Digital MEMS Vibration Sensor with
Embedded RF Transceiver
Data Sheet
ADIS16000/ADIS16229
Rev. 0 Document Feedback
Information furnished by Analog Devices is believed to be accurate and reliable. However, no
responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other
rights of third parties that may result from its use. Specifications subject to change without notice. No
license is granted by implication or otherwise under any patent or patent rights of Analog Devices.
Trademarks and registered trademarks are the property of their respective owners.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.
Tel: 781.329.4700 ©2013 Analog Devices, Inc. All rights reserved.
Technical Support www.analog.com
FEATURES
Wireless vibration system, 902.5 MHz to 927.5 MHz
Clear channel assessment and packet collision avoidance
Error detection and correction in radio frequency (RF)
protocol
Programmable RF output power
Gateway node (ADIS16000)
SPI to RF function
Manage up to 6 sensor nodes
Sensor node (ADIS16229)
Dual axis, ±18 g MEMS accelerometer
5.5 kHz sensor resonant frequency
Digital FFT range settings: 1 g, 5 g, 10 g, and 20 g
Sample rate up to 20 kSPS
Programmable wake-up capture, update cycle times
512-point, real valued FFT
Rectangular, Hanning, and flat top window options
Programmable decimation filter, 11 rate settings
Multirecord capture for selected filter settings
Manual capture mode for time domain data collection
Programmable FFT averaging of up to 255 averages
Record storage: 14 FFT records on two axes (x and y)
Programmable alarms, 6 spectral bands, 2 levels
Adjustable response delay to reduce false alarms
Internal self-test with status flags
Digital temperature and power supply measurements
Identification registers: serial number, device ID, user ID
37.8 mm × 22.8 mm × 8.8 mm MCML package (ADIS16000)
37.8 mm × 22.8 mm × 13.5 mm MCML package (ADIS16229)
Single-supply operation: 3.0 V to 3.6 V
Operating temperature range of −40°C to +85°C
APPLICATIONS
Vibration analysis
Condition monitoring
Machine health
Instrumentation and diagnostics
Safety shutoff sensing
GENERAL DESCRIPTION
The ADIS16000 and ADIS16229 enable the creation of a simple
wireless vibration sensing network for a wide variety of industrial
equipment applications. The ADIS16000 provides the gateway
function, which manages the network, while the ADIS16229
provides the remote sensing function.
The ADIS16229 iSensor® is a complete wireless vibration sensor
node that combines dual axis acceleration sensing with advanced
time domain and frequency domain signal processing. Time
domain signal processing includes a programmable decimation
filter and selectable windowing function. Frequency domain
processing includes a 512-point, real valued fast Fourier transform
(FFT); FFT magnitude averaging; and programmable spectral
alarms. The FFT record storage system offers users the ability to
track changes over time and capture FFTs with multiple
decimation filter settings.
The dynamic range, bandwidth, sample rate, and noise perfor-
mance of the ADIS16229 are well suited for a wide variety of
machine health and production equipment monitoring systems.
This device also provides a number of wireless configuration
parameters, enabling a wide level of flexibility in managing the
trade-off between battery life and communication frequency.
The ADIS16000 serial peripheral interface (SPI) provides simple
connectivity with most embedded processor platforms, and the
SMA connector interface enables the use of many different anten-
nas. This module supports up to six ADIS16229 devices at one
time, using a proprietary wireless protocol.
The ADIS16000 module is available in a 37.8 mm × 22.8 mm ×
8.8 mm multichip chip module laminate (MCML) structure, and
the ADIS16229 is available in a 37.8 mm × 22.8 mm × 13.5 mm
MCML structure. Both have an SMA connector for simple antenna
connection, have two mounting holes for simple installation, and
support operation over a −40°C to +85°C temperature range.
The ADIS16000 also includes a standard 1 mm, 14-pin connector
for connecting to an embedded processor system. The ADIS16229
provides a lead structure that enables simple connection with
battery leads.
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 2 of 32
TABLE OF CONTENTS
Features .............................................................................................. 1
Applications ....................................................................................... 1
General Description ......................................................................... 1
Revision History ............................................................................... 2
Functional Block Diagrams ............................................................. 3
Specifications ..................................................................................... 4
Timing Specifications .................................................................. 6
Absolute Maximum Ratings ............................................................ 7
Thermal Resistance ...................................................................... 7
ESD Caution .................................................................................. 7
Pin Configurations and Function Descriptions ........................... 8
Theory of Operation ...................................................................... 10
Sensing Element ......................................................................... 10
Signal Processing ........................................................................ 10
Sensor Communication ............................................................. 10
Gateway Communication .......................................................... 11
Basic Operation of the ADIS16000 .............................................. 12
SPI Write Commands ................................................................ 12
SPI Read Commands ................................................................. 12
Network Management ................................................................... 15
Sensor Node Recording Mode and Signal Processing ............... 18
Recording Mode ......................................................................... 18
Spectral Record Production ...................................................... 19
Sample Rate and Filtering ......................................................... 19
Dynamic Range and Sensitivity ................................................ 20
Dynamic Range Settings ............................................................ 21
PreFFT Windowing .................................................................... 21
FFT ............................................................................................... 21
FFT Averaging ............................................................................ 22
Recording Times ........................................................................ 22
Data Records ............................................................................... 22
FFT Record Flash Endurance ................................................... 22
Sensor Node Spectral Alarms ....................................................... 23
Alarm Definition ........................................................................ 23
Alarm Indicator Signals ............................................................. 24
Alarm Flags and Conditions ..................................................... 24
Alarm Status ................................................................................ 25
Worst-Case Condition Monitoring.......................................... 25
Reading Output Data ..................................................................... 26
Reading Data from the Data Buffer ......................................... 26
Accessing FFT Record Data ...................................................... 26
Data Format ................................................................................ 27
Real-Time Data Collection ....................................................... 27
Power Supply/Temperature ....................................................... 27
FFT Event Header ...................................................................... 28
Sy stem To ols .................................................................................... 29
Global Commands ..................................................................... 29
Device Identification .................................................................. 29
System Flags ................................................................................ 30
Self-Test ....................................................................................... 30
Flash Memory Management ..................................................... 30
Outline Dimensions ....................................................................... 31
Ordering Guide .......................................................................... 32
REVISION HISTORY
8/13—Revision 0: Initial Version
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 3 of 32
FUNCTIONAL BLOCK DIAGRAMS
Figure 1. ADIS16000 Block Diagram
Figure 2. ADIS16229 Block Diagram
DATA
RECORDS
SPI
ADIS16000
11483-001
CS
SCLK
DIN
DOUT
VDD RST DO1 DO2 ANTENNA
REGISTERS PROCESSOR WIRELESS
TRANSCEIVER
ADIS16229
VDD
PROCESSOR BUFFER FFT FILTER/
WINDOW
DATA
RECORDS CONFIGURATION
REGISTERS
ADC
MEMS
ACCELEROMETER
ANTENNA
WIRELESS
TRANSCEIVER
11483-002
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 4 of 32
SPECIFICATIONS
TA = −40°C to +85°C, VDD = 3.3 V, unless otherwise noted.
Table 1.
Parameter Test Conditions/Comments Min Typ Max Unit
ACCELEROMETERS (ADIS16229)
Measurement Range1 TA = 25°C ±18 g
Sensitivity, FFT TA = 25°C, 0 g to 20 g range setting 0.3052 mg/LSB
Sensitivity, Time Domain TA = 25°C 0.6104 mg/LSB
Sensitivity Error TA = 25°C ±0.3 %
−40°C ≤ TA ≤ +85°C ±6 %
Nonlinearity With respect to full scale ±0.2 ±1.25 %
Cross-Axis Sensitivity 4 %
Alignment Error With respect to package mounting holes 2.3 Degrees
Offset Error TA = 25°C ±0.01 g
−40°C ≤ TA ≤+85°C ±1 g
Offset Temperature Coefficient 2 mg/°C
Output Noise
A
11
m
g
rms
Output Noise Density TA = 25°C, 10 Hz to 1 kHz 0.248 mg/√Hz
Bandwidth ±5% flatness, see Figure 21 840 Hz
Sensor Resonant Frequency 5.5 kHz
LOGIC INPUTS2 (ADIS16000)
Input High Voltage, VINH 0.7 × VDD V
Input Low Voltage, VINL 0.2 × VDD V
Input Leakage Current 0.01 µA
RST, SCLK
0.1
mA
Input Capacitance, CIN 10 pF
DIGITAL OUTPUTS2 (ADIS16000)
Output High Voltage, V
OH
SOURCE
VDD 0.4
V
Output Low Voltage, VOL ISINK = 1 mA 0.36 V
FLASH MEMORY
Endurance
3
20,000
Cycles
Data Retention4 TJ = 85°C, see Figure 25 20 Years
START-UP TIME5
Initial Startup ADIS16000 80 ms
ADIS16229 150 ms
Reset Recovery6 ADIS16000 80 ms
ADIS16229 80 ms
Sleep Mode Recovery ADIS16229 25 ms
SAMPLE RATE Internal sample rate 20 kSPS
Clock Accuracy 3 %
TRANSCEIVER
Receiver Sensitivity −40°C ≤ TA ≤ +85°C −92 dBm
Transmission Power −18 −10.5 dBm
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 5 of 32
Parameter Test Conditions/Comments Min Typ Max Unit
POWER SUPPLY Operating voltage range, VDD 3.0 3.3 3.6 V
Power Supply Current, ADIS16000 Transmission mode, 10 dBm, 25°C 37 mA
Transmission mode, −1 dBm, 25°C 18 mA
Receive mode, 25°C 20 mA
Power Supply Current, ADIS16229
39
41
mA
Transmission mode, 10 dBm, −40°C to +85°C 37.5 mA
Transmission mode, −1 dBm, 25°C 19 26 mA
Transmission mode, −1 dBm, −40°C to +85°C 4 19.5 mA
Receive mode, 25°C 20.5 26 mA
Receive mode, −40°C to +85°C 21.5 mA
7.2
mA
Sleep mode, TA = 25°C 2.5 μA
1 The maximum range depends on the frequency of vibration.
2 The digital input/output signals are 5 V tolerant.
3 Endurance is qualified as per JEDEC Standard 22, Method A117, and measured at −40°C, +25°C, +85°C, and +125°C.
4 Retention lifetime equivalent at junction temperature (TJ) = 85°C as per JEDEC Standard 22, Method A117. Retention lifetime depends on junction temperature.
5 The start-up times presented reflect the time it takes for data collection to begin.
6 Applies to the reset line (RST = 0) and the software reset command (GLOB_CMD_G[7] = 1). The RST pin must be held low for at least 10 μs.
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 6 of 32
TIMING SPECIFICATIONS
TA = 25°C, VDD = 3.3 V, unless otherwise noted. See Figure 3 and Figure 4.
Table 2.
Parameter Description Min Typ Max Unit
fSCLK SCLK frequency 2.5 MHz
tSTALL Stall time 30 µs
tCS CS to SCLK edge 12.9 ns
tSL SCLK low pulse width, not shown in figures 500 ns
tSH SCLK high pulse width, not shown in figures 500 ns
tDAV Data output valid after SCLK edge 47.4 ns
tDSU Data input setup time before SCLK edge 25.8 ns
tDHD Data input hold time after SCLK edge 12.9 ns
t
DF
Data output fall time, not shown in figures
10.6
32.0
ns
tDR Data output rise time, not shown in figures 10.6 32.0 ns
tSR SCLK rise time 10.6 32.0 ns
tSF SCLK fall time 10.6 32.0 ns
tDOCS Data output valid after CS edge, not shown in figures 59.8 ns
t
SFS
CS high after SCLK edge
12.9
ns
Timing Diagrams
Figure 3. SPI Timing and Sequence
Figure 4. DIN Bit Sequence
CS
SCLK
DOUT
DIN
1 2 3 4 5 6 15 16
R/W A5A6 A4 A3 A2 D2
MSB DB14
D1 LSB
DB13 DB12 DB10DB11 DB2 LSBDB1
tCS tSFS
tDAV
tSR tSF
tDHD
tDSU
11483-003
CS
SCLK
t
STALL
11483-004
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 7 of 32
ABSOLUTE MAXIMUM RATINGS
Table 3.
