High Voltage Differential Probes HVD3605A, HVD3206A HVD310xA Key Features * kV, 2 kV, 6 kV CAT safety 1 rated models The HVD3000A series high voltage differential probes provide high CMRR * idest differential voltage W ranges available found in noisy, high common-mode power electronics environments. * Exceptional common-mode rejection ratio (CMRR) across a broad frequency range floating measurements. * * 1% gain accuracy over a broad frequency range to simplify the measurement challenges The probe's design is easy-to-use and enables safe, precise high voltage Exceptional Common-Mode Rejection Ratio Widest Application Coverage The HVD3000A series cover the fullest The CMRR for the probes is igh offset capability at both H high and low attenuation range of applications, from 120/240 V exceptional out to very high switch-mode power supplies through * * frequencies, greatly improving 600 V class and 5 kV class electrical AC and DC coupling measurement capability in the noisy, apparatus, Each model has the best roBus active probe interface P with automatic scaling high common-mode environments available gain accuracy, widest differ- found in power electronics. The high ential and offset voltage range, and * utoZero with auto A disconnect switch CMRR combined with low probe superior CMRR. * Wide oscilloscope compatibility noise and high offset capability provides superior measurement capabilities compared to other HV differential probes. High Precision Measurements HVD3000A series probes provide 1% gain accuracy enabling precise voltage measurements. AutoZero capability ensures further measurement precision by allowing small offset drifts to be calibrated out of the measurement. This feature can be applied without removing the HVD3000A leads from the device under test. 1500 Vdc CAT III Rating The HVD3206A supports 1500 Vdc solar photovoltaic (PV) measurements per the IEC/EN 61010- 031:2015 standard. It has a 1500 Vdc CAT III and a 2000 V (DC+peakAC) CAT I rating. Complete Probe Integration The ProBus interface provides power and communication to the probe eliminating the need for a separate power supply or batteries. Attenuation is automatically selected based on oscilloscope gain range (V/div) setting. WIDE APPLICATION COVERAGE - 600 V TO 5 KV APPARATUS 600 V Class, 3-phase Electrical Apparatus The HVD310xA probes are available in a range of bandwidths from 25 to 120 MHz with standard 2 meter cable (6 meter cable model available). All probes have industry best CMRR for best noise-immunity, are guaranteed to 1% gain accuracy, and have the industry's widest differential voltage range in a 1 kV rated high voltage differential probe. * nly 1 kV safety rated probe that O serves the full 600 V class requirement * 500 V differential range with 1 industry's best overshoot measurement capability (to 2000 Vpk) * Industry's best offset capability (1500 V) when used with HDO Series oscilloscopes * * Bandwidth rating up to 120 MHz * meter cable model available 6 (HVD3106A-6M) odels available without accessories M for a lower cost (HVD310xA-NOACC) Sinusoidal AC Line Input DC Bus Voltage Vpk-pk Vrms or Vac Line-Line 400 480 600 690 1131 1358 1697 1952 LineNeutral 653 784 980 1127 Vdc 566 679 849 976 Drive / Inverter Pulse-Width Modulated (PWM) Output Vpk (Rated) Line-Line 566 679 849 976 LineNeutral 327 392 490 563 Vpk (with Overshoot)* Line-Line 849 1019 1274 1464 LineNeutral 491 588 735 845 *Based on 50% overshoot condition 5 kV Class, 3-phase Electrical Apparatus The HVD3605A probe is safety-rated for 6000 Vrms and 8485 V(DC + peak AC) for full coverage of 5 kV class apparatus. The probe has ample 100 MHz