GPFM1200 Medical
Single Output Power Factor Corrected
• Eliminates isolation transformer requirement
• Ideal for distributed power applications
• Allows the use of commercial o the shelf dc/dc
converts in medical system
• Active Soft-Start, 0.99 PF minimizes ac line disturbance
• Overtemperature Shutdown with advance warning
• External Inhibit, Enable, Power Good
• Auxiliary Voltage for peripheral logic
• Meets all medical EMC requirements
• 2 year warranty
1200 WATT GLOBAL PERFORMANCE SWITCHERS
All specications are typical at nominal input, full load at 25OC unless otherwise stated
FEATURES:
DC Power Supplies
• marked to LVD
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
Conducted RF Susceptibility
Radiated RF Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Voltage Sags & Surges
EN55011, Class B; FCC Part 15, Class B
EN61000-4-2, Level 3
EN61000-4-6, Level 3
EN61000-4-3, Level 3
EN61000-4-4, Level 3
EN61000-4-5, Level 3
EN61000-4-11, Level 4
Ac Input
85 - 264 Vac, 47 to 63 Hz.
Input Current
17 A line current maximum, at 90 Vac, 60 Hz with full rated load, power factor
.99 typical, .96 minimum. Input current harmonic content meets the require-
ments of IEC 1000-3-2 A14 for all load conditions.
Input Protection
Internal ac fuse provided on all units blows only on unit failure. (Additional
external fusing required for some medical applications.)
Turn-on
1.0 second maximum; overshoot < 2% at turn-on; < 1% at turn-o.
Hold-Up Time
21 ms minimum from loss of ac input at full load to the output decreasing
4%. Ac fail warning signal occurs at least 5 ms before the output decreases
4%.
Inrush Current
Inrush at 240 Vac is less than 37 A, averaged over the rst ac half-cycle under
cold start conditions. Limiting provided by internal thermistors.
Temperature Coecient
0.03%/oC, typical.
Transient Response
300 μs max. response time for return to within 0.5% of nal value for a 50%
load change within the load range of 25% to 100%, with ∆i/∆t < 0.2 A/μs.
Maximum voltage deviation is 0.3%.
Overload Protection
Fully protected against short circuit and output overload. Power limiting is
cycling type with automatic recovery.
Reverse Voltage
Outputs protected against momentary reverse current not to exceed 20 A
peak for 10 ms max. with 0.5 A avg. Sustained reverse current at high levels
may damage unit.
Eciency
Minimum value at full rated load, nominal input voltage. 48 V units 80%.
Medical Leakage Current
The maximum leakage current is as follows;
Conditions Normal Single Fault
132 Vac, 60 Hz 90 μA 80 μA
264 Vac, 60 Hz 200 μA 400 μA
EMI / EMC Compliance
All models include built-in EMI ltering to meet the EMC requirements of
IEC60601-1-2.
Safety
All medical models comply with the latest editions of UL2601-1, CSA-C22.2 No.
601.1, EN60601-1. CB certicate to EN60601 standard is available.
Signal Pin Description
+12 V
Fan Fail
Over Temp
AC Fail
Inhibit
RTN
Master Enable
1
2
3
4
5
6
7
Auxiliary +12 V output, 0.3 A load maximum
Output, indicating that the fan has failed
Output, indicating excessive internal temperature
Output, indicating failure of the input AC source
Input, to shut down the power supply from an external
command
Common return for all the logic signals
Input, to control the power supply
SPECIFICATIONS
Condor DC Power Supplies Inc. 2311 Statham Parkway, Oxnard CA, 93033, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 Email: condor@condordc.com Rev. 10/05.
Data Sheet © 2005 Condor DC Power Supplies Inc. The information and specications contained in this data sheet are believed to be correct at time of publication.
However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specications are subject to change without notice.
ENVIRONMENTAL SPECIFICATIONS OPERATING NON-OPERATING
Temperature (A) 0 to 5 -40 to +85°C
Humidity (A) 0 to 95% RH 0 to 95% RH
Shock (B) 20 gpk 40 gpk
Altitude -500 to 10,000 ft -500 to 40,000 ft
Vibration (C) 1.5 g rms’ 0.003 g2/Hz 5 g rms’ 0.026 g2/Hz
GPFM1200 MECHANICAL SPECIFICATIONS
A. Units should be allowed to warm up/operate under non-condensing
conditions before application of power. Cooling provided by internal
fan— heatsink temperatures should not exceed 90oC for extended
periods in the installation. Derated maximum output power by 2.5% per
degree C above 50oC.
B. Random vibration—10 to 2000Hz, 6dB/octave roll-o from 350 to
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1
hr./axis non-operating.
C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3
orthogonal axes, total 6 shocks.
Medical Model Output Voltage (1) Output Current (2) Total Regulation (3) Ripple and Noise (4) OVP Set Point
GPFM1200-48 48 V 25 A 2% 480 mV p-p 51.0-59.0 V
Notes:
1.Output voltage is manually adjustable +5%.
2. Maximum continuous current rating is with integral fan.
3. Combined initial voltage set point, line, load and temperature regulation.
4.Dierential mode noise measured with scope probe (20 MHz bandwidth) directly across output connector of the
power supply with load terminated with 0.1 μF capacitor.
5.Output is oating (500 Vac or +700 Vdc); either +Vout or Vout Return may be connected to chassis ground.
Dimensions: Inches