Keysight Technologies
34410A and 34411A Multimeters
Data Sheet
Product Discontinuance Notice
The 34410A/11A DMM and all associated options will
be discontinued December 1, 2016. The last date this
product can be ordered is November 30, 2016.
For new product purchases, Keysight recommends
the 34465A, 6½ digit Truevolt DMM.
For more information, as well as to access to de-
tailed migration guides, please visit www.keysight.
com/nd/nextgenDMMs
To contact a product selection expert, visit
www.keysight.com/nd/contactus
Introduction
Keysight 34410A 6½-Digit High-Performance DMM
10,000 readings/s @ 5½-digits direct to PC
1,000 readings/s @ 6½-digits direct to PC
30 PPM 1-year Basic DC accuracy
LAN, USB & GPIB standard
DCV, ACV, DCI, ACI, 2-wire and 4-wire Resistance, Frequency, Period, Continuity, and Diode Test
Capacitance & Temperature measurements
Expanded measurement ranges
Data logger with 50 k reading non-volatile memory
Keysight 34411A 6½-Digit Enhanced-Performance DMM
All the features of the 34410A, plus:
50,000 readings/s @ 4½-digits direct to PC
1 million volatile reading memory
Analog level triggering
Programmable Pre/Post triggering
The Best Just Got Better
The Keysight Technologies, Inc. 34410A and 34411A 6½-Digit DMMs build on the phenomenal
success of the industry-standard Keysight 34401A. These new meters offer improved accuracy,
expanded measurement capability, dramatically improved measurement speed and throughput, and
modern computer interfaces including LAN and USB. The dual display offers both dual measurement
capabilities and ease of use when setting up and conguring the DMM. Improvements have been
made in every facet of the 34401A to make the best even better, whether you use it on the bench or in
a system.
Dramatic Speed Improvements
Whether it’s raw reading speed or fast system throughput, the 34410A sets a new benchmark in
performance. Using a new A/D technology, the 34410A achieves an impressive 10,000 readings a
second at 5½-digits, and can stream readings to your computer at this same speed! Triggering is fast
and precise, with both trigger latency and trigger jitter less than 1 µs, while bus query response is less
than 500 µs. ACV measurements are faster as well thanks to a digital measurement technique that
additionally improves accuracy at high and low frequencies. For even greater reading speeds, select
the 34411A, which achieves 50,000 readings a second at 4½-digits.
Enhanced Measurement Performance
The 34410A and 34411A offer Temperature and Capacitance capabilities, in addition to those
measurements you have come to expect, such as DCV, ACV, DCI, ACI, 2-wire and 4-wire Resistance,
Frequency, Period, Continuity and Diode Test. You also get Offset Compensated Ohms, allowing
you to accurately measure resistance in the presence of voltages. Measurement ranges have been
expanded as well; for example, DC and AC Current Ranges now go down to 100 µA, resulting in
100 pA resolution. Real-time math and statistics are included, and a peak-detect capability allows
you to capture peaks as short as 20 µs.
03 | Keysight | 34410A and 34411A Multimeters - Data Sheet
Even Greater Performance
with the 34411A
The 34411A has all the features
of the 34410A, plus additional
performance that makes it even
more powerful. With the ability to
make 50,000 readings per second at
41/
2-digits, Analog Level Triggering,
programmable Pre- and Post-Trigger
and 1 million readings of volatile
memory in addition to 50,000
reading of non-volatile memory,
you now have the ability to capture
low-frequency waveforms, char-
acterize device performance and
transfer results for analysis on your
computer.
Data Logger Function
A front panel data logger function
allows you to set the meter up to
make unattended, paced measure-
ments over a xed time or number of
events, then pull up the results later
for review or transfer to a computer
for analysis. Set the meter up to take
measurements every 10 seconds for
an hour, go have lunch, and check
the results upon your return. The
contextual front panel sequences
make setup and read back a breeze.
