Keysight Technologies
33500B Series
Waveform Generators
Data Sheet
2
Trueform Technology


33500B Series waveform generators with exclusive Trueform signal
generation technology offer more capability, fidelity and flexibility than
traditional DDS generators. Use them to accelerate your development
process from start to fi nish.
Over the past two decades, direct digital synthesis (DDS) has been the
waveform generation technology of choice in function generators and
economical arbitrary waveform generators. DDS enables waveform
generators with great frequency resolution, convenient custom
waveforms, and a low price.
As with any technology, DDS has intrinsic downsides and limitations
as well. Engineers with exacting requirements have had to either work
around the compromised performance or spend up to 10 times more
for a high-end, point-by-point waveform generator.
The Keysight Technologies, Inc. Trueform technology offers a
new alternative that blends the best of DDS and point-by-point
architectures, giving you the benefits of both without the limitations
of either. Trueform technology uses an exclusive digital sampling
technique that delivers unmatched performance at the same low price
you are accustomed to with DDS.
The table below highlights the revolutionary capabilities of
Trueform technology.
DDS:
Traditional
25 MHz
waveform
generator
Trueform:
Keysight
30 MHz 33511B
waveform
generator
Improvement
Edge jitter 500 ps 40 ps 12x better
Custom
waveform
replication
Skips
waveform
points
100% point
coverage
Exact
waveform
replication
Total
harmonic
distortion
0.2% 0.04% 5x better
Anti-alias
filtering
Must provide
externally
Always
anti-aliased
No anti-aliasing
artifacts
Sequenced
arb
Not possible Standard Easily create
complex wave-
form sequences
For more information about Keysight Trueform technology
please visit:
www.keysight.com/find/trueform


Trueform technology DDS technology
Lower
Distortion
0.04%
THD
Trueform at 200 kHz
DDS at 200 kHz
Better
Signal
Integrity
<40 ps 500 ps
Reduced
Jitter
3
Key attributes
Jitter Risetime Total harmonic distortion Standard memory
600
500
400
300
200
100
0
Keysight
Trueform
40 ps
500 ps
Typical
DDS
30
20
10
0
20 ns
8.7 ns
Keysight
Trueform
Typical
DDS
0.25
0.20
0.15
0.10
0.05
0
0.04%
0.20%
Keysight
Trueform
Typical
DDS
1500
1000
500
0Keysight
33500B Series
1 M
128 k
Older
generators
Easily generate the full range of signals
you need for the most demanding
measurements
Test your devices with confidence that
the waveform generator is
outputting the signals you expect
Select just the capabilities you need now,
then upgrade easily when your needs
change
Unique features of the 33500B Series waveform generators
Full BandwIdth Pulse Full-bandwidth pulse to 20 or 30 MHz
Set leading and trailing edge times independently
2 Channels Dual channel coupling, frequency and amplitude, equal and inverted
Set start phase for each channel, set phase shift between channels
Sum Modulate Sum two signals together, frequency and amplitude independent
2-tone, square-sine, noise on pulse
Point-By-Point Arb Create up to 1 million samples standard, 16 million optional
Connect arbs together, create up to 512 sequences
Voltage Settings Lowest voltage range at 1 mVpp, a 10x improvement
Set high and low voltage limits to prevent overload on DUT
PRBS Patterns Provides standard PRBS patterns, PN7 … PN23
Select PN type, set bit rate, set edge time

4
The 33500B Series waveform generators offer the common signals
and features you expect, such as modulation, sweep, and burst.
But the 33500B Series offers many more features that give you the
capabilities and flexibility you need to get your job done. Like an
intuitive front-panel user interface that makes it easy to quickly relearn
it when you’ve been focused elsewhere. Like built-in LAN, USB
and GPIB interfaces that make it easy to control your instruments or
transfer your waveforms to your instrument.
But the 33500B Series doesn’t stop there. It offers a variety of
capabilities you won’t find elsewhere—capabilities that help you
accelerate your testing and get your project wrapped up faster:

capability
Easily add noise to your signal for margin and distortion testing
using only a single channel. You can create dual-tone
multifrequency signals without a dual-channel generator,
which means you can preserve your budget for other test
needs. On a two-channel model, you can sum and combine
up to four signals.

