010-1, CAN/CSA-C22.2 No. 61010-1,
UL 61010-031, CAN/CSA-C22.2 No. 61010-031
EMC: EN61326-1 Class B, EN55011 Class B Group 1, EN61326-2-1
Standard Accessories :AA (R6) dry cells: 4, Test lead set (98015): 1, Fuse (installed) 440mA/1000V and 10A/1000V, Instruction manual: 1
+
--
Name Model Specification
DMM communication package 92015 USB communication adapter + USB
communication cable + Application software
Communication package for printer 97016 Printer adapter + Printer cable
Test leads 98073 1000V CAT #, 600V CAT $ Red/black (1 set)
Test leads with Alligator Clip 99014 1000V CAT #, 600V CAT $ Red/black (1 set)
Fuse 99015 440mA/1000V (1 piece/1 unit)
99016 10A/1000V (1 piece/1 unit)
TC-K temperature probe 90050 -50 to 600°C (For liquids)
TC-K temperature probe 90051 -50 to 600°C (For liquids)
90055 -20 to 250°C (For surfaces)
90056 -20 to 500°C (For surfaces)
Current clamp probe 96001 For 400A,AC Output: 10mV/A, AC
96030 200A,AC
96031 500A,AC
96032 700A,AC
96033 50A,AC
96034 3000A,AC
96035 3000A,AC
Carrying case 93029
Hard type (Houses the DMM, the test leads and communication cable)
Range Resolution Accuracy Testing Current (Vf=0.6V) Open-circuit Voltage Input Protection Voltage
2V 0.001V 1+2 Approx. 0.5mA <3.5V 1000V rms
Range Resolution Accuracy
Testing Current (Vf=0.6V)
Open-circuit Voltage
Input Protection Voltage
6000.1Buzzer sounds at 50+30 or less Approx. 1.2mA <3.5V 1000V rms
Range (auto-ranging) Resolution Accuracy Input Voltage Range
10.00 - 99.99Hz 0.01Hz 0.2 - 600V rms
90.0 - 999.9Hz 0.1Hz 0.02+1
0.900 - 9.999Hz 0.001kHz 0.4 - 600V rms
9.00 - 99.99kHz 0.01kHz 0.8 - 100V rms
Frequency Measurement (Hz) AC coupling, Maximum effective display: 9999
Diode Test( )
Continuity Check( )
Range Resolution Accuracy
Input Protection Voltage
-50 - 600°C 0.1°C 2+2°C 1000V rms
Temperature probe: Type K thermocouple sensor (optional)
Temperature Measurement (TEMP)
2+5 1000V rms
3+5
Range Resolution Accuracy
Input Protection Voltage
10nF 0.01nF 2+10*
1
100nF 0.1nF
1µF 0.001µF
10µF 0.01µF
100µF 0.1µF
1000µF1µF
*1: Accuracy after zero calibration for 10nF to 1µF range.
Capacitance( )
AC Current Measurement (~A) RMS detection crest factor: 3
Range Resolution Accuracy
Voltage Drop Maximum Input
Current
600µA 0.1µA<0.12mV/µA
6000µA1µA 440mA
60mA 0.01mA 0.75+5 1.5+5 <3.3mV/mA
fuse-protected
600mA
0.1mA
6A 0.001A <0.1V/A 10A
10A 0.01A fuse-protected
Shown above is the accuracy at 5 to 100% of range (2 to 10A for 10A range). Response time: 3 seconds or less
Add accuracy = ± (2% of reading + 2% of F.S.), except for sinusoidal wave. 4 counts or less is corrected to 0.
50/60Hz 40Hz - 1kHz
AC Voltage Measurement (~V)
AC coupling, RMS detection (TY530, TY520) crest factor: 3/mean-value detection (TY530 only) sinusoidal wave
Range Resolution Accuracy
Input Impedance Maximum Input
Voltage
600mV 0.1mV 10M, <200pF
6V 0.001V 1.5+5 11M, <50pF 1000V rms AC
60V 0.01V 0.5+5 1+5 1000V DC
600V 0.1V 10M, <50pF
1000V 1V -
Shown above is the accuracy at 5 to 100% of range (200 to 1000V for 1000V range, peak 1500V or less). Response time: 2 seconds or less
Add accuracy = ±(2% of reading + 2% of F.S.), except for sinusoidal wave. CMRR: 60dB or greater for DC to 60Hz (Rs = 1k). 4 counts or less
is corrected to 0.
