6WAC-DC Single Output MedicalAdaptor GSM06U series
Description
GSM06U is a highly reliable, 6W single-output green medical adaptor series. This product is equipped
with a 2-pin (no FG) standard US AC power plug, adopting the input range from 85VAC to 264VAC. The
entire series supplies different output voltages between 5VDC and 24VDC that can satisfy the demands
for various kinds of miniature medical devices. The circuitry design meets the international medical standards
(2 x MOPP), having an ultra low leakage current (<50 A), fitting the medical devices in direct electrical
contact with the patients.
With the efficiency up to 82% and the extreme low no-load power consumption below 0.3W, the design of
GSM06U observes the latest energy regulation (level ); the supreme feature allows the adaptor to save
the energy when it is either under the operating mode or the standby mode. The entire series utilizes the
94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock.
GSM06U is approved with the international medical safety certificates.
Universal AC input / Full range
2 pole USA plug
Low leakage current <50 A
Protections: Short circuit / Overload / Over voltage
Fully enclosed plastic case
Medical safety approved (2 x MOPP between primary to secondary)
Class power (without earth pin)
No load power consumption <0.3W
ErP step2 compliant (level )
Meet EISA 2007 (Energy Independence and Security Act)
Optional lock type DC plug
3 years warranty
Blood glucose meter
Blood pressure meter
Nebulizer
Inhaler
Portable medical device
Features Applications
μ
μ
File Name:GSM06U-SPEC 2014-07-28
Model Encoding
-U05GSM P1J06
Series name
Output wattage
US 2 pin AC inlet
Output voltage
DC plug type
{
Standard : P1J (without lock)
Optional : P2S (lock type)
SPECIFICATION
6WAC-DC Single Output MedicalAdaptor GSM06U series
File Name:GSM06U-SPEC 2014-07-28
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE Note.2
RATED CURRENT
CURRENT RANGE
RATED POWER
RIPPLE & NOISE (max.) Note.3
OUTPUT
SETUP, RISE, HOLD UP TIME
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY (Typ.)
AC CURRENT
INPUT
ENVIRONMENT
INRUSH CURRENT (max.)
LEAKAGE CURRENT(max.)
SAFETY STANDARDS
SAFETY &
EMC
(Note. 8)
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
PROTECTION
VOLTAGE TOLERANCE Note.4
LINE REGULATION Note.5
LOAD REGULATION Note.6
1000ms, 50ms, 12ms at full load
85 ~ 264VAC 120 ~ 370VDC
47 ~ 63Hz
68% 74% 74% 76% 77% 79% 80% 82%
0.18A / 100VAC
30A / 230VAC
IEC60950-1, ANSI/AAMI ES60601-1-11 approved
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P:5656VDC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
500Khrs min. MIL-HDBK-217F(25 )
0 ~ +40 (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-20 ~ +85 , 10 ~ 95% RH
±℃0.04% / (0 ~ 40 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
32*66*42.5mm (L*W*H)
100g ; 110pcs / 12Kg / CARTON
CONNECTOR
PLUG
CABLE
Standard type P1J: 2.1 * 5.5 * 11mm, turning fork type center positive for stock ; Other type available by customer requestedψψ
See page3 ; Other type available by customer requested
1.All parameters are specified at 115VAC input, rated load, 25 70% RH ambient.
2.DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4.Tolerance: includes set up tolerance, line regulation, load regulation.
5.Line regulation is measured from low line to high line at rated load.
6.Load regulation is measured from 0% to 100% rated load
7.The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives.
Compliance to FCC part 15 Class B
OVERLOAD
OVER VOLTAGE
>105% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
>120% rated output voltage
Protection type : Clamp by zener diode
EMC EMISSION
5V 7.5V6V 9V 12V 15V 18V 24V
GSM06U05 GSM06U06 GSM06U07 GSM06U09 GSM06U12 GSM06U15 GSM06U18 GSM06U24
1.2A 0.8A1.0A 0.66A 0.5A 0.4A 0.33A 0.25A
0 ~ 1.2A 0 ~ 0.8A0 ~ 1.0A 0 ~ 0.66A 0 ~ 0.5A 0 ~ 0.4A 0 ~ 0.33A 0 ~ 0.25A
6W 6W6W 6W 6W 6W 6W 6W
50mVp-p 80mVp-p50mVp-p 80mVp-p 100mVp-p 120mVp-p 150mVp-p 180mVp-p
±5.0%
±5.0%
±5.0%±5.0%
±5.0%±5.0%
±5.0%
±5.0%
±5.0% ±5.0% ±5.0%
±0.5% ±0.5%±0.5% ±0.5% ±0.5% ±0.5% ±0.5% ±0.5%
±4.0%
GSM06U05-P1J GSM06U06-P1J GSM06U07-P1J GSM06U09-P1J GSM06U12-P1J GSM06U15-P1J GSM06U18-P1J GSM06U24-P1J
Touch current < 50 A/264VACμ
±3.0% ±3.0% ±3.0% ±2.0%
Derating Curve
85 100 110 135 155 264230
100
90
80
70
60
50
40
INPUT VOLTAGE (V) 60Hz
LOAD(%)
Ta ( )
LOAD (%)
0102030 5040
70
100
Static Characteristics
Mechanical Specification Plug Assignment
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
NAME PLATE
12.7
18.2
6.3
24.2
1.5
17.1 0.5±
48 1±
32 1±
66 1±
Unit:mm
Plug Assignment
Standard plug: P1J
Optional lock type plug: P2S
42.5 1±
SWITCHCRAFT S761K plug equivalent
P/N OUTPUT
P1J
CENTER +
UL2468 22AWG 1220 +150/-0mm for 5~9V
UL2468 24AWG 1830 +150/-0mm for 12~24V
ID 2.1 x OD 5.5
11
C"+"
6WAC-DC Single Output MedicalAdaptor GSM06U series
File Name:GSM06U-SPEC 2014-07-28
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Mean Well:
GSM06U06-P1J GSM06U05-P1J GSM06U12-P1J GSM06U15-P1J GSM06U07-P1J GSM06U18-P1J GSM06U09-
P1J GSM06U24-P1J