You can adjust the bandwidth of the built-in noise generator to
control the frequency content of your signal. Apply just the frequency
stimulus you need so you concentrate the energy of your waveform
in frequency bands of interest.
Trueform Technology


The images at right show
approximately 10 dB increase
in amplitude at 50 kHz when
the bandwidth is reduced 10x.
You can see how the signal
energy is increased in the
frequencies of interest when
the bandwidth is reduced,
instead of being spread over
a very wide bandwidth
with lower amplitude at all
frequencies.
Dual-tone signal created
by summing waveforms
using the modulation
type, “Sum.”
5
DDS technology may skip points at higher
frequencies


Define any waveform shape and any waveform length using point-by
-point arbitrary waveform capability. Your waveforms are always
anti-aliased for exceptional accuracy, and you can play them at any
rate you select. Play your signals as defined, at your exact sample
rate, without the chance of missing short-duration anomalies that are
critical for testing device reliability.
Waveform sequencing
Waveform sequencing lets you create multiple configured waveforms
with several common segments and lets you build long, complex
waveforms using minimal instrument memory.
Scope display
of waveform.
Create the waveform in the 33503A
Waveform Builder Pro and download it
to the waveform generator.
Waveform generator display
with downloaded waveform.
Trueform at 50 kHz
Trueform at 200 kHz
DDS at 50 kHz
DDS at 200 kHz misses details
DDS skips points at
higher bandwidths
6

(PRBS) pattern generation
Test your digital serial buses by streaming standard PRBS patterns—
like PN7 and PN19—without the need for a separate pulse
generator. With fewer instruments, setting up your tests is easier.
You won’t find these built-in PRBS patterns in competitive
waveform generators.

There are five ways to create arbitrary waveforms for use
with the 33500B generator.
1. Use the included Waveform Builder Basic software to edit
and download a waveform file to the generator
2. Use 33503A Waveform Builder Pro software to create
more complex waveforms and sequencing
3. Capture a waveform from an oscilloscope and download
it to the generator
4. Create a waveform in MATLAB, Excel, etc. and download
it to the generator
5. Use the generator’s front panel to edit a waveform once
it is in the generator
You have lots of flexibility to choose the way you want to work.
BUILT-IN
USB
drive
Direct link
or USB
CSV or ARB
file
CSV
file
Embedded
waveform editor
Waveforms
from
Oscilloscopes
Waveforms
from
MATLAB & Excel
drive, COMPATIBLE
FILES
Waveform
Builder
Basic
Waveform
Builder
Pro
Easily generate
PRBS waveforms on
the 33500B Series
waveform generators.
You can select multiple sequence lengths
(such as PN15) and bit rates up to
50 Mbit/sec to create PRBS signals.


Need a quick answer? Get instant access to instrument
documentation in seven different languages in
smart-phone-friendly WebHelp format. You can access
all user documentation in the palm of your hand—
no PC or hardcopy manuals required. Another feature
you won’t find in competitive function/arb generators.
Trueform Technology Capabilities continued
7
Built-in Web browser
Easily set up and control your 33500B Series generator remotely
over a LAN connection using the built-in LXI Web browser. You can
monitor your tests and adjust settings from another office or
room, or even from home.

If you want to test your design with long, complex waveforms
with a variety of anomalies, you need to make sure your waveform
generator has sufficient memory. The 33500B Series’ standard
memory is 1 MSamples deep.
Typical DDS generators
offer only a fraction of that
amount but with the
33500B Series there is even
a 16 MSample memory
option available.
Use the optional high-stability timebase
for even better accuracy
Get improved timebase stability and frequency accuracy using the
optional high-stability timebase. The optional timebase offers
0.1 ppm stability, which is 20x more stable than the standard
timebase over one year.
1500
1000
500
0Keysight
33500B Series
1 M
128 k
Older
generators
8
33500B Series waveform generators offer the

Lowest jitter
With 12x better jitter than anything in their class, 33500B Series
waveform generators offer unparalleled edge stability. You can
even use them as a system clock for timing and triggering your other
instruments. With better jitter performance, you can place edges more
accurately, helping you reduce timing errors in your circuit design.