50/60Hz 40-500Hz 500Hz - 1kHz
Range Resolution Accuracy Input Resistance Maximum Input Voltage
TY520, TY530
600mV 0.1mV 10M
6V 0.001V 0.09+2 11M
1000V DC
60V 0.01V 1000V rms AC
600V 0.1V 10M
1000V 1V 0.15+2
NMRR: 60dB or greater for 50/60Hz ± 0.1%
CMRR: 120dB or greater for 50/60Hz (Rs = 1k) Response time: 1 second or less
DC Voltage Measurement( V)
Range Resolution Accuracy
Voltage Drop Maximum Input
Current
600µA 0.1µA<0.12mV/µA
6000µA1µA 0.2+2
60mA 0.01mA <3.3mV/mA
600mA
0.1mA
6A 0.001A 0.5+5 <0.1V/A 10A
10A 0.01A fuse-protected
Maximum testing current at 600mA of range is 440mA. Response time: 1 second or less.
440mA
fuse-protected
DC Current Measurement ( A)
Range Resolution Accuracy
Maximum Testing
Open-circuit
Input Protection
Current Voltage Voltage
6000.1<1.2mA
6k0.001k<110µA
60k0.01k<13µA
600k0.1k<1.3µA
6M0.001M0.5+1
60M0.01M1+2(0-40M)
2+2(40-60M)
*1: Accuracy after zero calibration for 600 to 6k range. Response time: 2 seconds or less for 600 to 600k, 10
seconds or less for 6M to 60M.
<1.3V 1000V rms
0.4+1
*
1
<130nA
<3.5V
Resistance Measurement ()
Features
A compact, light, and portable device with 12-mm caliber useful for tangled wiring.
When used with this probe*1, the DMM can measure and display current (which it otherwise cannot do by
itself). The TY500 series can directly read up to 60 A when used with the probe (in sensor mode).
Specifi cations
Model 96095
Diameter of measurable conductor 12 mm max.
Basic
performance
Current to measure Output voltage Accuracy (at 23ºC ± 5ºC)
AC 0.1 to 130 A Output: 10 mV/A AC (AC 1 to 1300 mVrms) 50/60 Hz 40 Hz to 1 kHz
1.2%+0.4 mV 2.5%+0.4 mV
DC 0 to ± 180 A Output: DC10 mV/A (DC 0 to ± 1800 mV) 1.2% + 0.4 mV
General specifi cations
Operating temperature and humidity –10 to 55ºC, 80%RH or less (no condensation)
Storage temperature and humidity –30 to 70ºC, 85%RH or less (no condensation)
Power supply
AAA alkaline cell × 2
Power alert: LED light on at 2.2 V ± 0.2 V
Auto power off at 1.9 V ± 0.2 V
Battery life Approx. 35 hours (continuous) (until LED light on)*2
Dimensions and weight
127(L) × 42(W) × 22(D) mm
Cable length: 1200 mm
Weight: Approx. 140 g (including cells)
Safety standard EN61010-1: CAT III 300V, pollution degree 2, operation at maximum altitude of 2,000 m, EN61010-2-032
EN61326-1: Class B, EN61326-2-2, EN55011 Class B Group 1
Accessories Soft carrying case (93040), Battery, User’s manual
*1 Readings must be converted when used with the DMM.
*2 After the battery alert, approx. 5 hours remain to automatic power-off.
Accessory AC/DC clamp-on probe (Model 96095)
NEW
7
Communication Functions and Application Software Allow Analyses and
Management of Measurement Data
Data management by dedicated application software
Optional Accessories*
Data layout on
Excel spreadsheet
Data saved in the DMM can be managed by the dedicated
application software (Model 92015).
Saved data can be transmitted from the internal memory to a PC.
Data collected in SAVE-memory mode or logging-memory mode
Measurements by the DMM can be monitored on a PC in real
time.
Large amounts of data that cannot be saved in the DMM internal
memory can be transmitted to a PC in real time.
Data can be written to an Excel* spreadsheet.
Maximum number of real-time data transmission: 32767
Measurement data can be laid out in an Excel spreadsheet.
Graphs can be automatically created on a spreadsheet.
