Technical Data
The Fluke Calibration 525B
Temperature/ Pressure Calibrator
gives you a workhorse combina-
tion of high accuracy and broad
functionality for temperature and
pressure instrument calibration.
Compact and economical, the
525B has an interface for auto-
mated calibration, providing wide
workload coverage in instrument
shops and calibration labs, as
well as in ATE applications.
The 525B is the most
accurate Fluke Calibration
temperature calibrator, sourc-
ing and measuring a complete
range of RTDs, thermocouples,
and the YSI 400 thermistor. It
also measures pressure cover-
ing common ranges from 1 inch
(6900 Pa) of water up to 10,000
PSI (69 MPa) using the Fluke 700
Series and 525A-P Series Pressure
Modules. Plus, the dc voltage and
current specifications of the 525B
enable you to calibrate other
process calibrators and a wide
variety of other instruments with
accuracy that rivals any calibrator
in its price range.
Key features:
Simulates and measures all
ANSI thermocouples, as well
as L and U types, and provides
cold junction compensation to
enable calibration of a wide
variety of thermocouple instru-
mentation.
Direct input for storage of
ITS-90 RTD constants.
RTD source uncertainties to
0.03 °C.
Direct measurement of all
Fluke 700 Series and 525A-P
Series Pressure Modules cover-
ing the most common pressure
ranges from 1˝ H20
to 10,000 PSI.
Converts easily to any pres-
sure unit directly from the
front panel or through remote
communications.
Sources dc voltage and current
for multifunction workload
coverage, enabling calibration
of data loggers, strip chart
recorders, multimeters, and
other industrial instruments.
Supports automation using Fluke
Calibration’s MET/CAL® Plus
Calibration Management
Software or custom automated
programs with standard
RS-232-C and IEEE-488
interface.
Eight user programmable
setpoints allow quick recall
of values for Zero, Span and
linearity checks during
calibration.
Programmable setpoint dwell
times for automated calibration
and troubleshooting.
Selectable internal or external
CJC (Cold junction compen-
sation) provides you with
the ability to make exacting
thermocouple measurements
using remote junctions.
Easy five-way binding posts for
low-loss convenient hookup.
Banana plug, screw terminal,
spade lug,
wraparound terminal, wire
pass-through terminal.
NIST-traceable calibration with
data is included.
At home in the cal lab
The 525B is a compact bench
instrument well suited for
traditional calibration labora-
tories with a large temperature
workload. Its 0 to 100 V output
and 0 to 100 mA current capa-
bilities—plus its ability to serve
as a highly-accurate pressure
standard—make it a versatile
performer.
Filling the gap in the
instrument shop
In process plant environments,
instrument shop technicians
responsible for maintaining the
equipment pool need to calibrate
and repair a wide range of instru-
ments, transmitters, field calibra-
tors, P/l converters, transducers,
and temperature probes. With
its ability to simulate almost any
temperature sensor, precise pres-
sure monitoring capability, and
dc voltage and current accuracy
rivaling any calibrator in its price
range, the 525B fills the gap
between lower accuracy hand-
held field calibrators and more
expensive high-end multifunction
calibrators.
525B Temperature/
Pressure Calibrator
Superior accuracy and functionality in
an economical benchtop package
2 Fluke Calibration 525B Temperature/Pressure Calibrator
A good fit for ATE test
applications
With precise calibration-grade
instrument control enabled
via an IEEE bus, the 525B is a
hardworking addition to auto-
mated test systems in quality and
manufacturing test applications
requiring capabilities such as
voltage and current sourcing,
sensor simulation, and tempera-
ture monitoring and validation.
Powerful functionality
that is easy to use
From start to finish, the 525B
is designed for ease of opera-
tion. The intuitive front panel
design features large keypads
and display that help reduce
training time and make the 525B
comfortable to use even for long
periods of time. Plus, you can
store frequently used constants
for a variety of probes in memory
for faster setup on the job.