Improved Ease of Use
From the inclusion of a second
display to conguring setups for
each of the measurement functions,
these new DMMs offer signicant
improvements in usability. Simple
things are simple; more complicated
setups are easier than ever. There
is even a new probe set designed to
more easily probe today’s ne-pitch
components. Finally, there is a
built-in Graphical Web Interface that
allows you to interactively control
the DMM without the hassle of
programming at all!
Modern I/O for Improved
Connectivity
When connecting to a computer,
select LAN, USB or GPIB inter-faces;
all three are standard on the 34410A
and 34411A. Concerned about the
viability of your existing software
programs? These new DMMs
respond to Standard Commands
for Programmable Instrumentation
(SCPI), and there is even a 34401A
Emulation Mode to ensure the
easiest upgrade possible. Keysight’s
I/O Library Suite ships with the
34410A and 34411A to help you
quickly establish an error-free
connection between your PC and
instrument. It provides robust
instrument control and works
with the software development
environment you choose.
LXI – Class C
LAN Extensions for Instruments (LXI)
provides the next generation I/O for
system applications requiring
the highest throughput. Transfer
rates of over 250,000 readings/s are
attainable ensuring even the most
data intensive measurements are
fast, without the overhead cost
of an instrument cardcage. Both the
34410A and the 34411A are LXI –
Class C compliant.
Built To Last
Our new DMMs were designed
to high standards of ruggedness and
reliability. From the robust package
with its shock absorbing bumpers to
careful selection of components and
conservative circuit design, these
meters are built to last. Calculated
Mean Time Between Failure (MTBF)
is in excess of 100,000 hours.
Backed by a worldwide network of
service centers, you can buy with
condence.
Go to the Web
For the latest information on
these or other Keysight DMMs,
go to www.keysight.com/nd/dmm
Accessories Included:
Test Lead Kit with probes and
SMT attachments.
Test report, power cord, and
USB interface cable.
Product Reference CD-ROM with
soft documentation and software:
Programmer’s Reference Help
Quick Start Tutorial
User’s Guide
Service Guide
Programming Examples
IntuiLink for Multimeters
LabVIEW and IVI-COM drivers
Optional Printed Documentation:
Quick Start Tutorial
User’s Guide
Service Guide
Bench Dimensions:
System Dimensions:
261.2 mm
103.8 mm
88.3 mm
303.2 mm
212.8 mm
272.3 mm
04 | Keysight | 34410A and 34411A Multimeters - Data Sheet
Accuracy Specifications ± (% of reading + % of range)1
Function Range3 Frequency, 24 Hour2 90 Day 1 Year
Temperature Coefficient/°C
Test Current or Tcal ± 1 °C Tcal ± 5 °C Tcal ± 5 °C 0 °C to (Tcal -5 °C)
Burden Voltage (Tcal +5 °C) to 55 °C
DC Voltage 100.0000 mV 0.0030 + 0.0030 0.0040 + 0.0035 0.0050 + 0.0035 0.0005 + 0.0005
1.000000 V 0.0020 + 0.0006 0.0030 + 0.0007 0.0035 + 0.0007 0.0005 + 0.0001
10.00000 V 0.0015 + 0.0004 0.0020 + 0.0005 0.0030 + 0.0005 0.0005 + 0.0001
100.0000 V 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001
1000.000 V4 0.0020 + 0.0006 0.0035 + 0.0006 0.0040 + 0.0006 0.0005 + 0.0001