* Excel is a registered trademark of Microsoft Corporation in the United States.
Specifications
Communication cable
Communication cable: IR communication adapter, USB
communication cable: 1
Cable length: 2 m
Interface: USB 1.1
Supported models: TY710, TY720, TY530
Application software
System requirements of PC
Operating system: WindowsXP/Vista(*)/7
CPU: Pentium 133 MHz or higher
Memory: 64 MB or larger
Storage device: Hard disk with 10 MB or more free space
CD-ROM drive: 1
Excel: Excel2003 or later (*)
Contents: CD-ROM software: 1
Communication cable
(communication adapter included): 1
User’s manual
Example of document windows in DMM application software
Data layout example on Excel spreadsheet
TY720
PC
USB
communication
Enlarged view
92015 Communications Package Specifications
92015 DMM Communication Package
* Windows and Excel is a registered trademark of Microsoft Corporation in the United States.
* For TY710, TY720, and TY530 only
Item Model Specifi cation
DMM communication
package 92015 USB communication cable (adaptor
included), application software
8
Low-cost Handheld DMM
• Compact size, ideal for carrying
• Large display for easy viewing
• Safe design allows measurement in excess of 20 A
(excluding 73204)
• Special model for voltage measurement (73204)
• Simple auto hold function
• Capacitors can be checked (73202/73203)
General Specifications
73201 / 02 / 03 / 04
:Auto hold, overvoltage and current warning
:Digital display: 4300-count digital reading
:Digital display: Approx. 2 times/sec
:0 to 50˚C; 80% RH or less at 0˚C to 40˚C, or 70% RH or less at
40˚C to 50˚C (no condensation)
:-20˚C to 60˚C, 70 RH or less (no condensation)
:Add accuracy x 0.1/ ˚C to the basic accuracy at a temperature
within 0˚C to 18˚C and 28˚C to 50˚C
:3.7 kV AC for 1 minute
(between input terminals and casing, for 73201,73202, 73203)
5.55 kV AC for 1 minute
(between input terminals and casing, for 73204)
:Two AAA (LR03 or R03) dry cells
:Approx. 600 hours
(for continuous DC voltage measurement with alkaline cells)
:The power is automatically turned off when no operation is made
for approx. 20 minutes (can be disabled). N/A for 73204
:74 (W) x 155 (H) x 31 (D) mm
:Approx. 240 g (including batteries)
:User's manual: 1
Test lead set (RD031): 1
AAA (LR03/R03) dry cells (built in): 2
Spare fuse F05 (500 mA/250 V, not included for 73204): 1
Spare fuse F02 (15 A/250 V, not included for 73204): 1
Additional Functions
Display
Measuring Rate
Operating Temp.and Humidity
Storage Temp.and Humidity
Temperature Coefficient
Withstanding Voltage
Power Supply
Battery Life
Auto Power Off
Dimensions
Weight
Standard Accessories
Series
73201
73202
73203
73204
Model
3.5
digits
3.5
digits
4300
count
4300
count
Optional Accessories
Name Model Specifi cation
Fuse F05 500 mA/600 V
F02 15 A/600 V
Test leads RD031 Red / black (1 set)
Carrying case (hard) B9646GB Houses the DMM and test leads
Rubber case 93007 For DMM
Performance
Test conditions: Temperature and humidity = 23˚C ± 5˚C, 80% RH or less;Accuracy = ±(% of reading + digits).
Note: Response time is the time required for achieving accuracy specifi ed for the corresponding range.
DC Voltage Measurement ( V)
Range Accuracy Input Resistance Maximum Input
Voltage73201 73202/04 73203
400.0 mV 0.5% + 1
0.5% + 1 0.3% + 1
>100 M
600 V
4.000 V 11 M
40.00 V
0.75% + 1 10 M
400.0 V
600 V
Response time: 1.5 seconds or less for 400 mV range, 1 seconds or less for all other ranges
AC Voltage Measurement ( V) Mean-value detection and RMS-value calibration
Range Accuracy Input Resistance Maximum Input
Voltage73201 73202 73203/04
4.000 V
1% + 5 0.75% + 5
>11 M, <50 pF
600 Vrms
40.00 V
>10 M, <50 pF400.0 V
600 V
Response time: 2 seconds or less
DC Current Measurement ( A)
Not available with 73204
Range Accuracy Voltage Drop Maximum Input Current
73201 73202 73203
400.0 µA *1
1% + 2
<0.17 mV/µA 400 mA
(500 mA/600 V fuse-protected)
4000 µA
40.00 mA *1
<3 mV/mA
400.0 mA
4.000 A 2% + 2 <0.04 V/A 10 A
(15 A/600 V fuse-protected)10.00 A *2
*1: Drift in the least signifi cant digit may occur.