of bandwidth, is standard with a 6 meter cable, is guaranteed to 1% gain accuracy, has excellent CMRR, and has the industry's widest differential voltage range. * nly probe that permits AC Line, O DC Bus, and Drive/Inverter output voltage probing through 4160 V apparatus ratings * Industry's best overshoot measurement capability (to 7600 Vpk) * Industry's best offset capability (6000 V) when used with HDO Series oscilloscopes * Standard 6 meter cable Sinusoidal AC Line Input DC Bus Voltage Vpk-pk Vrms or Vac Line-Line 2400 3300 4160 6788 9334 11766 LineNeutral 3920 5388 6794 Vdc 3395 4666 5884 Drive / Inverter Pulse-Width Modulated (PWM) Output Vpk (Rated) Line-Line 3395 4666 5884 LineNeutral 1960 2694 3397 *Based on 25% overshoot condition 2 Vpk (with Overshoot)* Line-Line 4244 5833 7355 LineNeutral 2450 3368 4246 WORLD'S ONLY 2 KV SAFETY RATED MODEL Solar PV Inverters The HVD3206A probe is rated to 1500 Vdc (2000 V DC + peak AC). This makes the probe ideal for testing single-phase or three-phase inverters or newer 1500 Vdc rated string-inverters (per IEC/EN 61010-031:2015). The HVD3206A has the same excellent 1% gain accuracy, industry-leading CMRR, industry-best offset range, standard 2 meter cable (6 meter cable model available), and a guaranteed 2000 Vpk differential voltage range. Solar String DC Bus Voltage Sinusoidal AC Output Vpk (Rated) LineLine-Line Neutral 849 490 849 490 Vdc 1000 1500 * Only 1500 Vdc safety rated probe per * uaranteed 2000 Vpk differential G voltage range * I ndustry's best offset capability (1500 V) when used with HDO Series oscilloscopes * * Bandwidth rating up to 120 MHz Vpk (with Overshoot)* Line-Line 1104 1104 IEC/EN 61010-031:2015 meter cable model available 6 (HVD3206A-6M) LineNeutral 637 637 *Based on 30% overshoot condition DC-DC Converters High-power DC-DC converters can operate at substantial voltages, 500 Vdc or higher. The HVD310xA models provide up to 1000 Vdc common-mode (HVD3206A models provide up to 1500 Vdc) and high precision (1% gain accuracy) DC voltage measurements. Automatic switchable attenuation keeps the probe in the optimum measurement range. Multiple probes can be used to understand complex device switching performance. * ommon mode range up to 2000 V C (DC + peak AC) with HVD3206A * * High precision - 1% accuracy utomatic gain switching for optimum A performance 3 SUPERIOR PERFORMANCE AT LOWER VOLTAGES 120 V / 240 V Half-Bridge Circuit Topologies Single-phase switch-mode power supplies and other devices utilizing half-bridge topologies need test and validation at DC bus voltages up to 340 Vdc with up to 680 Vp-p on the input voltage. HVD310xA probes are cost-effective solutions for probing the wide range of high voltage signals present in these systems. "No Accessory (NOACC)" versions of these probes permit additional cost savings. Use your existing accessories or purchase just the ones you want. * * High precision - 1% accuracy utomatic gain switching for optimum A performance Input Voltage (Vrms or Vac) Input Voltage (Vpk-pk) DC Bus Voltage (Vdc) Output PWM Voltage (Vpk) 120 240 340 680 170 340 170 340 MOSFET/IGBT Devices Upper-side Gate Drive Voltage Measurements Typically from 3 to 20 volts and "floating" at up to the DC Bus voltage. HVD-A Series probes can provide very good results, but the HVFO High Voltage Fiber Optic probe is optimized for the best upper-side gate drive voltage measurements with 140 dB CMRR, reduced DUT loading, and better pulse response. HVFO103 High Voltage Fiber Optically-Isolated Probe Device Analysis Conduction loss or Rds(on) measurements require a voltage probing solution that has high CMRR, fast