MET/CAL® Plus automates
525B calibration and docu-
mentation management
You can use the 525B with
MET/CAL Plus calibration software
(V.6.11 or later) to meet the strin-
gent documentation and reporting
requirements imposed by quality
standards such as ISO 9000.
MET/CAL Plus is a powerful
software environment for creating,
editing, testing, and documenting
calibration procedures, and for
performing automated calibrations.
At the core of MET/CAL Plus
is MET/TRACK® software, a
database management system
for test and measurement assets
created specifically to manage
equipment. It enables you to
track the information you need to
maintain quality calibrations. And
it supports the traceability and
record-keeping requirements of
modern quality and accreditation
standards.
MET/TRACK provides you with a
variety of standard report formats
prepared with Crystal Reports
Professional, which is included
with MET/CAL Plus. These report
formats cover a wide range of
information to help you meet
many of your documentation
requirements, including those
for ISO 9000 and similar quality
standards. They report on forward
and reverse traceability, allowing
you to easily document traceabil-
ity from any asset to the equip-
ment calibrated. It even includes
measurement uncertainty report-
ing to meet the requirements of
ISO/IEC 17025.
Pressure modules
An optional set of external pres-
sure modules provides pressure
measurement capability. The
525B can accept either the Fluke
700 Series or the 525A-P pres-
sure modules. Both module types
plug directly into the calibrators
front panel Lemo connector; the
525B firmware auto-detects the
type and value of the attached
module.
World-class temperature
and pressure calibration
solutions
In addition to the 525B Tempera-
ture/Pressure Calibrator, Fluke
Calibration offers a broad range
of solutions in temperature and
humidity calibration as well as
pressure and flow calibration.
Fluke Calibration makes every-
thing you need for calibrating
temperature sensors including
SPRTs, PRTs, thermistors, and
thermocouples. The constant
temperature baths and dry-block
calibrators offer unmatched stabil-
ity, and the primary temperature
standards are used in national
metrology institutes around the
world. For precision thermometry,
Fluke Calibration thermometer
readouts and probes are excep-
tionally accurate and easy to use,
and the 1620A “DewK” tempera-
ture and humidity data logger
handles your environmental
monitoring needs.
Fluke Calibration also offers a
range of high-end products focus-
ing on pressure and flow calibra-
tors and standards. More than 30
national measurement institutes
rely on Fluke Calibration pressure
standards to establish their pres-
sure references. Fluke Calibration
pressure standards, transfer stan-
dards, pressure controller/calibra-
tors and monitors are recognized
for their advanced features and
the realistic specifications that are
the hallmark of a true metrology
company.
Calibration service, repair,
and support
Fluke Calibration provides exten-
sive calibration support and
service to maximize the value of
your calibration investment. Our
worldwide network of Calibration
Centers has traceability to national
standards. Service and support
are just a telephone call or fax
away. We offer fast, quality repair
and calibration services includ-
ing a module exchange program,
comprehensive training, and full
support in setting up your lab.
General specifications
Warm-up time Twice the time since last warmed up, to a
maximum of 30 minutes
Settling time Less than 5 seconds for all functions and
ranges except as noted
Standard interface Standard, RS-232; IEEE-488 (GPIB)
Temperature
performance
Operating: 0 °C to 50 °C
Calibration (tcal): 15 °C to 35 °C
Storage: -20 °C to 70 °C
Electromagnetic
compatibility
Meets CE requirements, including
susceptibility
Temperature
coefficient
Temperature coefficient for temperatures
outside tcal +5 °C is 10 % of the 90 day speci-
fication (or 1 year, as applicable) per °C
Relative humidity Operating: < 80 % to 30 °C, < 70 % to 40 °C,
< 40 % to 50 °C
Storage: < 95 %, non-condensing
Altitude Operating: 3,050 m (10,000 ft) maximum
Non-operating: 12,200 m (40,000 ft) maxi-
mum
Safety EN 61010 Second, ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low
isolation
20 V
EMC Designed to comply with IEC 61326-1/1997
(EMC)
Line power Line voltage (selectable): 100 V/120 V or
220 V/240 V
Line frequency: 47 Hz to 63 Hz
Line voltage variation: ± 10 % about line
voltage setting
Power
consumption
15 VA maximum
Dimensions Height: 13.3 cm (5.25 in) plus 1.5 cm (0.6 in)
four feet on bottom
Width: 0.75 in standard rack width
Depth: 47.3 cm (18.6 in) overall
Weight (without
options)
4 kg (9 lb)
525B Temperature/Pressure Calibrator Fluke Calibration 3
General Specifications
Warm up time................................................................Twice the time since last warmed up, to a maximum of 30 minutes.
