HIGH VOLTAGE
PPE 20 kV
GENERAL CHARACTERISTICS
Operating temperature: 0 °C to 50 °C
Cable length: 3 m
Weight of probe + accessories: 495 gr
Coupling: 1 MDC/AC
Warranty: One year
Overvoltage category: CAT I
Pollution degree: 1
CLEANING
The exterior of the probe only should be
cleaned with a soft cloth moistened with
either water or isopropyl alcohol.
USING AND MAINTENANCE
The probe user should note that the protection
provided by the equipment may be impaired,
if the equipment is used in a manner not
specified by the manufacturer.
To guarantee accurate performance
characteristics, machanical shocks should be
avoided, as well as damage to the cable
through excessive bending. Handle with care,
especially when the extra thin spring-tip is
fitted. All maintenance should be referred to
qualified personnel.
SALES & SERVICE LECROY
Corporate Headquarters
U.S.A.: Chestnut Ridge
Phone (1) 914 578 6020
Fax (1) 914 578 5985
European Headquarters
Germany: Heidelberg
Phone (49) 6221 827 00
Fax (49) 6221 834 655
SAFETY INFORMATION
The probe is designed to be safe under the
following conditions:
Indoor use
Altitude to 2000 m
Temperature range from 0 °C to 50 °C
Maximum relative humidity 80 % for tem-
peratures up to 31 °C, decreasing linearly
to 40 % relative humidity at 50 °C
WARNING
Do not disassemble or disconnect probes while
they are connected to a voltage source. Use only
grounded oscilloscopes.
Attention: Refer to this manual before
using the probe.
Danger. High Voltage. ELECTRICAL CHARACTERISTICS
System attenuation: 1000:1
System input resistance: 100 M
Input capacitance: < 3 pF
Compensation range: 10 pF to 65 pF
System BW (-3 dB): 100 MHz typical
Input voltage:
1. Max. voltage: 20 kV DC incl. peak AC
2. AC: 14 kV AC rms 50/60 Hz Sinus
3. HF-Sinus-AC: See derating chart
4. Pulse peak: 40 kV (see peak pulse
derating chart
LF PROBE ADJUSTMENT (1000:1)
Connect the probe to a 1 kHz square wave
signal. Adjust compensation trimmer in the
BNC-box for optimum square wave response.
Undershoot
Correct
Overshoot
HF PROBE ADJUSTMENT
If necessary the probe can be adjusted for high
frequency. Connect the probe to a 1 MHz square
wave signal and adjust the trimme rin order to
receive an optimum square wave response.
Undershoot
Correct
Overshoot
STANDARD ACCESSORIES SUPPLIED
WITH THE PROBE
Accessories Quantities
1. Trimming tool 1
2. Ground lead with a crocodile clip 1
3. Hook 1
HIGH VOLTAGE PROBE KIT PK104
Accessories Quantities
1. Trimming tool 1
2. Ground lead with a crocodile clip 1
3. Hook 1
4. BNC Adapter 1
1
3
4
2
Derating PPE 20 kV
1000
10000
100000
0,01 0,1 1 10 100
Frequency [MHz]
Maximum Impuls Derating - PPE20kV
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
10 100 1000
Duration [msec]
Duty Cycles