PN 942735 July 1994 Rev.12, 11/01
©1994-2001 Fluke Corporation. All rights reserved. Printed in USA
All product names are trademarks of their respective companies
700 Series Pressure Modules
Instruction Sheet
15 PSID/G
PRESSURE MODULE
700
P
04
1 bar100 kPa
RANGE
Low
High
Differential
100 PSIG
PRESSURE MODULE
700
P
06
7 bar700 kPa
RANGE
Gage
yz1f.eps
Figure 1.
Introduction
The Fluke 700 Series Pressure M odul es allow you to measure
pressure with the Fluke 700 S eri es Process Calibrators. Read this
sheet before you use the pres sure module. This sheet contai ns
specifications and informat i on about how to avoid damaging the
pressure modul e. See the Calibrator Users Manual for operating
instructions.
The pressure modul e m easures press ure using an internal
microproc essor. It receives operating power from and sends digit al
informat i on to the 700 Series calibrator.
Gage pressure modul es have one press ure fitting and m easure
pressure with respect t o atmospheric pressure. Differential pressure
modules hav e two pressure fi ttings and measure the diff erence
between the applied pressure on the high fitting versus the l o w f i tting.
Figure 1 shows the two types . A differential press ure m odul e f unctions
like a gage modul e when the low fitting is open.
Compatibility of Pressure Modules with Fluke
Process Calibrators
Fluke-701, 702, 741, 743, and 744 Documenting Process
Calibrators: The Fluk e-700P00 and P27 Press ure M odul es are
completely compatible with these calibrators.
Fluke-716, 717, and 718 P ressure Cal ibrators: The Fluke-700P00
and P27 Press ure Modules will display pressure readings correctly on
716 Press ure Cal i brators with f i rm ware V1.2 or higher, or 717 and 718
Press ure Cal i brators with f i rm ware V1.3 or higher. To view the sof tware
revisi on, power-on the cali brat or while holding down the MAX key.
Read the REV num ber i n the display. Contact your Fluke Serv i ce
Center for a free update of cali brat o rs with lower revision numbers .
Reference P CN 4647.
Fluke-725 Multifuncti on P r ocess Cal ibrators: The Fluke-700P00 and
P27 Press ure Modules will display correc t pressure readings when
used with Fl uke-725 Multif unction Proc ess Calibrators at internal
software revision 1. 9 and l at er. To view the s oftware revis i on, power-on
the cali brator while holding down the 0% key. Read t he REV number in
the displ ay. Contact your Fluke Service Center f or a f ree update of
calibrat ors with lower rev i sion numbers. Reference PCN 4578.
Box Contents
Pressure m odul e, strap, 1/ 4 NPT to 1/4 IS O m etric adapters , i nstructi on
sheet.
Protecting Yourself from Pressure Releases
To avoid a violent release of pressure in a pressurized system, shut off
the isol at i on valve and sl owl y bleed off t he pressure before y ou at tach
or remove the pressure module f rom the pressure line.
Avoiding Overpressure Damage
Applyi ng pressure in exc ess of t he B URST PRESSURE specifi ed on
the press ure m odul e can destroy the pressure modul e. Burst pressure
is 3X f ul l scale (2X for 700P09, 30X f or 700P 00).
