NEVO+1200M
MEDICAL AC/DC MODULAR CONFIGURABLE POWER SUPPLY
DATA SHEET
_______________________________________________________________________________________________
NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016
Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com
6”x6”x1.61”
SMALL
1200W
POWERFUL
1.2kg
LIGHT
The NEVO+1200M is the smallest in its class and the ultimate power solution for medical applications where size, weight, low
standby power and primary side inhibit are vital factors and delivers up to 1200 Watts from a 1.2kg 6” x 6” x 1.61” package. Each
configured unit consists of an input module with up to eight output modules, where any combination of outputs can be fitted
to create a power solution with up to sixteen isolated outputs.
Standard features include intelligent fan control, wide output voltage adjust capability and primary side shutdown with standby
power consumption of less than 3 Watts. A low noise fan option with virtually silent operation is also available, which allows
you to use this innovative power supply in even the quietest of environments. The series carries full 3rd Edition 60601 safety
approvals and complies with EN55022-B EMC Standards and features market leading specifications and design in application
support.
MAIN FEATURES
Up to 1200 Watts of output power
6 x 6 x 1.61” footprint
Efficiency up to 89%
Primary side remote on/off
function
Standby power ≤ 3 Watts
Field configurable
UL60601 3rd Edition approved
Industry leading power density
(21W/in³)
Lightest modular design only 1.2kg
1000Watts/kg
Parallel & series connection of modules
Remote current / voltage
programming
Accurate current sharing
2 x 5V 1A bias supply
Low noise fan option
3 Year warranty
NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016
Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com
SPECIFICATIONS
Parameter
Details
Min
Typ
Max
Units
AC Input Voltage
Nominal range is 100V to 240V
85
264
Vrms
AC Input Frequency
Contact factory for 400Hz operation.
47
50/60
63
Hz
DC Input Voltage
Standard
120
370
Vdc
Power Rating
See graphs for de-rating
1200
Watts
Input Current
1200Watts output at 120Vrms input
12
Amps
Inrush Current
265Vrms (cold start)
40
Amps
Fusing
5x20 Fast acting
12.5
Amps
Input Current Limit
14
Amps
Efficiency
See graphs
86
89
%
Idle Power
All outputs fitted and enabled
46
Watts
Idle Power
All outputs fitted and Disabled
32
Watts
Standby Power
Latched off state, 120Vrms
2.5
Watts
Power Factor
0.99
0.99
Holdup
1200Watts output at 120Vrms input
17
20
21
mS
UVLO
Turn on only
78
84
Vrms
Over temperature
Internally monitored. Latching
115
125
°C
Reliability
40°C 80% load
2
FPMH
Leakage Current
Normal condition, 264V, 63Hz
190
µAmps
S i g n a l s
Output Bias voltage
Two isolated Bias Outputs available
4.8
5
5.2
V
Output Bias current
Hiccup type current limit
0
1
A
Power Good voltage
PNP open collector with internal 10k pull down resistor
8
10
15
V
Power Good current
0
20
mA
Inhibit voltage
2
15
V
Inhibit current
10k ohm input impedance
0.2
1.5
mA
Global inhibit voltage
3
15
V
Global inhibit current
5k ohm input impedance
0.6
3
mA
AC_OK voltage
High output
4.7
5.2
V
Low output
0
0.1
V
AC_OK current
-10
10
mA
AC_OK warning
See user manual for exceptions
5
mS
Primary Bias voltage
Medically Isolated
4.8
5
5.2
V
Primary Bias current
Hiccup type current limit
0.5
A
Primary Remote On/Off
Negative Edge Triggered, Refer to User Manual
5
V
Component Details Min Max Units
Fan Mag Lev Std (2 Fans per unit) 3.8 FPMH
Input Excluding FAN
2 FPMH
Output See individual output datasheets
