Amprobe® Test Tools www.Amprobe.com
Data Sheet
Specifications
Measuring Range CO2 0 ~ 9999 ppm
2001 ~ 9999 ppm (out of scale range)
Resolution 1ppm - CO2 concentration, 0.1°C / °F - Temp, 0.1% RH - Humidity
Accuracy ±75ppm±5% of reading (0 ~ 2000 ppm) Others range accuracy are not specified
Temperature ±0.6°C / ±0.9°F
Temp. Range -10°C to 60°C (14°F to 140°F)
RH Range 0% ~ 95%
Accuracy ±3% RH (at25°C, 10-90%) others ±5% for other RH range
DP (Dew Point Temp.) -20°C to 59.9°C (-4°F to 139.8°F)
WB (Wet Bulb Temp.) -5.0°C to 59.9°C (23°F to 139.8°F)
Warm-up Time 30 seconds
Response Time About 30 seconds
Current Consumption < 50mA average, < 310mA peak current ( ~55ms)
Operating Condition 0 ~ 50°C (32°F to 122°F), 0 ~ 95% RH (avoid condensation) for CO2
Dimension 130(L) x 85(W) x 60(H) mm (5.1 x 3.3 x 2.4 in)
Storage condition -20°C to 60°C (-4°F to 140°F), 0~99%RH (avoid condensation)
Power Supply 12 VDC adaptor
Standard packing includes Meter, Manual, Power adapter 100-240vac
Weight Approx. 193g (0.43 lb)
CO2-200 CO2 Wall-Mounted Meter
Amprobe® Test Tools
website: www.Amprobe.com
email: info@amprobe.com
Everett, WA 98203
Tel: 877-AMPROBE
Amprobe® Test Tools Europe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
©2009 Amprobe Test Tools. All rights reserved.
7/2009 3520597 Rev A