Si7034-A10
14 Rev. 1.0
4.2. Hysteresis
The moisture absorbent film (polymeric dielectric) of the humidity sensor will carry a memory of its exposure
history, particularly its recent or extreme exposure history. A sensor exposed to relatively low humidity will carry a
negative offset relative to the factory calibration, and a sensor exposed to relatively high humidity will carry a
positive offset relative to the factory calibration. This factor causes a hysteresis effect illustrated by the solid trace
in Figure 5. The hysteresis value is the difference in %RH between the maximum absolute error on the decreasing
humidity ramp and the maximum absolute error on the increasing humidity ramp at a single relative humidity
setpoint and is expressed as a bipolar quantity relative to the average error (dashed trace). In the example of
Figure 5, the measurement uncertainty due to the hysteresis effect is +/-1.0%RH.
4.3. Prolonged Exposure to High Humidity
Prolonged exposure to high humidity will result in a gradual upward drift of the RH reading. The shift in sensor
reading resulting from this drift will generally disappear slowly under normal ambient conditions. The amount of
shift is proportional to the magnitude of relative humidity and the length of exposure. In the case of lengthy
exposure to high humidity, some of the resulting shift may persist indefinitely under typical condition s. It is generally
possible to substantially reverse this affect by baking the device (see section “4.6. Bake/Hydrate Procedure” ).
4.4. PCB Assembly
4.4.1. Soldering
Like most ICs, Si7034 devices are shipped from the factory vacuum-packed with an enclosed desiccant to avoid
any RH accuracy drif t durin g storage and to prevent any mo isture-related issues during solder reflo w. The following
guidelines should be observed during PCB assembly:
Si7034 devices are compatible with standard board assembly process. Devices should be soldered using
reflow per the recommended card reflow profile. See Section “10. PCB Land Pattern and Solder Mask
Design” for the recommended card reflow profile.
A “no clean” solder process is recommended to minimize the need for water or solvent rinses after
soldering. Cleaning after soldering is possible, but must be done carefully to avoid impacting the
performance of the sensor. See AN607 for more information on cleaning.
It is essential that the exposed polymer sensing film be kept clean and undamaged. This can be
accomplished by careful handling and a clean, well-controlled assembly process. When in doubt or for
extra protection, a heat-resistant, protective cover such as Kapton(R) KPPD-1/8 polyimide tape can be
installed during PCB assembly.
4.4.2. Rehydration
The measured humidity value will generally shift slightly after solder reflow. A portion of this shift is permanent and
is accounted for in the accuracy specifications in Table 4. After soldering, an Si7034 should be allowed to
equilibrate under controlled RH conditions (room temperature, 45–55%RH) for at least 48 hours to eliminate the
remainder of the shift and return the device to its specified accuracy performance.
4.4.3. Rework
To maintain the specified sensor performance, care must be taken during rework to minimize the exposure of the
device to excessive heat and to avoid damage/contamination or a shift in the sensor reading due to liquids, solder
flux, etc. Manual touch-up using a soldering iron is permissible un der the following guidelines:
The exposed polymer sensing film must be kept clean and un damaged. A Kapton ® tape protective cover is
recommended during any rework operation.
Flux must not be allowed to contaminate the sensor; liquid flux is not recommended even with a cover in
place. Conventional lead-free solder with rosin core is acceptable for touch-up as long as a cover is in
place during th e rework.
If possible, avoid water or solvent rinses after touch-up. Cleaning after soldering is possible, but must be
done carefully to avoid impacting the performance of the sensor. See AN607 for more information on
cleaning.