Datasheet Evaluation Kit EK-H4 For SHTxx Humidity and Temperature Sensors Ready to use Hard and Software Suitable for data-logging experiments Programmable Interface Dew Point and Energy Calculation Multiple Multiplexer boxes with Sensors Product Summary The evaluation kit EK-H4 is made for running Sensirion's humidity and temperature sensors and to display measured values. It is able to communicate with SHT1x, SHT7x, SHT2x, SHT3x, SHTC1 and SHTW2. The measured values will be displayed on the display of the EK-H4. The main purpose of the evaluation kit is to demonstrate the unique performance of Sensirion's humidity and temperature sensors without any own hard or software design work. Together with a computer and a dedicated Viewer software (see sensirion.com/ek-h4) it may also be used as a data-logger for simple experimental measurements. The EK-H4 is connected to the computer via USB interface. The software may record and display humidity and temperature values as well as calculated values of dew point and absolute humidity. It is possible to connect multiple EK-H4 to different USB ports on the same computer and operate them through the same Viewer Software. EK-H4 Material Contents The EK-H4 contains the following components, enclosed in a handy suitcase: EK-H4 (Order no. 1-100661-02) 4 x SHT31 sensors on PCB with connector 4 x SHT21 sensors on PCB with connector. 4 x SHT71 sensors 4 x sensor cables with RJ45 plug and pin-type connector for connecting sensors with multiplexer box of 3m length. EK-H4 Flex (Order no. 1-101068-03) 4 x SHT31 sensors on FPCB 4 x SHTC1 sensors on FPCB 4 x SHTW2 sensors on FPCB 4 x sensor cables with RJ45 plug and FPCB connector for connecting sensors on FPCB with multiplexer box of 3m length. 1 EK-H4 multiplexer box: to connect up to 4 SHTxx Sensors Figure 1 Schematic of the Evaluation Kit EK-H4. The Evaluation Kit works without connection to a computer it only needs a connection to the power socket. Optionally, the EK-H4 can be connected to a computer and operated through the Viewer Software. 1 USB cable for connection to computer. 1 Power supply unit with US, EU, AU adapter EK-H4 Viewer Software is available from Sensirion's Webpage (www.sensirion.com/ek-h4). Device Certification The Evaluation Kit EK-H4 is RoHS and WEEE compliant. www.sensirion.com Version 5 - August 2016 1/8 Datasheet EK-H4 Specification Electrical specification: Table 1 Parameter min 8 typ 9 max Unit Input Voltage 10 V External Supply Input Voltage Ripple 0.5 V Supply Current 170 280 mA Input Voltage 4.75 5 5.25 V USB Supply Supply Current 170 280 mA Range 1.8 3.0 3.6 V 0.1 V Sensor Supply Resolution Voltage Accuracy 2 5 % Output current 40 110 1 mA Temperature Resolution 11 14 Bit Humidity Resolution 8 12 Bit Measurement interval 1 1440 s Communication USB (via Virtual Com Port (VCP)) Operating 0 to 50C (multiplexer box) Temperature -20 to 70C (cables, sensor) DC Power Supply: Pin Diagram EK-H4 Circuit Board 4 sensor channels Display 2 buttons for display modes Micro Controller Unit Debug interface User field connected to MCU USB interface Reset button Power supply Default Configuration After power-up two different voltages ranges (2.1V to 3.6V & 1.8V) can be chosen through the push button on the top of the evaluation kit. Parameter Sensor Supply Voltage Temperature Resolution Humidity Resolution Measurement interval Heater SHT7x OTP reload SHT2x Default value 3.0 V 12 Bit 12 Bit 1s off on off Power Supply Used Connection DC DC and USB USB Power LED EK-H4 Pin Assignment 8 1234 1234 2134 RJ45 Pin Sensor Pin Name Description 1 3 VSS Ground 3 1 SCL Serial clock 5 2 VDD Supply voltage 7 4 SDA Serial data EK-H4 supplied from DC (9V) DC (9V) USB (5V) The EK-H4 may be supplied by DC Power Supply Unit at 9V or USB connection at 5V. Power supply by USB is only possible if USB driver is installed correctly. 1 Table 2 Pin Assignments: Female RJ45 pin diagram on the left - SHTxx on PCB, SHT71 and SHTxx on FPCB sensors on the right of the upper panel. LED Functions + 1 short circuit current www.sensirion.com Pin Name Comment - GND Ground + Positive Voltage 9V Supply LED Power LED RJ45 Connector Version 5 - August 2016 Green LED Orange LED Function LED off denotes low supply voltage (DC < 8V, USB < 4.75V) or not connected. Sensor detected Measurement is in progress 2/8 EK-H4 Users Guide 1 Evaluation Kit Features The Evaluation Kit EK-H4 may be used as a stand-alone device, displaying measured values on the screen on the multiplexer box or - connected to a computer - as a datalogger run by the Sensor-viewer software. 1. Select multiplexer box, power supply cable, sensors with sensor cables as many as needed. 