Models 2534, 2540 & 2542 dual channel
Digital Storage Oscilloscopes deliver an
unmatched combination of performance and
value. Analog style controls combined with an
Auto measurement function make these
oscilloscopes easy to use. Advanced features
such as FFT function, digital filtering, waveform
recorder, delayed sweep/zoom, mask testing
and automatic measurements provide you with
powerful tools to debug your circuits.
Features
60 MHz & 100 MHz bandwidth, 1 GSa/s
real time sample rate
4000 point record length for each channel
Color LCD display
USB front panel host port for USB flash
drives standard
USB device interface standard
Advanced features include digital filter with
adjustable limits, mask testing and waveform
recorder/replay mode
24 automatic measurements
FFT standard plus 3 additional math
functions
Extensive Trigger capabilities including
pulse width and line-selectable video trigger
Multiple language interface
PC Software that lets you remotely control
the oscilloscope and capture, save and
analyze waveform data
Digital Storage Oscilloscopes
Models 2534, 2540 & 2542
FULL FEATURED OSCILLOSCOPES
THAT WON’T BREAK YOUR BUDGET
Technical data subject to change
© B&K Precision Corp. 2008
v040109
Data sheet
wwwwww..bbkkpprreecciissiioonn..ccoomm
Tel.: 714.921.9095
The oscilloscopes come with PC Software that
lets you easily capture, save and analyze
waveforms and measurement results. Unlike
other DSOs in this price category, each model
includes two 150 MHz high performance passive
probes that will not limit the bandwidth of your
measurement system.
The 2534, 2540 & 2542 are ideal
oscilloscopes for use in education and training,
design and debug, service and repair.
FFT spectrum analysis screen
2534 60 MHz 400 MSa/s
2540 60 MHz 1 GSa/s
2542 100 MHz 1 GSa/s
Model Bandwidth Sample Rate
22wwwwww..bbkkpprreecciissiioonn..ccoomm
11))UUSSBB hhoosstt ppoorrtt
Connect your USB flash drive to conveniently
store and recall waveform data (binary or csv),
setups and screen shots (bmp format). You can
also update the oscilloscope’s firmware from
this port.
22))EEaassyy sseettuupp aanndd uussee
The Auto button identifies the input signal and
automatically sets up the vertical, horizontal
and trigger controls to produce a useable
display.
You can automatically adjust the timebase to
view the waveform as single cycle or multiple
cycle.
33))PPrriinntt bbuuttttoonn
Simply press the Print button to save a screen
shot in bitmap format to a USB flash drive
44))MMeennuu OOnn//OOffff bbuuttttoonn
Configure the menu parameters and hide the
menu with the push of a button to view your
signal in full screen (12 divisions).
55))AAddvvaanncceedd ttrriiggggeerriinngg
Isolate the signal with advanced triggering
including pulse width and selectable video
trigger. Use the alternate trigger function,
typically only found in analog oscilloscopes, for
a stable display of signals unrelated in time.
66))WWaavveeffoorrmm aannaallyyssiiss wwiitthh mmaatthh
aanndd FFFFTT
Analyze your signals with add, subtract and
multiply functions. View the signal’s
frequency spectrum and perform harmonic
distortion analysis.
77))TTiimmee aanndd ddaattee ssttaammpp
Save files to external memory complete with
time and date stamp to help you stay
organized.
88))AAuuttoo ccaalliibbrraattiioonn
Automatically calibrate the instrument’s vertical
and horizontal system for optimal measurement
accuracy
Front panel features
7 3 2 8
1 4 6 5
Digital Storage Oscilloscopes
Models 2534, 2540 & 2542
CCoonnvveenniieenntt SSttoorraaggee CCoommppaarrttmmeenntt
Store accessories in the oscilloscope’s storage
compartment and keep your work bench clutter
free
33wwwwww..bbkkpprreecciissiioonn..ccoomm
Filter out unwanted signal components, such
as various types of noise, with the built in
digital filter. Select from lowpass, highpass,
bandpass or notch filter. The limits are
adjustable over a wide range. The available
range varies with each timebase settings. (e.g.
the lowpass filter corner frequency can be set
as low as 40 Hz when selecting a
timebase of 5 ms/div).
Create a user defined mask (pass/fail limits)
and automatically compare it against the input
signal from CH1 or CH2. This feature is ideal
for manufacturing test applications that
require instant go/no go test results.
Navigate your USB flash drive directory and
files with ease. Store and retrieve waveform
data, screen shots and setups complete with
time and date stamp and user defined names.
Monitor and analyze long term signal behavior
by recording data continuously over an
extensive period of time and playing it back for
post acquisition analysis. Data is recorded in a
sequence of up to 1000 frames of 4 k data
points each and the time interval between each
frame is adjustable from 1 ms 100 s. The
data can be saved in a single file to internal
memory or USB flash drive.
Display and measure the input signal’s
frequency spectrum. Select one of the 5 FFT
windows: Rectangular, Hanning, Hamming,
Blackman and Flattop. Use cursors to
measure the spectral component’s magnitude
and frequency.
