WWW.OWON.COM.CN
..……..….….……..….….……..….………..….…...…..…..…..…..………………………………………………………………………………………………………………..……..….….……..….….……..….………..….…...…..…..…..…..…………………………………………………………………………………………………………………………………………………
..……..….….……..….….……..….………..….…...…..…..…..…..………………………………………………………………………………………………………………..……..….….……..….….……..….………..….…...…..…..…..…..…………………………………………………………………………………………………………………………………………………
OWON BRANCH OFFICE : LILLIPUT ELECTRONICS INC 1-626-369-3088
Data-logger Mode
during recording the measurement value, possible to set the logging duration (min. 5ms), and length,
then get access to chart or table result
Performance Specifications
XDM3041
Measurement Range
Frequency Range
Accuracy: 1 Year ±
(% of reading +% of range)
DC Voltage
600mV, 6V, 60V, 600V, 1000V
/
0.02±0.01
True RMS AC Voltage
600mV, 6V, 60V, 600V, 750V
20 Hz - 50 Hz
2 + 0.10
50 Hz - 20 kHz
0.2 + 0.06
20 kHz - 50 kHz
1.0 + 0.05
50 kHz - 100 kHz
3.0 + 0.08
DC Current
600.00 μA
/
0.06 + 0.02
6.0000 mA
0.06 + 0.02
60.000 mA
0.1 + 0.05
600.00 mA
0.2 + 0.02
6.000 A
0.2 + 0.05
10.0000 A
0.250 + 0.05
True RMS AC Current
60.000 mA, 600.00 mA,
6.0000 A, 10.000 A
20 Hz - 45 Hz
2 + 0.10
45 Hz - 2 kHz
0.50 + 0.10
2 kHz - 10 kHz
2.50 + 0.20
Resistance
600.00 Ω
/
0.040 + 0.01
6.0000 kΩ
0.030 + 0.01
60.000 kΩ
0.030 + 0.01
600.00 kΩ
0.040 + 0.01
6.0000 MΩ
0.120 + 0.03
XDM3041 Bench-type Digital Multimeter
+ 4 inch 480 x 320 pixels high resolution LCD
+ resolutions up to 4 1/2 digits
+ reading rates up to 150 readings/s
+ true RMS AC voltage / current measurement
+ dual line display supported
+ the change trend analysis accessible via special chart mode
+ SCPI supported - remote control, and data-sharing
possible via LAN, USB, RS232 port, and WiFi*
* WiFi module is optional
+ multi- IO interface: USB Device / Host, RS232, LAN, and ext.
trigger input
WWW.OWON.COM.CN
.……..….…...…..…..…..…..………………………………………………………………………………………………………………………………………………………
………………………………………………………………………………………………………………………………………………………………………………………………
Fujian Lilliput Optoelectronics Technology Co., Ltd.
60.000 MΩ
0.90 + 0.03
100.00 MΩ
1.75 + 0.03
Diode Test
3.0000 V
0.5 + 0.01
Continuity
1000 Ω
0.5 + 0.01
Frequency
Period
200 mV - 750 V
0.01 + 0.003
0.01 + 0.003
0.01 + 0.003
0.01 + 0.006
20 mA - 10 A
0.01 + 0.003
0.01 + 0.003
Test Current
Capacitance
2.000 nF
3 + 1.0
20.00 nF
1 + 0.5
200.0 nF
1 + 0.5
2.000 μF
1 + 0.5
200 μF
1 + 0.5
10000 μF
2 + 0.5
Temperature
temperature sensors under 2 categories supported -
thermocouple (ITS-90 conversion between B / E / J / K / N / R / S / T type), and thermal
resistance (RTD sensor conversion between Pt100 and Pt385 type)
Other functions
Bar meter, histogram, trend chart, Max. Min. value, average value, standard deviation,
DB / DBm, Pass / Fail
Data-logger Function
Logging Duration
5ms
Logging Length
1M points
General
Dimension (W x H x D)
235 x 110 x 295 (mm)
Weight (without
package)
3.0 kg
Specifications subject to change without prior notice.
Accessories The accessories subject to final delivery.
Power Cord Manual USB Fuse Multimeter Lead Alligator Clip