2. Connect sensors via sensor cables to the multiplexer box. 3. Choose appropriate voltage range for the connected sensors 4. Connect multiplexer box via power supply cable to power (voltage supply also may occur via USB cable from a computer). Please note: The specific plug needs to be connected to the power supply cable - the plastic protection cap on cable side must be removed before connecting the plug. 5. As soon as the green LED light is shining, the measurement starts automatically. The stand-alone device displays temperature and humidity values of each connected sensor plus dew-point derived from the two measured values. Besides the above mentioned values the data-logger software displays calculated absolute humidity and power consumption. Furthermore, it is possible to change the acquisition interval (see section 4.7), switch the in-built heater on and activate a fast conversion mode (see section 4.9). Eventually measured data can be exported for data processing. 2 EK-H4 Suitcase Figure 3 Stand-Alone Evaluation Kit. Voltage supply to be set in the range of AC 80 ... 250V. 3.2 Display Information After powering the EK-H4 as standalone measurement tool the screen shows following information: Sensirion Evaluation Kit for SHTxx EKH4 V3.0 Figure 2 content 3 Schematic of the EK-H4 suitcase highlighting its The first screen shows an overview over the connected sensors. The second screen (accessed through the push buttons) shows a detailed information of the connected sensor. In the first row the sensor number, which can be matched with the printed number on the cover, is visible. In the second column the temperature is displayed in C. The third column shows the relative humidity. Not connected sensor ports are displayed with "--.-- " . 1 2 3 4 Stand-Alone Evaluation Kit 3.1 Getting Started Set up and application of the stand-alone device is straight forward: www.sensirion.com Figure 4 Version 5 - August 2016 25.45C --.--C 24.98C 25.71C 45.32%RH --.--%RH 46.27%RH 45.46%RH Overview Screen of the EK-H4 multiplexer box 3/8 Datasheet EK-H4 By pressing the buttons the displayed information can be switched from all sensors to the respective single sensors: Right button: Sensor 1 2 3 4 all 1 Left button: Sensor 4 3 2 1 all 4 The displayed information by choosing only one sensor is as following: Sensor 1: Temperature: Humidity RH: Dew Point: Figure 5 SHT7x 25.45C 45.32% 16.76C Individual screen for each connected sensor If no sensor is connected to the chosen port, the sensor designation is replaced with "no sens". To execute a self test for testing the sensor ports and voltage supply both buttons must be pressed for longer than one second. For a successful self test each port requires a sensor to be connected. If no sensor is connected the corresponding port will show an error. Sens 1: OK Sens 2: OK Sensor Supply EKH4 Supply 3: 4: : : OK OK OK OK By pressing one of the buttons the display information will switch back to the measurement. 4 Figure 6 EK-H4 configuration with connection to read-out software on computer. No additional power supply is needed. 4.3 Starting a Measurement Connect the EK-H4 multiplexer box with a USB cable to the computer and then start the software by doubleclicking on EK-H4_Viewer_VXY_Installer.exe. The connect dialog displays the COM ports to which EK-H4 multiplexer boxes are connected - see Figure 7. After clicking OK, a connection to the multiplexer box is established and the sensors are listed in the Sensor Select and Measurement Data sections - see Figure 8. Data-Logger with Software Using the EK-H4 system as a data-logger requires the installation of the read-out software on a computer and connecting multiplexer box and computer via USB cable. Prior to this a driver installation may be necessary. 4.1 EK-H4 Driver Installation Connect the EK-H4 multiplexer box to the computer via USB cable. EK-H4 uses a Silicon Laboratories CP210x USB to UART bridge controller. If a driver for this controller is not installed yet, a dialog pops up which helps you to search for a driver on the internet. Alternatively, the driver software may be installed from the EK-H4 memory stick. Figure 7 Connect dialog of read-out software. Window in the center displays COM ports with multiplexer boxes connected. 4.2 Read-Out Software Installation 4.4 Multiple EK-H4 Devices EK-H4 read-out software requires the .NET framework 4.5.2 or later. If it is not already installed on the computer, it may be downloaded from EK-H4 read-out software is able to read sensors from various EK-H4 multiplexers. The different systems are selected via connect dialog at start-up or using the menu item Connection. To disconnect EK-H4 systems use the menu item Connection/Disconnect. The connected Evaluation Kits are displayed under the Sensor Select section. Microsoft .net Web Installer Version 4.5.2. To install the EK-H4 software, simply copy the EK-H4 folder from the memory stick to the hard drive of the computer. www.sensirion.com Version 5 - August 2016 4/8 Datasheet EK-H4 4.6 Calculated Values Relative humidity and temperature data are measured by the sensors. From this data dew point and absolute humidity values are derived. 4.7 Settings In the settings section the measurement interval can be chosen. The default configuration is 1 second. 