Clamp-on Testers 300 Series and CL Series of Clamp-on Testers Leakage Clamp-on Tester with a sharp low pass filter function Leakage Clamp-on Tester capable of measuring from 1 mA AC Series Rich lineup for various current measurements 30032A 30031A CL Series Bulletin CL-E Selection Guide For Leakage Current Model Diameter of measurable conductor Method of detection Frequency characteristics Range AC current Resolution AC voltage Other measurement DC voltage functions Continuity check Frequency Temperature Data hold Peak hold Recorder output Mean value display Filter Switch Waveform monitor output Page 30031A 40mm Mean value 50/60Hz 3/30mA, 30/60A 0.001mA 2 30032A 40mm Mean value 50/60Hz 3/30mA, 30/60A 0.001mA * 2 CL320 24mm Mean value 40Hz to 400Hz 20/200mA,200A 0.01mA 7 CL340 40mm Mean value 20Hz to 1kHz 40/400mA,400A 0.01mA 8 CL360 CL345 68mm 40mm Mean value True RMS 40Hz to 1kHz 20Hz to 1kHz 40/400mA,400A 200mA/2/20/200/1000A 0.1mA 0.01mA 9 8 *Page1: Description of Harmonic Filter Function For AC Current Model Diameter of measurable conductor Method of detection Frequency characteristics AC current DC current Other measurement functions Range Resolution Range Resolution AC voltage DC voltage Continuity check Frequency Temperature Data hold Peak hold Recorder output Waveform monitor output Page CL120 24mm Mean value 40Hz to 1kHz 20/200A 0.01A 3 CL130 30mm Mean value 40Hz to 1kHz 200/600A 0.1A 3 CL135 30mm True RMS 40Hz to 1kHz 200/600A 0.1A 4 CL150 54mm Mean value 40Hz to 1kHz 400/2000A 0.1A 4 CL155 54mm True RMS 40Hz to 1kHz 400/2000A 0.1A 5 For AC/DC Currents Model Diameter of measurable conductor Method of detection Frequency characteristics AC current DC current Other measurement functions Page Range Resolution Range Resolution AC voltage DC voltage Continuity check Frequency Temperature Data hold Peak hold Recorder output Waveform monitor output CL220 24mm Mean value 20Hz to 1kHz 40/300A 0.01A 40/300A 0.01A 5 CL235 33mm True RMS 40Hz to 1kHz 400/600A 0.1A 400/1000A 0.1A 6 CL250 55mm Mean value 40Hz to 1kHz 400/2000A 0.1A 400/2000A 0.1A 6 CL255 55mm True RMS 30Hz to 1kHz 400/2000A 0.1A 400/2000A 0.1A 7 Description of Harmonic Filter Function Harmonic Filter Function (Only Available in the 30032A) 1. What is a Harmonic? Waveform Image Harmonic refers to sinusoidal quantity having a frequency that is an integral multiple of the fundamental frequency (for example, the commercial frequency). When a harmonic is superimposed on the fundamental frequency, the waveform is distorted. Fundamental frequency 2. Why is it Necessary to Accurately Measure the Leakage Current of the Fundamental Frequency (Commercial Frequency)? One problem when measuring the leakage current to check the isolation of electrical circuits in power distribution equipment is that the electrical isolation cannot be correctly understood due to the influence of a harmonic current. That is, the leakage current flowing from the electrical circuit to ground is very small so that, in order to check the isolation of electrical circuits by means of the leakage current, it is necessary to remove the harmonic component of the leak current and measure only the current of the fundamental frequency (commercial frequency). Distorted waveform Harmonic *Waveforms obtained when measuring a distribution board of a Yokogawa Meters & Instruments 3. The 30032A Employs a Harmonic Filter Corporation office Conventional leakage clamp-on testers could not sufficiently remove harmonic current components so measured leakage current values were often larger than the specified value due to the influence of a harmonic current. In this case, retesting with an insulation tester was required, resulting in increased effort and cost for the test. Under these circumstances, Yokogawa Meters & Instruments Corporation has developed the leakage clamp-on tester 30032A, which employs a high-performance harmonic filter that can accurately measure just the fundamental frequency component of the leakage current. Characteristics of Harmonic Filter 1. Filter Characteristic of the 30032A 2. Filter Comparison (between the On and Off States) When the frequency is more than 60 Hz, the sharp filter removes the harmonic component, leaving the fundamental frequency. For example, the level of 100 Hz is attenated to approx. 