Settling time..................................................................Less than 5 seconds for all functions and ranges except as noted.
Standard interface.......................................................RS-232, IEEE-488 (GPIB)
Temperature performance
Operating ....................................................................0 °C to 50 °C
Calibration (tcal) ........................................................18 °C to 28 °C
Storage ........................................................................-20 °C to 70 °C
Electromagnetic compatibility..................................CE: Conforms to EN61326
Temperature coefficient.............................................Temperature coefficient for temperatures outside tcal ± 5°C is
10 % of the 90 day specification (or 1 year if applicable) per °C
Relative humidity
Operating ....................................................................< 80 % to 30 °C, < 70 % to 40 °C, < 40 % to 50 °C
Storage ........................................................................< 95 % noncondensing
Altitude
Operating ....................................................................3,050 m (10,000 ft) maximum
Nonoperating..............................................................12,200 m (40,000 ft) maximum
Safety...............................................................................EN 61010 Second, ANSI/ISA-S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low isolation....................................................20 V
Line power
Line Voltage (selectable) ..........................................00 V/120 V or 220 V/240 V
Line Frequency ..........................................................47 to 63 Hz
Line Voltage Variation ..............................................± 10 % about line voltage setting
Power consumption.....................................................15 VA maximum
Dimensions
Height..........................................................................13.3 cm (5.25 in) plus 1.5 cm (0.6 in) four feet on bottom
Width.......................................................................... standard rack width
Depth...........................................................................47.3 cm (18.6 in) overall
Weight (without options)..............................................4 kg (9 lb)
Electrical Specifications
DC Voltage Specifications, Output
Absolute Uncertainty,
tcal ± 5 °C
± (ppm of output + µV) Stability
Ranges[1]
90 days 1 year 24 hours, ± 1 °C
± (ppm of output + µV)
Resolution Maximum
Burden[2]
0 to 100.000 mV 25 3303 5 + 2 1µV10 mA
0 to 1.00000 V 25 10 30 10 4 + 10 10 µV10 mA
0 to 10.0000 V 25 10030100 4 + 100 100µV10 mA
0 to 100.000 V 25 1 mV 30 1 mV 5 + 1 mV 1 mV 1 mA
TC Output
-10 to 75.000 mV 25 3303 5 + 21µV10
Notes:
[1] All outputs are positive only.
[2] Remote sensing is not provided. Output resistance is < 1.
Noise
Ranges Bandwidth 0.1 to 10 Hz
± (ppm of output + µV p-p) Bandwidth 10 Hz to 10 kHz
(µV rms)
0 to 100.000 mV 1µ
V6
µV
0 to 1.00000 V 10 µ
V6
0µV
0 to 10.0000 V 100µ
V6
00 µV
0 to 100.000 V 10 ppm+1 mV 20 mV
DC Current Specifications, Output
Absolute Uncertainty
tcal ± 5 °C± (ppm of output + µA)
Ranges[1]
90 days 1 year
Resolution
Maximum
Compliance
Voltage
Maximum
Inductive Load
0 to 100.000 mA 40 1501 1μA12 V 100 mH
Note:
[1] All outputs are positive only.