With total harmonic distortion of just 0.04%, the 33500B Series offers
5x better fidelity than other generators. Clean, spurious-free signals
don’t introduce noise or artifacts. See your design’s characteristics, not
the waveform generator’s, in your measurements.

The 33500B Series’ 8.4-ns rise and fall times are more than twice as
fast as you’ll find in typical waveform generators. You can place edges
with more confidence and more accurately set trigger points. Because
of the faster transition, higher harmonic content is created, which
helps you expand your understanding of your circuit.
Trueform Technology

that your signal generator is outputting
the signals you expect
Keysight 33500B Series waveform generators offer the lowest total
harmonic distortion (THD) in its class.
Typical DDS generator has a higher noise floor and greater
harmonics.
<40 ps 500 ps
Trueform technology significantly improves jitter performance.
Trueform technology. Traditional DDS technology.
If your generator is introducing spurious signals or harmonics, you’ll
have a hard time producing reliable designs. To be successful, you
need to test with clean, precise, low-noise signals. Keysight 33500B
Series waveform generators offer the highest signal fidelity so you can
generate the exact waveforms you need for your most challenging
measurements. You can be confident you are seeing your design’s
characteristics, not your waveform generator’s, in your measurements.
9

Today’s ultra-low-power products such as pacemakers, hearing aids
and remote sensors use very low voltages. The 33500B Series lets
you create signals as low as 1 mVpp. That’s 10x better amplitude
resolution than typical waveform generators offer.

Eliminate the need for an external modulation source. The 33500B
Series has a modulation frequency up to the frequency of the
waveform being modulated. Existing DDS-based generators have
a much lower internal modulation frequency. Now you can create
your complex signals all within a single generator.

The 33500B Series’ 16 bits of resolution is 4x that of most waveform
generators. You can make output changes all the way down to
1 V—exactly what you need for testing today’s low-voltage circuits
and designs.

Create pulses up to 30 MHz with the 33500B Series. Most DDS-
based generators offer reduced bandwidth when generating pulses.
With a broader operating range, you have the frequency you need for
a wider range of applications.
14-BIT 16-BIT
4X improvement
in vertical resolution.
10
Trueform Technology



Investment protection
With most waveform generators, you get only what you pay for when
you buy your instrument. But with 33500B Series waveform genera-
tors, there are eight different models to choose from so you can
purchase the capability you need now and upgrade later when your
project needs change. Your investment in test equipment is protected.
If you need to generate 30 MHz waveforms or arbitrary waveforms,
or if you need deeper memory for generating more complex signals,
you can easily add the capability after the fact with software
upgrades. And there’s no price penalty for adding the capability later.

If you are doing simple experiments in digital communications,
use the optional IQ player to play IQ files on your 2-channel
function generator.
Select from eight models to get the capability that fits your budget
now—then take advantage of easy software upgrades to expand
your instrument’s capabilities when you are ready.
Optional IQ player allows you to play Q files
on your 2-channel Arb function generator.
20 MHz
NO ARB 30 MHz
NO ARB
20 MHz
ARB 30 MHz
ARB
33509B
1 CHANNEL
33521B
1 CHANNEL
33510B
2 CHANNEL
33519B
1 CHANNEL
33520B
2 CHANNEL
33522B
2 CHANNEL
33511B
1 CHANNEL
33512B
2 CHANNEL
11

Choose the model with the capability you
need now, knowing you can upgrade later.
All models come with a rich set of built-in,
standard features, including LAN, USB and
GPIB interfaces, 1 MSample of memory,
an external timebase input, and basic
waveform generation software. You get ev-
erything you need to generate clean, precise,
low-noise signals for testing your designs.
Large, color, graphical display offers
simultaneous parameter setup, signal viewing
and editing for easy operation.
Supports remote operation using
a web browser to connect to a
built-in web page.
LAN (LXI Class C), USB and GPIB
connectivity for quick and easy connectivity
to a PC or network .
Front-panel
USB port for
file management.
True point-by-point arbitrary waveforms
with sequencing for more accurate
representation of user-defined signals.
Dual-channel mode with
independent or coupled channels.
Model No. Description ARB
33509B 20 MHz, 1-channel
33510B 20 MHz, 2-channel
33511B 20 MHz, 1-channel
33512B 20 MHz, 2-channel
33519B 30 MHz, 1-channel
33520B 30 MHz, 2-channel
33521B 30 MHz, 1-channel
33522B 30 MHz, 2-channel
12
Easily create custom waveforms with