Table 1. Pressure Module Specifications1
MODEL RANGE2GAGE OR
DIFFERENTIAL ISOLATED OR
NONISOLATED REFERENCE
UNCERTAINTY
(23° ± 3°C)
STABILITY
(1 YEAR) TEMP
(0° to 50°C) TOTAL UNCERTAINTY
(% of Full Scale)3 1 YEAR
FLUKE-700P00 0 to 1.000 in H2O
0 to 0.249 kPa
0 to 0.002 bar
Differential Nonisolated 0.300% 0.025% 0.025% 0.350%
FLUKE-700P01 0 to 10.00 in H2O
0 to 2.49 kPa
0 to 0.02 bar
Differential Nonisolated 0.200% 0.050% 0.050% 0.300%
FLUKE-700P02 0 to 1.0000 psi
0 to 6.8900 kPa
0 to 6.89 E2 bar
Differential Nonisolated 0.150% 0.070% 0.080% 0.300%
FLUKE-700P03 0 to 5.0000 psi
0 to 34.000 kPa
0 to 0.3400 bar
Differential Nonisolated 0.050% 0.020% 0.030% 0.100%
FLUKE-700P04 0 to 15.000 psi
0 to 100.00 kPa
0 to 1.0000 bar
Differential Nonisolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P05 0 to 30.000 psi
0 to 200.00 kPa
0 to 2.0000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P06 0 to 100.00 psi
0 to 700.00 kPa
0 to 7.0000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P27 0 to 300.00 psi
0 to 2000.0 kPa
0 to 20.000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P07 0 to 500.00 psi
0 to 3400.0 kPa
0 to 34.000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P08 0 to 1000.0 psi
0 to 7000.0 kPa
0 to 70.000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
FLUKE-700P09 0 to 1500.0 psi
0 to 10000.0 kPa
0 to 100.000 bar
Gage Isolated 0.025% 0.010% 0.015% 0.050%
1. Use of pressure zero function is required to achieve these specifications. Contact your Fluke Service Center for upgrade of your Fluke 701 or Fluke 702 V1.0, V1.1, or
V1.2 calibrator.
2. Available pressure units (psi, kPa, bar, inHg, mmHg, inH20, ftH20., kg/cm2, mmH20, cmH20) are determined by the calibrator being used.
3. Accuracy specifications apply for 0 to 100% of full scale, 0 to 50°C. 1% FS typical, -10 to 0°C. On 700P00 only, specification applies from 15 to 35°C.
4. Maximum Non-Destructive Pressure: 3X maximum rated pressure, including common mode pressure. (700P09: 2X, 700P00: 30X)
5. Maximum Common Mode Pressure: 3X maximum rated pressure. (700P09: 2X, 700P00: 30X)
6. Specifications reflect a confidence interval of 95%.
Avoiding Mechanical Damage
To avoid mec hani cally damagi ng the pressure module, never apply
more than 10 ft.-lbs. of torque between the pressure module fittings or
between the fi ttings and t he body of the module. Always appl y
appropriate torque bet ween the pressure modul e f itting and c onnecting
fitt i ngs or adapters. Fi gure 2 shows the correct way and incorrect
ways t o use a wrench when applyi ng t orque to the press ure m odul e
fitting.
15 PSID/G
PRESSURE MODULE
700
P
04
1 bar
100 kPa
RANGE
yz2f.eps
Figure 2.
Avoiding Corrosion Da mage
To avoid damaging the pressure modul e f rom corrosion, use it only
with specified medi a as shown below:
Isolat ed modules: any m edi um that is compatible with type 316
stainl ess steel .
Nonisolat ed m odul es: dry, noncorrosive gasses only.
Recommended Measurement Technique
For best res ul t s, it is recommended t hat the module be pressurized to
full s cale and then vent ed to zero press ure (at mosphere) prior to
zeroing and mak i ng measurements wi th the cali brat or.
Note
Low range pressure modules may be sensitive to
gravity. For best results, pressure modules 30 psi and
below should be held at the same physical orientation
from the time they are zeroed until the measurement is
complete.
Performance Test
If you need to check that the pressure module meets i ts acc uracy
specification, use a dead weight t ester or suit able pressure cal ibrator.
The accurac y of the dead weight tester or pressure calibrato r should
be at least 4X better than t he 700 Series pressure specif i cation.
Proceed as f ol l o ws to verif y that a pressure module is operating within
specification:
1. Read the press ure value with no externally appli ed pressure to
make sure the 0% of scale is c orrect. When reading t he pressure
with a 700 Series, press t he ZERO key to rem ove any zero
offset.