Noise
Ranges Bandwidth
0.1 to 10 Hz p-p
Bandwidth
10 Hz to 10 kHz rms
0 to 100.000 mA 2000 nA 20 μA
Resistance Specifications, Output
Absolute Uncertainty
tcal ± 5°C, ±Ω
Ranges[1]
90 days 1 year
Resolution Allowable Current[2]
5 to 400.00 Ω0.012 0.015 0.001 Ω1 to 3 mA
5 to 4.0000 kΩ0.25 0.3 0.01 Ω100μA to 1 mA
Notes:
[1] Continuously variable from 0 to 4 kΩ.
[2] For currents lower than shown, the floor adder increases by Floor(new) = Floor(old) x Imin/Iactual.
For example, a 500 μA stimulus measuring 100 Ω has a floor uncertainty of 0.015 Ω x 1 mA/500 μA = 0.03 Ω.
Resistance Specifications, Input
Absolute Uncertainty
tcal ± 5°C, ± (ppm of output + Ω)
Ranges[1]
90 days 1 year
Resolution Stimulus Current
5 to 400.00 Ω20 0.0035 20 0.004 0.001 Ω1mA
5 to 4.00000 kΩ20 0.0035 20 0.004 0.01 Ω0.1 mA
Note:
[1] 4-wire mode.
4 Fluke Calibration 525B Temperature/Pressure Calibrator
Thermocouple Specification, Output and Input
Absolute Uncertainty
tcal ± 5 °C, ± (°C)[1]
Range (°C)
Output/Input
TC Type
Minimum Maximum 90 days 1 year
B
600°C
800°C
1000 °C
1550 °C
800°C
1000 °C
1550 °C
1820 °C
0.42 °C
0.39 °C
0.40 °C
0.44 °C
0.46 °C
0.39 °C
0.40 °C
0.45 °C
C
0°C
150°C
650°C
1000 °C
1800 °C
150°C
650°C
1000 °C
1800 °C
2316 °C
0.25 °C
0.21 °C
0.23 °C
0.38 °C
0.63 °C
0.30 °C
0.26 °C
0.31 °C
0.50 °C
0.84 °C
E
-270°C
-100 °C
-25°C
350°C
650°C
-100 °C
-25°C
350°C
650°C
1820 °C
0.38 °C
0.16 °C
0.14 °C
0.14 °C
0.16 °C
0.50 °C
0.18 °C
0.15 °C
0.16 °C
0.21 °C
J
-270 °C
-100 °C
-25 °C
120°C
1000 °C
-100 °C
-30 °C
150°C
760°C
1200 °C
0.20 °C
0.18 °C
0.14 °C
0.14 °C
0.18 °C
0.27 °C
0.20 °C
0.16 °C
0.17 °C
0.23 °C
K
-270 °C
-100 °C
-25°C
120°C
1000 °C
-100 °C
-25 °C
120°C
1000 °C
1372 °C
0.25 °C
0.19 °C
0.14 °C
0.19 °C
0.30 °C
0.33 °C
0.22 °C
0.16 °C
0.26 °C
0.40 °C
L
-200 °C
-100°C
800°C
-100 °C
800°C
900°C
0.37 °C
0.26 °C
0.17 °C
0.37 °C
0.26 °C
0.17 °C
N
-270 °C
-100 °C
-25 °C
120°C
410°C
-100 °C
-25 °C
120°C
410°C
1300 °C
0.33 °C
0.20 °C
0.16 °C
0.14 °C
0.21 °C
0.40 °C
0.24 °C
0.19 °C
0.18 °C
0.27 °C
R
-50°C
250°C
400°C
1000°C
250°C
400°C
1000°C
1760°C
0.58 °C
0.34 °C
0.31 °C
0.30 °C
0.58 °C
0.35 °C
0.33 °C
0.40 °C
S
0°C
250°C
1000 °C
1400 °C
250°C
1000 °C
1400 °C
1750 °C
0.56 °C
0.36 °C
0.30 °C
0.35 °C
0.56 °C
0.36 °C
0.37 °C
0.46 °C
T
-270°C
-150 °C
0°C
120°C
-150°C
0°C
120°C
400°C
0.51 °C
0.18 °C
0.13 °C
0.12 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
U-200 °C
0°C0°C
600°C0.56 °C
0.27 °C0.56 °C
0.27 °C
mV -10 to 75.000 mV
Note:
[1] Does not include thermocouple wire error.
