FEATURES
• High accuracy linear output
• Externally triggered automatic or manual calibration
• Can be calibrated in fresh air (20.7% O2) or to any
other known O2 concentration
• Selectable output filtering allows adaptive, fast and
dynamic or slow and stable output
OUTPUT VALUES
Oxygen range (analogue output)2 0.11—25% O2
or 0.11—100% O2
Oxygen range (RS232 output) 0.11 and 100% O2
Accuracy after calibration3, 4 1% O2
Repeatability after calibration3 0.5% O2
Output resolution
Analogue 4—20mA 0.01mA
Analogue 0—10VDC 0.01V
Digital RS232 0.01% O2
Response time < 15s
Warm up time (prior to sensor operation) 60s
Output stabilisation time ~ 180s
Need help? Ask the expert
Tel: + 44 (0)1236 459 020
and ask for “Technical”
BENEFITS
• Cycling 3.3VDC logic output allows direct monitoring of the
O2 sensor pump cycle for diagnostic purposes
• No reference gas required
TECHNICAL SPECIFICATIONS
Supply voltage 24VDC ± 10%
Supply current 500mA max. at 24VDC
Digital output RS232
Analogue output 4—20mA; load 600Ω max.
or 0—10VDC; load 10kΩ min.
Housing temperature limits
Storage: -10°C to +85°C
Operating: -10°C to +85°C
Permissible gas temperatures (probe tip)
Standard: -100°C to +250°C
High: -100°C to +400°C
Gas flow rate 0 to 10 m/s
Permissible acceleration
Repetitive 5g
Incidental 30g
1) Prolonged operation below 0.1% O2 can damage the sensing element.
2) Range selectable by altering the position of the jumper links on the PCB; refer to PCB Layout on page 3.
3) Assuming barometric pressure (BP) remains constant.
4) As the O2 sensor measures the partial pressure of oxygen (PPO2) within the measurement gas deviations in the BP from
that present during calibration will cause readout errors proportional to the change. EG. if the sensor reads 21% O2 at
1013.25mbar and the BP increases by 1%, the sensor readout will also increase by 1% to 21.21% O2.
NOTES
DATA SHEET
Zirconia O2 Sensors
OXY-Flex Oxygen Analyser
Supply
Voltage
Housing Gas Temp Digital
Output
Response
Time
Analogue
Output