All EOS power supplies have UL, CSA and Nemko safety compliance and medical versions meet UL60601-1 standards. All our products are RoHs
compliant. Visit us at www.eospower.com. EOS brand power supplies and patents are owned exclusively by EOS Power. All Rights Reserved.
4EM-10-0xx 39-DE60-44226-002/A2
Medical 40 Watt AC/DC Power Supplies
Features :
Low leakage current
<250 μA
EN55022-B, CISPR22-B
FCC Part 15 Level B
Small 4.0”x2.0”x1.07
inch package
Ultra high efficiency
>82%
NEMKO, UL, and CSA
approvals to IEC60601
Cover Kit accessory
available.
MVLT40 Series
AC Input Auto-ranging 90 to 132V
180 to 264V
Efficiency
Typical MVLT40-3203
Other Models 75%
82%
Input Frequency 47-63 Hz
Input Current Full Load at low line 1.1Arms , max
Inrush Current High Line , cold start 75A, max
INPUT SPECIFICATION
Leakage Current <250 μA
Output Power 40W max
Voltage Set Point See table
Hold-Up Time 115V/230V 6ms./15ms
Voltage Set Point
Accuracy V1
V2 and V3 +3%
+5%
Line Regulation Low line to high line +0.3%
Load Regulation Min to 50% &
50% to Max load
V1
V2 and V3 +0.5%
+5%
Transient Response 50% to full load
Voltage Deviation
Recovery Time 10%, max
<5 ms
OVP
(For 3 o/p models only) V1 only 3.3 V
V1 only 5V 3.95V, +0.25V
6.2V, +0.4V
Overload Protection Primary limited Yes, 130%
Short Circuit Protection Auto recovery <6s Short Term
Ripple and Noise 3.3V and 5V
Other voltages 50mV
1%
Output Rise Time All outputs <100ms
OUTPUT SPECIFICATIONS
* Maximum current per output channel. Do not exceed total output power rating
Model
Number Output Voltage /
Current * Voltage
Set Point Min.
Load
MVLT40-1200 V1 5V@ 8.0A 5. 1V 0. 2A
MVLT40-1201 V1 12V@ 3.5A 12V 0. 1A
MVLT40-1202 V1 15V@ 2.7A 15V 0. 1A
MVLT40-1203 V1 24V@ 1.7A 24V 0. 05A
MVLT40-3200 V1
V2
V3
5V@6.0A
12V@2.0A
-12V@0.5A
5. 2V
12. 5V
-12. 8V
0. 5A
0. 1A
0. 0A
MVLT40-3201 V1
V2
V3
5V@6.0A
24V@1.0A
-12V@0.5A
5. 2V
24V
-12. 8V
0. 5A
0. 1A
0. 0A
MVLT40-3202 V1
V2
V3
5V@6.0A
15V@1.5A
-15V@0.5A
5. 2V
14. 6V
-14. 8V
0. 5A
0. 1A
0. 0A
MVLT40-3203 V1
V2
V3
3.3V@6.0A
5V@3.0A
-12V@0.5A
3. 3V
5. 2V
-12. 8V
0. 1A
0. 1A
0. 0A
VLT60CK Metal cover kit accessory
MVLT40-1204 V1 48V@ 0.83A 48V 0.05A
This industry and designer’s fa-
vorite is used to power a variety
of medical applications to include
endoscopy systems and video,
pumps/drips, test and lab gear,
sleep disorder, If space is a pre-
mium and thermal management
integral to your design we sug-
gest the MVLT40 as the optimum
solution. The MVLT40 family is
RoHs compliant and our factory is
ISO 14001 environmentally ap-
proved. Worldwide medical safety
standards apply and the power
supply is fitted with OVP and
OCP standard features. Applica-
tions include:
Endoscopy equipment
Pumps, IVs and drips
Laboratory gear
Fluid analyzers
Respiratory systems
Diagnostic systems
Low profile in a 4.0”x2.0”x1.07” package and over 82% efficient the EOS MVLT40 medical
power supply family comes in single and triple DC output configurations. The low leakage
current MVLT40 offers a full 40 watts output power with convection cooling and more with
optional air flow. A cover kit is available for easy drop in design and manufacturability.
Page 1
All EOS power supplies have UL, CSA and Nemko safety compliance and medical versions meet UL60601-1 standards. All our products are RoHs
compliant. Visit us at www.eospower.com. EOS brand power supplies and patents are owned exclusively by EOS Power. All Rights Reserved.
4EM-10-0xx 39-DE60-44226-002/A2
Medical 40W AC/DC Power Supplies
MVLT40 Series
EMC AND SAFETY SPECIFICATIONS
CE Mark Complies with the LVD
EMC EN5022-B,CISPR22-B, FCC
Part 15 Class B, EN50082-1
EN60601-1-2ND ED
Agency Approval VDE , UL, c-UL
Safety Standard IEC60601-1, EN60601-1-1,
UL60601-1, CSA C22.2 No
601 Class 1 SELV.
Safety file number NEMKO:P06206742
UL: E173812
Isolation Voltage 4000 VAC
Operating Temperature No derating
Linear derate to 50% 0 to 50oC
51 to 70oC
Storage Temperature -40 to 70oCC
Cooling Convection
Relative humidity Non condensing 95%, max
MTBF MIL - HDBK-217F > 100,000 hrs
ENVIRONMENTAL SPECIFICATION
MECHANICAL SPECIFICATION
AC Input Connector J1 Molex 3 position, 0.156 center
Header 26-60-4030 or equivalent
Ground Connector Spade connector
DC Output connector J2 Molex 6 position, 0.156 center
Header 26-60-4060 or equivalent
Size 4.0 “X 2.0” x 1.07”
101.6 x 50.8 x 27.18 mm
Weight 5oz (150g)
OUTPUT CONNECTOR J2
Single Output
Pin1 V1
Pin2 V1
Pin3 Common
Pin4 Common
Triple Output
V1
V1
Common
Common
Pin5 NC V3
Pin6 NC V2
Pin1 AC Neutral
Pin2 AC Line
INPUT CONNECTOR J1
INPUT SPADE CONNECTOR
GND AC Ground
Page 2