1%. This is the filter characteristic in the On and Off states. When the filter is in the On or Off state Amplitude ratios in the range between the fundamental frequency and the third frequency Amplitude ratio [dB] 10 Second harmonic amplitude ratio Fundamental frequency at 50 Hz Amplitude ratio at 100 Hz: -41 dB (average) Fundamental frequency at 60 Hz Amplitude ratio at 120 Hz: -56 dB (average) 0 -10 -20 -30 -40 Filter state Harmonic On Off Fundamental 0 dB 0 dB Second Third -56 dB -1.3 dB -80 dB or more -6.7 dB Figure 2: Harmonic Filter Characteristic 2 Amplitude ratio [dB] Figure 1: Harmonic Filter Characteristic 1 Harmonic frequency Fundamental frequency: 50 Hz 60 Hz -50 -60 -70 -80 40 60 80 100 120 140 160 10 0 -10 -20 -30 -40 Comparison between the ON and OFF states Fundamental frequency: 60 Hz Difference: approx. -50 dB at 120 Hz -50 -60 -70 Filter ON Filter OFF -80 10 180 100 Freq[Hz] 1,000 Freq[Hz] Measurement Example : Measurement location of leakage current A: Measurement of the grounding wire for the transformer class B grounding work B: Measurement of the electrical circuit C: Measurement of the grounding wire of electrical equipment Measurement method of leakage current (1) 3-phase 4-wire electrical circuit Load device (2) Single-phase 3-wire electrical circuit B B B C IG E1 B A A IG E2 Example of load measurement with the CL Series E1 In the case of load current Current Load device Current *Be careful of the polarity in the case of direct current. 1 Load device Leakage Clamp-on Testers 30032A 30031A AC Leakage AC Leakage 40 40 AC/3mA~60A AC/3mA~60A Filter Can measure leakage AC currents of 1mA Filter function can cut off harmonics-currents components from 2nd order Filter function select on or off Can measure leakage AC currents of 1mA Standard AC Leakage model AC current measurement AC current measurement Range 3mA 30mA 30A 60A Accuracy: (% of reading + digits) Accuracy: (% of reading + digits) Filter function OFF Resolution 0.001mA 0.01mA 0.01A 0.1A Accuracy 0.010 I 32.70mA: 1.0%+5 0.05 I 50.0A:1.0%+5 50.0 I 60.6A:5.0%+5 Maximum Allowable Current 3.270mA 32.70mA 32.70A 60.6A Range 3mA 30mA 30A 60A 3mA 30mA 30A 60A Resolution 0.001mA 0.01mA 0.01A 0.1A Accuracy 0.010 I 32.70mA: 1.5%+5 0.05 I 50.0A:1.5%+5 50.0 I 60.6A:5.5%+5 0.001mA 0.01mA 0.01A 0.1A Accuracy 0.010 I 32.70mA: 1.0%+5 0.05 I 50.0A:1.0%+5 50.0 I 60.6A:5.0%+5 Maximum Allowable Current 3.270mA 32.70mA 32.70A 60.6A Zero correction 3 mA range: Displays 0.000 mA (zero) when 0.010 mA < I 30 A range: Displays 0.00 A (zero) when 0.05 A < I Filter function ON Range Resolution Maximum Allowable Current 3.270mA 32.70mA 32.70A 60.6A General Specifications Parameter Method Display Specification Mean-value detection and rms-value calibration LCD (Digital reading 3200 counts) Bar graph (32 segments) Range swiching Range selection Auto or Manual Data Hold On all Range Operating temperature and humidity 0 to 50C, 80% RH or less (no condensation) Temperature coefficient Following values must be added in the temperature range of either 0 to 18C or 28 to 50C 0 I 50.0A: (0.08% of reading/C + 0.5 digits/C) 50.0 < I 60.6A: (0.3% of reading/C + 0.5 digits/ C) 0.0005%typical value (in terms of the magnitude of Effect of external magnetic fields: current in adjacent wires) Conforms EN 61010-1, EN 61010-2-032 Safety standards CAT. III 300V 300 Vrms or less Circuit voltage 3.7 kV AC for one minute Withstanding voltage CR2032 lithium battery 1 Power supply 6mW maximum Power consumption Approx. 90 hours Battery life Power approx. 10 minutes after the last switch operation. Automatic power-off Approx 70 (W) 178 (H) 25 (D) (mm) Dimensions Approx 200 g (including the battery) Weight User's manual, Battery, Soft carrying case (RB057) Accessories Note: Input current of 2nd-order and higher harmonics : 150 mA rms maximum in the 3 mA/30 mA range : 62 A rms maximum in the 30 A/60 A range Filter specifications (3 mA and 30 mA ranges and 30 A and 60 A ranges) Amplitude ratio at 100 Hz: -38 dB (1.26%) or less (typical: -41 dB) Amplitude ratio at 120 Hz: -53 dB (0.22%) or less (typical: -56 dB) Zero correction 3 mA range: Displays 0.000 mA (zero) when 0.010 mA < I 30 A range: Displays 0.00 A (zero) when 0.05 A < I General Specifications Parameter Method Display Specification Mean-value detection