DIGITAL MULTIMETERS & SYSTEMS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY (U.S. only)
2001
2002 7½-Digit High Performance Multimeter
8½-Digit High Performance Multimeter
2001 Condensed Specifications (continued)
AC AMPS
ACI ACCURACY1, 2
90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range, ±(% of reading + % of range)
RANGE
20Hz–
50Hz
50Hz–
200Hz
200Hz–
1kHz
1kHz–
10kHz
10kHz–
30kHz3
30kHz–
50kHz3
50kHz–
100kHz3
200 µA 0.35 + 0.015 0.2 + 0.015 0.4 + 0.015 0.5 + 0.015
2 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
20 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.12 + 0.015 0.25 + 0.015 0.3 + 0.015 0.5 + 0.015
200 mA 0.3 + 0.015 0.15 + 0.015 0.12 + 0.015 0.15 + 0.015 0.5 + 0.015 1 + 0.015 3 + 0.015
2 A 0.35 + 0.015 0.2 + 0.015 0.3 + 0.015 0.45 + 0.015 1.5 + 0.015 4 + 0.015
AC AMPS NOTES
1. Specifications apply for sinewave input, AC+DC coupling, 1 power line
cycle, digital filter off, following 55 minute warm-up.
2. Add 0.005% of range uncertainty for current above 0.5A rms for self-
heating.
3. Typical values.
FREQUENCY COUNTER
AC VOLTAGE INPUT: 1Hz–15MHz.
ACCURACY: ±(0.03% of reading).
DC IN-CIRCUIT CURRENT
TYPICAL RANGES: Current: 100µA to 12A. Trace Resistance: 1mW to 10W typical.
ACCURACY: ±(5% + 2 counts). For 1 power line cycle, Auto Zero on, 10 reading digital filter,
TCAL ±5°C, after being properly zeroed. 90 days, 1 year or 2 years.
TEMPERATURE
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90 and 100W platinum
RTDs DIN 43 760 or IPTS-68.
GENERAL
POWER: Voltage: 90–134V and 180–250V, universal self-selecting. Frequency: 50Hz, 60Hz,
or 400Hz self-identifying. Consumption: <55VA.
ENVIRONMENTAL: Operating Temperature: 0° to 50°C. Storage Temperature: –40° to
70°C. Humidity: 80% R.H., 0° to 35°C, per MIL-T-28800E1 Para 4.5.5.1.2.
PHYSICAL: Case Dimensions: 90mm high × 214mm wide × 369mm deep (3½ in. × 8½ in.
× 14½ in.). Net Weight: <4.2kg (<9.2 lbs.). Shipping Weight: <9.1kg (<20 lbs.).
STANDARDS
EMI/RFI: Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2. Meets FCC part 15 Class
B, CISPR-22 (EN55022).
Safety: Conforms to IEC348, CAN/CSA-C22.2. No. 231, MIL-T-28800E1. Designed to
UL1244.
Note 1: For MIL-T-28800E, applies to Type III, Class 5, Style E.
For complete specifications, refer to the 2001 Technical Data book.