Parameter Rating
Acceleration
Any Axis, Unpowered
2000 g
Any Axis, Powered 2000 g
VDD to GND −0.3 V to +3.96 V
Digital Input Voltage to GND −0.3 V to +3.96 V
Digital Output Voltage to GND −0.3 V to +3.96 V
Temperature
Operating Temperature Range
40°C to +85°C
Storage Temperature Range −65°C to +150°C
Stresses above those listed under Absolute Maximum Ratings
may cause permanent damage to the device. This is a stress
rating only; functional operation of the device at these or any
other conditions above those indicated in the operational
section of this specification is not implied. Exposure to absolute
maximum rating conditions for extended periods may affect
device reliability.
THERMAL RESISTANCE
θJA is specified for the worst-case conditions, that is, a device
soldered in a circuit board for surface-mount packages.
Table 4. Package Characteristics
Package Type θJA θJC Device Weight
14-Pin MCML 31°C/W 11°C/W 6.5 grams
ESD CAUTION
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 8 of 32
PIN CONFIGURATIONS AND FUNCTION DESCRIPTIONS
Figure 5. ADIS16000 Pin Locations
Figure 6. ADIS16000 Pin Configuration
Table 5. ADIS16000 Pin Function Descriptions
Pin No. Mnemonic Type 1 Description
1, 2 VDD S Power Supply, 3.3 V.
3, 4 GND S Ground.
5
DO2
O
Digital Output Line 2.
6, 8, 10 DNC N/A Do not connect to these pins.
7 DOUT O SPI, Data Output. When CS is low, DOUT is an output, and when CS is high, DOUT goes into a
three-state, high impedance mode.
9 SCLK I SPI, Serial Clock.
11 CS I SPI, Chip Select.
12 DIN I SPI, Data Input.
13 DO1 O Digital Output Line 1.
14 RST I Reset, Active Low.
1 S = supply, O = output, I = input, and N/A = not applicable.
PI N 1 PI N 2
PI N 14
11483-005
VDD
GND
CS
RST
DNC
DNC
DNC
SCLK
VDD
GND
DO2
DOUT
DIN
DO1
13
14
11
12
9
10
7
8
5
6
3
4
1
2
ADIS16000
TOP VI EW
(No t t o Scal e)
NOTES
1. THIS REPRESENTATION DISPLAYS THE TOP VI EW W HEN THE
CONNE CTOR IS V ISIBL E AND FACI NG UP.
2. M ATING CO NNE CTO R: M L-14- 4 OR EQUIV ALENT.
3. DNC = DO NO T CO NNE CT.
11483-006
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 9 of 32
Figure 7. ADIS16229 Pin Locations
Table 6. ADIS16229 Pin Function Descriptions
Pin No. Mnemonic Type 1 Description
1 VDD S Power Supply, 3.3 V
2 GND S Ground
1 S = supply.
PI N 1
PI N 2
11483-107
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 10 of 32
THEORY OF OPERATION
The ADIS16000 is the gateway node, and the ADIS16229 serves as
the remote sensor node in a wireless vibration monitoring system.
Using a proprietary wireless protocol, one ADIS16000 can support
up to six ADIS16229 nodes at one time in a local star network
configuration (see Figure 9). As the gateway node, the SPI
interface of the ADIS16000 provides access to an addressable
register map that manages configuration parameters (gateway
and sensor node), remote alarm flags, and remote vibration
data. The SPI interface of the ADIS16000 enables simple
connection to most embedded processors, and its standard
SMA connector supports direct connection to a wide variety of
antennas. The ADIS16229 requires only an antenna and battery
to start up and connect with the ADIS16000 to begin operation.
SENSING ELEMENT
Digital vibration sensing in the ADIS16229 starts with a MEMS
accelerometer core on two different axes. Accelerometers translate
linear changes in velocity into a representative electrical signal,
using a micromechanical system like the one shown in Figure 8.
The mechanical part of this system includes two different frames,
one fixed and one moving, that have a series of plates to form a
variable, differential capacitive network. When experiencing
the force associated with gravity or acceleration, the moving
frame changes its physical position with respect to the fixed
frame, which results in a change in capacitance. Tiny springs
tether the moving frame to the fixed frame and govern the
relationship between acceleration and physical displacement.
A modulation signal on the moving plate feeds through each
capacitive path into the fixed frame plates and into a demodulation
circuit, which produces the electrical signal that is proportional
to the acceleration acting on the device.
Figure 8. MEMS Sensor Diagram
SIGNAL PROCESSING
Figure 10 offers a simplified block diagram for the ADIS16229.
The signal processing stage includes time-domain data capture,
digital decimation/filtering, windowing, FFT analysis, FFT
averaging, and record storage. See Figure 18 for more details
on the signal processing operation.
SENSOR COMMUNICATION
The ADIS16000 provides access to the ADIS16229 through
dedicated pages in the register structure. When the ADIS16000
communicates with a remote ADIS16229, it copies all config-
uration information in these registers to their respective locations
in the ADIS16229 and acquires all of the data in the output
registers and data records of the ADIS16229.
Figure 9. Star Wireless Network Example
Figure 10. ADIS16229 Simplified Block Diagram
MOVABLE
FRAME
ACCELERATION
UNIT
FORCING
CELL
UNIT SENS ING
CELL
MOVING
PLATE
FIXED
PLATES
PLATE
CAPACITORS
ANCHOR
ANCHOR
11483-007
11483-009
EMBEDDED
PROCESSOR ADIS16000 ANTENNA
GATEWAY NODE SENS OR NODE S
ANTENNAADIS16229 BATTERY
ANTENNAADIS16229 BATTERY
11483-010
MEMS
SENSOR ADC PROCESSING DATA
BUFFER
RECORDS
SPI AND
REGISTERS WIRELESS
INTERFACE ANTENNA
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 11 of 32
GATEWAY COMMUNICATION
SPI Interface
The data collection and configuration command uses the SPI,
which consists of four wires. The chip select (CS) signal activates
the SPI interface, and the serial clock (SCLK) synchronizes the
serial data lines. Input commands clock into the DIN pin, one
bit at a time, on the SCLK rising edge. Output data clocks out of
the DOUT pin on the SCLK falling edge. Because the ADIS16000
serves only as an SPI slave, the DOUT contents reflect the
information requested using a DIN command.
Register Organization
The memory map for the ADIS16000 contains seven pages of user
accessible registers, which enable simple organization of both
local (gateway) and remote (sensor) functions. Each page has a
page control register (PAGE_ID), which is located at Address 0x00.
The contents in this location contain the number that represents
the active page. For example, if Address 0x00 has 0x05 in it, Page 5
is the active page. Change the active page assignment by writing a
new value (between 0x00 and 0x06) to Address 0x00. For
example, write 0x02 to Address 0x00 (DIN = 0x8002) to gain
access to the registers in Page 2.
Each 16-bit register has its own unique bit assignment and two
addresses: one for its upper byte and one for its lower byte. Table 10
and Table 11 provide more details on these memory maps, which
list each register, along with its function and lower byte address.
Table 7. ADIS16000 Register Map Page Organization
PAGE_ID Function Reference
0x0000 Gateway configuration Table 10
0x0001 Sensor Node 1 Table 11
0x0002
Sensor Node 2
Table 11
0x0003 Sensor Node 3 Table 11
0x0004 Sensor Node 4 Table 11
0x0005 Sensor Node 5 Table 11
0x0006 Sensor Node 6 Table 11
Dual Memory Structure
The user registers provide addressing for all input/output oper-
ations in the SPI interface. The control registers use a dual memory
structure (see Figure 11). The controller uses static random access
memory (SRAM) registers for normal operation, including user
configuration commands. The flash memory provides nonvolatile
storage for control registers that have flash backup (see Table 10
and Table 11). When the device powers on or resets, the flash
memory contents load into the SRAM, and the device starts
producing data according to the configuration in the control
registers. Storing configuration data in the flash memory requires
a manual flash update command. For the ADIS16000, set
DIN = 0x8000 (access Page 0), then set DIN = 0x9240 (set
GLOB_CMD_S[6] = 1). For a remote ADIS16229, use the
following steps to update its flash:
1. Turn to its page (for example, set DIN = 0x8001 to access
Sensor Node 1).
2. Set DIN = 0xB640 (GLOB_CMD_S[6] = 1).
3. Set DIN = 0x8000 (turn to Page 0).
4. Set DIN = 0x9202 (GLOB_CMD_G[1] = 1).
Figure 11. SRAM and Flash Memory Diagram
NONVOLATILE
FLASH MEMORY
(NO SPI ACCESS)
MANUAL
FLASH
BACKUP
START-UP
RESET
VOLATILE
SRAM
(SPI ACCESS)
11483-008
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 12 of 32
BASIC OPERATION OF THE ADIS16000
When the ADIS16000 has appropriate power on the VDD pin, it
automatically begins a self-initialization process. When this
process is completed, the SPI interface is activated and provides
access to its register structure. The SPI interface supports
connectivity with most embedded processor platforms, using
the connection diagram in Figure 12. The factory default
configuration for DO1 provides a busy indicator signal that
indicates when to avoid SPI communication requests.
Figure 12. Electrical Hook-Up Diagram
Table 8. Generic Master Processor Pin Names and Functions
Pin Name Function
SS Slave select
SCLK Serial clock
MOSI Master output, slave input
MISO
Master input, slave output
IRQ1, IRQ2 Interrupt request inputs (optional)
The ADIS16000 SPI interface supports full duplex serial
communication (simultaneous transmit and receive) and uses
the bit sequence shown in Figure 16. Table 9 provides a list of
the most common settings that require attention to initialize
a processor serial port for the ADIS16000 SPI interface.
Table 9. Generic Master Processor SPI Settings
Processor Setting Description
Master
ADIS16000 operates as slave
SCLK Rate ≤ 2.5 MHz Bit rate setting
SPI Mode 3 Clock polarity/phase
(CPOL = 1, CPHA = 1)
MSB First Bit sequence
16-Bit Shift register/data length
Table 10 and Table 11 provide lists of user registers with their lower
byte addresses. Each register consists of two bytes, each of which
has its own unique 7-bit address. Figure 13 relates the bits of
each register to their upper and lower addresses.
Figure 13. Generic Register Bit Definitions
SPI WRITE COMMANDS
User control registers govern many internal operations. The
DIN bit sequence in Figure 16 provides the ability to write to
these registers, one byte at a time. Some configuration changes
and functions require only one write cycle. For example, set
PAGE_ID[7:0] = 1 (DIN = 0x8001) to select Page 1 of the
register map.
Figure 14. SPI Sequence for Selecting Page 1 for Access (DIN = 0x8001)
SPI READ COMMANDS
A single register read requires two 16-bit SPI cycles that also use
the bit assignments shown in Figure 16. The first sequence
sets R/W = 0 and communicates the target address
(Bits[A6:A0]). Bits[D7:D0] are don’t care bits for a read DIN
sequence. DOUT clocks out the requested register contents
during the second sequence. The second sequence can also use
DIN to set up the next read. Figure 15 provides a signal diagram
for all four SPI signals while reading the PROD_ID_G. In this
diagram, DIN = 0x1600 and DOUT reflects the decimal
equivalent of 16,000.
