Embedded Power for Business-Critical Continuity Rev. 7.9.09_#98 LPS200-M Series of 3 LPS200-M Series 250 Watts Total Power: Input Voltage: # of Outputs: 125 - 250 Watts 90 - 264 Vac Single Special Features * * * * * * * * * * * * * Medical and ITE safeties Active power factor correction 3" x 5" footprint Less than 1U high EN61000-3-2 compliant Remote sense Power fail Adjustable main output Built-in Class B EMI filter Overvoltage protection Overload protection Thermal overload protection Isolated 12V Fan output Electrical Specifications Input Input range: Frequency: Inrush current: Efficiency: EMI/RFI: Power factor: Safety ground leakage current: 90 - 264 Vac; 120 - 300 Vdc 47-63 Hz 50 A max., cold start @ 25 C 86% typical at full load FCC Class B conducted; CISPR22 Class B conducted; EN55022 Class B conducted; VDE0878PT3 Class B conducted 0.99 typical 275 uA @ 50/60 Hz, 264 Vac input Output 125 W for convection (100 W for LPS202-M) ; 250 W (200 W for LPS202-M) with 30CFM forced air Adjustment range: 10% minimum on the main outputs Fan output: 12 V @ 0.5 A convection, @ 1 A forced air isolated, 10% Hold-up time: 16 ms @ 250 W load, 120 Vac input Short circuit protection on all outputs. Case overload protected @ 110-160% Overload protection: above rating Overvoltage protection: 15-50% above nominal output Maximum power: Logical Control Power failure: Safety * * * * * TUV 60950, 60601-1 UL 60950, 60601-1 cULus 60950, 60601-1 CB Certificate & report CE Mark (LVD) Remote sense: Open collector logic signal goes high 100-500 msec after main output; it goes low at least 6 msec before loss of regulation Compensates for 0.5 V lead drop min. Will operate without remote sense connected. Reverse connection protected. Embedded Power for Business-Critical Continuity Rev. 7.9.09_#98 LPS200-M Series of 3 Environmental Specifications Operating temperature: Storage temperature: Electromagnetic susceptibility: Humidity: Vibration: MTBF demonstrated: 0 to 50 C ambient derate each output as 2.5% per degree from 50 to 70 C. -20 C start up -40 C to +85 C Designed to meet EN61000-4; -2, -3, -4, -5, -6, -8, -11 Level 3 Operating; non-condensing 10% to 95% RH IEC68-2-6 to the levels of IEC721-3-2 >550,000 hours at full load and 25 C ambient conditions Ordering Information Model Number Output Voltage Minimum Load Maximum Load with Convection Cooling Maximum Load with 30CFM Forced Air LPS202-M 5V 0A 20 A 40 A 44 A 2% 50 mV LPS203-M 12 V 0A 10.3 A 20.8 A 22 A 2% 120 mV LPS204-M 15 V 0A 8.3 A 16.6 A 18 A 2% 150 mV LPS205-M 24 V 0A 5.2 A 10.4 A 11.5 A 2% 240 mV LPS208-M 48 V 0A 2.6 A 5.2 A 5.8 A 2% 480 mV Peak Load Regulation2 Ripple P/P (PARD)3 1. Peak current lasting <30 seconds with a maximum 10% duty cycle. 2. At 25C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 mHz bandwidth and 10 F (tantalum capacitor) in parallel with a 0.1 F capacitor at rated line voltage and load ranges. Pin Assignments Mating Connectors Connector LPS200-M SK1 Pin 1 Neutral Pin 3 Line SK2 TB-1 Common TB-2 Main output SK3 Pin 1 +V1 Remote sense Pin 2 -V1 Remote sense Pin 3 N/C Pin 4 N/C Pin 5 +Power fail Pin 6 Common Pin 7 N/C Pin 8 Common Pin 9 +12 V Fan Pin 10 +12 Fan Return (isolated) AC Input (SK1): Molex 09-50-8031 (connector) PINS: 08-52-0113 AC Ground: Molex 01-90020001 DC Output (SK2): Control Signals (SK3): Molex 19141-0058/0063/0083 Spade lug based on Cable Ampacity/AWG Molex 90142-0010 (USA) PINS: 90119-2110 or Amp: 87977-3 PINS: 87309-8 Emerson Network Power Connector Kit #70-841-020, includes all of the above. 1. Specifications subject to change without notice. 2. All dimensions in inches (mm), tolerance is 0.02" (0.5mm) 3. Mounting holes MH1 and MH2 should be grounded for EMI purposes. 4. Mounting hole MH1 is safety ground connection. 5. Specifications are for convection rating at factory settings at 115 VAC input, 25 C unless otherwise stated. 6. This power supply requires mounting on metal standoffs 0.20" (5m) in height. 7. Warranty: 2 years 8. Weight: 0.75 lb/0.34 kg Embedded Power for Business-Critical Continuity Americas Mechanical Drawing 3.00 (76.2) 0.15 Dia. (4) 4 Places 2.55 (64.8) MH1 SK1 3 1 5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698 Rev. 7.9.09_#98 LPS200-M Series of 3 Europe (UK) Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355 Asia (HK) 14/F, Lu Plaza 2 Wing Yip Street Kwun Tong, Kowloon Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888 5.00 (127.0) 4.55 (115.6) For global contact, visit: www.PowerConversion.com techsupport.embeddedpower @emerson.com While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. Emerson Network Power. The global leader in enabling business-critical continuity. AC Power Connectivity DC Power 2 MH2 SK2 Embedded Computing SK3 1 Embedded Power Monitoring Outside Plant Power Switching & Controls 1.32 (33.6) 1 SK3 VR1 Output Voltage Adjust Pot Precision Cooling Racks & Integrated Cabinets Services 5 10 Surge Protection 0.197" (5.0) Minimum Standoff Height EmersonNetworkPower.com Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. (c)2009 Emerson Electric Co.