4.8 Data Logging and Data Export Figure 8 EK-H4 software window before measurement starts. Sensors are displayed in different colors. 4.5 Temperature and Humidity Measurement Measurements are started and stopped by the Start/Stop button. The measured data are displayed in the Measurement Data table. The measurements which are displayed in this table can be selected in the context menu which appears when the right mouse button is clicked on the table. Two plots are available, which display the progress of a measurement over time. The measurement can be selected in the list below the plot. The scale of the axes can be altered by pressing the left mouse button on the axis and moving the mouse. The plot can be dragged using the left mouse button. The right mouse button can be used to display a context menu, which offers functions to print or copy the plot, or to set auto-scale options for the axes. The data displayed in the measurement plots is kept in memory. The measured values can be exported to a CSV file using the menu item File/Log to file. The measurement data can be cleared under the menu item View/Clear Graph. 4.9 Heater and Fast Conversion Behind every sensor there are two tick boxes, one to activate the in-built heater and one to activate a fast conversion mode. The heater can be used as a simple functional test. After switching on the heater, the temperature of the sensor should increase, while the humidity signal should drop. Please note that the heater power is different for every SHTxx sensor. Additionally, the temperature elevation that can be achieved through the in-built heater depends on the temperature itself (at higher ambient temperature a smaller temperature increase is achieved), the operating voltage (higher operating voltages lead to higher temperature increases) and on the thermal coupling of the sensor to the environment. The fast conversion mode activates a different measurement command which has a higher noise level. For SHT1x, SHT7x and the SHT2x series a measurement with a lower resolution is started. While for the SHT3x series the repeatability is changed to the low repeatability mode. The default command (tick box not active) issues a measurement command with the smallest noise level (high resolution or high repeatability mode), except for the SHT1x/SHT7x series, where the low resolution mode is active per default. 4.10 Firmware Update Figure 9 Software surface during measurement. Displayed values may be selected via menus at the bottom of the window. www.sensirion.com Starting from firmware version 2.4 (see section 3.2 on how to read out the firmware version) it is possible to update the firmware through the Viewer Software. If you have a EK-H4 box with version below 2.4 you can use the Viewer software Version 2, which is available from our webpage. It does support the functions (Heater, Fast Conversion and Firmware Update Function). Version 5 - August 2016 5/8 Datasheet EK-H4 In Order to open the update window go to the Help Tab and click on "Update Firmware EK-H4 box". To update the firmware perform the following steps: 1. 2. Check the firmware version of the EK-H4 box Download EK-H4 firmware (.hex file) from Sensirion's Webpage (sensirion.com/ek-h4) and store on the desktop Power the EK-H4 with the external power supply Disconnect the EK-H4 that you want to update (go to the connection tab and click on Disconnect, then choose the correct Com Port ) Open the Update window. Go to the Help Tab and click on "Update Firmware EK-H4 box". Choose the correct com Port of the EK-H4 which connects to the EK-H4 that you want to update Select the new EK-H4 firmware (.hex-File) with the Browse" Button from the desktop Press "Update" 3. 4. 5. 6. 7. 8. Error handing: In the case that the EK-H4 was not disconnected prior to opening the update window an error message will be shown, when the com port is selected (step 6). If this happens close the Update window and perform step 4 or unplug the USB cable and plug it in again. 4.11 Firmware compatibility Firmware 2.4 and higher operates with Viewer 3.0 Firmware 2.3 and lower can be operated with the Viewer V2 4.12 Sensor and EK-H4 Properties Information about the sensors such as the serial number is shown behind each sensor in the Measurement Data Section. www.sensirion.com 4.13 License Information The EK-H4 viewer software uses the NPlot charting library, to which the following license applies: ************************************************************************* NPlot - A charting library for .NET Copyright (C) 2003-2006 Matt Howlett and others. All rights reserved. Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met: 1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer. 2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution. 