ME10
10W-12W Single Output External Power
Medical Series
www.slpower.com ME10 Rev 2.7 02-MAY-16 2
Model Selection (continued)
Number Volts Current Power Noise
Regulation Regulation
ME10A0503B01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades (North American
Blade included)²
ME10A0603B01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703B01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903B01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203B01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503B01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403B01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
ME10A0503C01 5.0V 2.0A 10W 75mV pk-pk ±1% ±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North American
Blades³
ME10A0603C01 5.9V 1.6A 10W 75mV pk-pk ±1% ±5%
ME10A0703C01 7.5V 1.3A 10W 75mV pk-pk ±1% ±5%
ME10A0903C01 9.0V 1.1A 10W 90mV pk-pk ±1% ±5%
ME10A1203C01 12.0V 1.0A 12W 120mV pk-pk ±1% ±5%
ME10A1503C01 15.0V 0.8A 12W 150mV pk-pk ±1% ±5%
ME10A2403C01 24.0V 0.5A 12W 240mV pk-pk ±1% ±5%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and 6V models, values listed are
typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅Turn On Time Less than 700mS @115Vac, full load
Input Current 115Vac: 0.45A, 230Vac: 0.22A Hold-up Time 20mS min., at full Load, 100Vac input
Inrush Current 264Vac, cold start: will not exceed 40A Overtemperature
Protection Will shutdown upon an overtemperature
condition, auto-recovery.
Input Fuses F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models Overload Protection 130 to 180% of rating, Hiccup Mode
Earth Leakage
Current Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC Short Circuit
Protection Hiccup Mode, auto recovery.
Efficiency >87%, typical Overvoltage
Protection 130 to 150% of output voltage, hiccup mode
Output Power 10 to 12W continuous – See models chart for
specific voltage model ratings. Isolation Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
No Load Input
Power <0.1W per DoE Efficiency Level VI
Requirements Safety Standards EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
Ripple and Noise See models chart Operating
Temperature -20°C to +70°C
Output Voltage See models chart Temperature
Derating See derating chart
Transient
Response
500
s response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Storage Temperature -40°C to +85°C
Regulation See models chart Altitude Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Drop Test 1.4m from table top to wooden platform, 6
faces. Relative Humidity 5% to 95%, non-condensing
Vibration
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
E-Cap Life >10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day. MTBF >1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.