overdrive recovery, voltage clamping (so the oscilloscope is not overdriven), compensation flatness, gain/amplification to see small signal details, and precise offset generation to see the switching device's turn-off performance. For this type of analysis, the Teledyne LeCroy DA1855A Differential Amplifier is the best solution. It provides 100 MHz bandwidth, excellent common mode rejection ratio (CMRR) of 100,000:1 (typical), and overdrive recovery to within 100 mV from a 400 V input signal in less than 100 ns. DA1855A Differential Amplifier 4 UNDERSTANDING HV DIFFERENTIAL PROBE RATINGS AC line voltages and power electronics systems can operate at very high voltages, and all parts of the measurement circuit are not necessarily connected to ground, requiring HV safety ratings. Additionally, a variety of terms may be used to describe the same voltage. Below is a simple tutorial to help you understand the various types of voltage terms and how they relate to each other. Understanding these voltage terms is necessary to ensure the correct probe selection. Single-phase AC Line Input Ratings are provided in Vrms (also referred to as Vac) referred Line-Neutral. AC Line inputs are usually grounded and these voltages can be assumed to have a 0V reference voltage. Vpeak is calculated as 2*Vrms and Vpk-pk is calculated as 2*Vpeak. The example to the right is for a 120 Vrms rating. Three-phase AC Line Input Ratings are provided in Vrms (also referred to as Vac) referred Line-Line (L-L), which can be converted in magnitude to a Line-Neutral basis by dividing by 3. Vpeak (L-L) is calculated 2*Vrms and Vpk-pk is calculated as 2*Vpeak. The example on the left is for a 480 Vrms three-phase rated system with signals shown as Line-Line. Rectified AC (DC Bus/Link Voltage) For three-phase inputs, the three Line-Neutral (L-N) AC Line inputs are rectified and summed to provide a "stiff" DC (bus/link) voltage for the input to an inverter subsection. The example shown to the right is for a 600 Vrms system. The Line-Neutral RMS voltages are 600V/3, and each Vpeak(L-N) is calculated as 2*Vrms. The DC bus voltage after rectification and filtering is equal to 3*Vpeak(L-N), or 847 Vdc in this case. Inverter or Drive H-Bridge or Cascaded H-Bridge Outputs The nominal peak voltage (+ or -) of the PWM signals without accounting for signal overshoot is equal to the DC bus voltage. The measured 0V differential voltage value is not ground-referenced, and therefore a differential probe suitably rated to the DC bus voltage is recommended. Note that the +Vpeak and -Vpeak plus overshoot safety margin (typically ~50%) is what the HV differential probe must measure - Vpk-pk (2*Vpeak) will never occur in the circuit. 5 SPECIFICATIONS HVD3102A HVD3106A Bandwidth 25 MHz Rise Time (10-90) 14 ns Differential Voltage 1500 V (DC + peak AC) from 7 to 500 V/ Range (High Attenuation) div with up to 1500 V offset. 1750 V maximum typical measurable differential voltage before saturation. HVD3106A-6M 120 MHz 80 MHz 2.9 ns 4.4 ns 1500 V (DC + peak AC) from 7 to 500 V/div with up to 1500 V offset. 2000 V maximum typical measurable differential voltage before saturation Differential Voltage Range (Low Attenuation) 27.6 V (DC + peak AC) from 100 mV/div to 6.9 V/div with up to 150 V offset Common Mode Voltage Range Maximum Input Voltage to Earth Max Safe Input Voltage 1500 V (DC + peak AC), 1000 Vrms Sensitivity Gain Accuracy Slew Rate Attenuation Input Impedance 1500 Vpk (nominal, either input to ground) 1000 Vrms CAT III 100 mV/div to 6.9V/div (100X) 7V/div to 500V/div (1000X) 100 mV/div to 100 mV/div to 6.9V/div (50X) 6.9V/div (50X) 7V/div to 