DATA SHEET MBC150 SERIES MEDICAL 150W AC/DC FEATURES 110 W convection cooled -20 to 50 deg C full load operation 2 x MOPP Isolation 90-264 VAC input 2" x 4" x 1.3" (50.8 x 101.6 x 33.6 mm) Fan output, 12 Vdc @ 0.5A standard No minimum load required Conducted EMI EN 55022-B, FCC Part 15 Level B Medical Safety Agency Approvals APPLICATIONS Dialysis Surgical Devices Monitoring Home Health Pumps Portable Devices BCD.00288_AA_December-3-2013 1 www.power-one.com DATA SHEET TECHNICAL DATA: Input PARAMETER DESCRIPTION/CONDITION Input Voltage Range Universal Input Input Frequency Range Input Surge Current Safety Ground Leakage Current 47-63 Hz 230 Vac (cold start) 264 Vac 50 / 60 Hz 120 Vac @ Full load 230 Vac @ Full load Input Current 90 - 264 Vac 120 - 390 Vdc 65 A max. < 250 A max 1.7 A 0.85 A Output PARAMETER DESCRIPTION/CONDITION Voltage Adjustment V1 3% Transient Response Main output 50 to 100% load change, 50 Hz, 50% duty cycle, 0.1A / uSec < 10%, recovery time < 5 mSec Over Voltage Protection Over Current Protection Short Circuit Protection Efficiency Rise Time Set Point Accuracy (Main Output) V1 Rated output current Automatic recovery >86% < 100 mSec 1% 110 to 150% rated max 110% Typical Ordering Information PRODUCT FAMILY MBC150-1005G MBC150-1005G-2 MBC150-1T05G MBC150-1T05G-2 MBC150-1012G MBC150-1012G-2 MBC150-1T12G MBC150-1T12G-2 MBC150-1015G MBC150-1015G-2 MBC150-1T15G MBC150-1T15G-2 MBC150-1024G MBC150-1024G-2 MBC150-1T24G MBC150-1T24G-2 MBC150-1048G MBC150-1048G-2 MBC150-1T48G MBC150-1T48G-2 Vfan (all models) MBC150 Series VOLTS MAX LOAD MAX LOAD (VDC) CONVECTION (1) 300 LFM (1) 5.0 5.0 5.0 5.0 12 12 12 12 15 15 15 15 24 24 24 24 48 48 48 48 12 16.0 A 16.0 A 16.0 A 16.0 A 8.33 A 8.33 A 8.33 A 8.33 A 6.67 A 6.67 A 6.67 A 6.67 A 4.17 A 4.17 A 4.17 A 4.17 A 2.08 A 2.08 A 2.08 A 2.08 A 0.5 A (3) MINIMUM LOAD (A) RIPPLE & NOISE (4) CONNECTOR 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% 1% JST JST Screw Terminal Screw Terminal JST JST Screw Terminal Screw Terminal JST JST Screw Terminal Screw Terminal JST JST Screw Terminal Screw Terminal JST JST Screw Terminal Screw Terminal 16 A 16 A 20 A 20 A 12.5 A 12.5 A 12.5 A 12.5 A 10.0 A 10.0 A 10.0 A 10.0 A 6.25 A 6.25 A 6.25 A 6.25 A 3.13 A 3.13 A 3.13 A 3.13 A 0.5 A (3) 2 TOTAL REGULATION 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 2.5% 20% www.power-one.com DATA SHEET Notes: 1. Combined power from main output and Vfan should not exceed total power rating. 2. 3. 4. Fan output tolerance is 20%. Peak current for fan output is 1A. Ripple is 2% up to 20% load and less than 1% above 20% load. Output noise measurement is made with a 20 MHz 5. 6. bandwidth using a 6" twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor. Class 1 models have Earthing tab J4. Class 2 models (-2 suffix) have no Earthing tab. Specifications are for nominal input voltage, 25 C and max load unless otherwise stated. 7. 8. 9. 10. Air flow over length of supply recommended (either direction) for forced air rating. Derate output power per chart below. Specifications subject to change without notice. Warranty 2 years. General Specifications PARAMETER Hold Up Time MTBF Switching Frequency Isolation Voltage Weight DESCRIPTION/CONDITION 120 Vac 230 Vac >200 khrs PFC converter variable 35 to 250 kHz, 90 kHz typical Input to Output Min 5900 Vdc 150 g (0.33 lbs) 6 mSec 10 mSec Bellcore TR-332 Resonant converter: Variable 35 to 250 kHz, 90 kHz typical Input to Ground 2120 Vdc Environmental PARAMETER Operating Temperature Altitude Humidity Conducted Emissions Radiated Emissions Electromagnetic Susceptibility Harmonic Current DESCRIPTION/CONDITION Operating Storage Operating 10,000 ft. 95% EN55022, FCC part 15 Level B EN55022, FCC part 15 Level B EN61000-4 EN61000-3-2, Class D -20 to +70 C. See derating charts below. -40 to +85 C Non-operation 40,000 ft. Non-condensing To be controlled in end system 2, 3, 4, 5 Level 3 Safety PARAMETER DESCRIPTION/CONDITION EN / UL / CSA 60601-1 3rd Edition BCD.00288_AA_December-3-2013 3 www.power-one.com DATA SHEET Figure 1 Output Power Vs. Temperature Figure 2 Dimension Drawing (Top and Side View) Mechanical INPUT = J1 Pin 1: AC Line Pin 2: Removed Pin 3: AC Neutral Mating Connector: Molex: 09-50-3031 Pins: 08-50-0106 EARTHING TAB = J4 DC OUTPUT = J2 Molex: 19705-4301 Pin 1 = V1 Pin 2 = V1 Pin 3 = RTN Pin 4 = RTN Mating Connector: Molex: 190030001 Mating Connector: JST VHR-4M, Pins: SVH-41T-P1.1 AWG #20 to #16 FAN= J3 Pin 1 = +12 V @ 0.5 A Pin 2 = Fan return (isolated from DC output) Mating Connector: Tyco: 640440-2 Copyright (c) 2010 Power-One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherwise, the trademarks and service marks of Power-One Inc. in the U.S. and other countries. All other product or service names are the property of their respective holders. Power-One products are protected under numerous U.S. and foreign patents and pending applications, maskwork rights, and copyrights. Power-One reserves the right to make changes to any products and services at any time without notice. Power-One assumes no responsibility or liability arising out of the application or use of any information, product, or service described herein except as expressly agreed to in writing by Power-One Inc. NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not designed, intended for use in, or authorized for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president of Power-One, Inc. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice. MBC150 Series 4 www.power-one.com