2002 Condensed Specifications
DC VOLTS
DCV INPUT CHARACTERISTICS AND ACCURACY
Enhanced Accuracy 1 – 10PLC, DFILT 10 Relative Accuracy ±(ppm of reading + ppm of range)
Range Full Scale Resolution
Input
Resistance Transfer 5 24 Hours 2 90 Days 3 1 Year 3 2 Years 3
20 0 mV 4 ±210.000000 mV 1 nV >10 0 GW0.4 + 1.5 3.5 + 3 15 + 8 19 + 9 23 + 10
2 V 4 ±2.10000000 V 10 nV >10 0 G W0.2 + 0.15 1.2 + 0.3 6 + 0.8 10 + 0.9 14 + 1
20 V ±21.0000000 V 100 nV >10 0 G W0.1 + 0.05 1.2 + 0.1 6 + 0.15 10 + 0.15 14 + 0.15
200 V ±210.000000 V 1 µV 10 MW ±1% 0.5 + 0.08 5 + 0.4 14 + 2 22 + 2 30 + 2
1000 V 6 ±1100.00000 V 10 µV 10 MW ±1% 1 + 0.05 5 + 0.08 14 + 0.4 22 + 0.4 30 + 0.4
Normal Accuracy 7 – 1PLC, DFILT off
Relative Accuracy
±(ppm of reading + ppm of range)
Range Full Scale Resolution
Input
Resistance 24 Hours290 Days31 Year32 Years3
20 0 mV 4 ±210.00000 mV 10 nV >10 0 G W3.5 + 6 15 + 11 19 + 12 23 + 13
2 V 4 ±2.1000000 V 100 nV >10 0 G W1.2 + 0.6 6 + 1.1 10 + 1.2 14 + 1.3
20 V ±21.000000 V 1 µV >10 0 GW3.2 + 0.35 8 + 0.4 12 + 0.4 16 + 0.4
200 V ±210.00000 V 10 µV 10 MW ±1% 5 + 1.2 14 + 2.8 22 + 2.8 30 + 2.8
1000 V 6 ±1100.0000 V 100 µV 10 MW ±1% 5 + 0.4 14 + 0.7 22 + 0.7 30 + 0.7
DC VOLTS NOTES
1. Specifications are for 10 power line cycles, synchronous autozero,
10-reading repeat digital filter, autorange off, except as noted.
2. For TCAL ±1°C, following 4-hour warm-up. TCAL is ambient temperature
at calibration (23°C at the factory). Add 0.5ppm of reading uncertainty
if the unit is power cycled during this interval.
3. For TCAL ±5°C, following 4-hour warm-up.
4. Care must be taken to minimize thermal offsets due to operator cables.
5. Specifications apply for 20-reading repeat digital filter, TREF ±0.5°C
(TREF is the initial ambient temperature), and for measurements within
10% of the initial measurement value and within 10 minutes of the
initial measurement time.
6. Add 20ppm × (VIN/1000V)2 additional uncertainty for inputs above
200V, except in transfer accuracy specifications.
7. Specifications are for 1 power line cycle, normal autozero, digital filter
off, autorange off.
AC VOLTS Normal Mode RMS1
90 Days, 1 Year or 2 Years, ±2°C from last AC self-cal, for 1% to 100% of range 2
±(% of reading + % of range)
Range 20–50Hz 50–100Hz 0.1–2kHz 2–10kHz 10–30kHz 30–50kHz 50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV 0.25 + 0.015 0.07 + 0.015 0.02 + 0.02 0.02 + 0.02 0.025 + 0.02 0.05 + 0.01 0.3 + 0.015 0.75 + 0.025 2 + 0.1 5 + 0.2
2 V 0.25 + 0.015 0.07 + 0.015 0.02 + 0.02 0.02 + 0.02 0.025 + 0.02 0.05 + 0.01 0.3 + 0.015 0.75 + 0.025 2 + 0.1 5 + 0.2
20 V 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.04 + 0.015 0.05 + 0.015 0.07 + 0.015 0.3 + 0.015 0.75 + 0.025 4 + 0.2 7 + 0.2 4
200 V 3 0.25 + 0.015 0.07 + 0.015 0.03 + 0.015 0.04 + 0.015 0.05 + 0.015 0.07 + 0.015 0.3 + 0.015 0.75 + 0.025 4 4 + 0.2 4
750 V 3 0.25 + 0.015 0.1 + 0.015 0.05 + 0.015 0.06 + 0.015 0.08 + 0.015 0.1 + 0.015 4 0.5 + 0.015 4
AC VOLTS NOTES
1. Specifications apply for sinewave input, AC +
DC coupling, 1 power line cycle, autozero on,
digital filter off, following 55-minute warm-up.
2. For 1% to 5% of range below 750V range,
and for 1% to 7% of 750V range, add 0.01% of
range uncertainty. For inputs from 200kHz
to 2MHz, specifications apply above 10% of
range.
3. Add 0.001% of reading × (VIN/100V)2 addi-
tional uncertainty for inputs above 100V rms.
4. Typical values.
Model 2001 and 2002 specifications
Model 2001 and 2002 specifications