525B Temperature/Pressure Calibrator Fluke Calibration 5
RTD and Thermistor Specification, Output
Range (°C) Absolute Uncertainty
tcal ± 5 °C, ± (°C)[1]
RTD Type
Minimum Maximum 90 days 1 year
Pt 385, 100
-200°C
-80°C
0°C
100°C
300°C
400°C
630°C
-80 °C
0°C
100°C
300°C
400°C
630°C
800°C
0.03 °C
0.04 °C
0.04 °C
0.03 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.05 °C
0.04 °C
0.04 °C
0.05 °C
0.05 °C
Pt 3926, 100
-200°C
-80°C
0°C
100°C
300°C
400°C
-80°C
0°C
100°C
300°C
400°C
630°C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
Pt 3916, 100
-200°C
-190 °C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-190 °C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.03 °C
0.04 °C
0.04 °C
0.03 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.05 °C
Pt 385, 200
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.31 °C
0.32 °C
0.33 °C
0.33 °C
0.36 °C
0.36 °C
0.42 °C
0.42 °C
0.38 °C
0.38 °C
0.39 °C
0.39 °C
0.43 °C
0.43 °C
0.50 °C
0.50 °C
Pt 385, 500
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.13 °C
0.13 °C
0.13 °C
0.14 °C
0.14 °C
0.15 °C
0.16 °C
0.16 °C
0.15 °C
0.15 °C
0.16 °C
0.17 °C
0.17 °C
0.18 °C
0.19 °C
0.19 °C
Pt 385, 1000
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.06 °C
0.06 °C
0.07 °C
0.07 °C
0.07 °C
0.07 °C
0.08 °C
0.08 °C
0.07 °C
0.08 °C
0.08 °C
0.08 °C
0.09 °C
0.09 °C
0.09 °C
0.09 °C
Ni 120, 120
-80°C
0°C
100°C
0°C
100°C
260°C
0.02 °C
0.02 °C
0.01 °C
0.02 °C
0.02 °C
0.02 °C
Cu 427, 10 [2]-100 °C260 °C0.30 °
C0
.38 °C
YSI 400 15 °C50°C0.005 °C0.007 °C
Notes:
[1] 2-wire output.
[2] Based on MINCO Application Aid No. 18.
6 Fluke Calibration 525B Temperature/Pressure Calibrator
RTD and Thermistor Specification, Input
Range (°C) Absolute Uncertainty
tcal ± 5 °C, ± (°C)[1]
RTD Type
Minimum Maximum 90 days 1 year
Pt 385, 100
-200°C
-80°C
0°C
100°C
300°C
400°C
630°C
-80°C
0°C
100°C
300°C
400°C
630°C
800°C
0.011 °C
0.018 °C
0.018 °C
0.027 °C
0.031 °C
0.042 °C
0.050 °C
0.012 °C
0.020 °C
0.020 °C
0.030 °C
0.035 °C
0.047 °C
0.057 °C
Pt 3926, 100
-200°C
-80°C
0°C
100°C
300°C
400°C
-80°C
0°C
100°C
300°C
400°C
630°C
0.011 °C
0.014 °C
0.018 °C
0.026 °C
0.031 °C
0.041 °C
0.011 °C
0.015 °C
0.019 °C
0.029 °C
0.034 °C
0.046 °C
Pt 3916, 100
-200°C
-190 °C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-190 °C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.006 °C
0.011 °C
0.014 °C
0.018 °C
0.025 °C
0.026 °C
0.031 °C
0.040 °C
0.042 °C
0.006 °C
0.012 °C
0.015 °C
0.019 °C
0.028 °C
0.029 °C
0.034 °C
0.045 °C
0.047 °C
Pt 385, 200
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.008 °C
0.012 °C
0.015 °C
0.020 °C
0.050 °C
0.053 °C
0.070 °C
0.071 °C
0.009 °C
0.013 °C
0.017 °C
0.022 °C
0.053 °C
0.057 °C
0.075 °C
0.076 °C
Pt 385, 500
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.007 °C
0.019 °C
0.023 °C
0.030 °C
0.032 °C
0.037 °C
0.047 °C
0.048 °C
0.008 °C