True RMS 100.0000 mV 3 Hz – 5 Hz 0.50 + 0.02 0.50 + 0.03 0.50 + 0.03 0.010 + 0.003
AC Voltage5 to 750.000 V 5 Hz – 10 Hz 0.10 + 0.02 0.10 + 0.03 0.10 + 0.03 0.008 + 0.003
10 Hz – 20 kHz 0.02 + 0.02 0.05 + 0.03 0.06 + 0.03 0.005 + 0.003
20 kHz – 50 kHz 0.05 + 0.04 0.09 + 0.05 0.10 + 0.05 0.010 + 0.005
50 kHz – 100 kHz 0.20 + 0.08 0.30 + 0.08 0.40 + 0.08 0.020 + 0.008
100 kHz – 300 kHz 1.00 + 0.50 1.20 + 0.50 1.20 + 0.50 0.120 + 0.020
Resistance6 100.0000
1 mA 0.0030 + 0.0030 0.008 + 0.004 0.010 + 0.004 0.0006 + 0.0005
1.000000 k
1 mA 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001
10.00000 k
100 µA 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001
100.0000 k
10 µA 0.0020 + 0.0005 0.007 + 0.001 0.010 + 0.001 0.0006 + 0.0001
1.000000 M
5 µA 0.0020 + 0.0010 0.010 + 0.001 0.012 + 0.001 0.0010 + 0.0002
10.00000 M
500 nA 0.0100 + 0.0010 0.030 + 0.001 0.040 + 0.001 0.0030 + 0.0004
100.0000 M
500 nA || 10 M
0.200 + 0.001 0.600 + 0.001 0.800 + 0.001 0.1000 + 0.0001
1.000000 G
500 nA || 10 M
2.000 + 0.001 6.000 + 0.001 8.000 + 0.001 1.0000 + 0.0001
DC Current 100.0000 µA < 0.03 V 0.010 + 0.020 0.040 + 0.025 0.050 + 0.025 0.0020 + 0.0030
1.000000 mA < 0.3 V 0.007 + 0.006 0.030 + 0.006 0.050 + 0.006 0.0020 + 0.0005
10.00000 mA < 0.03 V 0.007 + 0.020 0.030 + 0.020 0.050 + 0.020 0.0020 + 0.0020
100.0000 mA < 0.3 V 0.010 + 0.004 0.030 + 0.005 0.050 + 0.005 0.0020 + 0.0005
1.000000 A < 0.8 V 0.050 + 0.006 0.080 + 0.010 0.100 + 0.010 0.0050 + 0.0010
3.000000 A < 2.0 V 0.100 + 0.020 0.120 + 0.020 0.150 + 0.020 0.0050 + 0.0020
True RMS 100.0000 µA to 3 Hz – 5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
AC Current7 3.00000 A 5 kHz – 10 kHz 0.20 + 0.04 0.20 + 0.04 0.20 + 0.04 0.030 + 0.006
Frequency 100 mV to 3 Hz – 5 Hz 0.070 + 0.000 0.070 + 0.000 0.070 + 0.000 0.005 + 0.000
or Period 750 V 5 Hz – 10 Hz 0.040 + 0.000 0.040 + 0.000 0.040 + 0.000 0.005 + 0.000
10 Hz – 40 Hz 0.020 + 0.000 0.020 + 0.000 0.020 + 0.000 0.001 + 0.000
40 Hz – 300 kHz 0.005 + 0.000 0.006 + 0.000 0.007 + 0.000 0.001 + 0.000
Capacitance8 1.0000 nF 500 nA 0.50 + 0.50 0.50 + 0.50 0.50 + 0.50 0.05 + 0.05
10.000 nF 1 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.05 + 0.01
100.00 nF 10 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
1.0000 µF 10 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
10.000 µF 100 µA 0.40 + 0.10 0.40 + 0.10 0.40 + 0.10 0.01 + 0.01
Temperature9
RTD
-200 °C to 600 °C 0.06 °C 0.06 °C 0.06 °C 0.003 °C
Thermistor -80 °C to 150 °C 0.08 °C 0.08 °C 0.08 °C 0.002 °C
Continuity 1000.0
1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
Diode Test10 1.0000 V 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
1. Specifications are for 90 minute warm-up and 100 PLC.
2. Relative to calibration standards.
3. 20% overrange on all ranges, except DCV 1000 V, ACV 750 V, DCI and ACI 3 A ranges.
4. For each additional volt over ± 500 V add 0.02 mV of error.
5. Specifications are for sinewave input > 0.3% of range and > 1 mVrms. Add 30 µV error for frequencies below 1
kHz. 750 VAC range limited to 8 x 107 Volts-Hz. For each additional volt over 300 Vrms add 0.7 mVrms of error.