500V/div (500X) 7V/div to 500V/div (500X) 1% (LF, guaranteed) 100 V/ns (maximum) 400 V/ns (maximum) 270 V/ns (maximum) 100x / 1000x 50x / 500x 50x / 500x 10 M || 2.5 pF (between inputs), 5 M || 5.0 pF (either input to ground) Input Coupling Output Coupling Output Termination Interface Input Lead Length Cable Length (input lead to oscilloscope connection) Noise and Rejection CMRR (Typical) Noise (Probe) Environmental Temperature (Operating) Temperature (Non-Operating) Humidity (Operating) Humidity (Non-Operating) Altitude (Operating) Altitude (Non-Operating) Pollution Degree Certifications CE (LVD Directive 2006/95/EC) CE (EMC Directive 2004/108/EC) UL Listed cUL Listed 6 2.25 m DC only AC or DC coupling 1 M ProBus 40 cm input lead length 6.8 m DC - 60 Hz: 85 dB 1 MHz: 65 dB 5 MHz: 40 dB 20 MHz: 30 dB DC - 60 Hz: 85 dB 1 MHz: 65 dB 5 MHz: 40 dB 20 MHz: 30 dB 100 MHz: 30 dB DC - 60 Hz: 85 dB 1 MHz: 65 dB 5 MHz: 40 dB 20 MHz: 30 dB 80 MHz: 30 dB 100X: <15 mVrms 1000X: <85 mVrms (referred to input) 50X: <30 mVrms 500X: <150 mVrms (referred to input) 50X: <30 mVrms 500X: <150 mVrms (referred to input) 0C to 50C -40C to 70C 5% to 80% RH (Non-Condensing) up to 30C, decreasing linearly to 45% RH at 50C 5% to 95% RH (Non-Condensing), 75% RH above 30C, 45% RH above 40C 3000 m maximum When used with clip accessories, 2000 m maximum 10,000 m 2, Indoor use only IEC/EN 61010-031:2015 IEC/EN 61326-1:2013 UL 61010-031 (Second Edition) CAN/CSA-C22.2 No. 61010-031-15 SPECIFICATIONS HVD3206A HVD3206A-6M HVD3605A Bandwidth 120 MHz Rise Time (10-90) 2.9 ns Differential Voltage 2000 V (DC + peak AC) from 7 to 500 Range (High Attenuation) V/div with up to 1500V offset. 80 MHz 100 MHz 4.4 ns 4.3 ns 2000 V (DC + peak AC) from 7 to 500 7000 V (DC + peak AC) from 28V/div to V/div 2000 V/div with up to 1500V offset. with up to 6000V offset. (7600 V maximum measurable differential voltage before saturation) Differential Voltage 27.6 V (DC + peak AC) from 100 mV/div 27.6 V (DC + peak AC) from 100 mV/div 700 V (DC + peak AC) from 300mV/div Range (Low Attenuation) to 6.9 V/div to 6.9 V/div to 27.5 V/div with up to 150V offset. with up to 150V offset. with up to 600V offset. Common Mode Voltage 2000 V (DC + peak AC) 2000 V (DC + peak AC) 7600V (DC + peak AC), 6000 Vrms Range Maximum Input Voltage 2000 Vpk 2000 Vpk 7600 Vpk to Earth (nominal, either input to ground) (nominal, either input to ground) (nominal, either input to ground) Max Safe Input Voltage 2000 V (DC + peak AC) CAT I * 2000 V (DC + peak AC) CAT I * 8485 V (DC + peak AC) CAT I * 1500 Vdc CAT III 1500 Vdc CAT III 6000 Vrms CAT I * 1000 Vrms CAT III 1000 Vrms CAT III 1500 Vdc CAT III 1000 Vrms CAT III Sensitivity 100 mV/div to 6.9V/div (50X) 100 mV/div to 6.9V/div (50X) 300 mV/div to 27.5 V/div (200X) 7V/div to 500V/div (500X) 7V/div to 500V/div (500X) 28 V/div to 2000 V/div (2000X) Gain Accuracy Slew Rate Attenuation Input Impedance Input Coupling Output Coupling Output Termination Interface Input Lead Length Cable Length (input lead to oscilloscope connection) Noise and Rejection CMRR (Typical) Noise (Probe) Environmental Temperature (Operating) Temperature (Non-Operating) Humidity (Operating) Humidity (Non-Operating) Altitude (Operating) Altitude (Non-Operating) Pollution Degree Certifications CE (LVD Directive 2006/95/EC) CE (EMC Directive 2004/108/EC) UL Listed cUL Listed 400 V/ns (maximum) 50x / 500x 10 M || 2.5 pF (between inputs) 5 M || 5.0 pF (either input to ground) 2.25 m 1% (LF, guaranteed) 270 V/ns (maximum) 1000 V/ns (typical) 50x / 500x 200x / 2000x 10 M || 2.5 pF (between inputs) 48 M || 2.5 pF (between inputs) 5 M || 5.0 pF (either input to