PyroUSB PC Congurable
Non-Contact Temperature Sensor with
4 to 20 mA Output
Congurable temperature range, emissivity setting
etc. from a PC via CalexSoft software and USB
cable (supplied)
Features max; min; average and instantaneous
readings; peak or valley hold; reected energy
compensation
OPC Server capabilities
Temperature range –40 to 1000°C
Emissivity: 0.1 to 1.0
Response time: 240 ms to 90%
Stainless steel housing, sealed to IP65
Field of view: 15:1, 30:1 or Close Focus
4 to 20 mA output
Quick and easy installation
Optional air/water cooled housing, air purge collar,
laser sighting tool and mounting brackets
+- -
+
Display/
Controller
Red
Black
(Blue)
Power Supply
+- -
+
Display/
Controller
Red
Black
(Blue)
Power Supply
PC
PC
N/C
USB
USB
The PyroUSB Series measures temperatures from
-40°C to 1000°C accurately and consistently, with an
outstanding response time of 240 ms. The 4 to 20
mA output is compatible with almost any indicator,
controller, recorder, data logger etc. without the need
for special interfacing or signal conditioning.
Model PU151 has 15:1 optics making it suitable for
most applications. PU301 is specially designed for
distant targets and has an optical resolution of 30:1.
PUCF is designed for small targets and measures a
spot size of ø5mm at 100mm distance.
All PyroUSB Series sensors are fully congurable
from a PC using the CalexSoft software and USB
cable supplied. This user friendly software enables
the user to set the range and emissivity, compensate
for reected energy; apply ltering; select max, min,
average or instantaneous readings; and peak or
valley hold processing. These features can also be
monitored and adjusted by an OPC Client. Other
features include Data Acquisition, Alarms and a
Scrolling Graphical Display.
The sensor will operate with either the 4 to 20 mA
cable connected, the USB cable connected, or both.
The USB cable has an IP65 connector at the sensor
end. An IP65 cap protects the sensor when the USB
cable is not connected.
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
15 32 48
0.6 1.2 1.9
0 19.7 39.4
0 500 1000
Spot Dia.
(inches)
Spot Dia.
(mm)
Spot Dia.
(inches)
Spot Dia.
(mm)
Spot Dia.
(inches)
Spot Dia.
(mm)
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
15 48 82
0.6 1.9 3.2
0 19.7 39.4
0 500 1000
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
15 515
0.6 0.2 0.6
0 3.9 7.9
0 100 200
D:S 30:1D:S 15:1
PU151 PU301 PUCF
Temperature Range
-40°C to 1000°C
Field-of-View
15:1 (PU151)
30:1 (PU301)
ø5mm @ 100mm (PUCF)
Output
4 to 20 mA (linear with temperature)
Conguration
Via PC port conforming to USB 2.0
Accuracy
±1% of reading or ±1°C whichever is greater
Repeatability
±0.5% of reading or ±0.5°C whichever is greater
Emissivity
0.1 to 1.0
Response Time, t90
240 ms (90% response)
Spectral Range
8 to 14 μm
Supply Voltage
24 V DC (28Vdc max)
Sensor Voltage
6 V DC min
Max Loop Impedance
900 Ω
Maximum Span
1000°C
Minimum Span
100°C
MECHANICAL
Construction
Stainless Steel
Dimensions
25 mm diameter x 106.5 mm long
Thread mounting
M20 x 1 mm pitch
Weight with Output Cable
175 g
Output Cable Length
1 m
USB Cable Length
1 m
ENVIRONMENTAL
Environmental Rating
IP65
Ambient Temperature Range
0°C to 70°C
Relative Humidity
95% max. non-condensing
GENERAL SPECIFICATIONS
DIAMETER OF TARGET SPOT MEASURED VERSUS DISTANCE FROM SENSING HEAD
AIR PURGE COLLAR
The air purge collar
is used to keep dust,
fumes, moisture and
other contaminants
away from the lens. Air
ows into the tting on
the side and out of the
aperture at the front.
FIXED MOUNTING BRACKET
The L-shaped xed
mounting bracket oers
a rigid support for the
sensor and allows ne
adjustment in a single
plane.
ADJUSTABLE MOUNTING BRACKETS
The adjustable mount-
ing bracket consists
of a xed mounting
bracket plus another L-
shaped bracket. When
assembled as shown
the adjustable mounting
bracket oers a rigid
support for the sensor
and allows ne adjustment in two planes.
AIR/WATER COOLED HOUSING
The air/water cooled
housing allows the
sensor to withstand
ambient temperatures
which exceed the
normal 70°C limit. Air
or water (depending on
the degree of cooling
required) ows into one of the ttings on the side and
out of the other. To prevent condensation forming on the
lens, the air/water cooled housing is supplied complete
with an air purge collar. Please note, the air/water
cooled housing must be ordered with the sensor and
cannot be tted by the user.
LASER SIGHTING TOOL
The Laser Sighting
Tool screws onto the
front of the sensor
during installation and
indicates precisely
where the sensor
is aiming. Once the
sensor has been
aimed at the centre
of the target and locked in position the Laser Sighting
Tool can be removed. The laser is activated by means
of a push button on the front of the tool which has a
latching mechanism.
ACCESSORIES
Calex Electronics Limited
PO Box 2, Leighton Buzzard, Bedfordshire, England LU7 4AZ
Tel: +44 (0)1525 373178/853800 Fax: +44 (0)1525 851319 Lo-call Tel: 0845 3108053
E-mail: mail@calex.co.uk Online: http://www.calex.co.uk
Issue C - March 14
Specications subject to change without notice
MODEL NUMBERS
Cooling
(blank) = Sensor without cooling
WJ = Air/water cooled jacket
with air purge collar
Field of view
151 = 15:1 divergent optics
301 = 30:1 divergent optics
CF = Close-focus optics (focal spot size 5 mm at
100 mm distance)
PUxxx xx