software
Get advanced signal creation/editing capability without tedious
programming with optional 33503A BenchLink Waveform Builder
Pro software. The Microsoft Windows-based program provides
easy-to-use creation tools, such as an equation editor, waveform
math and drawing tools, that make it easy to create custom signals.
It features a standard function library, waveform sequencer and
filters as well as windowing functions that allow you to easily
modify and further refine your waveform. A library of built-in signals
helps you quickly create more complex waveforms.
The result is quicker, easier creation of custom waveforms,
coupled with deeper analysis insight into your signals. For
additional information and to download a 30-day trial version of
the software, visit: www.keysight.com/find/33503trial.

Create and edit complex waveforms using 33503A Waveform Builder Pro software.
13



33500B Series waveform generators with Trueform technology
Bandwidth 20 MHz 20 MHz 30 MHz 30 MHz
Number of channels 1 2 1 2
Waveform generator 33509B 33510B 33519B 33520B
Waveform generator with arb capability 33511B 33512B 33521B 33522B
Application Order option
Additional memory for long waveforms MEM (only available on models with arb)
Baseband IQ Player with adjustments IQP (only available on 33512B/33522B)
Security features with NISPOM SEC
Ultra-high stability timebase OCX
Upgrade desired Order upgrade option
Increase bandwidth to 30 MHz 335BW1U on 1-channel models
335BW2U on 2-channel models
Add arbitrary waveform capability 335ARB1U on 1-channel models
335ARB2U on 2-channel models
Add 16M memory to arb 335MEM1U on 1-channel arb models
335MEM2U on 2-channel arb models
Add NISPOM and file security 335SECU
Add IQ baseband signal player to 2-channel arb 335IQPU
Add high-stability timebase 33500U-OCX (Must return to Keysight)
NOTE: Cannot upgrade a 1-channel generator to a 2- channel generator

14

Unless otherwise stated, all specifications apply with a 50 Ω resistive load and auto range ON.
Instrument characteristics
Models & options
33509B/11B/19B/21B 1-channel
33510B/12B/20B/22B 2-channel
Option MEM Increases arbitrary waveform memory to 16 MSa/channel*
Option OCX OCXO timebase for ultra-high stability
Option IQP IQ player (Only available on 33512B and 33522B)
Waveforms
Standard Sine, square, ramp, pulse, triangle, Gaussian noise,
PRBS (Pseudorandom Binary Sequence), DC
Built-in arbitrary* Cardiac, exponential fall, exponential rise, Gaussian pulse,
Haversine, Lorentz, D-Lorentz, negative ramp, sinc
User-defined arbitrary* Up to 1 MSa (16 MSa with Option MEM) with multi-segment sequencing
Operating modes & modulation types
Operating modes Continuous, modulate, frequency sweep, burst, output gate
Modulation types AM, FM, PM, FSK, BPSK, PWM, Sum (carrier + modulation)
Waveform characteristics
Sine
Frequency range 1 Hz to 20 MHz or 30 MHz, 1 Hz resolution
Amplitude flatness (spec)1, 2
(relative to 1 kHz)
< 100 kHz: ± 0.10 dB
100 kHz to 5 MHz: ± 0.15 dB
5 to 20 MHz: ± 0.30 dB
20 to 30 MHz:** ± 0.40 dB
Harmonic distortion (typ) 2, 3 < 20 kHz: < -70 dBc
20 to 100 kHz: < -65 dBc
100 kHz to 1 MHz: < -50 dBc
1 to 20 MHz: < -40 dBc
20 to 30 MHz:** < -35 dBc
THD (typ) 20 Hz to 20 kHz: < 0.04%
Non-harmonic spurious (typ) 2, 3 Standard: < -75 dBc, increasing +20 dB/decade above 2 MHz
Option OCX: < -75 dBc, increasing +20 dB/decade above 10 MHz
(or < -100 dBm, whichever is greater, below 500 MHz)
Phase noise (SSB) (typ)
1 kHz offset:
10 kHz offset:
100 kHz offset:
Standard Option OCX
-105 -110 dBc/Hz
-115 -125 dBc/Hz
-125 -135 dBc/Hz
***Only available on 33511B/12B/21B/22B
***Only available on 33519B/20B/21B/22B
NOTE: See page 22 for footnotes 1 through 10
15