Figure 15. Example SPI Read, PROD_ID_G (Page 0), Second Sequence
Figure 16. Example SPI Read Sequence
SYSTEM
PROCESSOR
SPI MASTER ADIS16000
SCLK
CS
DIN
DOUT
SCLK
SS
MOSI
MISO
3.3V
IRQ2
IRQ1
DO2
VDD I/O LINES ARE COMPATIBLE WITH
3.3V OR 5V LOGIC LEVELS
11
9
12
7
5
DO1
13
12
34
11483-011
UPPER BYTE
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
LOWER BYTE
11483-012
CS
DIN
SCLK
11483-013
DOUT = 0011 1110 1000 0000 = 0x3E80 = 16,000 = P ROD_I D_G
SCLK
CS
DIN
DOUT
11483-014
R/W R/W
A6 A5 A4 A3 A2 A1 A0 D7 D6 D5 D4 D3 D2 D1 D0
DB0DB1DB2DB3DB4DB5DB6DB7DB8DB9DB10DB11DB12DB13DB14DB15
NOTES
1. DOUT BITS ARE BASED ON THE PREVIOUS 16-BIT SEQUENCE (R/W = 0).
CS
SCLK
DIN
DOUT
A6 A5
DB13DB14
DB15
11483-015
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 13 of 32
Table 10. User Register Memory Map, PAGE_ID = 0x0000
Register Name Access1 Flash Backup1 Address Default1 Function Reference
PAGE_ID Read/write N/A 0x00 0x0000 Page identifier Table 7
NETWORK_ID Read/write Yes 0x02 0x1234 Network identifier, unique to a network Table 14
FLASH_CNT_G Read only Yes 0x04 N/A Flash update counter Table 87
NW_ERROR_STAT Read only No 0x06 0x0000 Network error indicators Table 26
TX_PWR_CTRL_G Read/write Yes 0x08 0x0008 Transmission power control, gateway Table 18
RSSI_G Read only No 0x0A N/A Received signal strength Table 16
TEMP_OUT_G Read only No 0x0C N/A Output, temperature Table 67
SUPPLY_OUT_G Read only No 0x0E N/A Output, supply voltage Table 64
TEST_MODE Read/write Yes 0x10 0x0000 FCC test support modes Table 23
GLOB_CMD_G
Write only
No
0x12
N/A
System commands
Table 13
CMD_DATA Read/write No 0x14 0x0000 Data to sensor nodes Table 12
PROD_ID_G Read only Yes 0x16 0x3E80 Product identifier, ADIS16000 Table 80
Reserved N/A No 0x18 N/A Reserved
LOT_ID1_G Read only No 0x1A N/A Lot Identifier 1 Table 76
LOT_ID2_G Read only No 0x1C N/A Lot Identifier 2 Table 78
Reserved
N/A
No
0x1E
N/A
Reserved
Reserved N/A No 0x20 N/A Reserved
Reserved N/A No 0x22 N/A Reserved
SERIAL_NUM_G Read only No 0x24 N/A Serial number, lot specific Table 82
Reserved N/A No 0x26 N/A Reserved
Reserved
N/A
No
0x28
N/A
Reserved
GPO_CTRL Read/write Yes 0x2A 0x0008 General-purpose output control Table 84
DIAG_STAT_G Read only No 0x2C 0x0000 Diagnostics, status flags Table 85
1 N/A = not applicable.
Table 11. User Register Memory Map, PAGE_ID ≥ 0x0001 (Registers Representing Remote ADIS16229 Units)
Register Name Access1
Flash
Backup1 Address Default1, 2 Function Reference
PAGE_ID Read/write N/A 0x00 N/A Page identifier Table 7
SENS_ID
Read only
Yes
0x02
N/A
Sensor identifier
Table 15
FLASH_CNT_S Read only Yes 0x04 N/A Status, flash memory write count Table 88
X_BUF Read only No 0x06 N/A Output, buffer for x-axis acceleration data Table 60
Y_BUF Read only No 0x08 N/A Output, buffer for y-axis acceleration data Table 61
TEMP_OUT_S Read only No 0x0A N/A Output, temperature during capture Table 68
SUPPLY_OUT_S Read only No 0x0C N/A Output, power supply during capture Table 65
FFT_AVG1 Read/write Yes 0x0E 0x0108 Control, FFT average, SR0 and SR1 Table 35
FFT_AVG2 Read/write Yes 0x10 0x0101 Control, FFT average, SR2 and SR3 Table 36
BUF_PNTR Read/write No 0x12 0x0000 Control, buffer address pointer Table 58
REC_PNTR Read/write No 0x14 0x0000 Control, record address pointer Table 59
X_SENS Read/write No 0x16 N/A Control, x-axis scale correction factor Table 33
Y_SENS
Read/write
No
0x18
N/A
Control, y-axis scale correction factor
Table 34
REC_CTRL1 Read/write Yes 0x1A 0x1102 Record control register Table 28
REC_CTRL2 Read/write Yes 0x1C 0x00FF Record control register Table 31
Reserved N/A No 0x1E N/A Reserved
ALM_F_LOW Read/write Yes 0x20 0x0000 Spectral alarm band, low frequency Table 44
ALM_F_HIGH Read/write Yes 0x22 0x0000 Spectral alarm band, high frequency Table 45
ALM_X_MAG1
Read/write
Yes
0x24
0x0000
Spectral alarm band, X-Axis Alarm Trigger Level 1
magnitude
Table 46
ALM_Y_MAG1 Read/write Yes 0x26 0x0000 Spectral alarm band, Y-Axis Alarm Trigger Level 1
magnitude
Table 47
ALM_X_MAG2 Read/write Yes 0x28 0x0000 Spectral alarm band, X-Axis Alarm Trigger Level 2
magnitude
Table 48
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 14 of 32
Register Name Access1
Flash
Backup1 Address Default1, 2 Function Reference
ALM_Y_MAG2 Read/write Yes 0x2A 0x0000 Spectral alarm band, Y-Axis Alarm Trigger Level 2
magnitude
Table 49
ALM_PNTR Read/write No 0x2C 0x0000 Spectral alarm band pointer Table 43
ALM_S_MAG Read/write Yes 0x2E 0x0000 Alarm, system alarm trigger level Table 50
ALM_CTRL Read/write Yes 0x30 0x0080 Alarm, control register Table 42
AVG_CNT Read/write Yes 0x32 0x9630 Sample rate control (average count) Table 29
DIAG_STAT_S Read only No 0x34 0x0000 System status register Table 86
GLOB_CMD_S Write only No 0x36 N/A Global command register Table 75
ALM_X_STAT
Read only
No
0x38
0x0000
Alarm, x-axis status register
Table 51
ALM_Y_STAT Read only No 0x3A 0x0000 Alarm, y-axis status register Table 52
ALM_X_PEAK Read only No 0x3C 0x0000 Alarm, x-axis peak level Table 53
ALM_Y_PEAK Read only No 0x3E 0x0000 Alarm, y-axis peak level Table 54
TIME_STMP_L Read only No 0x40 0x0000 Time stamp, low integer Table 73
TIME_STMP_H Read only No 0x42 0x0000 Time stamp, high integer Table 74
ALM_X_FREQ
Read only
No
0x44
0x0000
Alarm frequency of x-axis ALM_X_PEAK
Table 55
ALM_Y_FREQ Read only No 0x46 0x0000 Alarm frequency of y-axis ALM_Y_PEAK Table 56
PROD_ID_S Read only Yes 0x48 0x3F65 Product identification register Table 81
REC_FLSH_CNT Read only No 0x4A N/A Flash write cycle count Table 40
REC_INFO1 Read only No 0x4C 0x0008 Record Settings 1 Table 71
REC_INFO2 Read only No 0x4E 0x0000 Record Settings 2 Table 72
REC_CNTR Read only No 0x50 0x0000 Record counter Table 38
PKT_TIME_L Read only No 0x52 N/A Received packet time stamp, low integer Table 24
PKT_TIME_H Read only No 0x54 N/A Received packet time stamp, high integer Table 25
PKT_ERROR_STAT Read only No 0x56 0x0000 Missed packets/error indicator Table 27
TX_PWR_CTRL_S Read/write Yes 0x58 0x0008 Transmission power control Table 19
RSSI_S
Read only
No
0x5A
N/A
Received signal strength indicator
Table 17
RF_MODE Read/write Yes 0x5C 0x0000 Wireless communication configuration Table 20
UPDAT_INT Read/write Yes 0x5E 0x0FA0 Update interval Table 21
INT_SCL Read/write Yes 0x60 0x0001 Update interval scale Table 22
Reserved N/A N/A 0x62 N/A Reserved
USER_SCR Read only Yes 0x64 0x0000 User scratch register Table 83
Reserved N/A N/A 0x66 N/A Reserved
LOT_ID1_S Read only Yes 0x68 N/A Lot Identification Code 1 Table 77
LOT_ID2_S Read only Yes 0x6A N/A Lot Identification Code 2 Table 79
Reserved N/A N/A 0x6C N/A Reserved
Reserved N/A N/A 0x6E N/A Reserved
1 N/A = not applicable.
2 All registers in Page 1, Page 2, Page 3, Page 4, Page 5, and Page 6 read 0x0000 prior to connecting with the ADIS16229.
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 15 of 32
NETWORK MANAGEMENT
When the ADIS16000 and the ADIS16229 have an appropriate
supply voltage across their VDD and GND pins, they both self-
initialize and prepare themselves for connecting. After completing
this process, the ADIS16229 starts sending connection requests
to any available ADIS16000 devices within range. The system
microcontroller manages the response of the ADIS16000 to
these requests, using the CMD_DATA (see Table 12) and
GLOB_CMD_G registers (see Table 13), which are both in
Page 0 of the ADIS16000. Adding an ADIS16229 network
requires the following two steps:
1. Set GLOB_CMD_G[0] = 1 (DIN = 0x9201, in Page 0).
2. Wait 500 µs and then write the node number (between 0
and 6) to the CMD_DATA register.
After this second step, the connection process can take up to
10 seconds after writing this code. Removing a sensor from the
network uses a similar two-step process: write the sensor node
number to the CMD_DATA register, and then set
GLOB_CMD_G[8] = 1 (DIN = 0x9301, Page 0).
Table 12. CMD_DATA,
Page 0, Low Byte Address = 0x14, Read/Write
Bits Description (Default = 0x0000)
[15:4] Not used
[3:0] Sensor node for GLOB_CMD_G[1] and
GLOB_CMD_G[0] commands, range = 000 (0) to 110 (6)
Table 13. GLOB_CMD_G,
Page 0, Low Byte Address = 0x12, Write Only
Bits Description (Default = Not Applicable)
15 Remove the sensor node identified by the CMD_DATA
register
14 Not used
13 Request current configuration and payload from the
sensor node identified by the CMD_DATA register
12 Manual update of alarm registers for the sensor node
identified by the CMD_DATA register
11 Read alarm registers from the sensor node identified
by the CMD_DATA register
10 Transfer alarm configuration from ADIS16000 page to
remote sensor registers
9 Not used
8
Remove sensor node in CMD_DATA from the network
7 Software reset
6 Save registers to flash memory
5 Flash test, compare sum of flash memory with factory
value
4 Clear DIAG_STAT_G register
3 Restore factory register settings, including capture
buffer and alarm registers
2 Not used
1 Update sensor node in CMD_DATA register, in one of
the manual modes
0
Add sensor node in CMD_DATA to the network
Set GLOB_CMD_G[1] = 1 (DIN = 0x9202) to initialize an
update of all of the registers, except those associated with the
spectral alarms. Set GLOB_CMD_G[12] = 1 (DIN = 0x9310) to
update all of the alarm registers, after configuring them. Separating
this function helps manage the flash memory endurance. The
NETWORK_ID register (see Table 14) provides a user configurable
identification number in its lower byte and provides a separate
8-bit number for application-specific information that does not
influence the operation of the ADIS16000.
Table 14. NETWORK_ID,
Page 0, Low Byte Address = 0x02, Read/Write
Bits Description (Default = 0x1234)
[15:8]
User configurable (no internal function)
[7:0] User configurable, 8-bit network identification number
The SENS_ID register (see Table 15) contains the value 0x0000
when a particular page in the ADIS16000 memory map is not
managing a specific ADIS16229 connection. When a page is
managing a connection with an ADIS16229, the lower byte
indicates the assigned node number within the network.
Table 15. SENS_ID,
Page 1 to Page 6, Low Byte Address = 0x02, Read Only
Bits Description (Default = Not Applicable)
[15:8] When connected, it is 0xAD.