and rms-value calibration LCD (Digital reading 3200 counts) Bar graph (32 segments) Range switching Range selection Auto or Manual Data Hold On all Range Operating temperature and humidity 0 to 50C, 80% RH or less (no condensation) Temperature coefficient Following values must be added in the temperature range of either 0 to 18C or 28 to 50C 0 I 50.0A: (0.08% of reading/C + 0.5 digits/C) 50.0 < I 60.6A: (0.3% of reading/C + 0.5 digits/ C) 0.0005%typical value (in terms of the magnitude of Effect of external magnetic fields current in adjacent wires) Conforms EN 61010-1, EN 61010-2-032 Safety standard CAT. III 300V 300 Vrms or less Circuit voltage 3.7 kV AC for one minute Withstanding voltage CR2032 lithium battery 1 Power supply 6mW maximum Power consumption Approx. 90 hours Battery life Power approx. 10 minutes after the last switch operation. Automatic power-off Approx 70 (W) 178 (H) 25 (D) (mm) Dimensions Approx 200 g (including the battery) Weight User's manual, Battery, Soft carrying case (RB057) Accessories 2 Clamp-on Testers CL120 CL130 AC A AC A 24 30 AC/20~200A AC/200~600A AC V/ Light weight & compact design Mean value display Data hold function Approved for comformity to safety standards EN61010-1, Mean value display Data hold function Approved for comformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. III 600 V) EN61010-2-032 (CAT. III 300 V) Specifications Parameter AC current Specifications At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Range 20A 200A Accuracy 2.0+7 (501kHz) 2.0+5 (50/60Hz) 3.0+10 (401kHz) Parameter AC current Range 200A 600A General Specifications Parameter Method of detection Display Response time Range switching Data hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field Effect of conductor position Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Auto power-off Diameter of measurable conductor Dimensions Weight Accessories At 23;5;, 75%RH or less Accuracy:(% of reading + digits) AC voltage Specification Mean value LCD(Digital display:1999 counts) Approx. 2 seconds Manual-range On all range 0-40;, 85% RH or less (no condensation) 0.8A or less at 400A/m 2% or less Conforms EN 61010-1, EN61010-2-032 300Vrms or less 3.7kV AC for one minute LR-442(3V) or SR-442 Approx. 100 hours (continuous) Approx. 1mA Approx. 10 minutes 24mm diameter max. Approx. 59(W)148(H)26(D)mm Approx. 100g User's manual, batteries, carrying case(93033) Resistance 200V/600V 200 Accuracy 1.5+6 (50/60Hz) 2.0+5 (401kHz) 1.0+3 (50/60Hz) 2.0+5 (401kHz) 1.0+2 (50/60Hz) 1.5+4 (401kHz) 1.2+4, Beeps at below 30 (continuity check) General Specifications Parameter Method of detection Display Response time Specification Mean value LCD(Digital display:1999 counts) Approx. 1 second (approx. 2 seconds on resistance renge) Range switching Manual-range Data hold On all range Operating temperature and humidity -10-50;, (no condensation) up to 30;, 90%RH or less up to 40;, 75%RH or less up to 50;, 45%RH or less Temperature coefficient 2A or less at 400A/m Effect of external magnetic field 2% or less Effect of conductor position Conforms EN 61010-1, EN61010-2-031, EN61010-2-032 Safety standard 600Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage 6F22(006P)9V1 or 6LR612 Power supply Approx. 200 hours (continuous) Battery life Approx. 2mA Consumed current Approx. 10 minutes Auto power-off 30mm diameter max. Diameter of measurable conductor Approx. 93(W)210(H)40(D)mm Dimensions Approx. 400g Weight User's manual, batteries, carrying case(93032) Accessories 3 Clamp-on Testers CL135 CL150 AC A AC A 30 54 AC/200~600A AC/400~2000A AC V/DC V/ RMS AC V/ True RMS Display Data hold function Approved for comformity to safety standard EN61010-1, Mean value display DC output function Data hold function Sleep function Approved for comformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. III 600 V) Specifications Parameter AC current At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Range 200A 600A AC voltage Crest factor Resistance EN61010-2-031, EN61010-2-032 (CAT. III 600 V, CAT. II 1000 V) Specifications 200V/600V 200 At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Accuracy 1.5+4 (50/60Hz) 2.0+5 (401kHz) 1.5+4 (50/60Hz) 2.0+5 (401kHz) 1.0+2 (50/60Hz) 1.5+4 (401kHz) 3 (50/60Hz) 1.2+4 , Beeps at below 30(continuity check) Parameter AC current Accuracy 1.0+3 (50/60Hz) 2.0+3 (401kHz) 1.0+3 (50/60Hz) 2000A(01500A) 3.0+3 (401kHz) 3.0(50/60Hz) 2000A(15002000A) AC voltage 1.0+2 (50/60Hz) 40/400/750V 1.5+3 (401kHz) DC voltage 1.0+2 40/400/1000V 1.5+2 , Beeps at below 5035(continuity check) Resistance 400/4k/40k/400k 1.5% rdg 0.5mV (50/60Hz) DC output 400A(0400mV) 2.5% rdg 0.5mV (401kHz) 1.5% rdg 0.5mV (50/60Hz) 2000A(0150mV/01500A) 3.5% rdg 0.5mV (401kHz) 2000A(150200mV/15002000A) 3.5% rdg (50/60Hz) General Specifications Parameter Method of detection Display Response time Specification True RMS LCD(Digital display:1999 counts) Approx. 