1GHz SPECTRUM ANALYZER
SPECIFICATIONS
GSP-810 (150kHz~1GHz)
FEATURES
* Options:Tracking Generator,
Romote Control
Power Meter,
Software
* RS-232C Interface and Software to get Trace
from GSP-810 to PC
* 9 Memories of Save/Recall
* Functions: Max. Hold, Average(2 ~ 32 Traces),
Freeze, Peak Search, Marker to Center Functions
* Two Markers for Absolute and Relative
Measurement
@30kHz, -100dBm Typical @220kHz RBW
* Good Noise Floor Performance: -95dBm
* Reference Level Range : -30dBm ~ +20dBm
* RBW : 3k, 30k, 220k, 4MHz
* High Resolution of Span to Measure the More
Zero, 2kHz~100MHz/divDetailed Signal :
10ppm: StabilityFrequency High *
* Fully Digital Phase Locked Loop Technique
Design
* Frequency Range : 150kHz ~ 1GHz
* High Input Protection Level : +30dBm,
25VDC
Demodulator WB FM, 120kHz deviation
MB FM, 75kHz deviation
NB FM, 30kHz deviation
AM
Outputs : Internal speaker, 3.5mm stereo jack, wired for mono operation
Calibrate Signal 80MHz, -30dBm
POWER SOURCE
INTERFACE
ACCESSORIES :
DIMENSIONS & WEIGHT
RS-232C standard & remote display software
(The software will be downloaded from GW Website.)
AC 100V/120V/220V/230V 10%, 50/60Hz
User manual x 1, Power cord x 1
310(W) x 150(H) x 455(D) mm, Approx. 8.5kg
FREQUENCY
Frequency Range
Aging Rate
Span Range
Phase Noise
Frequency Resolution
Frequency Display
Frequency Control
150kHz~1GHz
10ppm, 0~50 C, 2ppm/yr
Zero, 2kHz~100MHz/div in 1-2-5 sequence
-77dBc/Hz @ 1GHz 30kHz offset
1kHz C.F. entry, 40Hz sweep resolution at 2kHz/div
6 1/2 digit setting
Digital phase locked
RESOLUTION BANDWIDTH
RBW Range
RBW Accuracy
Video Bandwidth Range
AMPLITUDE
Measurement Range
Overload Protection
Reference Level Range
Amplitude Display Range
Amplitude Accuracy
Frequency Flatness
Amplitude Level Linearity
3kHz, 30kHz, 220kHz, 4MHz
15%
1.6kHz/90kHz couple with RBW
-100dBm~+20dBm
+30dBm continuous, 25VDC
-30dBm~+20dBm
75dB
1.5dB typical @ 0dBm, 80MHz
1.5dB over 100MHz, 2.5dB typical over entire band/ 3dB:150kHz~10MHz
1.5dB over 70dB
DYNAMIC RANGE
Average Noise Floor -95dBm @30kHz RBW, -100dBm typical /-75dBm: 150kHz~10MHz
Third Inter-Modulation
Harmonic Distortion
Non-Harmonic Spurious
DISPLAY SYSTEM
Display Device
Display Function
FUNCTIONS
Marker Mode
Number of Markers
Marker Resolution
Marker Accuracy
Memory
Trace
Setup
<-70dBc, @-40dBm input, 2tones, 2MHz apart/
<-45dBc:150kHz~10MHz
<-40dBc, RF input < selected reference
<-60dBc typical down from reference level, average, 5MHz/div
CRT Display, 8 x 10 graticule, 6-inch waveform screen
LCD Display, 4 line x 20 character data screen
Center Frequency Control, Bandwidth, Reference Level, Span Range, Amplitude
Absolute, relative, PK->marker, marker->center
2
0.1dB, 1kHz
0.1dB amplitude accuracy
10 memorise of save/recall
Max. hold, average(2~32 traces), freeze(Hold)
Access parameters
Opt. 01
Opt. 02
Opt. 03
REMOTE CONTROL SOFTWARE
Connecting PC to get the trace and provide the control for setting
Type N female
TRACKING GENERATOR (Factory Installed)
Frequency Range
Power Level Range
Power Level Overload
Return Loss
Readout Resolution
Accuracy
Readout
POWER METER (Factory Installed)
10MHz ~ 2 GHz, usable to 2.7GHz
-20 dBm ~ +23 dBm, usable to +30 dBm
+40 dBm < 10% duty cycle, < 10 mS duration
< 1:1.35 VSWR into 50 ohms, < 1:1.25 typical
0.2 mW, 100 mW scale, 2 W, 1 mW scale;
0.1dB, Log scale
(10% rdg 1digit)