3. Neither the name of NPlot nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission. THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE. Version 5 - August 2016 6/8 Datasheet EK-H4 Revision History Date August 2009 May 2010 May 2015 June 2015 July 2015 August 2016 www.sensirion.com Version 1.0 1.1 2 3 4 5 Page(s) 1-7 2 all 4 1,2 1 Changes Initial release Operating temperature ranges are defined - replacement for storage conditions Added SHT3x, Updated description to Viewer Software 3.0 Updated .net framework instructions Combined EK-H4 and EK-H4 Flex Replaced SHTW1 on FPCB by SHTW2 on FPCB Version 5 - August 2016 7/8 Datasheet EK-H4 Important Notices Warning, Personal Injury Do not use this product as safety or emergency stop devices or in any other application where failure of the product could result in personal injury. Do not use this product for applications other than its intended and authorized use. Before installing, handling, using or servicing this product, please consult the data sheet and application notes. Failure to comply with these instructions could result in death or serious injury. If the Buyer shall purchase or use SENSIRION products for any unintended or unauthorized application, Buyer shall defend, indemnify and hold harmless SENSIRION and its officers, employees, subsidiaries, affiliates and distributors against all claims, costs, damages and expenses, and reasonable attorney fees arising out of, directly or indirectly, any claim of personal injury or death associated with such unintended or unauthorized use, even if SENSIRION shall be allegedly negligent with respect to the design or the manufacture of the product. ESD Precautions The inherent design of this component causes it to be sensitive to electrostatic discharge (ESD). To prevent ESD-induced damage and/or degradation, take customary and statutory ESD precautions when handling this product. See application note "ESD, Latchup and EMC" for more information. Warranty SENSIRION warrants solely to the original purchaser of this product for a period of 12 months (one year) from the date of delivery that this product shall be of the quality, material and workmanship defined in SENSIRION's published specifications of the product. Within such period, if proven to be defective, SENSIRION shall repair and/or replace this product, in SENSIRION's discretion, free of charge to the Buyer, provided that: notice in writing describing the defects shall be given to SENSIRION within fourteen (14) days after their appearance; such defects shall be found, to SENSIRION's reasonable satisfaction, to have arisen from SENSIRION's faulty design, material, or workmanship; the defective product shall be returned to SENSIRION's factory at the Buyer's expense; and the warranty period for any repaired or replaced product shall be limited to the unexpired portion of the original period. This warranty does not apply to any equipment which has not been installed and used within the specifications recommended by SENSIRION for the intended and proper use of the equipment. EXCEPT FOR THE WARRANTIES EXPRESSLY SET FORTH HEREIN, SENSIRION MAKES NO WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH RESPECT TO THE PRODUCT. ANY AND ALL WARRANTIES, INCLUDING WITHOUT LIMITATION, WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY EXCLUDED AND DECLINED. SENSIRION is only liable for defects of this product arising under the conditions of operation provided for in the data sheet and proper use of the goods. SENSIRION explicitly disclaims all warranties, express or implied, for any period during which the goods are operated or stored not in accordance with the technical specifications. SENSIRION does not assume any liability arising out of any application or use of any product or circuit and specifically disclaims any and all liability, including without limitation consequential or incidental damages. All operating parameters, including without limitation recommended parameters, must be validated for each customer's applications by customer's technical experts. Recommended parameters can and do vary in different applications. SENSIRION reserves the right, without further notice, (i) to change the product specifications and/or the information in this document and (ii) to improve reliability, functions and design of this product. Copyright(c) 2016, SENSIRION. CMOSens(R) is a trademark of Sensirion All rights reserved Headquarters and Subsidiaries SENSIRION AG Laubisruetistr. 50 CH-8712 Staefa ZH Switzerland Sensirion Inc., USA phone: +1 805 409 4900 info-us@sensirion.com www.sensirion.com Sensirion Korea Co. Ltd. phone: +82 31 337 7700~3 info-kr@sensirion.com www.sensirion.co.kr phone: +41 44 306 40 00 fax: +41 44 306 40 30 info@sensirion.com www.sensirion.com Sensirion Japan Co. Ltd. phone: +81 3 3444 4940 info-jp@sensirion.com www.sensirion.co.jp Sensirion China Co. Ltd. phone: +86 755 8252 1501 info-cn@sensirion.com www.sensirion.com.cn Sensirion Taiwan Co. Ltd. phone: +41 44 306 40 00 info@sensirion.com To find your local representative, please visit www.sensirion.com/contact www.sensirion.com Version 5 - August 2016 8/8