1 second (approx. 2 seconds on resistance renge) Range switching Manual-range Data hold On all range Operating temperature and humidity -10-50;, (no condensation) up to 30;, 90%RH or less up to 40;, 75%RH or less up to 50;, 45%RH or less Temperature coefficient 2A or less at 400A/m Effect of external magnetic field 3% or less Effect of conductor position Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 600Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage 6F22(006P)9V1 or 6LR612 Power supply Approx. 200 hours (continuous) Battery life Approx. 2mA Consumed current Approx. 10 minutes Auto power-off 30mm at maximum Diameter of measurable conductor Approx. 93(W)210(H)40(D)mm Dimensions Approx. 400g Weight User's manual, batteries, carrying case (93032) Accessories Range 400A General Specifications Parameter Method of detection Display Response time Range switching Specification Mean value LCD(Digital display:4000 counts) Approx. 2 seconds Manual-range(on AC current range)/ Auto-range(on AC voltage range, resistance range) Data hold On all range Peak hold On AC current range Operating temperature and humidity 0-40;, 85% RH or less (no condensation) Temperature coefficient 1A or less at 400A/m Effect of external magnetic field (2.0% rdg + 3dgt) or less Effect of conductor position Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 1000Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage R6P(SUM-3)2 or LR62 Power supply Approx. 150 hours (continuous) Battery life Approx. 5mA Consumed current Automatically powered down in about 10 minutes Sleep function after the last switch operation 54mm at maximum Diameter of measurable conductor Approx. 105(W)247(H)49(D)mm Dimensions Approx. 470g Weight User's manual, batteries, carrying case (93034) Accessories 4 Clamp-on Testers CL155 CL220 AC A AC A/DC A 54 24 AC/400~2000A AC/40~300A RMS DC/40~300A AC V/DC V/ Light weight & compact design Mean value display Sleep function Approved for comformity to safety standards EN61010-1, True RMS display DC output function Data hold function Sleep function Approved for comformity to safety standards EN61010-1, EN61010-2-032 (CAT. III 300 V) EN61010-2-031, EN61010-2-032 (CAT. IV 300V, CAT. III 600V, CAT. II 1000V) Specifications Specifications At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Parameter DC current Accuracy 1.0+3 (50/60Hz) 2.0+3 (401kHz) 1.0+3 (50/60Hz) 2000A(01500A) 3.0+3 (401kHz) 3.0(50/60Hz) 2000A(15002000A) 1.0+2 (50/60Hz) AC voltage 40/400/750V 1.5+3 (401kHz) 1.0+2 DC voltage 40/400/1000V 1.5+2 , Beeps at below 5035(continuity check) Resistance 400/4k/40k/400k 1.5% rdg 0.5mV (50/60Hz) DC output 400A(0400mV) 2.5% rdg 0.5mV (401kHz) 1.5% rdg 0.5mV (50/60Hz) 2000A(0150mV/01500A) 3.5% rdg 0.5mV (401kHz) 2000A(150200mV/15002000A) 3.5% rdg (50/60Hz) Parameter AC current Range 400A AC current At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Range 40A 300A(20200A) 300A(200300A) 40A 300A(20200A) 300A(200300A) Accuracy 1.0+4 1.5+4 3.0 1.0+4 (50/60Hz) 2.5+4 (201kHz) 1.5+4 (50/60Hz) 2.5+4 (201kHz) 3.5 (50/60Hz) 4.0 (201kHz) General Specifications Parameter Method of detection Display Response time Range switching Data hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field Effect of conductor position Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Sleep function General Specifications Parameter Method of detection Display Response time Range switching Specification True RMS LCD(Digital display:4000 counts) Approx. 