0.020 °C
0.025 °C
0.033 °C
0.035 °C
0.041 °C
0.052 °C
0.053 °C
Pt 385, 1000
-200°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
-80°C
0°C
100°C
260°C
300°C
400°C
600°C
630°C
0.011 °C
0.014 °C
0.019 °C
0.025 °C
0.027 °C
0.030 °C
0.041 °C
0.042 °C
0.012 °C
0.015 °C
0.020 °C
0.028 °C
0.030 °C
0.034 °C
0.045 °C
0.047 °C
PtNi 385, 120
(Ni120)
-80°C
0°C
100°C
0°C
100°C
260°C
0.009 °C
0.010 °C
0.011 °C
0.010 °C
0.011 °C
0.012 °C
Cu 427, 10 [2]-100 °C260 °C0.067 °C0.069 °C
YSI 400 15 °C50°C0.005 °C0.007 °C
SPRT -200°C660 °C0.05 °
C0
.06 °C
Notes:
[1] 4-wire mode. Uncertainties listed do not include probe uncertainties.
[2] Based on MINCO Application Aid No. 18.
525B Temperature/Pressure Calibrator Fluke Calibration 7
700 Series Pressure Modules Specifications
Model
Range/
Resolution
Range (approx)/
Resolution
Reference1
uncertainty
(23 ± 3 °C) High2 side media Low2 side media Fitting material
Differential Fluke 700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 % Dry Dry 316 SS
Fluke 700P01 10 in. H2O/0.01 2.5 kPa/0.002 0.200 % Dry Dry 316 SS
Fluke 700P02 1 psi/0.0001 6900 Pa/0.7 0.150 % Dry Dry 316 SS
Fluke 700P22 1 psi/0.0001 6900 Pa/0.7 0.100 % 316 SS Dry 316 SS
Fluke 700P03 5 psi/0.0001 34 kPa/0.001 0.050 % Dry Dry 316 SS
Fluke 700P23 5 psi/0.0001 34 kPa/0.001 0.025 % 316 SS Dry 316 SS
Fluke 700P04 15 psi/0.001 103 kPa/0.01 0.025 % Dry Dry 316 SS
Fluke 700P24 15 psi/0.001 103 kPa/0.01 0.025 % 316 SS Dry 316 SS
Gage Fluke 700P05 30 psi/0.001 207 kPa/0.01 0.025 % 316 SS N/A 316 SS
Fluke 700P06 100 psi/0.01 690 kPa/0.07 0.025 % 316 SS N/A 316 SS
Fluke 700P27 300 psi/0.01 2070 kPa/0.1 0.025 % 316 SS N/A 316 SS
Fluke 700P07 500 psi/0.01 3400 kPa/0.1 0.025 % 316 SS N/A 316 SS
Fluke 700P08 1000 psi/0.1 6900 kPa/0.7 0.025 % 316 SS N/A 316 SS
Fluke 700P09 1500 psi/0.1 10 M Pa/0.001 0.025 % 316 SS N/A 316 SS
Absolute Fluke 700PA3 5 psi/0.0001 34 kPa/0.001 0.050 % 316 SS N/A 316 SS
Fluke 700PA4 150 psi/0.001 103 kPa/0.001 0.050 % 316 SS N/A 316 SS
Fluke 700PA5 30 psi/0.001 207 kPa/0.01 0.050 % 316 SS N/A 316 SS
Fluke 700PA6 100 psi/0.01 690 kPa/0.001 0.050 % 316 SS N/A 316 SS
Vacuum Fluke 700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 % 316 SS Dry 316 SS
Fluke 700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 % 316 SS Dry 316 SS
Dual Fluke 700PD2 ± 1 psi/0.0001 ± 6900 Pa/0.7 0.150 % 316 SS Dry 316 SS
Fluke 700PD3 ± 5 psi/0.0001 ± 34 kPa/0.001 0.040 % 316 SS Dry 316 SS
Fluke 700PD4 ± 15 psi/0.001 ± 103 kPa/0.01 0.025 % 316 SS Dry 316 SS
Fluke 700PD5 -15/30 psi/0.001 -100/207
kPa/0.01
0.025 % 316 SS N/A 316 SS
Fluke 700PD6 -15/100 psi/0.01 -100/690
kPa/0.07
0.025 % 316 SS N/A 316 SS
Fluke 700PD7 -15/200 psi/0.01 -100/1380
kPa/0.1
0.040 % 316 SS N/A 316 SS
High Fluke 700P29 3000 psi/0.1 20.7 M Pa/0.001 0.050 % C276 N/A C276
Fluke 700P30 5000 psi/0.1 34 M Pa/0.001 0.050 % C276 N/A C276