Waveform characteristics, continued
Square & pulse
Frequency range 1 Hz to 20 MHz or 30 MHz, 1 Hz resolution
Rise and fall times (nom) Square: 8.4 ns, fixed
Pulse: 8.4 ns to 1 s, independently variable, 100-ps or 3-digit resolution
Overshoot (typ) < 2%
Duty cycle 0.01% to 99.99% 8
Pulse width 16 ns minimum, 100-ps resolution
Jitter (cycle-to-cycle, typ) < 40 ps rms
Ramp & triangle
Frequency range 1 Hz to 200 kHz, 1 Hz resolution
Ramp symmetry 0.0% to 100.0%, 0.1% resolution
(0% is negative ramp, 100% is positive ramp, 50% is Triangle)
Nonlinearity (typ) < 0.05% from 5% to 95% of the signal amplitude
Gaussian noise
Bandwidth (typ) 1 mHz to 20 MHz or 30 MHz, variable
Crest factor (nom) 4.6
Repetition period > 50 years
Pseudorandom binary sequence (PRBS)
Bit rate 1 mbps to 50 Mbps, 1 mbps resolution
Sequence length 2m -1, m=7, 9, 11, 15, 20, 23
Rise and fall times (nom) 8.4 ns to 1 s, variable, 100-ps or 3-digit resolution
Arbitrary waveform characteristics
IQ Player Characteristics
General
Waveform length 8 Sa to 1 MSa per channel (16 MSa with Option OCX) in increments of 1 sample
Sample rate 1 Sa/s to 250 MSa/s, 1 Sa/s resolution
Voltage resolution 16 bits
Bandwidth (-3 dB, nom) Filter Off: 40 MHz
“Normal” Filter On: 0.27 x (Sample Rate)
“Step” Filter On: 0.13 x (Sample Rate)
Rise and fall time 0.35 / Bandwidth (10 ns min) with "Normal" or "Step" filter On
Settling time (typ) < 200 ns to 0.5% of final value
Jitter (typ) Filter Off: < 40 ps rms
“Normal” or “Step” filter On: < 5 ps
Balance Adjust
Amplitude Gain
(channel to channel amplitude balance)
-30% to +30%
Delta Offset Ch 1 and Ch 2 ± (5 VDC - Peak AC) into 50 Ω
± (10 VDC - Peak AC) into open circuit
IQ Signal Skew
(adjusts channel to channel skew)
-4.00 ns to +4.00 ns
Views
Amplitude vs. Time
X-Y (constellation diagram)
16

Output characteristics
Isolation
Outputs Connector shells for channel output(s), Sync, and Mod In are connected
together but isolated from the instrument’s chassis. Maximum
allowable voltage on isolated connector shells is ±42 Vpk
Signal output
Output impedance (nom) 50 Ω
On, off, inverted User-selectable for each channel
Voltage limit User-definable VMAX and VMIN limits
Overload protection Output turns off automatically when an overload is applied
Instrument will tolerate a short-circuit to ground indefinitely
Amplitude
Range 1 mVpp to 10 Vpp into 50 Ω
2 mVpp to 20 Vpp into open circuit
Resolution 4 digits
Units Vpp, Vrms, or dBm, selectable
Accuracy 1, 2 (spec) ±1% of setting ±1 mVpp at 1 kHz
DC offset
Range 5±(5 VDC - Peak AC) into 50 Ω
±(10 VDC - Peak AC) into open circuit
Resolution 4 digits
Units VDC
Accuracy 1, 2 (spec) ±1% of Offset setting ±0.25% of Amplitude setting ±2mV
F 
Standard frequency reference (spec)
1 year, 23° C ± 5° C ±1 ppm of setting ± 15 pHz
1 year, 0° C to 55° C ±2 ppm of setting ± 15 pHz
High-stability frequency reference (spec): Option OCX
1 year, 0° C to 55° C ±0.1 ppm of setting ± 15 pHz