2. Connect the pressure module t o a dead wei ght tester.
3. Set the dead wei ght tester t o 20% of the pressure modules full
scale value.
4. Make sure the reading agrees wit h t he dead wei ght tester value
within the Total Uncertainty speci fication i n Tabl e 1.
5. Set the dead wei ght tester t o 40, 60, 80, and 100% of full scale
and compare the respectiv e readi ngs.
6. If tem perature sensit i vity i s of concern, repeat steps 1 through 5
at various controlled temperatures.
Pressure Calibration Kit
The Fluke-700PCK P ressure Calibrati on Kit makes it poss i bl e to
calibrat e pressure modules at ambient tem perature with a preci sion
pressure c al i brator or dead weight tes ter at least 4x better than the
module spec i fication. A 386 or better P C and Wi ndows® 3.1 or better
are required. The ki t i s an optional ac cessory available from your
distributor or Fluke.
Certifications
P
Conforms to European Standard EN61010-1, EN61326.
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and
workmanship under normal use and service. The warranty period is one year and
begins on the date of shipment. Parts, product repairs and services are warranted
for 90 days. This warranty extends only to the original buyer or end-user customer
of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or
to any product which, in Flukes opinion, has been misused, altered, neglected or
damaged by accident or abnormal conditions of operation or handling. Fluke
warrants that software will operate substantially in accordance with its functional
specifications for 90 days and that it has been properly recorded on non-defective
media. Fluke does not warrant that software will be error free or operate without
interruption.
Fluke authorized resellers shall extend this warranty on new and unused products
to end-user customers only but have no authority to extend a greater or different
warranty on behalf of Fluke. Warranty support is available if product is purchased
through a Fluke authorized sales outlet or Buyer has paid the applicable
international price. Fluke reserves the right to invoice Buyer for importation costs
of repair/replacement parts when product purchased in one country is submitted
for repair in another country.
Flukes warranty obligation is limited, at Flukes option, to refund of the purchase
price, free of charge repair, or replacement of a defective product which is
returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center
or send the product, with a description of the difficulty, postage and insurance
prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke
assumes no risk for damage in transit. Following warranty repair, the product will
be returned to Buyer, transportation prepaid (FOB Destination). If Fluke
determines that the failure was caused by misuse, alteration, accident or
abnormal condition of operation or handling, Fluke will provide an estimate of
repair costs and obtain authorization before commencing the work. Following
repair, the product will be returned to the Buyer transportation prepaid and the
Buyer will be billed for the repair and return transportation charges (FOB Shipping
Point).
THIS WARRANTY I S BUYERS SOLE AND EXCLUSIVE REMEDY AND IS IN
LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
BUT NOT LIM ITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES OR LO SSES, INCLUDING LOSS OF DATA, WHETHER ARISING
FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT,
RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied
warranty, or exclusion or limitation of incidental or consequential damages, the
limitations and exclusions of this warranty may not apply to every buyer. If any
provision of this Warranty is held invalid or unenforceable by a court of competent
jurisdiction, such holding will not affect the validity or enforceability of any other
provision.
In Case of Difficulty
For service or calibrati on, call your nearest authoriz ed Fl uke Servic e
Center. To contact Fluk e call one of the following telephone num bers:
USA: 1-888-99-FLUK E (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in t he worl d: +1-425-446-5500
Or, visit Flukes Web site at www.fluke. com.
Fluke Corporation Fluke Europe B .V.
P.O. B ox 9090 P.O. B ox 1186
Everett, WA 5602 B.D. Eindhoven
98206-9090 The Netherlands
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Fluke:
FLUKE-700P05EX FLUKE-700P06EX FLUKE-700P09EX FLUKE-700P01EX FLUKE-700P03 FLUKE-700P09
FLUKE-700P04 FLUKE-700P06 FLUKE-700P08 FLUKE-700P01 FLUKE-700P00 FLUKE-700P05 FLUKE-700P02