[7:0] Sensor node number. When connected, the range is 0 to 6.
Receiver Signal Strength
The RSSI_G and RSSI_S registers (see Table 16 and Table 17)
provide tools for tuning the transmission power control at each
location. Keep the transmission power high enough to maintain at
least −92 dBm in these registers for best communication reliability.
Table 16. RSSI_G,
Page 0, Low Byte Address = 0x0A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Received signal strength. This is a 16-bit twos comple-
ment number, where 0x0000 represents 0 dBm and
each LSB represents 1 dBm of change in the received
signal strength. Examples include the following:
0x0001 = +1 dBm.
0xFFFF = −1 dBm.
0xFFA0 = −92 dBm.
Table 17. RSSI_S,
Page 1 to Page 6, Low Byte Address = 0x5A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Received signal strength. This is a 16-bit twos comple-
ment number, where 0x0000 represents 0 dBm and
each LSB represents 1 dBm of change in the received
signal strength. Examples include the following:
0x0001 = +1 dBm.
0xFFFF = −1 dBm.
0xFFA0 = −92 dBm.
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 16 of 32
Transmission Power Control
Both the ADIS16000 and ADIS16229 units provide controls for
transmission power in the TX_PWR_CTRL_G (see Table 18)
and TX_PWR_CTRL_S (see Table 19) registers. The registers
provide users with the ability to optimize the transmission power
for battery optimization and to manage interference influence
on other networks. Note that compliance with FCC Part 15.247
involves limiting the transmission power to −1 dBm.
Table 18. TX_PWR_CTRL_G,
Page 0, Low Byte Address = 0x08, Read/Write
Bits
Description (Default = 0x0008)
[15:14] Not used (don’t care).
[13:8] Channel frequency. This is a 6-bit, offset binary number
that selects one of 51 channel frequencies. The lowest
setting, 000000 (binary for 0 LSB), represents
902.5 MHz, and each LSB represents an increase of
0.5 MHz. The maximum setting, 110010 (binary for
50 LSB), selects a channel frequency of 927.5 MHz.
[7:5] Not used (don’t care).
[4:0] Transmission power setting. This is a 5-bit, offset binary
number that selects the transmission power in the
gateway node. The lowest setting, 00000 (binary for
0 LSB), selects a transmission power of −15.5 dBm, and
each LSB increases the transmission power by
+1.7 dBm. The maximum setting, 01111, selects a
transmission power of 10 dBm.
Table 19. TX_PWR_CTRL_S,
Page 1 to Page 6, Low Byte Address = 0x58, Read/Write
Bits Description (Default = 0x0008)
[15:14] Not used (don’t care).
[13:8] Channel frequency. This is a 6-bit, offset binary number
that selects one of 51 channel frequencies. The lowest
setting, 000000 (binary for 0 LSB), represents
902.5 MHz, and each LSB represents an increase of
0.5 MHz. The maximum setting, 110010 (binary for
50 LSB), selects a channel frequency of 927.5 MHz.
[7:5] Not used (don’t care).
[4:0]
Transmission power setting. This is a 5-bit, offset binary
number that selects the transmission power in the
gateway node. The lowest setting, 00000 (binary for
0 LSB), selects a transmission power of −15.5 dBm, and
each LSB increases the transmission power by
+1.7 dBm. The maximum setting, 01111, selects a
transmission power of 10 dBm.
Wireless Configuration
The RF_MODE register (see Table 20) provides a number of
important wireless configuration parameters. Note that when
the transmission power exceeds −1 dBm and the update period
is less than 10 seconds, FCC Part 15.247 requires the use of
frequency hopping.
Table 20. RF_MODE,
Page 1 to Page 6, Low Byte Address = 0x5C, Read/Write
Bits Description (Default = 0x0000)
[15:12] Not used (don’t care)
[11] Continuous transmission mode (single-channel, per
TX_PWR_CTRL_S, see Table 19)
[10] Continuous frequency hopping mode
[9:7] Not used (don’t care)
[6] Frequency hopping (0 = disable, 1 = enable)
[5:2] Not used (don’t care)
[1] Update gateway on alarm only (0 = disable, 1 = enable)
[0]
Not used (don’t care)
The UPDAT_INT and INT_SCL registers (see Table 21 and
Table 22) establish the time between wake-up events, where the
remote ADIS16229 captures data, analyzes it, and communicates
the information.
Table 21. UPDAT_INT,
Page 1 to Page 6, Low Byte Address = 0x5E, Read/Write
Bits Description (Default = 0x0FA0)
[15:0] Offset binary number, scale factor set by INT_SCL register
Table 22. INT_SCL,
Page 1 to Page 6, Low Byte Address = 0x60, Read/Write
Bits Description (Default = 0x0001)
[15:2]
Not used (don’t care)
[1:0] Scale factor
00 = 30.52 µs/LSB, maximum = 2 seconds
01 = 0.488 ms/LSB, maximum = 31.98 seconds
10 = 1/128 sec/LSB, maximum = 512 seconds
11 = 1 sec/LSB, maximum = 18.2 hours
The TEST_MODE register (see Table 23) provides specific test
modes for FCC testing.
Table 23. TEST_MODE,
Page 0, Low Byte Address = 0x10, Read/Write
Bits Description (Default = 0x0000)
[15:12] Not used
11 Continuous transmission mode (single-channel, per
TX_PWR_CTRL_G, see Table 18)
10
Continuous frequency hopping mode
[9:0] Not used
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 17 of 32
Communication Tools
The PKT_TIME_H (upper word) and PKT_TIME_L (lower
word) registers provide a 32-bit timer for tracking the relative
times associated with packet transmission times. When the timer
reaches its maximum value of 0xFFFFFFFF, it automatically
starts over at 0x00000000.
Table 24. PKT_TIME_L,
Page 1 to Page 6, Low Byte Address = 0x52, Read Only
Bits Description (Default = Not Applicable)
[15:0] Offset binary number, lower word, 1 LSB = 0.488 ms
Table 25. PKT_TIME_H,
Page 1 to Page 6, Low Byte Address = 0x54, Read Only
Bits Description (Default = Not Applicable)
[15:0] Offset binary number, upper word
The NW_ERROR_STAT register (see Tabl e 26) provides all of
the error flags associated with the wireless communication.
Table 26. NW_ERROR_STAT,
Page 0, Low Byte Address = 0x06, Read Only
Bits Description (Default = 0x0000)
[15:11] Not used
10 Invalid packet received
9 Received packet from an unknown device
8 Packet synchronization failure from the most recent
received packet
[7:6] Not used
5 Packet length mismatch
4 Missing packet
3 Packets received out of sync
2 Failure to receive acknowledgment from a sensor node
1 Low signal strength from a sensor node, read RSSI_S
register for power level of this signal (see Table 17)
0 CRC mismatch error associated with the most recent
packet from the sensor node packet
Table 27. PKT_ERROR_STAT,
Page 1 to Page 6, Low Byte Address = 0x56, Read Only
Bits Description (Default = 0x0000)
[15:10] Not used
9 Received packet from an unknown device
8 Packet synchronization failure from the most recent
received packet
[7:6] Not used
5 Packet length mismatch
4 Missing packet
3 Packets received out of sync
2 Failure to receive acknowledgment from a sensor node
1 Low signal strength from a sensor node, read RSSI_S
register for power level of this signal (see Table 17)
0 CRC mismatch error associated with the most recent
packet from the sensor node packet
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 18 of 32
SENSOR NODE RECORDING MODE AND SIGNAL PROCESSING
The ADIS16229 provides a complete sensing system for recording
and monitoring vibration data. Figure 17 provides a simplified
block diagram for the signal processing associated with spectral
record acquisition on both axes (x and y). User registers provide
controls for data type (time or frequency), trigger mode (manual or
automatic), collection mode (real time or capture), sample rates
and filtering, windowing, FFT averaging, spectral alarms, and
input/output management.
RECORDING MODE
The recording mode selection establishes the data type (time or
frequency domain), trigger type (manual or automatic), and data
collection (captured or real time). The REC_CTRL1[1:0] bits
(see Table 28) provide four operating modes: manual FFT, auto-
matic FFT, manual time capture, and real time. After REC_CTRL1
is set, the manual FFT, automatic FFT, and manual time capture
modes require a start command to start acquiring a spectral or
time domain record. All of these modes automatically trigger
when the sensor receives the configuration packet from the
gateway. Set GLOB_CMD_S[11] = 1 to halt the operation and
wait for further instructions from the ADIS16000.
Manual FFT Mode
Set REC_CTRL1[1:0] = 00 to place the device in manual FFT
mode, which triggers a single FFT cycle. When the spectral record
is complete, the device transmits the data to the ADIS16000 and
waits for another start command.
Automatic FFT Mode
Set REC_CTRL1[1:0] = 01 to place the device in automatic FFT
mode. Use the UPDAT_INT and INT_SCL registers to establish
the period between wake-up times, which triggers data capture,
FFT computation, and analysis.
Manual Time Capture Mode
Set REC_CTRL1[1:0] = 10 to place the device into manual time
capture mode, which results in triggering a single time domain
data capture. When the device operates in this mode, 512 samples
of time domain data are loaded into the buffer for each axis.
This data goes through all time domain signal processing,
except the preFFT windowing, prior to loading into the data
buffer for user access. When the data record is complete, the
device transmits the data to the ADIS16000 and waits for
another start command.
Real-Time Mode
Set REC_CTRL1[1:0] = 11 to place the device into real-time
mode. In this mode, the device samples only one axis, at a rate
of 5 kSPS, and provides data on its output register at the SR0
sample rate setting in AVG_CNT[3:0] (see Table 29). Select
the axis of measurement in this mode by reading its assigned
register. For example, select the x-axis by reading X_BUF, using
DIN = 0x0600. See Table 60 or Tabl e 61 for more information
on the x_BUF registers. No other ADIS16229 nodes are able to
communicate with the ADIS16000 when one of the ADIS16229
nodes is in real-time mode.
Table 28. REC_CTRL1,
Page 1 to Page 6, Low Byte Address = 0x1A, Read/Write
Bits Description (Default = 0x1102)
[15:14] Not used (don’t care)
[13:12] Window setting;
00 = rectangular, 01 = Hanning, 10 = flat top, 11 = N/A1
11 SR3: 1 = enabled for FFT, 0 = disabled;
Sample rate = 20,000 ÷ 2AVG_CNT[15:12] (see Table 29)
10 SR2: 1 = enabled for FFT, 0 = disabled;
Sample rate = 20,000 ÷ 2AVG_CNT[11:8] (see Table 29)
9 SR1: 1 = enabled for FFT, 0 = disabled;
Sample rate = 20,000 ÷ 2AVG_CNT[7:4] (see Table 29)
8 SR0: 1 = enabled for FFT, 0 = disabled;
Sample rate = 20,000 ÷ 2AVG_CNT[3:0] (see Table 29)
7 Power-down between each recording, 1 = enabled
[6:4] Not used (don’t care)
[3:2] Storage method;
00 = none, 01 = alarm trigger, 10 = all, 11 = N/A1
[1:0] Recording mode;
00 = manual FFT, 01 = automatic FFT, 10 = manual time
capture, 11 = real-time sampling/data access
1 N/A = not applicable.
Figure 17. Simplified Block Diagram
11483-016
MEMS ADC PROCESSING DATA
BUFFER
RECORDS
SPI AND
REGISTERS
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 19 of 32
SPECTRAL RECORD PRODUCTION
The ADIS16229 produces a spectral record by taking a time record
of data on both axes, then scaling, windowing, and performing
an FFT process on each time record. This process is repeated for
a programmable number of FFT averages, with the FFT result
of each cycle accumulating in the data buffer. After completing
the selected number of cycles, the FFT averaging process completes
by scaling the data buffer contents. Then the data buffer contents
are available to the SPI and output data registers.
SAMPLE RATE AND FILTERING
The sample rate for each axis is 20 kSPS. The internal ADC samples
both axes in a time-interleaving pattern (x1, y1, x2, y2, and so on)
that provides even distribution of data across the data record.