Waveform sequencing (Available only on 33511B/12B/21B/22B)
Operation Individual arbitrary waveforms (segments) can be combined into user-defined lists (sequences)
to form longer, more complex waveforms. Each sequence step specifies whether to repeat
the associated segment a certain number of times, to repeat it indefinitely, to repeat it until a
Trigger event occurs, or to stop and wait for a Trigger event. Additionally, the behavior of the
Sync output can be specified in each step.
To improve throughput, up to 32 sequences totalling up to 1,024 segments can be pre-loaded
into volatile memory.
Segment length 8 Sa to 1 MSa per channel (16 MSa with Option MEM) in increments of 1 sample
Sequence length 1 to 512 steps
Segment repeat count 1 to 1x106, or infinite
17


Carrier AM FM PM FSK BPSK PWM Sum Burst Sweep
Sine and Square ●●●●● ●●●
Pulse
Triangle and Ramp
Gaussian Noise a
PRBS
Single ARB* bb
Sequenced ARB*
Modulating signals
Carrier Sine Square Triangle / Ramp Noise PRBS ARB* External
Sine
Square and Pulse
Triangle and Ramp
Gaussian Noise
PRBS
ARB*

Amplitude modulation (AM)
Source Internal or external, or either channel with 2-channel models
Type Full-carrier or double-sideband suppressed-carrier
Depth 10% to 120%, 0.01% resolution
Frequency modulation (FM) 7
Source Internal or external, or either channel with 2-channel models
Deviation 1 Hz to 15 MHz, 1 Hz resolution
Phase modulation(PM)
Source Internal or external, or either channel with 2-channel models
Deviation 0° to 360°, 0.1° resolution
Frequency shift key modulation (FSK) 7
Source Internal timer or ext trig connector
Mark & space Any frequency within the carrier signal's range
Rate 0 Hz to 1 MHz
Binary phase shift key modulation (BPSK)
Source Internal timer or ext trig connector
Phase shift 0° to 360°, 0.1° resolution
Rate 0 Hz to 1 MHz
Pulse width modulation (PWM)
Source Internal or external, or either channel with 2-channel models
Deviation 80% to 100% of pulse width, 0.01% resolution
Additive modulation (Sum)
Source Internal or external, or either channel with 2-channel models
Ratio 60% to 100% of carrier amplitude, 0.01% resolution
a. Gated burst only. b. Applies to sample clock, not whole waveform.
* Only applies to 33511B/12B/21B/22B
18

Burst 9
Type Counted or gated
Count 1 to 1x108 cycles, or infinite
Gated Produces complete cycles while Ext Trig is asserted
Start/stop phase 4 -360° to 360°, 0.1° resolution
Trigger source Internal Timer or Ext Trig connector
Marker Adjustable to any cycle; indicated by the trailing edge of the Sync pulse
Sweep 7
Type Linear, Logarithmic, List (up to 128 user-defined frequencies)
Operation Linear and Logarithmic sweeps are characterized by a Sweep time (during which the
frequency changes smoothly from Start to Stop), a Hold time (during which the frequency
stays at the Stop frequency), and a Return time (during which the frequency changes
smoothly from Stop to Start). Returns are always linear.
Direction Up (Start freq < Stop freq) or Down (Start freq > Stop freq)
Start and stop frequencies Any frequency within the waveform's range
Sweep time Linear: 1 ms to 3600 s, 1 ms resolution;
3601 s to 250,000 s, 1 s resolution
Logarithmic: 1 ms to 500 s
Hold time 0 s to 3600 s, 1 ms resolution
Return time 0 s to 3600 s, 1 ms resolution
Trigger source 10 Immediate (continuous), external, single, bus, or timer
Marker Adjustable to any frequency between Start and Stop for Linear
and Logarithmic types or any frequency in the list for List type;
indicated by the trailing edge of the sync pulse

Range 1 s to 8000s, 6-digit or 4 ns resolution
2-channel characteristics (Only applies to 33519B/20B/21B/22B)
Operating modes Independent, coupled parameter(s), combined (Ch 1 + Ch 2),
Equal (Ch 2 = Ch 1), or differential (Ch 2 = -Ch 1)
Parameter coupling None, frequency (ratio or difference) and/or amplitude and DC offset
Relative phase 0° to 360°, 0.1° resolution
Skew (typ) < 200 ps (when performing identical operations)
Crosstalk (typ) < -85 dB
19

