ME10
10W-12W Single Output External Power
Medical Series
www.slpower.com ME10 Rev 2.7 02-MAY-16 1
Features
Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*
Approved to EN/IEC/UL60601-1, 3rd edition
applications with and isolation levels which
satisfy the 2 MOPP requirements
Meets DoE Efficiency Level VI Requirements
No load input power
Average Efficiency
Up to 12W of AC-DC Power
Universal Input 90-264Vac Input Range
Desktop and Wall-Plug versions
Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin
E-cap life of >7 years
<1,000,000 Hours MTBF
3 Year Warranty
IP22 Rated Enclosure
Description
A high performance AC to DC external power supply family designed for medical applications. The ME10 Medical Series
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition, with isolation levels which satisfy the 2
MOPP requirements; and designed to UL/EN/IEC60601-1-2, 4th edition for EMC*. The ME10 Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up
to +40°C and applicable output power derating at 50°C. These models are available in desktop and wall-plug versions, include
an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Output
Output
Ripple &
Line
Load
Output
Input
Number Volts Current Power Noise
¹
Regulation Regulation
Connector
ME10A0503F01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
ME10A0603F01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703F01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903F01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203F01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503F01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403F01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
ME10A0503N01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
ME10A0603N01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703N01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903N01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203N01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503N01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403N01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
ME10A0503Q01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
ME10A0603Q01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703Q01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903Q01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203Q01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503Q01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403Q01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
LPS
ME10
10W-12W Single Output External Power
Medical Series
www.slpower.com ME10 Rev 2.7 02-MAY-16 2
Model Selection (continued)
Model
Output
Output
Ripple &
Line
Load
Output
Input
Number Volts Current Power Noise
¹
Regulation Regulation
Connector
ME10A0503B01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades (North American
Blade included)²
ME10A0603B01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703B01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903B01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203B01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503B01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403B01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
ME10A0503C01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North American
Blades³
ME10A0603C01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703C01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903C01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203C01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503C01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403C01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and 6V models, values listed are
typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input 100-240Vac, ±10%, 47-63Hz, 1Turn On Time Less than 700mS @115Vac, full load
Input Current 115Vac: 0.45A, 230Vac: 0.22A Hold-up Time 20mS min., at full Load, 100Vac input
Inrush Current 264Vac, cold start: will not exceed 40A Overtemperature
Protection Will shutdown upon an overtemperature
condition, auto-recovery.
Input Fuses F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models Overload Protection 130 to 180% of rating, Hiccup Mode
Earth Leakage
Current Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC Short Circuit
Protection Hiccup Mode, auto recovery.
Efficiency >87%, typical Overvoltage
Protection 130 to 150% of output voltage, hiccup mode
Output Power 10 to 12W continuous See models chart for
specific voltage model ratings. Isolation Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
No Load Input
Power <0.1W per DoE Efficiency Level VI
Requirements Safety Standards EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
Ripple and Noise See models chart Operating
Temperature -20°C to +70°C
Output Voltage See models chart Temperature
Derating See derating chart
Transient
Response
500
µ
s response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, i/t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Storage Temperature -40°C to +85°C
Regulation See models chart Altitude Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Drop Test 1.4m from table top to wooden platform, 6
faces. Relative Humidity 5% to 95%, non-condensing
Vibration
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
E-Cap Life >10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day. MTBF >1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.
ME10
10W-12W Single Output External Power
Medical Series
www.slpower.com ME10 Rev 2.7 02-MAY-16 3
EMI/EMC Compliance
Conducted Emissions: EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions: EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise: High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports: EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Electrical Fast Transients (EFT) /Bursts: EN55024/IEC61000-4-4, Level 4, +/- 4.4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode) EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
Conducted Disturbances induced by RF
Fields
EN55022/IEC61000-4-6, 3.6V/m Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5
Rated Power frequency magnetic fields EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions EN55011/EN61000-3-2, Class A
Flicker Test EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Mechanical Drawing
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
Notes: 1. All dimensions in mm.
2. Weight = 110g
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
3. The unit should not be covered or enclosed to protect against excessive case temperature rise.
ME10
10W-12W Single Output External Power
Medical Series
www.slpower.com ME10 Rev 2.7 02-MAY-16 4
Derating Chart:
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
0
2
4
6
8
10
12
14
-20 -10 0 10 20 30 40 50 60 70
Watts
Ambient Temperature C
5V
7V
12V
24V