The averaging/decimating filter provides a control for the final
sample rate in the time record. By averaging and decimating the
time domain data, this filter provides the ability to focus the
spectral record on lower bandwidths, producing finer frequency
resolution in each FFT frequency bin. The AVG_CNT register (see
Table 29) provides the settings for the four different sample rate
options in REC_CTRL1[11:8] (SRx in Table 28). All four options
are available when using the manual FFT, automatic FFT, and
manual time capture modes. When more than one sample rate
option is enabled while the device is in one of the manual modes,
the device produces a spectral record for one SRx at a time,
starting with the lowest number. After completing the spectral
record for one SRx option, the device waits for another start
command before producing a spectral record for the next SRx
option that is enabled in REC_CTRL1[11:8]. When more than one
sample rate option is enabled while the device is in the automatic
FFT mode, the device produces a spectral record for one SRx
option and then waits for the next automatic trigger, which
occurs based on the UPDAT_INT and INT_SCL registers (see
Table 21 and Table 22). See Figure 19 for more details on how
multiple SRx options influence data collection and spectral
record production. When real-time mode is used, the output
data rate reflects the SR0 setting.
Figure 18. Signal Flow Diagram, REC_CTRL1[1:0] = 00 or 01, FFT Analysis Modes
Figure 19. Spectral Record Production, with All SRx Settings Enabled
K = 1
x
K
N
A
N
a
1÷N
A
DUAL
AXIS
MEMS
ACCEL20.48kSPS
K
o
K
s
FFT FFT
AVERAGE
(N
F
)
SAMPLE RATE SETTING
REC_CTRL1[11:8]
WINDOW SETTING
REC_CTRL1[13:12]
RANGE - S CALE SE TT ING
K
s
= A
MAX
÷ 2
15
A
MAX
= PEAK FRO M RE C_CTRL 2[ 7: 0]
N
F
= # OF AVERAGES
N
F
= FFT_AVG
WINDOW
FFT
RECORD
1
FFT
RECORD
m
FFT
RECORD
13
FFT
RECORD
0
FFT RE CORDS— NONVOLATILE F L ASH MEMORY
m = REC_CNT R
11483-017
DATA
BUFFER
SPI
REGISTER
ACCESS
SENSITIVI T Y ADJUSTMENT
X_SE NS AND Y _S E NS
X1Y1Y2PWR2TEMP2
X512 Y512
X2DATA CAP TURE
FFT RECORD
RECORDS
TRIGGER
SPI/DO/TIMER TRIGGER
SPI/DO/TIMER
DATA RDY
TRIGGER
SPI/DO/TIMER
DATA RDY
TRIGGER
SPI/DO/TIMER
DATA RDY DATA RDY
RECO RD 1
SR0 RECORD 1
SR1 RECORD 1
SR2 RECORD 1
SR3
512 SAM PLES 512 SAMPLES 512 SAMPLES
11483-018
FFT1FFT2FFTNFFT
AVG TEMP
AVG
PWR
AVG
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 20 of 32
Table 30 provides a list of SRx settings available in the AVG_CNT
register (see Tabl e 29), along with the resulting sample rates, FFT
bin widths, bandwidth, and estimated total noise. Note that each
SRx setting also has associated range settings in the REC_CTRL2
register (see Table 31) and the FFT averaging settings that are
shown in the FFT_AVG1 and FFT_AVG2 registers (see Table 35
and Table 36, respectively).
Table 29. AVG_CNT,
Page 1 to Page 6, Low Byte Address = 0x32, Read/Write
Bits Description (Default = 0x9630)
[15:12] Sample Rate Option 3 (SR3), binary (0 to 10),
SR3 option sample rate = 20,000 ÷ 2AVG_CNT[15:12]
[11:8] Sample Rate Option 2 (SR2), binary (0 to 10),
SR2 option sample rate = 20,000 ÷ 2AVG_CNT[11:8]
[7:4] Sample Rate Option 1 (SR1), binary (0 to 10),
SR1 option sample rate = 20,000 ÷ 2AVG_CNT[7:4]
[3:0] Sample Rate Option 0 (SR0), binary (0 to 10),
SR0 option sample rate = 20,000 ÷ 2AVG_CNT[3:0]
Table 30. Sample Rate Settings and Filter Performance
SRx
Setting
Sample Rate, f
S
(SPS)
Bin Width
(Hz)
Bandwidth
(Hz)
0 20,000 39.1 10,000
1 10,000 19.5 5,000
2 5,000 9.8 2,500
3 2,500 4.9 1,250
4 1,250 2.4 625
5
625
1.2
313
6 313 0.6 156
7 156 0.3 78
8 78 0.2 39
9 39 0.1 20
10 20 0.0 10
DYNAMIC RANGE AND SENSITIVITY
The range of the ADIS16229 accelerometers depends on the
frequency of the vibration. The accelerometers have a self-
resonant frequency of 5.5 kHz, and the signal conditioning
circuit applies a single-pole, low-pass filter (2.5 kHz) to the
response. The self-resonant behavior of the accelerometer
influences the relationship between vibration frequency and
dynamic range. Figure 20 displays the response to peak input
amplitudes, assuming a sinusoidal vibration signature at each
frequency. The accelerometer resonance and low-pass filter also
influence the magnitude response, as shown in Figure 21.
Figure 20. Peak Magnitude vs. Frequency
Figure 21. Magnitude/Frequency Response
20
0
2
4
6
8
10
12
14
16
18
1000 2000 4000 5000 6000
PEAK M AGNITUDE (g)
FRE QUENCY ( Hz )
11483-019
2g PEAK RESPONSE
14g PEAK RESPONSE
16g PEAK RESPONSE
18g PEAK RESPONSE
100 1000 5000
MEAN
MAG NITUDE ( g)
FRE QUENCY ( Hz )
11483-020
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
–3σ
+3σ
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 21 of 32
DYNAMIC RANGE SETTINGS
REC_CTRL2 (see Table 31) provides four range settings that are
associated with each sample rate option, SRx. The range options
referenced in REC_CTRL2 reflect the maximum dynamic range,
which occurs at the lower part of the frequency range and does
not account for the decrease in range (see Figure 20). For example,
set REC_CTRL2[5:4] = 10 (DIN = 0x9C20) to set the peak acceler-
ation (AMAX) to 10 g on the SR2 sample rate option. These settings
optimize FFT precision and sensitivity when monitoring lower
magnitude vibrations. For each range setting in Table 31, this
stage scales the time domain data so that the maximum value
equates to 215 LSBs for time domain data and 216 LSBs for
frequency domain data.
Note that the maximum range for each setting is 1 LSB smaller
than the listed maximum. For example, the maximum number of
codes in the frequency domain analysis is 216 − 1, or 65,535. When
using a range setting of 1 g in one of the FFT modes, the maxi-
mum measurement is equal to 1 g times 216 − 1, divided by 216.
See Table 32 for the resolution associated with each setting and
Figure 18 for the location of this operation in the signal flow
diagram. The real-time mode automatically uses the 20 g
range setting.
Table 31. REC_CTRL2,
Page 1 to Page 6, Low Byte Address = 0x1C, Read/Write
Bits Description (Default = 0x00FF)
[15:8] Not used (don’t care)
[7:6] Measurement range, SR3;
11 = 1 g, 10 = 5 g, 01 = 10 g, 00 = 20 g
[5:4] Measurement range, SR2;
11 = 1 g, 10 = 5 g, 01 = 10 g, 00 = 20 g
[3:2] Measurement range, SR1;
11 = 1 g, 10 = 5 g, 01 = 10 g, 00 = 20 g
[1:0] Measurement range, SR0;
11 = 1 g, 10 = 5 g, 01 = 10 g, 00 = 20 g
Table 32. Range Settings and LSB Weights
Range Setting (g)
Time Mode
(mg/LSB)
FFT Mode
(mg/LSB)
0 to 1 0.0305 0.0153
0 to 5 0.1526 0.0763
0 to 10 0.3052 0.1526
0 to 20 0.6104 0.3052
Scale Adjustment
The x_SENS registers (see Table 33 and Table 34) provide a
fine-scale adjustment function for each axis. The following
equation describes how to use measured and ideal values to
calculate the scale factor for each register in LSBs:
SCFx =
1
XM
XI a
a × 218
where:
aXI is the ideal x-axis value.
aXM is the actual x-axis measurement.
These registers contain correction factors, which come from the
factory calibration process. The calibration process records accel-
erometer output in four different orientations and computes the
correction factors for each register.
These registers also provide write access for in-system adjustment.
Gravity provides a common stimulus for this type of correction
process. Use both +1 g and −1 g orientations to reduce the effect
of offset on this measurement. In this case, the ideal measurement
is 2 g, and the measured value is the difference of the accelero-
meter measurements at +1 g and −1 g orientations. The factory-
programmed values are stored in flash memory and are restored
by setting GLOB_CMD_S[3] = 1 (DIN = 0xB604). See Table 75.
Table 33. X_SENS,
Page 1 to Page 6, Low Byte Address = 0x16, Read/Write
Bits Description (Default = Not Applicable)
[15:0] X-axis scale correction factor (SCFx), twos complement
Table 34. Y_SENS,
Page 1 to Page 6, Low Byte Address = 0x18, Read/Write
Bits Description (Default = Not Applicable)
[15:0] Y-axis scale correction factor (SCFy), twos complement
PREFFT WINDOWING
REC_CTRL1[13:12] provides three options for preFFT windowing
of time data. For example, set REC_CTRL1[13:12] = 01 to use
the Hanning window, which offers the best amplitude resolution
of the peaks between frequency bins and minimal broadening of
peak amplitudes. The rectangular and flat top windows are also
available because they are common windowing options for
vibration monitoring. The flat top window provides accurate
amplitude resolution with a trade-off of broadening the peak
amplitudes.
FFT
The FFT process converts each 512-sample time record into
a 256-point spectral record that provides magnitude vs.
frequency data.
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 22 of 32
FFT AVERAGING
The FFT averaging function combines multiple FFT records to
reduce the variation of the FFT noise floor, enabling detection
of lower vibration levels. Each SRx option in the REC_CTRL1
register has its own FFT average control, which establishes the
number of FFT records to average into the final FFT record.
To enable this function, write the number of averages for each
SRx option that is enabled in the REC_CTRL1 register to the
FFT_AVGx registers. For example, set FFT_AVG2[15:8] = 0x10
(DIN = 0x9110) to set the number of FFT averages to 16 for the
SR3 sample rate option.
Table 35. FFT_AVG1,
Page 1 to Page 6, Low Byte Address = 0x0E, Read/Write
Bits Description (Default = 0x0108)
[15:8] FFT averages for a single record, SR1 sample rate,
NF in Figure 18; range = 1 to 255, binary
[7:0] FFT averages for a single record, SR0 sample rate,
NF in Figure 18; range = 1 to 255, binary
Table 36. FFT_AVG2,
Page 1 to Page 6, Low Byte Address = 0x10, Read/Write
Bits Description (Default = 0x0101)
[15:8] FFT averages for a single record, SR3 sample rate,
NF in Figure 18; range = 1 to 255, binary
[7:0] FFT averages for a single record, SR2 sample rate,
NF in Figure 18; range = 1 to 255, binary
RECORDING TIMES
When using automatic FFT mode, the automatic recording period
(UPDAT_INT, see Table 21) must be greater than the total
recording time. Use the following equations to calculate the
recording time:
Manual time mode
tR = tS + tPT + tST + tAST
FFT modes
tR = NF × (tS + tPT + tFFT) + tST + tAST
Table 37 provides a list of the processing times and settings used
in these equations.