Sync/marker output
Connector Front-panel BNC, isolated from chassis
Functions Sync, sweep marker, burst marker, or arbitrary waveform marker
Assignment Channel 1 or channel 2
Polarity Normal or inverted
Voltage level (nom) 3 Vpp into open circuit, 1.5 Vpp into 50 Ω
Output impedance (nom) 50 Ω
Minimum pulse width (nom) 16 ns
External trigger/gate
Connector Rear-panel BNC, chassis-referenced
Function Input or output
Assignment Channel 1, channel 2, or both (as input)
Channel 1 or channel 2 (as output)
Polarity Positive or negative slope
Voltage level (nom) 0 V to 0.4 V for low, > 2.3 V for high, 3.5 V maximum (as input)
3 Vpp (nom) into open circuit, 1.5 Vpp (nom) into 50 Ω (as output)
Impedance (nom) 10k Ω, DC-coupled (as input)
50 Ω (as output)
Minimum pulse width (nom) 16 ns
Input rate DC to 1 MHz
Minimum pulse width 100 ns (as input)
Duty cycle (nom) 50% (as output)
Trigger delay 0 s to 1000 s, 4 ns resolution; applies to all trigger events
Input latency (typ) < 135 ns with Trigger Delay set to zero
Input jitter (typ) < 2.5 ns, rms
Fanout ≤ 4 total Keysight 33500B Series waveform generators

Connector Rear-panel BNC, isolated
Assignment Channel 1, Channel 2, or both
Voltage level ±5 V full-scale
Input impedance (nom) 5k Ω
Bandwidth (-3 dB, typ) 0 Hz to 100 kHz
Frequency reference input
Connector Rear-panel BNC, isolated from chassis and all other connectors
Reference selection Internal, external, or auto
Frequency range Standard: 10 MHz ± 20 Hz
Option OCX: 10 MHz ± 1 Hz
Lock time (typ) <2 s
Voltage level 200 mVpp to 5 Vpp
Input Impedance (nom) 1k Ω || 20 pF, AC-coupled
20

Frequency reference output
Connector Rear-panel BNC, chasis-referenced
Frequency (nom) 10 MHz
Output impedance (nom) 50 Ω, AC-coupled
Level (nom) 0 dBm, 632 mVpp into 50 Ω

Set and read Year, month, day, hour, minute, second
Battery CR-2032 coin-type, replacable, >5-year life (typ)
Programming times (meas.)

LAN (socket) LAN (VXI-11) USB 2.0 GPIB
Change function 5 ms 6 ms 5 ms 5 ms
Change frequency 2 ms 3 ms 2 ms 3 ms
Change amplitude 20 ms 20 ms 19 ms 22 ms
Select user arb (16 k) 9 ms 11 ms 9 ms 9 ms
Arbitrary waveform download speed to volatile
(binary transfer) LAN (socket) LAN (VXI-11) USB 2.0 GPIB
4 k sample  6 ms 18 ms 8 ms 39 ms
1 M sample  1.3 s 2.6 s 13 s 9.1 s
Memory
Arbitrary waveform and instrument state memory
Volatile 1x106 samples per channel or 16x106 samples per channel (Option MEM)
512 sequence steps per channel
Non-volatile File sytem file space is limited to 64 MB
(~32 MSa of arbitrary waveform records)
Instrument state
Store / Recall User defined instrument states
Power Off Power Off state automatically saved
Power On Factory default settings or last power off settings
USB File System
Front-panel port USB 2.0 high-speed mass storage (MSC) class device
Capability Read or write instrument configuration settings, instrument states
and user arbitrary waveform and sequence files.
Speed 10 MB/s (nom)
21