6. Specifications are for 4-wire resistance measurements, or 2-wire using Math Null.
Without Math Null, add 0.2 Ω additional error in 2-wire resistance measurements.
7. Specifications are for sinewave input > 1% of range and > 10 µArms. Frequencies > 5 kHz are typical for all
ranges. For the 3 A range (all frequencies) add 0.05% of reading + 0.02% of range to listed specifications.
8. Specifications are for 1-hour warm-up using Math Null. Additional errors may occur for non-film capacitors.
9. For total measurement accuracy, add temperature probe error.
10. Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is typical.
Variation in the current source will create some variation in the voltage drop across a diode junction.
05 | Keysight | 34410A and 34411A Multimeters - Data Sheet
A-to-D Converter Noise Performance
Integration Time Resolution Normal Mode Readings/Second4
(NPLC) (ppm of range)1 Rejection (dB)2
0.0015 30 0 50,000
0.0025 15 0 25,000
0.006 6 0 10,000
0.02 3 0 3,000
0.06 1.5 0 1,000
0.2 0.7 0 300
1 0.3 55 60 (50)
2 0.2 1103 30 (25)
10 0.1 1103 6 (5)
100 0.03 1103 0.6 (0.5)
1. Resolution is defined as the typical DCV 10 V range RMS noise. Auto-zero on for NPLC ≥ 1.
See manual for additional noise characteristics.
2. Normal mode rejection for power line frequency ± 0.1%.
3. For power-line frequency ± 1% 75 dB and for ± 3% 55 dB.
4. Maximum rate with auto-zero off for 60 Hz and (50 Hz) operation.
5. Only available for the 34411A.
System Reading and Throughput Rates
DMM memory to PC (Maximum reading rate out of memory6
Drawing – Path B
Reading GPIB USB 2.0 LAN (VXI-11) LAN (Sockets)
Format Readings/s Readings/s Readings/s Readings/s
ASCII 4,000 8,500 7,000 8,500
32-bit Binary 89,000 265,000 110,000 270,000
64-bit Binary 47,000 154,000 60,000 160,000
Direct I/O Measurements (Single reading – measure and I/O time)6
6. 1/2 scale input signal, immediate trigger, trigger delay 0, auto-zero off, auto-range off,
no math, null off, 60 Hz line frequency. Specifications are for 34410A or (34411A). See
manual for performance on other functions.
System Performance
Function Range Auto- Maximum Maximum Internal
Change (ms)7 Change
(ms)8
range (ms)9 External Trigger Rate10
LAN/GPIB
Trigger Rate
DCV/2-wire 22 3.9/2.6 7.5 5,000/s 10,000/s
Resistance (50,000/s)
ACV/ 37 6.5/6.4 19 500/s 500/s
Frequency
7. Time to change from 2-wire Resistance to this specified function, or DCV to 2-wire
Resistance using the SCPI “FUNC” command.
8. Time to change from one range to the next higher range, ≤ 10 V, ≤ 10 MΩ.
9. Time to automatically change one range and be ready for the new measurement, ≤ 10 V,
10 MΩ.
10. Specifications are for 34410A or (34411A).
DMM input DMM memory
50 k readings
non-volatile
ADC
Path “A”
USB
LAN
GPIB
34410A /
34411A digital
multimeter
PC
Path “C” Path “B”
100
10
1
0.1
0.01
0.001 0.01 0.1 1 10 100
RMS noise (ppm of range)
Integration time (NPLC)
System Reading Architecture
Drawing - Path C Maximum Reading Rate
into Memory or to Direct
I/O (Readings/s) Drawing
– Path A or C
Function Resolution
(NPLC)
GPIB
ms
USB 2.0
ms
LAN
(VXI-11)
ms
LAN
(sock-
ets) ms
DCV/2-wire
Resistance
0.006
(0.001)
2.6 2.9 4.6 3.2 10,000 (50,000)
ACV/
Frequency
Fast Filter 1
ms gate
10.0 10.0 10.0 10.0 500
06 | Keysight | 34410A and 34411A Multimeters - Data Sheet
DC Voltage
Measurement Method:
Continuously integrating
multi-slope IV A/D converter
Linearity: 0.0002% of reading
(10 V range) + 0.0001% of range
Input Resistance:
0.1 V, 1 V, 10 V 10 MΩ or > 10 GΩ
Ranges (Selectable)
100 V, 1000 V 10 MΩ ± 1%
Ranges (Fixed)
Input Bias Current:
< 50 pA at 25 °C
Input Protection: 1000 V
DC CMRR: 140 dB1
True RMS AC Voltage
Measurement Method:
AC-coupled True RMS measurement.