1 FPMH
Warranty
3 Years
Parameter
Details
Min
Max
Units
Isolation voltage
Input to output (2 MOPP)
4000
Vac
Input to chassis (1 MOPP)
1500
Vac
Output to chassis
250
Vdc
Output to output
250
Vdc
Isolation clearance
Primary to secondary (reinforced)
7
mm
Primary to chassis (basic)
2.5
mm
Isolation creepage
Primary to secondary (reinforced)
12
mm
Primary to chassis (basic)
4
mm
Leakage current
Medical: 265Vac, 63Hz, 25°C
300 uA
Parameter
Details
Size
154.5mm (L) x 152.4mm (W) x 41.0 ± 1.0mm (H)
Weight
720 gram + 60 gram per output module
Mounting
Bottom (see diagram for details)
NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016
Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com
ENVIRONMENTAL
S t o r a g e
Parameter
Details
Min
Max
Units
Temperature
-40
+85
°C
Humidity
Relative, non-condensing
5
95
%
Altitude
-200
5000
m
Air Pressure
54
106
kPa
O p e r a t i o n
Temperature
Full power
-20
50
°C
Derate input and outputs at 2.5%/°C
50
70
°C
Humidity
Relative, non-condensing
5
95
%
Altitude
(-200 to 2000m for UL60601-1)
-200
3000
m
Air Pressure
78
106
kPa
Noise level
Measured 1m from fan intake
Unit at idle
Unit at full power, 25°C
42
61
dBA
dBA
Shock
3000 bumps at 10G (16ms) half sine wave
Vibration
1.5G 10 to 200Hz sine wave, 20G for 15min in 3 axes random vibration
EMC
Emissions
Parameter
Standard
Level
Radiated electric field
EN55011, EN55022, FCC
A (See Note)
Conducted emissions
EN55011, EN55022, FCC
B
Harmonic Distortion
EN61000-3-2
Compliant
Flicker & Fluctuation
EN61000-3-3
Compliant
Immunity
Electrostatic discharge
EN61000-4-2
(15kV air, 8kV contact)
4
Radiated RFI
EN61000-4-3 (10V/m)
3
Fast Transient burst
EN61000-4-4 (4kV)
4
Input line surges
EN61000-4-5 (1kV L-N, 2kV L-E)
3
Conducted RFI
EN61000-4-6 (10V)
4
Power Freq. Magnetic Field
EN61000-4-8 (10A/m)
3
Voltage Dips
EN61000-4-11 (EN55024)
Compliant
Note: To meet Class B radiated emissions the end user should add ferrites to I/P and O/P cables. Consult Vox Power for details.
AGENCY APPROVALS
Standard
Details
File
IEC/EN60601-1 3rd Edition UL: E316486
UL60601-1
3rd Edition
CAN/CSA-C22.2 No. 60601-1
(2008)
ANSI/AAMI ES60601-1
(2005 + C1:09 + A2:10)
CE MARK
LVD 2014/35/EU
CB certificate and report available on request
0
200
400
600
800
1000
1200
1400
80 100 120 140 160 180 200 220 240 260
Output Power (W)
Input Voltage (RMS)
Line Derating
Derate at 10W per Volt below 120Vrm s
No Derating for DC operation
0
200
400
600
800
1000
1200
1400
-20 -10 010 20 30 40 50 60 70
Output Power (W)
Ambient Temperature (Celcius)
Temperature Derating
Derate at 30W per degree celcius above 50 degree
0,66
0,68
0,70
0,72
0,74
0,76
0,78
0,80
0,82
0,84
0,86
0,88
0,90
0,92
100 200 300 400 500 600 700 800 900 1000 1100 1200
Efficiency
Output power (W)
Typical Load Effici ency (220Vrms)
OP1
OP2
OP3
OP4 0,80
0,81
0,82
0,83
0,84
0,85
0,86
0,87
0,88
0,89
0,90
80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260
Efficiency
Input Voltage(Vrms)
Ty p ical Line Efficiency (Pmax)
OP1
OP2
OP3
OP4
NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016
Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com
MECHANICAL DIMENSIONS AND MOUNTING SCREWS
SCREWS
LOCATION
DETAILS
PENETRATION
TIGHTENING
MOUNTING M4 x 4 4mm max, including chassis 0.55 NM (1)
OUTPUT MODULES M3 x 5, Countersink Posi, 16 Places Defined by screw 0.35 NM (1)
CHASSIS LID AND FACEPLATE M3 x 5, Countersink Posi, 11 Places Defined by screw 0.35 NM (1)
1. Torque settings are for general reference only. The torque settings shown in the datasheet are the insert manufacturers recommended values.