2 seconds Manual-range(on AC current range)/ Auto-range(on AC voltage range, resistance range) Data hold On all range Peak hold On AC current range Operating temperature and humidity 0-40;, 85% RH or less (no condensation) Temperature coefficient 1A or less at 400A/m Effect of external magnetic field (2.0% rdg + 3dgt) or less Effect of conductor position Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 1000Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage R6P(SUM-3)2 or LR62 Power supply Approx. 80 hours (continuous) Battery life Approx. 7mA Consumed current Automatically powered down in about 10 minutes Sleep function after the last switch operation 54mm at maximum Diameter of measurable conductor Approx. 105(W)247(H)49(D)mm Dimensions Approx. 470g Weight User's manual, batteries, carrying case (93034) Accessories Diameter of measurable conductor Dimensions Weight Accessories 5 Specification Mean value LCD(Digital display:4000 counts) Approx. 2 seconds Auto-range On all range 0-40;, 85% RH or less (no condensation) 1A or less at 400A/m (2.0% rdg + 5dgt) or less Conforms EN61010-1, EN61010-2-032 300Vrms or less 3.7kV AC for one minute LR-442(3V) or SR-442 Approx. 11 hours (continuous) Approx. 9mA Automatically powered down in about 5 minutes after the last switch operation 24mm at maximum Approx. 59(W)147(H)25(D)mm Approx. 100g User's manual, batteries, carrying case(93033) Clamp-on Testers CL235 CL250 AC A/DC A AC A/DC A 33 55 AC/400~600A AC/400~2000A RMS DC/400~2000A AC V/DC V/ DC/400~1000A DC V/AC V/ /Hz Mean value display Sleep function Data hold function Approved for comformity to safety standards EN61010-1, True RMS display Sleep function Data hold function Approved for comformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. IV 600 V, CAT. III 1000 V) EN61010-2-031, EN61010-2-032 (CAT. III 600 V) Specifications Parameter AC current DC current AC voltage DC voltage Crest factor Resistance Frequency Specifications At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Range 400/600A 400/1000A 40/400/600V 40/400/600V 400/4000 103000Hz Parameter DC current AC current Accuracy 1.5+5 (50/60Hz) 3.5+5 (401kHz) 1.0+5 1.5+5 (50/60Hz) 3.5+5 (401kHz) 1.0+5 3 1.0+5, Beeps at below 20(continuity check) 1.5+5 DC voltage AC voltage Resistance DC output General Specifications Parameter Specification Method of detection True RMS Display LCD(Digital display:3999 counts) Response time Approx. 2 second Range switching Auto-range Data hold On all range Peak hold On AC/DC current range, AC/DC voltage range Average measurement 0-40;, 90% RH or less (no condensation) Operating temperature and humidity Temperture coefficient Effect of external magnetic field 2% or less Effect of conductor position Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 600Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage 6F22(006P)9V1 or 6LR611 Power supply Approx. 15 hours (continuous) Battery life Approx. 15mA Consumed current Automatically powered down in about Sleep function 30 minutes after the last switch operation 33mm at maximum Diameter of measurable conductor Approx. 91(W)210(H)40(D)mm Dimensions Approx. 450g Weight User's manual, batteries, carrying case(93032) Accessories At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Accuracy 1.5+2 1.5+2 (50/60Hz) 400A/2000A(01000A) 3.0+4 (40500Hz) 5.0+4 (5001kHz) 3.0+2 (50/60Hz) 2000A(10012000A) 1.0+2 400/1000V 1.5+2 (50/60Hz) 400/750V 1.5+4 (401kHz) 1.5+2, Beeps at below 5035(continuity check) 400/4000 1.5% rdg 3mV DC400A(0400mV) 1.5% rdg 3mV DC2000A(0200mV) 1.5% rdg 3mV (50/60Hz) AC400A(0400mV) 3.0% rdg 3mV (40500Hz) AC2000A(0100mV/01000A) 5.0% rdg 3mV (5001kHz) AC2000A(100.1200mV/10012000A) 3.0% rdg 3mV (50/60Hz) Range 400/2000A General Specifications Parameter Method of detection Display Response time Range switching Specification Mean value LCD(Digital display:3999 counts) Approx. 2 seconds Manual-range(on current, voltage range) /Auto-range(on resistance range) Peak hold On all range Max hold On current/voltage range Operating temperature and humidity 0-40;, 85% RH or less (no condensation) Temperature coefficient 4A or less at 400A/m Effect of external magnetic field (1.5% rdg + 3dgt) or less Effect of conductor position Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 1000Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage R6P(SUM-3)2 or LR62 Power supply Approx. 100 hours (continuous) Battery life Approx. 