Digital sampling with anti-alias filter.
Crest Factor:
No additional error for crest factors
< 10. Limited by peak input and 300
kHz bandwidth.
Peak Input:
300% of range or 1100 V
Overload Ranging:
Will select higher range if peak input
overload is detected during auto
range. Overload is reported in manual
ranging.
AC CMR: 70 dB2
Maximum Input: 400 Vdc, 1100 Vpk
Input Impedance:
1 MΩ ± 2% in parallel with < 150 pF
Input Protection:
750 Vrms all ranges
Resistance
Measurement Method:
Selectable 2-wire or 4-wire.
Current source referenced to LO input.
Offset Compensation:
Selectable on the 100 Ω, 1 kΩ,
and 10 kΩ ranges
Max. Lead Resistance (4-wire):
10% of range per lead for 100
, 1 k
.
1 kΩ per lead on all other ranges
Input Protection:
1000 V on all ranges
DC Current
Current Shunt:
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Input Protection: 3 A, 250 V fuse
Measurement Characteristics
1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum
2. For 1 kΩ unbalanced in LO lead and < 60 Hz, ± 500 V peak maximum
3. Maximum rate for DCV, DCI, and resistance functions (using zero settling delay, autozero
off, manual range)
4. 34411A only
True RMS AC Current
Measurement Method:
AC-coupled True RMS measurement.
Directly coupled to the fuse and shunt.
Digital sampling with anti-alias filter.
Current Shunt:
200 Ω for 100 µA, 1 mA
2 Ω for 10 mA, 100 mA
0.1 Ω for 1 A, 3 A
Maximum Input:
The peak value of the DC + AC current must
be < 300% of range. The RMS current must
be < 3 A including the DC current content.
Input Protection: 3 A, 250 V fuse
Frequency and Period
Measurement Method:
Reciprocal-counting technique. AC-coupled
input
using the AC voltage measurement
function.
Input Impedance:
1 MΩ ± 2% in parallel with < 150 pF
Input Protection: 750 Vrms all ranges
Capacitance
Measurement Method:
Current input with measurement of
resulting ramp.
Connection Type: 2-wire
Temperature
Thermistor:
2.2 kΩ, 5 kΩ, and 10 kΩ
RTD: α = 0.00385
Ro from 49 Ω to 2.1 kΩ
Continuity/Diode Test
Response Time:
300 samples/sec with audible tone
Continuity Threshold: Fixed at 10 Ω
Operating Characteristics
Maximum readings/second
Digits
Function3 4.5 5.5 6.5
DCV 50 k4 10 k 1 k
2-wire
50 k4 10 k 1 k
DCI 50 k4 10 k 1 k
Frequency 500 90 10
Period 500 90 10
Filter setting fast med slow
ACV 500 150 50
ACI 500 150 50
Additional 34411A Specifications
Resolution: See table on page 4
Overall Bandwidth, DCV & DCI:
15 kHz typical @ 20 µs aperture (-3 dB)
Triggering: Pre/Post, Int/Ext, Pos/Neg
Timebase Resolution: 19.9524 µs 0.01% accuracy
Trigger Jitter:
2 µs (p-p), 20 µs (p-p) when pre-triggered
Spurious-Free Dynamic Range
& Signal to Noise Distortion Ratio
Function DCV Range Spur-Free SNDR
1 V -75 dB 60 dB
10 V1 -70 dB 60 dB
100 V -75 dB 60 dB
110 V range: 2 V (p-p) < signal < 16 V (p-p)
Triggering and Memory
Reading Hold Sensitivity:
1% of reading
Samples per Trigger:
1 to 50,000 (34410A)
1 to 1,000,000 (34411A)
Trigger Delay: 0 to 3600 s (20 µs step size)
External Trigger:
Programmable edge, Low-power TTL compatible
Delay: < 1 µs Max rate: 5,000/s
Jitter: < 1 µs Min Pulsewidth:
1 µs
Voltmeter Complete:
3 V Logic output,
2 µs pulse with programmable edge