Table 37. Typical Processing Times
Function Time (ms)
Sample Time, tS 1 ÷ fS, per AVG_CNT
Processing Time, t
PT
18.7
Storage Time, tST 120.0
Alarm Scan Time, tAST 2.21
Number of FFT Averages, NF Per FFT_AVG1, FFT_AVG2
FFT Time, tFFT 32.7
The storage time (tST) applies only when a storage method is
selected in REC_CTRL1[3:2] (see Table 28 for more details
about the record storage settings). The alarm scan time (tAST)
applies only when the alarms are enabled in ALM_CTRL[3:0]
(see Table 42 for more information). For systems that cannot use
DO1 to monitor the status of these operations, understanding
the recording time helps predict when data is available. Note that
when using automatic FFT mode, the automatic recording period
(UPDAT_INT, see Table 21) must be greater than the total
recording time.
DATA RECORDS
After the ADIS16229 finishes processing the FFT data, it stores
the data into the data buffer, where it is available for external access
using the SPI and x_BUF registers (see Table 60 and Table 61).
REC_CTRL1[3:2] (see Table 28) provides programmable condi-
tions for writing buffer data into the FFT records, which are in
nonvolatile flash memory locations. Set REC_CTRL1[3:2] = 01
to store buffer data into the flash memory records only when an
alarm condition is met. Set REC_CTRL1[3:2] = 10 to store every
set of FFT data into the flash memory locations. The flash memory
record provides space for a total of 14 records. Each record stored
in flash memory contains a header and frequency domain (FFT)
data from both axes, x and y. When all 14 records are full, new
records do not load into the flash memory. The REC_CNTR
register (see Table 38) provides a running count for the number
of records that are stored. Set GLOB_CMD_S[8] = 1 (DIN =
0xB701) to clear all of the records in flash memory.
Table 38. REC_CNTR,
Page 1 to Page 6, Low Byte Address = 0x50, Read Only
Bits Description (Default = 0x0000)
[15:5] Not used
[4:0] Total number of records taken; range = 0 to 13, binary
When used in conjunction with automatic trigger mode and record
storage, FFT analysis for each sample rate option requires no
additional inputs. Depending on the number of FFT averages, the
time between each sample rate selection may be quite large. Note
that selecting multiple sample rates reduces the number of records
available for each sample rate setting, as shown in Table 39.
Table 39. Available Records per Sample Rate Selected
Number of Sample Rates Selected Available Records
1 14
2 7
3 4
4 3
FFT RECORD FLASH ENDURANCE
The REC_FLSH_CNT register (see Table 40) increments when
all 14 records contain FFT data.
Table 40. REC_FLSH_CNT,
Page 1 to Page 6, Low Byte Address = 0x4A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Flash write cycle count; record data only, binary
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 23 of 32
SENSOR NODE SPECTRAL ALARMS
The alarm function offers six spectral bands for alarm detection.
Each spectral band has high and low frequency definitions, along
with two different trigger thresholds (Alarm 1 and Alarm 2) for
each accelerometer axis. Tabl e 41 provides a summary of each
register used to configure the alarm function.
Table 41. Alarm Function Register Summary
Register Address Description
ALM_F_LOW 0x20 Alarm frequency band, lower limit
ALM_F_HIGH 0x22 Alarm frequency band, upper limit
ALM_X_MAG1 0x24 X-axis Alarm Trigger Level 1 (warning)
ALM_Y_MAG1 0x26 Y-axis Alarm Trigger Level 1 (warning)
ALM_X_MAG2 0x28 X-axis Alarm Trigger Level 2 (fault)
ALM_Y_MAG2 0x2A Y-axis Alarm Trigger Level 2 (fault)
ALM_PNTR 0x2C Alarm pointer
ALM_S_MAG 0x2E System alarm trigger level
ALM_CTRL 0x30 Alarm configuration
DIAG_STAT_S
0x34
Alarm status
ALM_X_STAT 0x38 X-axis alarm status
ALM_Y_STAT 0x3A Y-axis alarm status
ALM_X_PEAK 0x3C X-axis alarm peak
ALM_Y_PEAK 0x3E Y-axis alarm peak
ALM_X_FREQ 0x44 X-axis alarm frequency of peak alarm
ALM_Y_FREQ 0x46 Y-axis alarm frequency of peak alarm
The ALM_CTRL register (see Table 42) provides control bits that
enable the spectral alarms of each axis, configures the system
alarm, sets the record delay for the spectral alarms, and configures
the clearing function for the DIAG_STAT_S error flags (see
Table 86).
Table 42. ALM_CTRL,
Page 1 to Page 6, Low Byte Address = 0x30, Read/Write
Bits Description (Default = 0x0080)
[15:12]
Not used.
[11:8] Response delay; range = 0 to 15. Represents the number
of spectral records for each spectral alarm before a
spectral alarm flag is set high.
7 Latch DIAG_STAT_S error flags. Requires a clear status
command (GLOB_CMD_S[4]) to reset the flags to 0.
1 = enabled, 0 = disabled.
6 Enable DO1 as an Alarm 1 output indicator and enable
DO2 as an Alarm 2 output indicator. 1 = enabled, 0 =
disabled.
5 System alarm comparison polarity.
1 = trigger when less than ALM_S_MAG[15:0].
0 = trigger when greater than ALM_S_MAG[15:0].
4 System alarm. 1 = temperature, 0 = power supply.
3 Alarm S enable (ALM_S_MAG). 1 = enabled, 0 = disabled.
2 Not used.
1 Alarm Y enable (ALM_Y_MAG). 1 = enabled, 0 = disabled.
0
Alarm X enable (ALM_X_MAG). 1 = enabled, 0 = disabled.
ALARM DEFINITION
The alarm function provides six programmable spectral bands, as
shown in Figure 22. Each spectral alarm band has lower and upper
frequency definitions for all of the sample rate options (SRx). It
also has two independent trigger level settings, which are useful
for systems that value warning and fault condition indicators.
Figure 22. Spectral Band Alarm Setting Example, ALM_PNTR = 0x03
Select the spectral band for configuration by writing its number
(1 to 6) to ALM_PNTR[2:0] (see Tabl e 43). Then select the sample
rate option using ALM_PNTR[9:8]. This number represents a
binary number, which corresponds to the x in the SRx sample
rates option associated with REC_CTRL1[11:8] (see Table 28).
For example, set ALM_PNTR[7:0] = 0x05 (DIN = 0xAC05) to
select Alarm Spectral Band 5, and set ALM_PNTR[15:8] = 0x02
(DIN = 0xAC02) to select the SR2 sample rate option.
Table 43. ALM_PNTR,
Pages 1 to 6, Low Byte Address = 0x2C, Read/Write
Bits Description (Default = 0x0000)
[15:10] Not used
[9:8] Sample rate option; range = 0 to 3 for SR0 to SR3
[7:3] Not used
[2:0] Spectral band number; range = 1 to 6
MAGNITUDE
FREQUENCY
1 2 3 4 5 6
ALM_F_HIGH
ALM_F_LOW
ALM_x_MAG1
ALM_x_MAG2
11483-021
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 24 of 32
Alarm Band Frequency Definitions
After the spectral band and sample rate settings are set, program
the lower and upper frequency boundaries by writing their bin
numbers to the ALM_F_LOW register (see Table 44) and
ALM_F_HIGH register (see Table 45). Use the bin width
definitions listed in Table 30 to convert a frequency into a bin
number for this definition. Calculate the bin number by dividing
the frequency by the bin width associated with the sample rate
setting. For example, if the sample rate is 5000 Hz and the lower
band frequency is 400 Hz, divide that number by the bin width
of 10 Hz to arrive at the 40th bin as the lower band setting. Then
set ALM_F_LOW[7:0] = 0x28 (DIN = 0xA028) to establish 400 Hz
as the lower frequency for the 5000 SPS sample rate setting.
Table 44. ALM_F_LOW,
Page 1 to Page 6, Low Byte Address = 0x20, Read/Write
Bits Description (Default = 0x0000)
[15:8] Not used
[7:0] Lower frequency, bin number; range = 0 to 255
Table 45. ALM_F_HIGH,
Page 1 to Page 6, Low Byte Address = 0x22, Read/Write
Bits Description (Default = 0x0000)
[15:8] Not used
[7:0]
Upper frequency, bin number; range = 0 to 255
Alarm Trigger Settings
The ALM_x_MAG1 and ALM_x_MAG2 registers (see Table 46 to
Table 49) provide two independent trigger settings for both axes of
acceleration data. They use the data format established by the range
settings in the REC_CTRL2 register (see Table 31) and recording
mode in REC_CTRL1[1:0] (see Table 28). For example, when using
the 0 g to 1 g mode for FFT analysis, 32,768 LSB is the closest
setting to 500 mg. Therefore, set ALM_Y_MAG2 = 0x8000
(DIN = 0xAB80, 0xAA00) to set the critical alarm to 500 mg,
when using the 0 g to 1 g range option in REC_CTRL2 for FFT
records. See Table 31 and Table 32 for more information about
formatting each trigger level. Note that the trigger settings
associated with Alarm 2 must be greater than the trigger
settings for Alarm 1. In other words, the alarm magnitude
settings must meet the following criteria:
ALM_X_MAG2 > ALM_X_MAG1
ALM_Y_MAG2 > ALM_Y_MAG1
Table 46. ALM_X_MAG1,
Page 1 to Page 6, Low Byte Address = 0x24, Read/Write
Bits Description (Default = 0x0000)
[15:0] X-Axis Alarm Trigger Level 1, 16-bit unsigned (see
Table 31 and Table 32 for the scale factor)
Table 47. ALM_Y_MAG1,
Page 1 to Page 6, Low Byte Address = 0x26, Read/Write
Bits Description (Default = 0x0000)
[15:0] Y-Axis Alarm Trigger Level 1, 16-bit unsigned (see
Table 31 and Table 32 for the scale factor)
Table 48. ALM_X_MAG2,
Page 1 to Page 6, Low Byte Address = 0x28, Read/Write
Bits Description (Default = 0x0000)
[15:0] X-Axis Alarm Trigger Level 2, 16-bit unsigned (see
Table 31 and Table 32 for the scale factor)
Table 49. ALM_Y_MAG2,
Page 1 to Page 6, Low Byte Address = 0x2A, Read/Write
Bits Description (Default = 0x0000)
[15:0] Y-Axis Alarm Trigger Level 2, 16-bit unsigned (see
Table 31 and Table 32 for the scale factor)
Table 50. ALM_S_MAG,
Page 1 to Page 6, Low Byte Address = 0x2E, Read/Write
Bits
Description (Default = 0x0000)
[15:0] System alarm trigger level, data format matches target
from ALM_CTRL[4] (see Table 42)
Enable Alarm Settings
Before configuring the spectral alarm registers, clear their current
contents by setting GLOB_CMD_S[9] = 1 (DIN = 0xB702).
After completing the spectral alarm band definitions, save the
settings by setting GLOB_CMD_S[12] = 1 (DIN = 0xB710).
The device ignores the save command if any of these locations
has already been written to.
ALARM INDICATOR SIGNALS
GPO_CTRL[5:0] (see Table 84) and ALM_CTRL[6] (see Table 42)
provide controls for establishing DO1 and DO2 as dedicated alarm
output indicator signals. Use GPO_CTRL[5:0] to select the alarm
function for DO1 and/or DO2, then set ALM_CTRL[6] = 1 to
enable DO1 to serve as an Alarm 1 indicator and DO2 as an
Alarm 2 indicator. This setting establishes DO1 to indicate
Alarm 1 (warning) conditions and DO2 to indicate Alarm 2
(critical) conditions.
ALARM FLAGS AND CONDITIONS
The FFT header (see Table 70) contains both generic alarm
flags (DIAG_STAT_S, see Table 86) and spectral band-specific
alarm flags (ALM_x_STAT, see Table 51 and Table 52). The FFT
header also contains magnitude (ALM_x_PEAK, see Table 53
and Table 54) and frequency information (ALM_x_FREQ, see
Table 55 and Table 56) associated with the highest magnitude of
vibration content in the record.
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 25 of 32
ALARM STATUS
The ALM_x_STAT registers (see Table 51 and Table 52) provide
alarm bits for each spectral band on the current sample rate option.