NEVO+1200 Medical System Datasheet | DOC6035r03 | Release date 24/03/2016
Vox Power Limited | Unit 2, Red Cow Interchange Estate, Ballymount, Dublin 22, D22 Y8H2, Ireland | T +353 1 4591161 | www.vox-power.com
All specifications are believed to be correct at time of publishing. Vox Power Ltd reserves the right to make changes to any of its products and to change or improve any part of the
specification, electrical or mechanical design or manufacturing process without notice. Vox Power Ltd does not assume any liability arising out of the use or application of any of its
products and of any information to the maximum extent permitted by law. No license, express or implied, by estoppel or otherwise, to any intellectual property rights is granted by this
document or by any products of Vox Power Ltd. VOX POWER LTD DISCLAIMS ALL WARRANTIES AND REPRESENTATIONS OF ANY KIND WHETHER EXPRESS OR IMPLIED, INCLUDING, BUT
NOT LIMITED TO, IMPLIED WARRANTIES OF SUITABILITY, FITNESS FOR PURPOSE, MERCHANTABILITY AND NONINFRINGEMENT.
Please consult your local distributor or Vox Power directly to ensure that you have the latest revision before using the product and refer to the latest relevant user manual for further
information relating to the use of the product. Vox Power Ltd products are not intended for use in connection with life support systems, human implantations, nuclear facilities or systems,
aircraft, spacecraft, military or naval missile, ground support or control equipment used for the purpose of guidance navigation or direction of any aircraft, spacecraft or military or naval
missile or any other application where product failure could lead to loss of life or catastrophic property damage. The user will hold Vox Power Ltd harmless from any loss, cost or damage
resulting from its breach of these provisions.
PINOUTS
v
J1
J2bJ2a
11
12
1
2
J6
6 1
N E L
J1
J1
Circuit
Details
1
Live
2
Earth
3
Neutral
J2a/b
Circuit
Details
1
Power Good
Slot
A and E
2
Inhibit
3
Power Good
Slot
B and F
4
Inhibit
5
Power Good
Slot
C and G
6
Inhibit
7
Power Good
Slot
D and H
8
Inhibit
9
Global Inhibit
10
AC OK
11
+5V 1A Bias Supply
12
COM
J6
1
Common
2
+5V 500mA Bias
3
Shut Down
4
Reserved
5
Reserved
6
Reserved
REF
DETAILS
MANUFACTURER
HOUSING
TERMINAL
J1
MAINS INPUT: 3 Pin, Barrier, 6-32 Steel Screws, 0.8 NM or 7IN LB Torque
Cable 14-18AWG, 300V, 16A, 105°C, use appropriately rated fork or ring
terminal.
MOLEX
J2a/b
GLOBAL SIGNALS: 12 Pin, 2mm, without Friction Lock, 24-30 AWG
MOLEX
511101251
503948051
J6
INPUT BIAS: OUTPUT SIGNALS: 6 Pin, 1.25mm, with Friction lock, 28-32
AWG
MOLEX 510210600 500588000
Notes
1. Direct equivalents may be used for any connector parts.
2. All cables must be rated 105°C min, equivalent to UL1015
NEVO Power Series
NEVO+1200 M - 1 1 2 2 3 3 4 4 - 0 0 0
Factory Use
Leakage Current
USE '0' for unused
slots. Blanking
plates will be
inserted at
factory.
Medical
Slot A - Output #
Slot H - Output #
Slot B - Output #
Slot G - Output #
Slot C - Output #
Slot F - Output #
Slot D - Output #
Slot E - Output #
Our design team will assist with value add requirement if an application requires standard/non-standard accessories or non-nominal voltage
settings. Once approved, the factory will issue a 3 or 4 digit code for your specific configuration which can be used for all future orders of the
same configuration. When ordering an input unit with no outputs inserted, simply order NEVO+1200M.