*2: Measurement of 11 to 20 A can be performed within 30 seconds. A warning buzzer sounds when 30 seconds have passed.
Response time: 1 second or less
AC Current Measurement ( A)
Not available with 73204 Mean-value detection and RMS-value calibration
Range Accuracy (40 - 500 Hz) Voltage Drop Maximum Input Current
73201 73202 73203
400.0 µA *1 2% + 20 <0.17 mV/µA 400 mA
(500 mA/600 V fuse-protected)
4000 µA 2% + 5
40.00 mA *1 2% + 20 <3 mV/mA
400.0 mA 2% + 5
4.000 A 2.5% + 20 <0.04 V/A 10 A
(15 A/600 V fuse-protected)10.00 A *2
*1: Drift in the least signifi cant digit may occur.
*2: Measurement of 11 to 20 A can be performed within 30 seconds. A warning buzzer sounds when 30 seconds have passed.
Response time: 2 second or less
Resistance Measurement ( )
Range Accuracy Maximum Testing
Current
Open-circuit
Voltage
Input Protection
Voltage73201 to 73204
400.0 0.75% + 2 <1 mA <3.4 V
600 V
4.000 k
0.75% + 1
<0.5 mA <1.0 V
40.00 k<70 µA
<0.7 V
400.0 k<7 µA
4.000 M2% + 1 <0.7 µA
40.00 M5% + 2 <70 µA
Response time: 1 second or less for 400 k range or less, 5 seconds or less for 4 M range, 15 seconds or less for 40 M range
Continuity Check ( )
Range Continuity Beeper Open-circuit Voltage Input Protection Voltage
73201 to 73204
400.0 Buzzer sounds at 50 ± 20 or less <3.4 V 600 V
Response time: 0.2 second or less (buzzer response)
Diode Test ( )
Range Accuracy Open-circuit Voltage Input Protection Voltage
73201 to 73204
2.00 V 1% + 1 (testing current 1 mA or less) <3.4 V 600 V
Response time: 1 second or less
Capacitor Check ( )
Range Accuracy Input Protection
73201/04 73202 73203
20.00 nF
Not available 2% + 5, typical
(20 nF range: Accuracy after zero calibration)
500 mA/250 V
fuse-protected
200.0 nF
2.000 µF
20.00 µF
200.0 µF
Response time: 1 second or less
9
Pocket DMM with Superb Portability
73101
Model
General Specifications
73101
:Auto hold
:Digital display: 4300-count digital reading
:Digital display: Approx. 2 times/sec
:0 to 50˚C; 80% RH or less (no condensation)
:-10˚C to 60˚C, 70 RH or less (no condensation)
:Two LR44 dry cell
:Approx. 200 hours
(for continuous DC voltage measurement)
:The power is automatically turned off when no operation is made
for approx. 20 minutes (can be disabled).
:76 (W) x 117 (H) x 18 (D) mm
:Approx. 110 g (including batteries)
:Safety EN61010-2-30, EN61010-031, EN61010-1
(300 V, CAT II; contamination level 2)
EMC EN61326-1, EN61326-2-2
EN55011 Class B Group 1
:User's manual: 1
LR44 dry cells: 2
Additional Functions
Display
Measuring Rate
Operating Temp.and Humidity
Storage Temp.and Humidity
Power Supply
Battery Life
Auto Power Off
Dimensions
Weight
Compliance with Standards
Standard Accessories
3.5
digits
3.5
digits
4300
count
4300
count
Performance
Test conditions: Temperature and humidity = 23˚C ± 5˚C, 80% RH or less;Accuracy = ±(% of reading + digits).