Table 51. ALM_X_STAT,
Page 1 to Page 6, Low Byte Address = 0x38, Read Only
Bits Description (Default = 0x0000)
15 Alarm 2 on Band 6; 1 = alarm set, 0 = no alarm
14 Alarm 1 on Band 6; 1 = alarm set, 0 = no alarm
13 Alarm 2 on Band 5; 1 = alarm set, 0 = no alarm
12 Alarm 1 on Band 5; 1 = alarm set, 0 = no alarm
11 Alarm 2 on Band 4; 1 = alarm set, 0 = no alarm
10 Alarm 1 on Band 4; 1 = alarm set, 0 = no alarm
9 Alarm 2 on Band 3; 1 = alarm set, 0 = no alarm
8 Alarm 1 on Band 3; 1 = alarm set, 0 = no alarm
7 Alarm 2 on Band 2; 1 = alarm set, 0 = no alarm
6
Alarm 1 on Band 2; 1 = alarm set, 0 = no alarm
5 Alarm 2 on Band 1; 1 = alarm set, 0 = no alarm
4 Alarm 1 on Band 1; 1 = alarm set, 0 = no alarm
3 Not used
[2:0] Most critical alarm condition, spectral band; range = 1 to 6
Table 52. ALM_Y_STAT,
Page 1 to Page 6, Low Byte Address = 0x3A, Read Only
Bits Description (Default = 0x0000)
15
Alarm 2 on Band 6; 1 = alarm set, 0 = no alarm
14 Alarm 1 on Band 6; 1 = alarm set, 0 = no alarm
13 Alarm 2 on Band 5; 1 = alarm set, 0 = no alarm
12 Alarm 1 on Band 5; 1 = alarm set, 0 = no alarm
11 Alarm 2 on Band 4; 1 = alarm set, 0 = no alarm
10 Alarm 1 on Band 4; 1 = alarm set, 0 = no alarm
9
Alarm 2 on Band 3; 1 = alarm set, 0 = no alarm
8 Alarm 1 on Band 3; 1 = alarm set, 0 = no alarm
7 Alarm 2 on Band 2; 1 = alarm set, 0 = no alarm
6 Alarm 1 on Band 2; 1 = alarm set, 0 = no alarm
5 Alarm 2 on Band 1; 1 = alarm set, 0 = no alarm
4 Alarm 1 on Band 1; 1 = alarm set, 0 = no alarm
3 Not used
[2:0] Most critical alarm condition, spectral band; range = 1 to 6
WORST-CASE CONDITION MONITORING
The ALM_x_PEAK registers (see Table 53 and Table 54) contain
the peak magnitude for the worst-case alarm condition in each
axis. The ALM_x_FREQ registers (see Table 55 and Table 56)
contain the frequency bin number for the worst-case alarm
condition.
Table 53. ALM_X_PEAK,
Page 1 to Page 6, Low Byte Address = 0x3C, Read Only
Bits Description (Default = 0x0000)
[15:0] Alarm peak, x-axis, accelerometer data format
Table 54. ALM_Y_PEAK,
Page 1 to Page 6, Low Byte Address = 0x3E, Read Only
Bits Description (Default = 0x0000)
[15:0]
Alarm peak, y-axis, accelerometer data format
Table 55. ALM_X_FREQ,
Page 1 to Page 6, Low Byte Address = 0x44, Read Only
Bits Description (Default = 0x0000)
[15:8] Not used
[7:0] Alarm frequency of x-axis peak alarm level,
FFT bin number; range = 0 to 255
Table 56. ALM_Y_FREQ,
Page 1 to Page 6, Low Byte Address = 0x46, Read Only
Bits Description (Default = 0x0000)
[15:8] Not used
[7:0] Alarm frequency of y-axis peak alarm level,
FFT bin number; range = 0 to 255
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 26 of 32
READING OUTPUT DATA
After the ADIS16229 updates the ADIS16000 with its data, the
data is available in the data buffer and FFT records, if selected.
In manual time capture mode, the record for each axis contains
512 samples. In manual and automatic FFT mode, each record
contains the 256-point FFT result for each accelerometer axis.
Table 57 provides a summary of registers that provide access to
processed sensor data.
Table 57. Output Data Registers (Sensor Nodes)
Register Address Description
TEMP_OUT_S 0x0A Internal temperature
SUPPLY_OUT_S 0x0C Power supply
BUF_PNTR
0x12
Data buffer index pointer
REC_PNTR 0x14 FFT record index pointer
X_BUF 0x06 X-axis accelerometer data
Y_BUF 0x08 Y-axis accelerometer data
GLOB_CMD_S 0x36 FFT record retrieve command
TIME_STMP_L
0x40
Time stamp, lower word
TIME_STMP_H 0x42 Time stamp, upper word
REC_INFO1 0x4C FFT record header information
REC_INFO2 0x4E FFT record header information
READING DATA FROM THE DATA BUFFER
After completing a spectral record and updating each data buffer,
the ADIS16000 loads the first data sample from each data buffer
into the x_BUF registers (see Table 60 and Table 61) and sets the
buffer index pointer in the BUF_PNTR register (see Table 58) to
0x0000. The index pointer determines which data samples load
into the x_BUF registers. For example, writing 0x009F to the
BUF_PNTR register (DIN = 0x9300, DIN = 0x929F) causes the
160th sample in each data buffer location to load into the x_BUF
registers. The index pointer increments with every x_BUF read
command, which causes the next set of capture data to load into
each capture buffer register automatically. This enables an efficient
method for reading all 256 samples in a record, using sequential
read commands, without having to manipulate the BUF_PNTR
register.
Figure 23. Data Buffer Structure and Operation
Table 58. BUF_PNTR,
Page 1 to Page 6, Low Byte Address = 0x12, Read/Write
Bits Description (Default = 0x0000)
[15:9] Not used
[8:0] Data bits; range = 0 to 255 (FFT), 0 to 511 (time)
ACCESSING FFT RECORD DATA
The FFT records can be stored in flash memory. The REC_PNTR
register (see Table 59) and GLOB_CMD_S[13] (see Table 75)
provide access to the FFT records, as shown in Figure 24. For
example, set REC_PNTR[7:0] = 0x0A (DIN = 0x940A) and
GLOB_CMD_S[13] = 1 (DIN = 0xB720) to load FFT Record 10 in
the FFT buffer for SPI/register access.
Table 59. REC_PNTR,
Page 1 to Page 6, Low Byte Address = 0x14, Read/Write
Bits Description (Default = 0x0000)
[15:4] Not used (don’t care)
[3:0] Data bits
Figure 24. FFT Record Access
256/512
INTERNAL SAM PLING SYSTEM SAM PLES, PRO CESSES, AND
STORES DATA IN DATA BUFFERS.
0
Y_BUF
TEMP_OUT
SUPPLY_OUT
BUF_PNTR X-AXIS
ACCELEROMETER
DATA
BUFFER
Y-AXIS
ACCELEROMETER
DATA
BUFFER
X_BUF
DATA IN BUFFERS LOAD INTO
USER O UTPUT REGISTERS
FFT ANALYSIS
11483-022
11483-023
m = REC_PNTR
GLOB_CM D_S [ 13] = 1
FFT
RECORD
0
FFT
HEADER
0
X Y X Y X Y X Y
FFT
RECORD
1
FFT
HEADER
1
FFT
RECORD
m
FFT
HEADER
m
DATA
BUFFER
X, Y
SPI
REGISTERS
FFT
HEADER
REGISTERS
FFT
RECORD
15
FFT
HEADER
15
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 27 of 32
DATA FORMAT
Table 60 and Table 61 list the bit assignments for the x_BUF
registers. The acceleration data format depends on the range
scale setting in REC_CTRL2 (see Table 31) and the recording
mode settings in REC_CTRL1 (see Table 28). Table 62 provides
some data formatting examples for the FFT mode, and Table 63
offers some data formatting examples for the 16-bit, twos
complement format used in manual time mode.
Table 60. X_BUF,
Page 1 to Page 6, Low Byte Address = 0x06, Read Only
Bits Description (Default = Not Applicable)
[15:0] Buffer for x-axis acceleration data
See Table 32 for scale sensitivity
Format = twos complement (time), binary (FFT)
Table 61. Y_BUF,
Page 1 to Page 6, Low Byte Address = 0x08, Read Only
Bits Description (Default = Not Applicable)
[15:0] Buffer for y-axis acceleration data
See Table 32 for scale sensitivity
Format = twos complement (time), binary (FFT)
Table 62. FFT Mode, 5 g Range, Data Format Examples
Acceleration (mg) LSB Hex Binary
4,999.9237 65,535 0xFFFF 1111 1111 1111 1111
100 × 5 ÷ 65,536 100 0x0064 0000 0000 0110 0100
2 × 5 ÷ 65,536 2 0x0002 0000 0000 0000 0010
1 × 5 ÷ 65,536 1 0x0001 0000 0000 0000 0001
0 0 0x0000 0000 0000 0000 0000
Table 63. Manual Time Mode, 5 g Range, Data
Format Examples
Acceleration (mg) LSB Hex Binary
4999.847 32,767 0x7FFF 1111 1111 1111 1111
~1000 6,554 0x199A 0001 0001 10011010
2 × 5 ÷ 32,768 2 0x0002 0000 0000 0000 0010
1 × 5 ÷ 32,768 1 0x0001 0000 0000 0000 0001
0 0 0x0000 0000 0000 0000 0000
−1 × +5 ÷ +32,768 −1 0xFFFF 1111 1111 1111 1111
−2 × +5 ÷ +32,768 −2 0xFFFE 1111 1111 1111 1110
~−1000 −6554 0xE666 1110 0110 0110 0110
−5000 −32,768 0x8000 1000 0000 0000 0000
REAL-TIME DATA COLLECTION
When using real-time mode, select the output channel by reading
the associated x_BUF register. For example, set DIN = 0x0880
to select the y-axis sensor for sampling. After selecting the channel,
use the data ready signal to trigger subsequent data reading of
the Y_BUF register. In this mode, use the time domain data
formatting for a range setting of 20 g, as shown in Table 32.
POWER SUPPLY/TEMPERATURE
Both the ADIS16000 and the ADIS16229 offer power supply
and temperature measurements. The ADIS16229 performs
these measurements at the end of each spectral record, while
the ADIS16000 does so continuously. The power supply
measurements (SUPPLY_OUT_x, see Table 64 and Table 65)
come from averaging a data record of 256 samples (50 kSPS).
The temperature measurements (TEMP_OUT_x, see Table 67
and Table 68) come from averaging a data record of 64 samples,
which cover a total time of 1.7 ms. When using real-time mode,
these registers update only when this mode starts.
Table 64. SUPPLY_OUT_G,
Page 0, Low Byte Address = 0x0E, Read Only
Bits Description (Default = Not Applicable)
[15:0] Power supply, binary, 3.3 V = 0x1D46, 0.44 mV/LSB
Table 65. SUPPLY_OUT_S,
Page 1 to Page 6, Low Byte Address = 0x0C, Read Only
Bits Description (Default = Not Applicable)
[15:0] Power supply, binary, 3.3 V = 0x1D46, 0.44 mV/LSB
Table 66. Power Supply Data Format Examples
Supply Level (V) LSB Hex Binary
3.6 8182 0x1FF6 0001 1111 1111 0110
3.3 + 0.00044 7501 0x1D4D 0001 1101 0100 1101
3.3 7500 0x1D4C 0001 1101 0100 1100
3.3 − 0.00044 7499 0x1D4B 0001 1101 0100 1011
3.15 7159 0x1BF7 0001 1011 1111 0111
Table 67. TEMP_OUT_G,
Page 0, Low Byte Address = 0x0C, Read Only
Bits Description (Default = Not Applicable)
[15:0] Temperature data, offset binary, 4584 LSB = 0°C,
−0.0815°C/LSB
Table 68. TEMP_OUT_S,
Page 1 to Page 6, Low Byte Address = 0x0A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Temperature data, offset binary, 4584 LSB = 0°C,
−0.0815°C/LSB
Table 69. Internal Temperature Data Format Examples
Temperature (°C) LSB Hex Binary
125 3050 0x0BEA 0000 1011 1110 1010
0 + 0.0815 4583 0x11E7 0001 0001 1110 0111
0 4584 0x11E8 0001 0001 1110 1000
0 − 0.0815 4585 0x11E9 0001 0001 1110 1001
−40 5075 0x13D3 0001 0011 1101 0011
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 28 of 32
FFT EVENT HEADER
Each FFT record has an FFT header containing information
that fills all of the registers listed in Table 70. The information
in these registers contains recording time, record configuration
settings, status/error flags, and several alarm outputs. The registers
listed in Table 70 update with every record event and also update
with record-specific information when using GLOB_CMD_S[13]
(see Table 75) to retrieve a data set from the FFT record in
flash memory.