Nonvolatile Memory: 50,000 readings
Volatile Memory:
50,000 readings (34410A)
1,000,000 readings (34411A)
Sample Timer:
Range: 0 to 3600 s (20 µs step sizes)
Jitter: < 100 ns
General Specifications
Power Supply:
100 V/120 V/220 V/240 V ± 10%
Power Line Frequency:
45 Hz to 66 Hz and 360 Hz to 440 Hz, Automatically
sensed at power-on
Power Consumption: 25 VA peak (16 W average)
Operating Environment: Full accuracy for
0 °C to 55 °C, 80% R.H. at 40 °C non-condensing
Storage Temperature: -40 °C to 70 °C
Weight: 3.72 kg (8.2 lbs)
Safety: IEC 61010-1, EN 61010-1, UL 61010-1,
CAN/CSA-C22.2 No. 61010-1, Refer to
Declarations of Conformity for current
revisions. Measurement CAT II 300 V,
CAT I 1000 V. Pollution Degree 2
EMC: IEC 61326, EN 61326, CISPR 11, ICES-001,
AS/NZS 2064.1, Refer to Declaration of Conformity for
current revisions.
Vibration & Shock: MIL-T-28800E,
Type III, Class 5 (Sine Only)
LXI Compliance: LXI Class C, ver. 1.0
Warranty: 3 year standard
07 | Keysight | 34410A and 34411A Multimeters - Data Sheet
Ordering Information
Keysight 34410A and 34411A Multimeters
Accessories included Test lead kit with probes and SMT attachments, calibration
certicate, power cord, and USB interface cable
Product reference CD with
soft documentation and software
Programmer’s reference help
Quick start tutorial
User’s guide
Service guide
Programming examples
IntuiLink for multimeters
LabVIEW and IVI-COM drivers
Options
Opt. A6J ANSI Z540 compliant calibration
Probes/leads/clip accessories
11059A Kelvin probe set
11060A Surface mount device (SMD) test probes
11062A Kelvin clip set
34133A Precision electronic test leads
34134A DC coupled current probe
34136A High voltage probe
34138A Test lead set
34171B Input terminal connector (sold in pairs)
34172B Input calibration short (sold in pairs)
34308A Thermistor kit
34330A 30 A current shunt
E2308A 5 k thermistor probe
Y1133A Low-thermal external digital multimeter scanning kit
Rack mount kits
34190A Rackmount kit: designed for use with only one instrument, mounted on
either the left or the right side of the rack
34191A 2U dual ange kit: secures the instrument to the front of the rack.
This kit can be used with the 34194A dual lock link kit to mount two
half-width, 2U height instruments side-by side.
34194A Dual lock link kit: recommended for side-by-side combinations and
includes links for instruments of different depths. This kit can be used
with the 34191A 2U dual ange kit to mount two half-width, 2U height
instruments side-by-side.
Other accessories
34131A Hard transit case
34162A Accessory pouch
E5810A LAN/GPIB gateway
08 | Keysight | 34410A and 34411A Multimeters - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, Inc. 2013 - 2015
Published in USA, May 28, 2015
5989-3738EN
www.keysight.com
www.keysight.com/find/DMM
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacic
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 11 2626
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-04-23-15)
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the
Web inside your test systems. Keysight is a founding member of the LXI
consortium.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.