9mA Consumed current Automatically powered down in about Sleep function 10 minutes after the last switch operation 55mm at maximum Diameter of measurable conductor Approx. 105(W)250(H)49(D)mm Dimensions Approx. 530g Weight User's manual, Test Lead(98011), Accessories Output plug(98012), batteries, carrying case(93034) 6 Leakage Clamp-on Testers Clamp-on Testers CL255 CL320 AC A/DC A AC Leakage 55 24 AC/400~2000A AC/20mA~200A DC/400~2000A RMS AC V/DC V/ /Hz True RMS display Sleep function Data hold function Approved for comformity to safety standards EN61010-1, Mean value display Auto power-off Manual range switching Approved for comformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. III 600 V, CAT. II 1000 V) Specifications Parameter DC current AC current 400/4000 103999Hz DC400A(0400mV) DC2000A(15200mV) Accuracy 1.5+2 1.5+3 (50/60Hz) 3.0+4 (301kHz) 3.5+3 (50/60Hz) 1.0+2 1.5+3 (50/60Hz) 2.0+4 (301kHz) 3 1.5+2, Beeps at below 20(continuity check) 1.5+5 1.5% rdg 3mV 1.5% rdg 3mV AC400A(0400mV) /AC2000A(15170mV/1501700A) 1.5% rdg 3mV (50/60Hz) 3.0% rdg 3mV (401kHz) AC2000A(170.1200mV/17012000A) 3.5% rdg 3mV (50/60Hz) 400A/2000A(1501700A) 2000A(17012000A) 40/400/1000V 40/400/750V Crest factor Resistance Frequency DC output Specifications At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Range 400/2000A DC voltage AC voltage EN61010-2-032 (CAT. III 300 V) At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Parameter Range AC current 20mA/200mA 200A(0100A) 200A(100.1200A) Accuracy WIDE(40400Hz) 50/60Hz 2.0+4 (50/60Hz) 3.0+5 (50/60Hz) 5.0+6 (40400Hz) 5.0+4 (50/60Hz) 5.0+5 (50/60Hz) General Specifications Parameter Method of detection Display Response time Range switching Data hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field Effect of conductor position General Specifications Parameter Method of detection Display Response time Specification True RMS LCD(Digital display:3999 counts) Approx. 1 second(on DC current/voltage range), Approx. 2 seconds(AC current/voltage range, resistance range) Auto-range Range switching On all range (without peak hold) Data hold On current/voltage range Peak hold On current/voltage range Average Measeurement 0-40;, 85% RH or less (no condensation) Operating temperature and humidity Temperature coefficient 4A or less at 400A/m Effect of external magnetic field (1.5% rdg + 3dgt) or less Effect of conductorposition Conforms EN61010-1, EN61010-2-031, EN61010-2-032 Safety standard 1000Vrms or less Circuit voltage 5.55kV AC for one minute Withstanding voltage 6F22(006P)9V1 or 6LR611 Power supply Approx. 15 hours (continuous) Battery life Approx. 15mA Consumed current Automatically powered down in about Sleep function 10 minutes after the last switch operation 55mm at maximum Diameter of measurable conductor Approx. 105(W)250(H)49(D)mm Dimensions Approx. 540g Weight User's manual, Test Lead(98011), Accessories Output plug(98012), batteries, carrying case(93034) Effect of residual current Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Auto power-off Diameter of measurable conductor Dimensions Weight Accessories 7 Specification Mean value LCD(Digital display:1999 counts) Approx. 2 seconds Manual-range On all range 0-40;, 85% RH or less (no condensation) 10mA or less in proximity to a 14.4mm-dia conductor carrying 100A Within 5dgt for 0 to 50A, or 2% for 50 to 200A (10mm-dia conductor at inside the jaw) 10mA or less in proximity to a 10mm-dia conductor carrying 50A Conforms EN61010-1, EN61010-2-032 300Vrms or less 3.7kV AC for one minute LR-442(3V) or SR-442 Approx. 15 hours (continuous) Approx. 5mA Approx. 10 minutes 24mm at maximum Approx. 60(W)149(H)26(D)mm Approx. 120g User's manual, batteries, carrying case(93033) Leakage Clamp-on Testers CL340 CL345 AC Leakage AC Leakage 40 40 AC/40mA~400A AC/40mA~400A RMS True RMS display Auto power-off Manual range switching Approved for comformity to safety standards EN61010-1, Mean value display Auto power-off Manual range switching Approved for comformity to safety standards EN61010-1, EN61010-2-032 (CAT. III 300 V) EN61010-2-032 (CAT. III 300 V) Specifications Specifications At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Parameter Range AC current 40mA/400mA 400A(0350A) 400A(350400A) Accuracy WIDE(20Hz) 50/60Hz 1.0+5 (50/60Hz) 2.5+10 (201kHz) 2.5+10 (401kHz) 1.0+5 (50/60Hz) 5.0 (401kHz) 2.0 (50/60Hz) Effect