Table 70. FFT Header Register Information
Register Address Description
DIAG_STAT_S
0x34
Alarm status
ALM_X_STAT 0x38 X-axis alarm status
ALM_Y_STAT 0x3A Y-axis alarm status
ALM_X_PEAK 0x3C X-axis alarm peak
ALM_Y_PEAK 0x3E Y-axis alarm peak
TIME_STMP_L 0x40 Time stamp, lower word
TIME_STMP_H 0x42 Time stamp, upper word
ALM_X_FREQ 0x44 X-axis alarm frequency of peak alarm
ALM_Y_FREQ 0x46 Y-axis alarm frequency of peak alarm
REC_INFO1 0x4C FFT record header information
REC_INFO2 0x4E FFT record header information
The REC_INFO1 register (see Tabl e 71) and the REC_INFO2
register (see Table 72) capture the settings associated with the
current FFT record.
Table 71. REC_INFO1,
Page 1 to Page 6, Low Byte Address = 0x4C, Read Only
Bits Description (Default = 0x0008)
[15:14] Sample rate option;
00 = SR0, 01 = SR1, 10 = SR2, 11 = SR3
[13:12] Window setting;
00 = rectangular, 01 = Hanning, 10 = flat top,
11 = not applicable
[11:10] Signal range;
00 = 1 g, 01 = 5 g, 10 = 10 g, 11 = 20 g
[9:8] Not used (don’t care)
[7:0] FFT averages; range = 0 to 255
Table 72. REC_INFO2,
Page 1 to Page 6, Low Byte Address = 0x4E, Read Only
Bits Description (Default = 0x0000)
[15:4] Not used (don’t care)
[3:0] AVG_CNT setting
The TIME_STMP_x registers (see Table 73 and Tabl e 74)
provide a relative time stamp that identifies the time for
the current FFT record.
Table 73. TIME_STMP_L,
Page 1 to Page 6, Low Byte Address = 0x40, Read Only
Bits Description (Default = 0x0000)
[15:0] Record time stamp, low integer, binary, seconds
Table 74. TIME_STMP_H,
Page 1 to Page 6, Low Byte Address = 0x42, Read Only
Bits Description (Default = 0x0000)
[15:0] Record time stamp, high integer, binary, seconds
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 29 of 32
SYSTEM TOOLS
GLOBAL COMMANDS
The GLOB_CMD_S register (see Table 75) provides an array of
single write commands for convenience. Setting the assigned bit
to 1 activates each function. When the function completes, the
bit restores itself to 0. For example, clear the capture buffers by
setting GLOB_CMD_S[8] = 1 (DIN = 0xB701). All of the com-
mands in the GLOB_CMD_S register require that the power
supply be within normal limits for the execution times listed in
Table 75.
Table 75. GLOB_CMD_S,
Page 1 to Page 6, Low Byte Address = 0x36, Write Only
Bits Description (Default = Not Applicable)
15 Clear autonull correction
14 Retrieve spectral alarm band information from the
ALM_PNTR setting
13 Retrieve record data from flash memory
12 Save spectral alarm band registers to SRAM
11 Record start/stop
10 Set BUF_PNTR = 0x0000
9 Clear spectral alarm band registers from flash memory
8
Clear records
7 Software reset
6 Save registers to flash memory
5 Flash test, compare sum of flash memory with factory
value
4 Clear DIAG_STAT_S register
3
Restore factory register settings and clear the capture
buffers
2 Self-test, result in DIAG_STAT_S[5]
1 Power-down
0 Autonull
DEVICE IDENTIFICATION
Table 76. LOT_ID1_G,
Page 0, Low Byte Address = 0x1A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Lot identification code
Table 77. LOT_ID1_S,
Page 1 to Page 6, Low Byte Address = 0x68, Read Only
Bits Description (Default = Not Applicable)
[15:0] Lot identification code
Table 78. LOT_ID2_G,
Page 0, Low Byte Address = 0x1C, Read Only
Bits Description (Default = Not Applicable)
[15:0] Lot identification code
Table 79. LOT_ID2_S,
Page 1 to Page 6, Low Byte Address = 0x6A, Read Only
Bits Description (Default = Not Applicable)
[15:0] Lot identification code
Table 80. PROD_ID_G,
Page 0, Low Byte Address = 0x16, Read Only
Bits Description (Default = 0x3E80)
[15:0] 0x3E80 = 16,000
Table 81. PROD_ID_S,
Page 1 to Page 6, Low Byte Address = 0x48, Read Only
Bits Description (Default = 0x3F65)
[15:0] 0x3F65 = 16,229
Table 82. SERIAL_NUM_G,
Page 0, Low Byte Address = 0x24, Read Only
Bits Description (Default = Not Applicable)
[15:0] Serial number, lot specific
Table 83 shows a blank register that is available for writing user-
specific identification.
Table 83. USER_SCR,
Page 1 to Page 6, Low Byte Address = 0x64, Read Only
Bits Description (Default = 0x0000)
[15:0] User scratch register
Table 84. GPO_CTRL,
Page 0, Low Byte Address = 0x2A, Read/Write
Bits Description (Default = 0x0008)
[15:6] Not used
[5:4] DO2 function selection
00 = general-purpose
01 = alarm indicator
10 = busy indicator/data ready (real-time mode)
11 = not used
[3:2] DO1 function selection
00 = general-purpose
01 = alarm indicator
10 = busy indicator/data ready (real-time mode)
11 = not used
1 DO2 polarity
1 = active high
0 = active low
0
DO1 polarity
1 = active high
0 = active low
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 30 of 32
SYSTEM FLAGS
Critical system flags are in the DIAG_STAT_x registers (see
Table 85 and Table 86) for each ADIS16229. These flags indicate
various indicators for monitoring the network. Multiple flags
in these registers can be high at one time, and the flags persist
(that is, go high again, after clearing) when the error conditions
continue to exist. The flags in DIAG_STAT_S[6:0] remain in a
latch condition until the problem clears or the flags are cleared
using GLOB_CMD_S[4]. All of the alarm flags in the upper
byte (DIAG_STAT_S[15:8] latch if ALM_CTRL[7] = 1 (see
Table 42).
Table 85. DIAG_STAT_G,
Page 0, Low Byte Address = 0x2C, Read Only
Bits Description (Default = 0x0000)
15 Not used
14 System alarm; 1 = alarm condition
13 Sensor Node 6 new data; 1 = new data received
12 Sensor Node 5 new data; 1 = new data received
11 Sensor Node 4 new data; 1 = new data received
10 Sensor Node 3 new data; 1 = new data received
9 Sensor Node 2 new data; 1 = new data received
8 Sensor Node 1 new data; 1 = new data received
7 Command completion; 1 = completed
6 Flash memory failure, from GLOB_CMD_S[5] test
(see Table 75)
[5:4] Not used
3 SPI communication; 1 = SCLKs ≠ multiple of 16
2 Flash update failure
1 Power supply > 3.625 V
0 Power supply < 2.0 V
Table 86. DIAG_STAT_S,
Page 1 to Page 6, Low Byte Address = 0x34, Read Only
Bits Description (Default = 0x0000)
15 Not used
14 System alarm; 1 = error condition exists, 0 = no error
13 Not used
12 Alarm 2, Y-axis; 1 = alarm condition, 0 = no alarm
11 Alarm 2, X-axis; 1 = alarm condition, 0 = no alarm
10 Not used
9 Alarm 1, Y-axis; 1 = alarm condition, 0 = no alarm
8 Alarm 1, X-axis; 1 = alarm condition, 0 = no alarm
7 Not used
6 Flash memory failure, from GLOB_CMD_S[5] test
(see Table 75)
5 Self-test (1 = fail, 0 = pass)
4 Record process interrupted
3 Not used
2 Flash update failure
1 Power supply > 3.625 V
0 Power supply < 2.0 V
SELF-TEST
Set GLOB_CMD_S[2] = 1 (DIN = 0xB602) (see Table 75) to
run an automatic self-test routine, which reports a pass/fail
result to DIAG_STAT_S[5] (see Table 85).
FLASH MEMORY MANAGEMENT
Set GLOB_CMD_S[5] = 1 (DIN = 0xB620) to run an internal
checksum test on the flash memory, which reports a pass/fail
result to DIAG_STAT_S[6]. The FLASH_CNT_S register (see
Table 87) provides a running count of flash memory write cycles
in each ADIS16229. This is a tool for managing the endurance
of the flash memory. The FLASH_CNT_G (see Table 87) register
provides this function for the ADIS16000 as well. Figure 25
quantifies the relationship between data retention and junction
temperature.
Table 87. FLASH_CNT_G,
Page 0, Low Byte Address = 0x04, Read Only
Bits Description (Default = Not Applicable)
[15:0] Binary counter for writing to flash memory
Table 88. FLASH_CNT_S,
Page 1 to Page 6, Low Byte Address = 0x04, Read Only
Bits Description (Default = Not Applicable)
[15:0] Binary counter for writing to flash memory
Figure 25. Flash Memory Data Retention vs. Junction Temperature
600
450
300
150
030 40
RETE NT I ON (Years)
JUNCTION TEMPERATURE (°C)
55 70 85 100 125 135 150
11483-024
OBSOLETE
Data Sheet ADIS16000/ADIS16229
Rev. 0 | Page 31 of 32
OUTLINE DIMENSIONS
Figure 26. 14-Pin Connector Multichip Chip Module Laminate [MCML]
(ML-14-4)
Dimensions shown in millimeters
06-13-2013-A
TOP VIEW
END VIEW
47.75
47.35
46.95
38.00
37.80
37.60
1.727
1.600
1.473
23.00
22.80
22.60
6.00
BCS
1.00 BS C
PITCH
0.50 BCS
0.30 BCS
SQ
1.00 BS C
DETAIL A
SIDE VIEW
8.77 BS C 7.54 BS C
9.00 BS C
4.50 BS C
14.60 BS C
11.40 BS C
11.40 BS C 5.37 BS C
Ø3.60 BS C
Ø7.14 BS C
DETAIL A
OBSOLETE
ADIS16000/ADIS16229 Data Sheet
Rev. 0 | Page 32 of 32
Figure 27. 2-Pin Multichip Chip Module Laminate [MCML]
(ML-2-1)
Dimensions shown in millimeters
ORDERING GUIDE
Model
1
Temperature Range
Package Description
Package Option
ADIS16000AMLZ −40°C to +85°C 14-Pin Connector Multichip Chip Module Laminate [MCML] ML-14-4
ADIS16229AMLZ
−40°C to +85°C
2-Pin Multichip Chip Module Laminate [MCML]
ML-2-1
1 Z = RoHS Compliant Part.
06-16-2013-A
END VIEW
47.75
47.35
46.95
38.00
37.80
37.60
1.727
1.600
1.473
23.00
22.80
22.60
13.50 BS C
3.50 BS C
30.80 BS C
3.50
BSC
3.94
BSC
2.54
BSC
1.27 BS C
15.80 BS C 10.60 BS C
7.50 BS C
11.40 BS C
SIDE VIEW
Ø2.70 BS C
(3 PL CS)
TOP VIEW
©2013 Analog Devices, Inc. All rights reserved. Trademarks and
registered trademarks are the property of their respective owners.
D11483-0-8/13(0)