DC Voltage Measurement ( V)
Range Accuracy Input Resistance Maximum Input Voltage
400.0 mV 1.2 + 2 >100 M
600 V DC
4.000 V 0.7 + 1 11 M
40.00 V
1.2 + 1 10 M
400.0 V
600 V
AC Voltage Measurement ( V) Mean-value detection and RMS-value calibration
Range Accuracy Input Resistance Maximum Input Voltage
4.000 V
2.0 + 5
>11 M, <50 pF
600 Vrms
40.00 V
>10 M, <50 pF400.0 V
600 V
Resistance Measurement ( )
Range Accuracy Maximum Testing Current Open-circuit Voltage Input Protection Voltage
400.0
1.2 + 2
<1 mA <3.4 V
600 V
4.000 k<0.5 mA <1.0 V
40.00 k<70 µA
<0.7 V
400.0 k<7 µA
4.000 M2.0 + 3 <0.7 µA
40.00 M5.0 + 3 <70 nA
Continuity Check ( )
Range Continuity Beeper Open-circuit Voltage Input Protection Voltage
400.0 50 ± 20 <3.4 V 600 V
Diode Test ( )
Range Accuracy Testing Current Open-circuit Voltage Input Protection Voltage
2.00 V 1.5 + 1 <1.0 mA <3.4 V 600 V
10
Optional Accessories and Spare Parts
Name Model Specifi cation Appl i cab le DMM Models Appearance
DMM communication
package 92015 USB communication adapter + USB communication cable +
Applicatio n software TY700 s eries
TY530
Test leads 98073 1000V CA T.II I 600V CAT.IV Red/black (1set) All m od els exce pt 73 101
RD031 L-plug, Red/black (1set) 732 series
Test leads with Alligator Clip 990141000V CAT.III 600 V C AT.IV Red/black (1set) All mod els exce pt 73101
Alli ga t or clips B9646HF Red/black(1set) All models
Fuse
F02 1 5A/250V (3pcs/1set) 73201/73202/73203
F05 500mA/250V(3pcs/1set)
99015 440mA/1000V(1pc/1set) TY700/TY500 ser ies
99016 10A/1000V(1pc/1set)
Rubber case 93007 732 series
Carrying case B9646GB Hard case
93029 Hard case (Houses the DMM, the test leads and
communication cable) TY700/TY 500 s eries
Temperature (thermocouple
typ e K) p rob e
90050 -50°C to 600°C(for liquid)
TY700/TY 500 s eries
90051 -50°C to 600°C(for liquid)
90055 -20°C to 250°C(for surface)
90056 -20°C to 500°C(for surface)
Current clamp probe
96001 For 400A AC; 10mV/A AC output All models e xcept 73101
(with TY500 series upto 60A
can b e read directly)
96095 For 130A AC/180A DC; 10m V/A AC/DC output
Current Clamp Probe:TY700/TY500 series (Direct reading is possible for TY500 series)
Name 96036 96033 96030 96031 96032 96034 96035
Current Clamp Probe
*9V battery-operated.
Can be AC adaptor-operated (optional)
Measurable
Conductor Diameter dia. 40mm dia. 18mm dia. 30mm dia. 30mm dia. 65mm dia. 65 x 1 00mm dia. 170mm
Measurement Range 2A,AC 50A,AC 200A,AC 500A,AC 700A,AC
(1000A for 5 minu tes) 1000A , 2000A,
3000A, AC 300 - 3000A,AC
Output Voltage 50mV,AC 500mV,AC 500mV,AC 500mV,AC 250mV,AC 500mV,AC 500mV,AC
Accuracy *varies accordin g
to input/Amplitude
±0.5% of rdg ±0. 5% of rdg ±0.5% of rdg ±0.5% of rdg ±1.0% of rdg ±1.0% of rdg ±1.0% of rdg
Freq ue ncy Ra nge 2 0H z - 5kH z 20Hz - 20kHz 20Hz - 20kHz 20Hz - 5kHz 45Hz - 66Hz 30Hz - 1.5kHz 10Hz - 20kHz
Maximum Circuit
Voltage 50V,AC 300V,AC 600V,A C 600 V,A C 600 V,AC 600V,AC 1000V,AC ( pri)
Note:UseACvoltagerangeoftheDMM.
Note:NeedtocovertthemeterreadingexceptTY500series.
Basic Usage Digital Multimeters
Measurement probe
Measuring a Voltage
and Resistance
Measurement probe Measurement probe
Measuring a Current
The circuit needs
to be open for connection.