Fluke 700P31 10000 psi/1 69 M Pa/0.007 0.050 % C276 N/A C276
1 Total uncertainty, one year for temperature range O °C to +5O °C. Total uncertainty, 1.0% of full span for temperature range -10 °C to O °C. For POO module only,
compensated temperature range is 15 °C to 35 °C. 2 “Dry indicates dry air or non-corrosive gas as compatible media.
“316SS” indicates media compatible with Type 316 Stainless Steel. “C276” indicates media compatible with Hastelloy C276. Use of pressure zero is required prior to
measurement or source. Max overpressure specification includes common mode pressure. Modules are CE rated. Metric adapter(s}: 1/4” NPr female to male BSP/ISO 1/4-
19, tapered thread, included with all modules except P29, P3O, and P31, all modules include a NIST traceable certificate and test data.
525A-P Series Precision Pressure Transducers
Type Model Range/Resolution Range/Resolution Reference Uncertainty (23 ± 3 °C)
Differential 525A-P02 1 psi/0.00001 6900 Pa/0.01 0.008 % FS
Gage 525A-P03 5 psi/0.00001 34 kPa/0.001 0.008 % FS
Gage 525A-P04 15 psi/0.001 103 kPa/0.001 0.008 % FS
Gage 525A-P05 30 psi/0.0001 207 kPa/0.001 0.008 % FS
Gage 525A-P06 100 psi/0.001 690 kPa/0.001 0.008 % FS
Gage 525A-P07 500 psi/0.001 3400 kPa/0.01 0.008 % FS
Gage 525A-P08 1000 psi/0.01 6900 kPa/0.01 0.008 % FS
Gage 525A-P29 3000 psi/0.01 20.7 M Pa/0.0001 0.008 % FS
Absolute 525A-PA4 15 psi/0.0001 103 kPa/0.001 0.008 % FS
Absolute 525A-PA5 30 psi/0.0001 207 kPa/0.001 0.008 % FS
Absolute 525A-PA6 100 psi/0.001 690 kPa/0.001 0.008 % FS
Absolute 525A-PA7 500 psi/0.001 3400 kPa/0.01 0.008 % FS
Absolute 525A-PA8 1000 psi/0.01 6900 kPa/0.01 0.008 % FS
Vacuum 525A-PV4 -15 TO 0 psi/0.0001 -34 kPa/0.001 0.008 % FS
8 Fluke Calibration 525B Temperature/Pressure Calibrator
Model Description
525B Temperature/Pressure Calibrator
Software
MET/CAL Plus Automated Calibration Software (IEEE-488 and RS-232)
Accessories
5520A-525A/
Leads
Thermocouple and test lead set 1 meter TC wire with male
mini-plugs, All ANSI, L and U types; TC shorting plug; 1,
80AK-8001 TC adapter; 1, bead thermocouple type K; 1 meter
shielded test lead with double banana plugs (M); 2, 30 A test
leads, banana type connectors
Y525 19” Rack Mount Kit
Fluke 700PD3 Dual Pressure Module ± 5 psi / ± 34 kPa
Fluke 700P31 High Pressure Module 10000 psi/69 M Pa
Fluke 700P22 Differential Pressure Module 1 psi / 6900 Pa
Fluke 700P24 Differential Pressure Module 15 psi / 103 kPa
Fluke 700PD2 Dual Pressure Module ± 1 psi / ± 6900 Pa
Fluke 700P08 Gage Pressure Module 1000 psi / 6900 kPa
Fluke 700PA4 Absolute Pressure Module 15 psi / 103 kPa
Fluke 700P30 High Pressure Module 5000 psi / 34 M Pa
Fluke 700PD7 Dual Pressure Module -15/200 psi / -100/1380 kPa
Fluke 700P05 Gage Pressure Module 30 psi / 207 kPa
Fluke 700P06 Gage Pressure Module 100 psi / 690 kPa