General characteristics
Computer interfaces
LXI- C (rev1.3) 10/100Base-T Ethernet (Sockets & VXI-11 protocol)
USB2.0 (USB-TMC488 protocol)
GPIB/IEEE-488.1, IEEE-488.2
Web user interface Remote operation and monitoring
Programming language SCPI-1999, IEEE-488.2
Keysight 33210A / 33220A compatibile
Graphical display 4.3" Color TFT WQVGA (480x272) with LED backlight
Mechanical
Size 261.1mm W x 103.8mm H x 303.2mm D (with bumpers installed)
212.8mm W x 88.3mm H x 272.3mm D (with bumpers removed)
2U x ½ rack width
Weight (nom) 3.3 kg (7.2 lbs)
Environmental
Storage temperature -40˚C to 70˚C
Warm-up time 1 hour
Operating environment EN61010, polution degree 2; indoor locations
Operating temperature 0˚C to 55˚C
Operating humidity 5% to 80% RH, non-condensing
Operating altitude up to 3000 meters
Regulatory
Safety Complies with European Low Voltage Directive and carries the CE-marking.
Conforms to UL 61010-1, CSA C22.2 61010-1, and IEC 61010-1:2001
EMC Complies with European EMC Directive for test and measurement products.
- IEC/EN 61326-1
- CISPR Pub 11 Group 1, class A
- AS/NZS CISPR 11
- ICES/NMB-001
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada
Acoustic Noise (nom) SPL 35 dB(A)
Line power
Voltage 100 V - 240 V 50/60 Hz -5%, +10%
100 V - 120 V 400 Hz ±10%
Power consumption (typ) < 45 W, < 130 VA
Warranty 1 year standard, 3 years optional
22

1. Add 1/10th of the output amplitude and offset accuracy specification per °C for operation
at temperatures beyond 23°C ± 5°C.
2. Auto range ON.
3. DC Offset set to zero.
4. limited to arbitrary waveforms that are < 1 million points; phase resolution limited by
number of points in arbitrary waveforms < 3,600 points.
5. Output noise is typically 20 dB lower when (DC + Peak AC) < 320 mV (into 50 Ω) or 640
mV
(into open circuit).
6. Subject to maximum output voltage limits.
7. All frequency changes are phase-continuous.
8. Subject to pulse width limits.
9. Counted Burst operation is not allowed for Gaussian Noise.
10. External trigger only for sweep time > 8000 sec.

The warranted performance of a calibrated instrument that has been stored for a
minimum of 2 hours within the operating temperature range of 0º C – 55º C and
after a 45-minute warm up period. All specifications include measurement uncer-
tainty and were created in compliance with ISO-17025 methods.
Data published in this document are specifications (spec) only where specifically
indicated.
Typical (typ)
The characteristic performance, which 80% or more of manufactured instruments
will meet. This data is not warranted, does not include measurement uncertainty,
and is valid only at room temperature (approximately 23ºC).
Nominal (nom)
The mean or average characteristic performance, or the value of an attribute that
is determined by design such as a connector type, physical dimension, or operating
speed.
This data is not warranted and is measured at room temperature (approximately
23ºC).

An attribute measured during development for purposes of communicating the
expected performance. This data is not warranted and is measured at room tem-
perature (approximately 23ºC).
Accuracy
Represents the traceable accuracy of a specified parameter. Includes measurement
error and timebase error, and calibration source uncertainty.
Random measuremet errors are combined using the root-sum-square method and
are multiplied by M for the desired Confidence Level. Syatematic errors are added
linearly and include time skew errors, trigger timing errors, and timebase errors as
appropriate for each measurement type.

For 99% Confidence use k= 2.5 in accuracy calculations.
For 95% Confidence use k= 2.0 in accuracy calculations.
myKeysight
www.keysight.com/find/mykeysight

www.axiestandard.org
®


ATCA®®

www.lxistandard.org


consortium.
www.pxisa.org


Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty



Keysight Assurance Plans
www.keysight.com/find/AssurancePlans

instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.

Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners


For more information on Keysight

services, please contact your local Keysight


Americas
 (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
 (800) 829 4444

Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100

Austria 0800 001122
Belgium 0800 58580
 0800 523252
France 0805 980333
Germany 0800 6270999
 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
 0800 0233200
Russia 8800 5009286
Spain 800 000154
 0200 882255
 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
 0800 0260637


(BP-09-23-14)
23 | Keysight | 33500B Series Waveform Generators - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, 2012 - 2014

5991-0692EN
www.keysight.com