V10 TR100M SERIES 100W MEDICAL SWITCHING ADAPTER Features * Universal Input Range 90~264VAC * Meets EN60601-1 and EN55011 Class B * Continuous Short Circuit Protection * Over Voltage Protection * Efficiency & Standby Power Meet Level V (TR100M120~150: Output Cable Length1500mm 14AWG /UL1185) (TR100M180~480: Output Cable Length1800mm 16AWG /UL1185) * Meets 2 MOPP Ordering information TR100MXXX- XX Model No. DC Plug Type MODEL TR100M120 X OVP E: with OVP XX or XXX DC Cable Length and Type 11: 720mm with Ferrite Core 12: 1220mm with Ferrite Core 36: 1500mm with Ferrite Core 13: 1800mm with Ferrite Core OUTPUT VOLTAGE OUTPUT CURRENT RIPPLE & NOISE VOLTAGE ACCURACY LINE REGULATION LOAD REGULATION % EFF. (Typ.) NOTE 2 NOTE 1 NOTE 3 NOTE 4 NOTE 5 12 V 8.34 A 1% 2% 1% 4% 88% 2% 1% 3% 88% TR100M150 15 V 6.67 A 1% TR100M180 18 V 5.56 A 1% 2% 1% 2% 88% TR100M190 19 V 5.27 A 1% 2% 1% 2% 88% TR100M200 20 V 5.0 A 1% 2% 1% 2% 88% TR100M240 24 V 4.17 A 1% 2% 1% 2% 89% TR100M480 48 V 2.1 A 1% 2% 1% 2% 89% Specifications are subject to change without notice. 1 V10 Specifications TR100M Series Derating Curve INPUT SPECIFICATIONS: Voltage .............................................................................. 90~264Vac Frequency ...................................................................... 47 to 63Hz Inrush Current ........................... Cold Start @25 100A max. @240Vac Conducted EMI .................................................. CISPR/FCC Class B Isolation ...................................................... Input to output = 4k Vac Leakage Current ............................................................ 0.1mA max. OUTPUT SPECIFICATIONS: Hold-up Time ...................................................... 16ms typ. @115Vac Short Circuit Protection ............................... Hiccup Mode (Auto Recover) Over Voltage Protection ................................. TVS Component to Clamp Temperature Coefficient .......................................................... 0.05%/ GENERAL SPECIFICATIONS: SAFETY AND EMC: Operating Temperature ............................... -20 ~ 60 (see derating curve) Emission and lmmunity ............ EN55011, EN60601-1-2, EN61000-3-2 Storage Temperature .........................................................-20 ~ 85 Humidity .................................................. 93% RH max. Non condensing Cooling ................................................................ Natural Convection EN61000-3-3 rd. Safety (Medical 3 )............................................. IEC60601-1, EN60601-1 UL ANSI/AAMI ES60601-1 Switching Frequency .................................................... 70KHz Typical NOTE: MTBF ......... MIL-HDBK-217F, GB, at 25/115VAC .......... 150Khrs min. Altitude ............................................................................... 3000m Dimensions .............. 5.591x2.283x1.457 inches (142.00x58.00x37.00mm) Weight ........................ ............................................................485g AC Inlet ..................................... ........................ ............IEC320/C8 1. Voltage setpoint at 60% full load. 2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for Ripple & Noise measuring @20MHz BW. 3. Line regulation measured from 100Vac to 240Vac, full load. 4. Load regulation measured from 60% to 100% full load and from 60% to 20% load (60% +/- 40% full load) 5. Typical efficiency with 230 VAC and max. load at 25C. Mechanical Specification Typical at 25C, nominal line and 75% load, unless otherwise Specified 2 Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: Cincon: TR100M120-01E02-LVL-V TR100M120-02E02-LVL-V TR100M120-11E02-LVL-V TR100M120-12E02-LVL-V TR100M150-01E02-LVL-V TR100M150-02E02-LVL-V TR100M150-11E02-LVL-V TR100M150-12E02-LVL-V TR100M180-01E03-LVL-V TR100M180-02E03-LVL-V TR100M180-11E03-LVL-V TR100M180-12E03-LVL-V TR100M190-01E03-LVL-V TR100M190-02E03-LVL-V TR100M190-11E03-LVL-V TR100M190-12E03-LVL-V TR100M200-01E03-LVL-V TR100M200-02E03-LVL-V TR100M200-11E03-LVL-V TR100M200-12E03-LVL-V TR100M240-01E03-LVL-V TR100M240-02E03-LVL-V TR100M240-11E03-LVL-V TR100M240-12E03-LVL-V TR100M480-01E03-LVL-V TR100M480-02E03-LVL-V TR100M480-11E03-LVL-V TR100M480-12E03-LVL-V TR100M240-11E13-LVL-V TR100M480-01E13-LVL-V TR100M190-02E13-LVL-V TR100M120-01E12-LVL-V TR100M240-12E13-LVL-V TR100M150-02E12-LVL-V TR100M150-01E12-LVL-V TR100M120-02E12-LVL-V TR100M190-01E13-LVL-V TR100M190-11E13-LVL-V TR100M480-11E13-LVL-V TR100M180-02E13-LVL-V TR100M180-11E13-LVL-V TR100M240-02E13-LVL-V TR100M150-12E12-LVL-V TR100M120-11E12-LVL-V TR100M120-12E12-LVL-V TR100M200-12E13-LVL-V TR100M480-12E13-LVL-V TR100M200-11E13-LVL-V TR100M480-02E13-LVL-V TR100M240-01E13-LVL-V TR100M180-12E13-LVL-V TR100M200-01E13-LVL-V TR100M190-12E13-LVL-V TR100M200-02E13-LVL-V TR100M180-01E13-LVL-V TR100M150-11E12-LVL-V