GPFM1200 Medical
Single Output Power Factor Corrected
• Eliminates isolation transformer requirement
• Ideal for distributed power applications
• Allows the use of commercial “o the shelf” dc/dc
converts in medical system
• Active Soft-Start, 0.99 PF minimizes ac line disturbance
• Overtemperature Shutdown with advance warning
• External Inhibit, Enable, Power Good
• Auxiliary Voltage for peripheral logic
• Meets all medical EMC requirements
• 2 year warranty
1200 WATT GLOBAL PERFORMANCE SWITCHERS
All specications are typical at nominal input, full load at 25OC unless otherwise stated
FEATURES:
DC Power Supplies
• marked to LVD
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
Conducted RF Susceptibility
Radiated RF Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Voltage Sags & Surges
EN55011, Class B; FCC Part 15, Class B
EN61000-4-2, Level 3
EN61000-4-6, Level 3
EN61000-4-3, Level 3
EN61000-4-4, Level 3
EN61000-4-5, Level 3
EN61000-4-11, Level 4
Ac Input
85 - 264 Vac, 47 to 63 Hz.
Input Current
17 A line current maximum, at 90 Vac, 60 Hz with full rated load, power factor
.99 typical, .96 minimum. Input current harmonic content meets the require-
ments of IEC 1000-3-2 A14 for all load conditions.
Input Protection
Internal ac fuse provided on all units blows only on unit failure. (Additional
external fusing required for some medical applications.)
Turn-on
1.0 second maximum; overshoot < 2% at turn-on; < 1% at turn-o.
Hold-Up Time
21 ms minimum from loss of ac input at full load to the output decreasing
4%. Ac fail warning signal occurs at least 5 ms before the output decreases
4%.
Inrush Current
Inrush at 240 Vac is less than 37 A, averaged over the rst ac half-cycle under
cold start conditions. Limiting provided by internal thermistors.
Temperature Coecient
0.03%/oC, typical.
Transient Response
300 μs max. response time for return to within 0.5% of nal value for a 50%
load change within the load range of 25% to 100%, with ∆i/∆t < 0.2 A/μs.
Maximum voltage deviation is 0.3%.
Overload Protection
Fully protected against short circuit and output overload. Power limiting is
cycling type with automatic recovery.
Reverse Voltage
Outputs protected against momentary reverse current not to exceed 20 A
peak for 10 ms max. with 0.5 A avg. Sustained reverse current at high levels
may damage unit.
Eciency
Minimum value at full rated load, nominal input voltage. 48 V units 80%.
Medical Leakage Current
The maximum leakage current is as follows;
Conditions Normal Single Fault
132 Vac, 60 Hz 90 μA 80 μA
264 Vac, 60 Hz 200 μA 400 μA
EMI / EMC Compliance
All models include built-in EMI ltering to meet the EMC requirements of
IEC60601-1-2.
Safety
All medical models comply with the latest editions of UL2601-1, CSA-C22.2 No.
601.1, EN60601-1. CB certicate to EN60601 standard is available.
Signal Pin Description
+12 V
Fan Fail
Over Temp
AC Fail
Inhibit
RTN
Master Enable
1
2
3
4
5
6
7
Auxiliary +12 V output, 0.3 A load maximum
Output, indicating that the fan has failed
Output, indicating excessive internal temperature
Output, indicating failure of the input AC source
Input, to shut down the power supply from an external
command
Common return for all the logic signals
Input, to control the power supply
SPECIFICATIONS