Noisy signal
Noise free signal after applying digital lowpass filter
The tools you need
DDiiggiittaall ffiilltteerriinngg
DDeellaayyeedd SSwweeeepp//ZZoooomm
PPoowweerrffuull mmeeaassuurreemmeenntt ffuunnccttiioonnss
MMaasskk tteessttiinngg
UUsseerr ffrriieennddllyy iinntteerrffaaccee ffoorr ffiillee hhaannddlliinngg
WWaavveeffoorrmm RReeccoorrddeerr
Use the oscilloscope’s delayed sweep feature
to zoom in on a particular area of a signal in
real time while viewing the entire captured
waveform simultaneously.
Digital Storage Oscilloscopes
Models 2534, 2540 & 2542
The included Comsoft PC software provides
full access to the oscilloscope’s display,
measurements waveform data and front panel
controls through the rear panel USB device
port.
The software provides a seamless synchroniza-
tion between the oscilloscope and PC,
effortlessly allowing quick imports of captured
waveform data and measurement results into
Microsoft Excel for further analysis.
All oscilloscope parameters can be easily
controlled via a PC without the need for
programming. Front panel knobs can be
emulated via a virtual panel. Alternatively,
parameters can be selected from a menu.
PPCC ccoonnnneeccttiivviittyy aanndd ddooccuummeennttaattiioonn
44wwwwww..bbkkpprreecciissiioonn..ccoomm
PPeerrffoorrmmaannccee CChhaarraacctteerriissttiiccss
Bandwidth 60 MHz 60 MHz 100 MHz
Real time sample rate 400 MSa/s 1 GSa/s
(2 channels interleaved)
Channels 2
Display 5.7 inch (145 mm) diagonal
Color LCD
Rise Time <5.83 ns <5.83 ns <3.50 ns
Record Length 4000 points
Vertical Resolution 8 bits
Vertical Sensitivity 2 mV - 5 V/div
DC gain accuracy ±3.0 %
Maximum Input Voltage 400 Vpk, CAT II (between signal and reference
BNC connector)
Position Range ± 8 divisions from center of screen
Bandwidth Limit 20 MHz
Time Base range 2.5 ns/div 50 s/div (2534)
2 ns/div 50 s/div (2540 & 2542)
Timebase accuracy 100 ppm
Input Coupling AC, DC, GND
Input Impedance 1 M in parallel with19 pf
Vertical and Horizontal Zoom Vertically or horizontally expand or compress a live
or stopped waveform
I/O interface USB host port on front panel supports USB flash
drives. USB device port for connection to PC
(Requires included Comsoft Software for use)
AAccqquuiissiittiioonn MMooddeess
Sample Display sample data only
Peak Detect
Average Waveform averaged, selectable from
2, 4, 16, 32, 64, 128, 256
Roll Mode For time base settings 500 ms/div-50 s/div
TTrriiggggeerr SSyysstteemm
Trigger Types Edge, Pulse Width, Video*
Trigger Modes Auto, Normal, Single
Trigger Coupling AC, DC, LF reject, HF reject
Trigger Source CH1, CH2, AC line, Ext, Ext/5, Alternate
*Support formats PAL/SECAM, NTSC. Triggers on odd or even field, all lines or line number
CCuurrssoorrss
Types Amplitude, Time
Measurements V, T, 1/ T
AAuuttoommaattiicc WWaavveeffoorrmm MMeeaassuurreemmeenntt
Time Rise time, Fall Time, Cycle Frequency, Period, Positive
Pulse Width, Negative Pulse width, Delay, Phase, X at
Min, X at Max
Voltage MAX, MIN, Peak-Peak, Average, Vrms, High, Low,
Amplitude, Cycle RMS, Cycle Average, Overshoot,
Preshoot
Frequency Hardware counter provides frequency readout of
trigger source with 5 digit resolution
WWaavveeffoorrmm MMaatthh
Math function FFT, add, subtract, multiply
FFT Windows: Hanning, Hamming, Blackman, Rectangular,
Flattop,
2048 sample points
AAuuttoosseettSingle button automatic setup of both channels for
vertical, horizontal and trigger systems
DDiissppllaayy
Display Mode 1/4 VGA (5.7”) 256 color LCD (320x240) with
adjustable contrast and inverse video
Display Types Point, Vector
Persistence Off, infinite
Waveform Interpolation Sin(x)/x, Linear
Format YT and XY
PPoowweerr RReeqquuiirreemmeennttss100-240 VAC, 50 VAmax, 47 Hz to 440 Hz
EEnnvviirroonnmmeennttaall
Temperature Operating: 0º C to +40º C
Nonoperating: -20º C to +55º C
Humidity Operating: 95 % RH, 40º C
Nonoperating: 90 % RH, 55º C
Altitude Operating to 3000 m
Pollution Degree Pollution degree 2 for indoor use only.
EElleeccttrroommaaggnneettiicc ccoommppaattiibbiilliittyy aanndd SSaaffeettyy
EMC This oscilloscope is in compliance with council EMC
directive 2004/108/EC
Safety EN61010-1:2001
GGeenneerraall
Dimensions 310 mm (W) x 147 mm (H) x 269 mm (D)
Width x Height x Depth 12.2 in x 5.8 in x 10.6 in
Weight 3.6 kg (8 lbs)
Accessories
Supplied: User Manual, Two 150 MHz 10:1 passive probes (model PR 37A),
Power cord , USB interface cable, Comsoft Software Installation disk
Optional: PR 32A Demodulator Probe, PR 55 High Voltage Probe
Digital Storage Oscilloscopes
Models 2534, 2540 & 2542
Specifications
2534 2540 2542
model
Two Year Warranty