mW or dBm
Frequency Range
Amplitude Resolution
Amplitude Accuracy
Attenuation Accuracy
Amplitude Flatness
Harmonics
Reverse Power
Impedance
Return Loss
Connector
150kHz ~ 1GHz
50 dBm ~ 0dBmAmplitude Range
1 dB
1 dB @ 0 dBm, 80 MHz
1 dB @ 50 MHz
1 dB @ 10 MHz/DIV, 1.5dB@0dB,entire band
< -30 dBc (<-25dBc, 150kHz ~ 10MHz)
< +30 dBm
50 nominal
< 10 dB (VSWR < 2)
+_
+_
+_ +_
ADP-002: adaptor, SMA(J/F) ~ N(P/M) x 2
ATN-100: 10dB attenuator, N(J/F) ~ N(P/M) x 1
GSC-002: Kit box x 1
ADP-001: adaptor, BNC(J/F) ~ N(P/M) x 2
ADP-101: adaptor,BNC(J/F)75 ~BNC(P/M)50 x 2
GTL-304: RF cable assembly(RG223,N(P/M)-N(J/F),300mm)x2
GSC-003: Kit box x 1
GAK-001: termination 50 , N(P/M) x 1
GAK-002: Cap with chain, N(P/M) x 1
GTL-302: RF cable assembly(RG223,N(P/M),300mm)x2
GSC-004: Kit box x 1
GKT-001 General Kit set
GKT-002
GKT-003
GKT-006
ATA-001
GTL-301
BNC Antenna
RF Cable
CATV Kit set
RLB Kit set
EMI Probe Kit set
RG 223 N(P/M), 1000mm
ANT-02: 3cm Loop, H-Field Probe x 1
ANT-01: 6cm Loop, H-Field Probe x 1
ANT-03 6mm Stub tip, E- Field Probe x 1
PR-03: Touch Passive Probe x 1 ADP-02: N(P/M)~SMA(J/F) Adapter x 1
ADP-01: N(P/M)~BNC(J/F) Adapter x 1
Test Lead: BNC(P/M)~BNC(P/M) RF Cable x 1
Test Lead: SMA(P/M)~SMA(P/M) RF Cable x 1
RLB-001 Return Loss Bridge
RLB Frequency Range 10MHz ~ 1GHz
1GHz Spectrum Analyzer
GSP-810
(An additional ADP-001 is needed for fitting GSP spectrum analyzers)
GSC-002
ADP-002
ATN-100
GTL-303
For:GSP-810/827/830
GSC-003
ADP-001
ADP-101
GTL-304
For:GSP-810/827/830
GSC-004
GAK-001
GAK-002
GTL-302
For:GSP-810/827/830
GKT-001 General Kit Set
GKT-002 CATV Kit Set
GKT-003 RLB Kit Set
For:GSP-810/827/830
GKT-006 EMI Probe Kit Set
ANT-02
ANT-01
ANT-03
PR-03
ADP-02
ADP-01
Test Lead: BNC(P)~BNC(P) RF Cable x 1
Test Lead: SMA(P)~SMA(P) RF Cable x 1
10MHz ~ 1GHz
For: GSP-810/827/830
RLB-001 Return Loss Bridge
The GSP-810 Spectrum Analyzer is a digitally synthesized, easy-to-use RF measurement instrument
with a frequency range up to 1GHz. It uses a state-of-the-art design with a phased-locked RF system
that combines measurement quality and performance with an affordable price. For flexibility and versa
tility, the GSP-810 has a number of options and optional accessories such as a factory installed
tracking generator or the extensive probe kit options. The GSP-810 is also designed for rugged use
with a 1-watt input protection. The small size and light weight of the GSP-810 make it easy to carry
and stow anywhere.
The GSP-810 includes free software (Windows 2000/XP compatible) for displaying, saving and
printing results via the standard RS232 interface. Optional software is also available for full PC remote
control at additional cost.
All these functions and features make the GSP-810 an affordable, expandable, and high quality
instrument suitable for a variety of applications.
Note: Need to Collocate the Optional Accessories.
ORDERING INFORMATION
OPTION
OPTIONAL ACCESSORIES
FREE DOWNLOAD
PC Software
Programming Labrary
Remote Display Software
DLL file for Programming Use
GTL-303: RF cable assembly(RD316),SMA(P/M),600mm) x 2
ATA-002
Near Field Probe (An additional ADP-001 & GTL-110 is needed for fitting GSP spectrum analyzers)
GTP-3000
Passive touch Probe Set
PR-03 Passive touch probe, SMA(F) ADP-002 Adaptor, SMA(J/F) ~ N(P/M)
GTL-303RF Cable Assembly, SMA(P/M), 600mm