Checking the Polarity
of a Diode
Current Current
AK
Anode Cathode
Voltage/Resistance Measurement
The COM terminal and V/ terminal are used. To
measure a voltage, set the dial to voltage measurement.
To measure a resistance, set the dial to resistance
measurement. Some DMM models can also display the
frequency and calculated decibel value at the same time
when measuring an AC voltage.
During resistance measurement, it is
possible to switch the function to
checking of the continuity of the
measured circuit.
Current Measurement
The COM terminal, and A, µA or mA terminal are used.
Some models have shutters for preventing erroneous
insertion into the current terminals and allow a contact of
a lead to a current terminal only when the dial is set to
current measurement. For these models, you cannot set
the dial to voltage measurement while
a lead is left inserted into a current
terminal. This feature provides greater
safety.
Diode Test
A current flows through a diode when the power supply is
connected as (1) below, while, almost no current flows
when the power supply is connected as (2). The diode
test function applies an adequate forward voltage across
a diode to make a constant current flow and measures
the voltage drop in the forward direction
to determine the forward and reverse
directions of the diode.
DC power
supply DC power
supply DC power
supply
AC power
supply AC power
supply
Load
Load (1)(2)
98073 99014
Represented by:
MIK-EM18
YOKOGAWA CORPORATION OF AMERICA
2 Dart Road, Newnan, GA. 30265-1094 U.S.A.
Phone: +1-770-253-7000 Facsimile: +1-770-254-0928
YOKOGAWA EUROPE B. V.
Euroweg 2 3825 HD Amersfoort, THE NETHERLANDS
Phone: +31-88-4641000 Facsimile: +31-88-4641111
YOKOGAWA ENGINEERING ASIA PTE. LTD.
5 Bedok South Road, Singapore 469270 SINGAPORE
Phone: +65-6241-9933 Facsimile: +65-6241-2606
YOKOGAWA AMERICA DO SUL LTDA.
Praca Acapulco, 31-Santo Amaro, Sao Paulo/SP, BRAZIL CEP-04675-190
Phone: +55-11-5681-2400
Facsimile: +55-11-5681-4434
YOKOGAWA ELECTRIC KOREA CO., LTD.
C&M Sales Seoul Office
1301-1305, 13rd floor, Kolon digital tower, 106-1,
Yangpyongdong-5Ga, Yeongdeungpo-Gu, Seoul,
150-105, Korea
Phone: +82-2-2628-3810 Facsimile: +82-2-2628-3899
YOKOGAWA AUSTRALIA PTY. LTD.
Tower A/112-118 Talavera Road Macquarie Park, NSW 2113
Australia
Phone: +61-2-8870-1100 Facsimile: +61-2-8870-1111
YOKOGAWA INDIA LTD.
Plot No. 96. Electronic City Complex, Hosur Road, Bangalore 560100, INDIA
Phone: +91-80-4158-6000 Facsimile: +91-80-2852-1441
YOKOGAWA SHANGHAI TRADING CO., LTD.
4F Tower D, Cartelo Crocodile Building, No.568 West Tianshan
Road, Shanghai, CHINA
Phone: +86-21-6239-6363 Facsimile: +86-21-6880-4987
YOKOGAWA MIDDLE EAST B. S. C.(C)
P.O.BOX 10070, Manama, Building 577, Road 2516,
Busaiteen 225, Muharraq, BAHRAIN
Phone: +973-17-358100 Facsimile: +973-17-336100
YOKOGAWA ELECTRIC CIS LTD.
Grokholskiy per. 13, Build. 2, 4th Floor, 129090, Moscow
RUSSIAN FEDERATION
Phone: +7-495-737-7868 Facsimile: +7-495-737-7869
Tachihi Bld. No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan
Phone: +81-42-534-1413 Facsimile: +81-42-534-1438
YOKOGAWA METERS & INSTRUMENTS CORPORATION
International Sales Dept.
World Wide Web site at
http://tmi.yokogawa.com
NOTICE
Before using the product, read the instruction manual
carefully to ensure proper and safe operation
Subject to change without notice. [Ed: 12/b] Printed in Japan: sep. 2013(C)/ (KP)
All Rights Reserved. Copyright © 2009, Yokogawa Meters & Instruments Corporation.