of conductor position Effect of residual current Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Auto power-off Diameter of measurable conductor Dimensions Weight Accessories Parameter Range AC current 40mA/400mA 400A(0300A) 400A(300400A) Accuracy WIDE(20Hz) 50/60Hz 1.0+5 (50/60Hz) 2.5+10 (201kHz) 2.5+10 (401kHz) 1.0+5 (50/60Hz) 5.0 (401kHz) 2.0 (50/60Hz) General Specifications General Specifications Parameter Method of detection Display Response time Range switching Data hold Peak hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Parameter Method of detection Display Response time Range switching Data hold Peak hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field Specification Mean value LCD(Digital display:3999 counts)* Approx. 2 seconds Manual-range On all range On all range 0-40;, 85% RH or less (no condensation) 10mA or less in proximity to a 15mm-dia conductor carrying 100A 40/400mA range:Within 5dgt at every part inside the jaw400A range, 0 to 250A:Within 0.5%rdg 5dgt at every part inside the jaw section 12mA or less in proximity to a 10mm-dia conductor carrying 100A Conforms EN61010-1, EN61010-2-032 300Vrms or less 3.7kV AC for one minute R0-3(UM-4) 2 or LR03 2 Approx. 40 hours (continuous) Approx. 13mA Approx. 10 minutes 40mm at maximum Approx. 81(W)185(H)40(D)mm Approx. 270g User's manual, batteries, carrying case(93030) Effect of conductor position Effect of residual current Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Auto power-off Diameter of measurable conductor Dimensions Weight Accessories Specification True RMS LCD(Digital display:4200 counts)* Approx. 2 seconds Manual-range On all range On all range 0-40;, 85% RH or less (no condensation) 10mA or less in proximity to a 15mm-dia conductor carrying 100A 40/400mA range:Within 5dgt at every part inside the jaw400A range, 0 to 250A:Within 0.5%rdg 5dgt at every part inside the jaw section 12mA or less in proximity to a 10mm-dia conductor carrying 100A Conforms EN61010-1, EN61010-2-032 300Vrms or less 3.7kV AC for one minute R0-3(UM-4) 2 or LR03 2 Approx. 24 hours (continuous) Approx. 21mA Approx. 10 minutes 40mm at maximum Approx. 81(W)185(H)32(D)mm Approx. 270g User's manual, batteries, carrying case(93030) *6000 counts (40/400mA range) *6000 counts (40/400mA range) 8 Leakage Clamp-on Testers Clamp-on Testers CL360 96001 AC Leakage 20Hz~20kHz 68 AC A AC/200mA~1000A 33 Compact and light with high-performance 20Hz to 20kHz wide frequency characteristics Can be connected to a digital multimeter Need not be connected to a power source Fit to waveform measurement using oscilloscopes and oscillographic Approved for conformity to safety standards EN 61010-1, EN 61010-2-032 (CAT. III 300 V, CAT. II 600 V) recorders Specifications Specifications Parameter At 23;5;, 75%RH or less Accuracy:(% of reading + digits) Measuring range Output voltage Accuracy (at 23;5;, sine wave input) 0 to 400 A ACrms (600 A peak) 0 to 4 V ACrms (10mV/A) 1.5% of rdg 0.4mV (20Hz to 40Hz) 1.0% of rdg 0.2mV (40Hz to 1kHz) (amplitude) (0.8+0.2 f kHz)% of rdg (0.2+0.04 f kHz)mV (1 to 20kHz) (phase) Within 3.0; (40Hz to 1kHz) Temperature coefficient 0.05% fs/; in a range of 0; to 18; and 28; to 50; Output impedance Approximately 30 Load impedance 100k min.//100 pF max. Effect of external magnetic field 2mV(0.2A)or less at 400A/m Voltage of circuit under measurement 600V ACrms maximum Diameter of applicable conductor Maximum diameter of 33 mm Operating temperature and humidity limits 0; to 50;, 80% R.H. max.(non-condensing) Storage temperature limits -20; to 60;(non-condensing) Withstanding voltage 3.7kV AC for one minute (core and case, and core and output terminals) Safety Conforms EN61010-1, EN61010-2-032 External dimensions Approximately 73(W)130(H)30(D)mm Weight Approximately 220g Output cord length Approximately 2.5 meters(with banana plug) Accessories One portable case and one User's manual Range Accuracy WIDE(401kHz) 50/60Hz AC current 1.0+2 (50/60Hz) 200mA/2A/20A 1.5+2 3.0+2 (401kHz) 1.5+2 (50/60Hz) 200A 2.0+2 3.5+2 (401kHz) 1.5+2 (50/60Hz) 1000A(0500A) 2.0+2 3.5+2 (401kHz) 5.0 (50/60Hz) 1000A(5011000A)* 5.5 10.0 (401kHz) AC output 200mA/2A/20A(0200mV) 2.0 2.0 200A(0200mV) 2.5 2.5 1000A/(050mV/0500A) 3.0 3.0 1000A/(50100mV/5011000A) 5.0 5.0 DC output 200mA/2A/20A(0200mV) 3.0 3.5 200A(0200mV) 3.5 4.0 1000A/(050mV/0500A) 5.0 5.5 1000A/(50100mV/5011000A) 7.0 7.5 *Measurement of 501 to 1000A can be performed within 10 minutes. General Specifications Parameter Method of detection Display Response time Range switching Data hold Peak hold Operating temperature and humidity Temperature coefficient Effect of external magnetic field Effect of conductor position Effect of residual current Safety standard Circuit voltage Withstanding voltage Power supply Battery life Consumed current Diameter of measurable conductor Dimensions Weight Accessories Specification Mean value LCD(Digital display:1999 counts) Approx. 