The MBC60 Series of open-frame medical power supplies, with its wide universal 90-264 VAC input range, is available at 60 W of output power and a variety of single and multiple output voltages. The MBC series was designed to 3rd edition medical approvals and provides 2x MOPP (Means of Patient Protection) isolation for class 1 and class 2 installations. 60 W Convection Cooled 90-264 VAC Input -20 to 50C Full Load Operation 2 x MOPP 4.0 x 2.0 x 1.2 inch (101.6 x 50.8 x 30.48 mm) No Minimum Load Required - Single Output Models No Load Power < 0.3 W IEC Protection Class Options: o Class I: Earthing Tab J4 (no suffix) o Class II: No Earthing Tab (-2 suffix) Conducted EMI EN 55022-B, FCC Part 15 Level B Medical Safety Agency Approvals CE marked These medical power supplies are ideal for, monitoring, home health equipment as well as surgical devices. Dialysis Monitoring Pumps Surgical Devices Home Health Ultrasound North America +1-866.513.2839 Asia-Pacific +86.755.29885888 Europe, Middle East +353 61 225 977 tech.support@psbel.com (c) 2015 Bel Power Solutions, Inc. BCD.00287_AB MBC60 Series OUTPUT CURRENT MAX (A) 10.0 MINIMUM LOAD (A) 3 0.0 RIPPLE & NOISE 2 MBC60-1005G OUTPUT VOLTAGE (VDC) 1 5.1 1% TOTAL REGULATION 0.8% MBC60-1012G 12 5.4 0.0 1% 0.8% 130% Typical MBC60-1015G 15 4.33 0.0 1% 0.8% 130% Typical MBC60-1024G 24 2.7 0.0 1% 0.8% 130% Typical MBC60-1048G 48 1.35 0.0 1% 0.8% 130% Typical 5.2 8.0 0.5 1% 0.8% 130% Typical MBC60-3000G 12.5 3.0 0.1 1% 5.3% -12.8 0.5 0.0 1% 5.3% 5.2 8.0 0.5 1% 0.8% MODEL MBC60-3001G MBC60-3002G MBC60-3003G 24.0 1.5 0.1 1% 5.3% -12.8 0.5 0.0 1% 5.3% 5.2 8.0 0.5 1% 0.8% 14.6 2.5 0.1 1% 5.3% -14.8 0.5 0.0 1% 5.3% 3.3 6.0 1.0 1.5% 0.8% 5.2 3.0 0.1 1% 5.3% -12.8 0.5 0.0 1% 5.3% OVP THRESHOLD 130% Typical 130% Typical 130% Typical 130% Typical Warranty 2 years. NOTES: 1 2 3 Maximum outputs for each output. Max power rating should not be exceeded. Output noise measurement is made with a 20 MHz bandwidth using a 6" twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor. Minimum load specified to meet cross regulation. 866.513.2839 tech.support@psbel.com (c) 2015 Bel Power Solutions, Inc. BCD.00287_AB MBC60 Series Specifications are for nominal input voltage, 25C and max load unless otherwise stated. PARAMETER DESCRIPTION / CONDITION CRITERION Input Voltage Universal Input 90 - 264 VAC 120 VAC: 230 VAC: Single output models Multi output models 120 VAC: 230 VAC: 120 VAC: 230 VAC: 47 to 400 Hz 1.5 A max. 0.75 A max. < 0.3 W < 0.5 W 30 A max. 60 A max. < 140 A < 250 A Input Frequency 4 Input Current No Load Power Inrush Current Leakage Current 4 Safety Approved: 47 to 63 Hz. PARAMETER DESCRIPTION / CONDITION CRITERION Efficiency Typical 85% Hold Up Time 120 / 230 VAC 6 ms Output Power 50 to 65 W Line Regulation Voltage Adjustment V1: V2 & V3: V1 +/-0.3% +/-0.5% +/-5% < 10%, recovery time < 5 ms < 100 ms 3% 5% 10% Remote Sense V1 0.5 VDC compensation Over Current Protection Typical above rating 130% Over Voltage Protection Typical for V1 only 130% Short Circuit Protection Short term, auto recovery PARAMETER DESCRIPTION / CONDITION Input to Output, 2M0PP Input to Ground, 1M0PP Output to Ground Typical CRITERION Min. 5700 VDC 1500 VAC 500 VAC 67 kHz MTBF according to Telcordia -SR332-Issue 3 Refer to derating curve Start-up is guaranteed 1.87 million hours -20 to 70C -20 to 0C -40 to 85C Load Regulation Transient Response V1: V2 & V3: Main output 50 to 100% load change, 50/60 Hz, 50% duty cycle, 0.1A / us Rise Time Set Point Tolerance Isolation Voltage Switching Frequency Reliability Operating Temperature Storage Temperature 866.513.2839 tech.support@psbel.com (c) 2015 Bel Power Solutions, Inc. BCD.00287_AB MBC60 Series PARAMETER DESCRIPTION / CONDITION Conducted Emissions EN55022-B, CISPR22-B, FCC PART15-B Radiated Emissions EN55022-B, CISPR22-B, FCC PART15-B To be controlled in end system Harmonic Current EN61000-3-2 Class A Static Discharge EN61000-4-2 Level 3 RF Field Susceptibility EN61000-4-3 Level-3 Fast Transients/Bursts EN61000-4-4 Level-3 Surge Susceptibility EN61000-4-5 Level-3 Humidity Non Condensing Operating: Non-Operating: 95% 10,000 ft. 40,000 ft. Altitude PARAMETER CRITERION DESCRIPTION/CONDITION Approved to the latest edition of the following standards: CSA/UL60601-1, EN60601-1 and IEC60601-1 Complies with LVD Directive Agency Approvals CE mark Figure 1 - Output Power Vs. Temperature CONNECTOR PIN AC Input Connector J1 DC Output Connector J2 Remote Sense J3 Earthing Tab J4 DESCRIPTION/CONDITION Pin 1 AC Neutral Pin 2 AC Line Pin 1,2 V1 Pin 3,4 RTN Pin 5 V3 Pin 6 V2 Pin 1 +V1 Sense Pin 2 -V1 Sense MANUFACTURER / PN Molex: 26-60-4030 or equivalent Mating: 09-50-3031; Pins: 08-50-0106 Tyco: 640445-6 or equivalent Mating: 647402-6; Pins: 3-647409-1 Molex: 22-23-2021 or equivalent Mating: 22-01-2021 Molex: 19705-4301 Mating: 190030001 866.513.2839 tech.support@psbel.com (c) 2015 Bel Power Solutions, Inc. BCD.00287_AB MBC60 Series PARAMETER DESCRIPTION/CONDITION Weight 150 g (0.33 lbs.) Dimensions 101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch) Figure 2 - Mechanical Drawing MBC60-1xxxG Figure 3 - Mechanical Drawing MBC60-3xxxG NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice. 866.513.2839 tech.support@psbel.com (c) 2015 Bel Power Solutions, Inc. BCD.00287_AB