ground) 24 M || 5.0 pF (either input to ground) DC only AC or DC coupling 1 M ProBus 40 cm input lead length 6.8 m 6.8 m DC - 60 Hz: 85 dB 1 MHz: 65 dB 5 MHz: 40 dB 20 MHz: 30 dB 100 MHz: 30 dB DC - 60 Hz: 85 dB 1 MHz: 65 dB 5 MHz: 40 dB 20 MHz: 30 dB 100 MHz: 30 dB 50X: <30 mVrms 500X: <150 mVrms (referred to input) 50X: <30 mVrms 500X: <150 mVrms (referred to input) DC - 60 Hz: 85 dB 10 kHz: 70 dB 1 MHz: 64 dB (200x) 1 MHz: 50 dB (2000x) 10 MHz: 40 dB (200x) 10 MHz: 30 dB (2000x) 100 MHz: 30 dB 200x: <65 mVrms 2000x: <320 mVrms (typical, referred to input) 0C to 50C -40C to 70C 5% to 80% RH (Non-Condensing) up to 30C, decreasing linearly to 45% RH at 50C 5% to 95% RH (Non-Condensing), 75% RH above 30C, 45% RH above 40C 3000 m maximum When used with clip accessories, 2000 m maximum 10,000 m 2, Indoor use only IEC/EN 61010-031:2015 IEC/EN 61326-1:2013 UL 61010-031 (Second Edition) CAN/CSA-C22.2 No. 61010-031-15 * CAT I per IEC/EN 61010-031/A1:2008. No Rated Measurement Category per IEC/EN 61010-031:2015. 7 ORDERING INFORMATION Product Description Product Code 1 kV, 25 MHz High Voltage Differential Probe with 2 m cable 1 kV, 120 MHz High Voltage Differential Probe with 2 m cable 1 kV, 80 MHz High Voltage Differential Probe with 6 m cable 1 kV, 25 MHz High Voltage Differential Probe with 2 m cable without tip Accessories 1 kV, 120 MHz High Voltage Differential Probe with 2 m cable without tip Accessories 2 kV, 120 MHz High Voltage Differential Probe with 2 m cable 2 kV, 80 MHz High Voltage Differential Probe with 6 m cable 6 kV, 100 MHz High Voltage Differential Probe with 6 m cable High Voltage Replacement Accessories Kit (Includes 2 each, 1 Black, 1 Red): Safety Alligator Clips, Plunger Pincer Clips, Plunger Hook Clips, Plunger Alligator Clips, Spade Terminals HVD3102A HVD3106A HVD3106A-6M HVD3102A-NOACC HVD3106A-NOACC HVD3206A HVD3206A-6M HVD3605A PK-HV-001 HVD310xA-NOACC HVD310X-NOACC High Voltage Differential Probe PK-HVA-01 PK-HVA-02 PK-HVA-03 PK-HVA-04 PK-HVA-06 PK-HVA-05 HVD-A Probe Accessories PK-HVA-01 Safety Alligator Clips (Red/Black) 1000 V CAT III PK-HVA-07 HVD310xA / 310xA-6M HVD310X/310X-6M PK-HVA-02 Plunger Pincer Clips (Red/Black) 1000 V CAT II High Voltage Differential Probe PK-HVA-01 PK-HVA-02 PK-HVA-03 PK-HVA-04 PK-HVA-06 PK-HVA-05 PK-HVA-03 Plunger Hook Clips (Red/Black) 1000 V CAT II PK-HVA-07 PK-HVA-04 Plunger Alligator Clips (Red/Black) 1000 V CAT III HVD3206A / 3206A-6M HVD3206/3206-6M High Voltage Differential Probe PK-HVA-01 PK-HVA-02 PK-HVA-03 PK-HVA-04 PK-HVA-06 PK-HVA-05 PK-HVA-07 PK-HVA-06 6 KV Safety Alligator Clips (Red/Black) 6000 V CAT I HVD3605A HVD3605 High Voltage Differential Probe PK-HVA-01 PK-HVA-02 PK-HVA-03 PK-HVA-04 PK-HVA-06 PK-HVA-05 PK-HVA-05 Spade Terminals (Red/Black) 1000 V CAT III PK-HVA-07 2 kV Plunger Alligator Clips (Red/Black) 2000 V CAT I PK-HVA-07 Customer Service Teledyne LeCroy oscilloscopes and probes are designed, built, and tested to ensure high reliability. In the unlikely event you experience difficulties, our digital oscilloscopes are fully warranted for three years and our probes are warranted for one year. This warranty includes: No charge for return shipping Long-term 7-year support Upgrade to latest software at no charge * * * 1-800-5-LeCroy teledynelecroy.com Local sales offices are located throughout the world. Visit our website to find the most convenient location. (c) 2018 by Teledyne LeCroy, Inc. All rights reserved. Specifications, prices, availability, and delivery subject to change without notice. Product or brand names are trademarks or requested trademarks of their respective holders. hvd3000a-probes-ds-14mar18