1 second Manual-range On all range On all range -10-50;, 80% RH or less (no condensation) 15mA or less in proximity to a 10mm-dia conductor carrying 100A 2% or less 10mA or less in proximity to a 10mm-dia conductor carrying 100A Conforms EN61010-1, EN61010-2-032 600Vrms or less 3.7kV AC for one minute 6F22(006P)9V1 or 6LR611 Approx. 60 hours (continuous) Approx. 5mA 68mm at maximum Approx. 129(W)248(H)55(D)mm Approx. 570g User's manual, batteries, carrying case(93031) 9 Accessories 99025 Specifications Item Measuring Range Ratio/Range Accuracy Allowable Measurement Time Conductor Size Frequency Response Safety Standard Withstand Voltage Dimensions Weight Accessories For use with following models Optional Accessories Specifications 0AC3000A 10:1 (input to output) 2% of input 0.5A 01000A(continuous),10001500A(10 minutes max.), 3000A(30 seconds max.) 100mm max. (100150mm) 50Hz/60Hz EN61010-1 CAT.III300V Pollution Degree 2 AC3700V for 1 minute 150(W)317(H)33(D)mm 40(W)45(H)10(D)mm Output coil Approx. 750g 93035(Carrying Case) Clamp Adapter : 99025 Output Cable for Terminal Screw : 91019 Item Output Cable for Terminal Screw Output Cable with Banana Plug Clamp Adapter Model Specification 91019 Cable length: approx.1.1m 91020 Cable length: approx.2.0m CL360 Ratio/Range = 10:1/3000A CL120,CL130,CL135,CL150, CL155,CL220,CL235,CL250, CL255,CL320,CL340,CL345 Specification Angle Plug type Straight Plug type 3pcs/set Hard type Soft type Soft type Soft type Soft type Hard type Soft type Applicable model CL130,CL135,CL235 CL150,CL155,CL250,CL255 CL150,CL155,CL250,CL255 CL340,CL345 CL360 CL130,CL135,CL235 CL120,CL220,CL320 CL150,CL155,CL250,CL255 99025 30031A/30032A 99025 Applicable model CL150,CL155 CL250,CL255 Supplementary Products Item Test Lead Test Lead Output Pllug Carrying Case Carrying Case Carrying Case Carrying Case Carrying Case Carrying Case Carrying Case Model 98010 98011 98012 93030 93031 93032 93033 93034 93035 RB057 Test lead : 98010 Test lead : 98011 Output Plug : 98012 Carrying case : 93030 Carrying case : 93031 Carrying case : 93032 Carrying case : 93033 Carrying case : 93034 Carrying case : 93035 Output Cable with Banana Plug : 91020 NOTICE World Wide Web site at http://www.yokogawa.com/ymi YOKOGAWA METERS & INSTRUMENTS CORPORATION Tachihi Bld. No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan Phone: +81-42-534-1413 Facsimile: +81-42-534-1426 International Sales Dept. YOKOGAWA CORPORATION OF AMERICA (U.S.A.) YOKOGAWA EUROPE B. V. (THE NETHERLANDS) YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE) YOKOGAWA AMERICA DO SUL LTDA (BRAZIL) YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA) YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA) YOKOGAWA INDIA LTD. (INDIA) YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA) YOKOGAWA MIDDLE EAST B. S. C.(C) (BAHRAIN) YOKOGAWA ELECTRIC CIS LTD. (RUSSIAN FEDERATION) Phone: +1-770-253-7000 Phone: +31-88-4641000 Phone: +65-6241-9933 Phone: +55-11-5681-2400 Phone: +82-2-551-0660 Phone: +61-2-8870-1100 Phone: +91-80-4158-6000 Phone: +86-21-6239-6363 Phone: +973-17-358100 Phone: +7-495-737-7868 Subject to change without notice. All Rights Reserved. Copyright (c) 1996, Yokogawa Meters & Instruments Corporation. Facsimile: +1-770-254-0928 Facsimile: +31-88-4641111 Facsimile: +65-6241-2606 Facsimile: +55-11-5681-4434 Facsimile: +82-2-551-0665 Facsimile: +61-2-8870-1111 Facsimile: +91-80-2852-1441 Facsimile: +86-21-6880-4987 Facsimile: +973-17-336100 Facsimile: +7-495-737-7869 [Ed: 10/b] Before using the product, read the instruction manual carefully to ensure proper and safe operation. Represented by: MIK-ES17 Printed in Japan: Mar. 2011(B)/5,000(KP)