Fluke 700P29 High Pressure Module 3000 psi / 20.7 M Pa
Fluke 700PD5 Dual Pressure Module -15 / 30 psi / -100 / 207 kPa
Fluke 700PV3 Vacuum Pressure Module -5 psi / -34 kPa
Fluke 700P23 Differential Pressure Module 5 psi / 34 kPa
Fluke 700PA5 Absolute Pressure Module 30 psi / 207 kPa
Model Description
Accessories continued
Fluke 700P03 Differential Pressure Module 5 psi / 34 kPa
Fluke 700PD6 Dual Pressure Module -15/100 psi / -100/690 kPa
Fluke 700P04 Differential Pressure Module 15 psi / 103 kPa
Fluke 700PA6 Absolute Pressure Module 100 psi / 6900 kPa
Fluke 700P01 Differential Pressure Module 10.00 in. H2O / 2.5 kPa
Fluke 700PD4 Dual Pressure Module ± 15 psi / ± 103 kPa
525A-P02 Differential 1 psi/6900 Pa
525A-P03 Gage 5 psi/34 kPa
525A-P04 Gage 15 psi/103 kPa
525A-P05 Gage 30 psi/207 kPa
525A-P06 Gage 100 psi/690 kPa
525A-P07 Gage 500 psi/3400 kPa
525A-P08 Gage 1000 psi/6900 kPa
525A-P29 Gage 3000 psi/20.7 M Pa
525A-PA4 Absolute 15 psi/103 kPa
525A-PA5 Absolute 30 psi/207 kPa
525A-PA6 Absolute 100 psi/690 kPa
525A-PA7 Absolute 500 psi/3400 kPa
525A-PA8 Absolute 1000 psi/6900 kPa
525A-PV4 Vacuum -15 to 0 psi/-34 kPa
Support
MET/SUPPORT
Gold
Software Support Program, one-year membership
MET/SUPPORT
GLDNW
MET/SUPPORT Gold Program, up to four workstations,
one-year membership
Pressure Measurement
The Calibrator can accept either the Fluke 700 or 525A-P series pressure modules. Pressure modules plug directly into the
front panel Lemo connector with the Calibrator firmware autodetecting the type and value of the module you are
attaching.
Range ............................................Determined by pressure module
Accuracy/Resolution.................Determined by pressure module
Units
PSI ..............................................pounds per square inch
inH2O4°C..................................inches of water at 4 degrees Celsius
inH2O20°C................................inches of water at 20 degrees Celsius
cmH2O4°C.................................centimeters of water at 4 degrees Celsius
cmH2O20°C..............................centimeters of water at 20 degrees Celsius
BAR ............................................bars
mBAR.........................................millibars
KPAL ..........................................kilopascals
inHG 0°C...................................inches of mercury at 0 degrees Celsius
mmHG 0°C................................millimeters of mercury at 0 degrees Celsius
Kg/cm2......................................kilograms per square centimeter
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40
2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1(425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2001-2012 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 6/2012 1611141D_EN
Pub_ID: 10884-eng Rev 02
Fluke Calibration. Precision, performance, confidence.