1586A Super-DAQ Precision Temperature Scanner Extended Specifications The 1586A Super-DAQ is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, dc voltage, dc current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab. * Measure thermocouples, PRTs, thermistors, dc voltage, dc current, and resistance * Best-in-class temperature measurement accuracy: - PRTs: 0.005 C (using external DAQ-STAQ Multiplexer) - Thermocouples: 0.5 C (using High-Capacity Module and internal CJC) - Thermistors: 0.002 C * Input Channels: Up to 40 isolated universal inputs * Flexible configuration: Internal HighCapacity Module and/or DAQ-STAQ Multiplexer * Selectable scan speed: Up to 10 channels per second * Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM) * Real-time color trending: Chart up to four channels simultaneously * Automated sensor calibration: Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines * Data storage: Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive. Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft (R) Excel * Data security: Administrator and user profiles for protecting settings and ensuring test traceability * Mx + B scaling and channel offset zero function * Alarms: Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded 1586A Super-DAQ at a glance Back lit main function keys-- always know the mode of operation and recording status. Color display with channel status indicators. Chart up to 4 channels simultaneously. Front panel Memory management-- for downloading and moving data and setup files. Five-way binding post for reference thermometers. Also measures dc voltage, dc current, and resistance. Measure/DMM function for quick measurements and recording with minimal setup. USB support for easy transfer of data and setup files. Function soft keys support fast setup and operation. Standby key turns the display and keypad off-- eliminates warm-up time. Navigation keys for fast selection of information. Back panel Line voltage and fuse selector. Two rear panel slots can be configured for the internal High-Capacity Module or the external DAQ-STAQ Multiplexer--or one of each. Power switch and line cord connector. Dedicated connector for control of Fluke Calibration dry-wells or temperature baths for automated tests. USB and Ethernet ports used for remote operation. 2 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner Digital external alarm trigger outputs. Totalizer and Digital I/O input ports. 1586A modules at a glance 1586-2586 High-Capacity Module Universal input channels designed to measure temperature, resistance, dc voltage, and dc current for data acquisition applications on the factory floor where channel count and scan speed are important. Clear top case with embedded channel numbers for easy wiring identification. Two dedicated dc current channels. Electronic ID with serial number and calibration date reports to the 1586A Super-DAQ. Twenty 2-wire or ten 3-, or 4-wire connections. Built-in cable strain relief to help secure connections. 1586-2588 DAQ-STAQ Multiplexer Designed for high-accuracy measurements in secondary temperature calibration labs. Easily connect/disconnect thermocouples, PRTs, and thermistors. Twenty mini-jack thermocouple inputs and up to ten four-wire, gold-plated PRT/thermistor connectors. Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 3 1586A Super-DAQ general specifications General Maximum input 50 V Offset voltage <2 V 3-wire internal resistance mismatch <50 m Basic CJC accuracy 0.25 C Mains Voltage 100 V Setting 90 V to 110 V 120 V Setting 108 V to 132 V 220 V Setting 198 V to 242 V 240 V Setting 216 V to 264 V Frequency 47 Hz to 440 Hz Power Consumption 36 VA peak (24 W average) Environment Temperature Operating 0 C to 50 C Full accuracy 18 C to 28 C Storage -20 C to 70 C Warm-up 1 hour to full accuracy specifications Relative Humidity (non-condensing) Operating 0 C to 30 C <80 % 30 C to 50 C <50 % Storage -20 C to 70 C <95 % Operating 2,000 m Storage 12,000 m Altitude Vibration and Shock Channel Capacity Input Protection Math Channels Triggers Memory USB Host Port USB Device Port 4 Fluke Calibration Complies with MIL-PRF-28800F Class 3 Total analog channels 45 Voltage/resistance channels 41 Current channels 5 Digital I/O 8 bits Totalizer 1 Alarm outputs 6 Trigger input 1 50 V all functions, terminals and ranges Number of channels 20 Operations sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum Interval, external (trigger input), alarm, remote (bus), manual, automated test Scan data RAM 75,000 readings with timestamp Data/Setup flash memory 20 MB Connector type Type A Function Memory File system FAT32 Memory capacity 32 GB Connector type Type B Class Instrument Function Control and data transfer Command protocol SCPI 1586A Super-DAQ Precision Temperature Scanner LAN RS-232 Dimensions Conformity Function Control and data transfer Network protocols Ethernet 10/100, TCP/IP Command protocol SCPI Connector D-sub 9 pin (DE-9) Baud rates 1200, 2400, 4800, 9600, 19200, 38400 Function Temperature source control output Height 150 mm Width 245 mm Depth 385 mm Weight 6 kg (typical configuration) Shipping weight 9.5 kg (typical configuration) CE, CSA, IEC 61010 3rd edition 1586-2586 High-Capacity Module specifications General Maximum input 50 V Offset voltage <2 V 3-wire internal resistance mismatch <50 m Basic CJC accuracy 0.6 C 1586-2588 DAQ-STAQ Multiplexer specifications General Maximum input 50 V Offset voltage <2 V 3-wire internal resistance mismatch <50 m Basic CJC accuracy 0.25 C Measurement specifications Accuracy specifications generally apply with medium and slow sample rates (unless otherwise noted), after a warm-up time of 1 hour, and within an environment temperature range of 18 C to 28 C, and may depend on the channel. The confidence level for accuracy specifications is 95 % within one year of calibration. Scan rate Display resolution Fast 10 channels per second max (0.1 seconds per channel) Medium 1 channel per second (1 second per channel) Slow 4 seconds per channel 4 1/2 to 6 1/2 digits, depending on function and Sample Rate (see Measurement Characteristics tables below to find the display resolution of temperature readings) PRT/RTD Temperature range -200 C to 1200 C (depending on the sensor) Resistance range 0 to 4 k Offset compensation 0 to 400 , 4-wire automatic current reversal 400 to 4000 or 3-wire none Fast sample rate 2 ms Medium sample rate 250 ms Slow sample rate 250 ms Source current reversal interval (0 to 400 range) Maximum lead resistance (4-wire ) 2.5 % of range per lead for 400 and 4 k ranges. Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 5 PRT/RTD resistance accuracy Accuracy is given as % of measurement or ohms, whichever is greater. Basic accuracy is for 4-wire PRT/RTD. When using 3-wire PRT/RTD add 0.013 to the accuracy specification for internal resistance mismatch and voltage offset if using Channel 1, or add 0.05 if using channels x01 through x20. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. Range 0 to 400 400 to 4 k Sample Rate DAQ-STAQ Multiplexer and Channel 1 High-Capacity Module T.C./ C Outside 18 C to 28 C Slow 0.002 % or 0.0008 0.003 % or 0.003 0.0001 % or 0.0008 Medium 0.002 % or 0.002 0.003 % or 0.003 0.0001 % or 0.0008 Fast 0.002 % or 0.005 0.003 % or 0.006 0.0001 % or 0.0008 Slow 0.004 % or 0.06 0.006 % or 0.06 0.0001 % or 0.008 Medium 0.004 % or 0.1 0.006 % or 0.1 0.0001 % or 0.008 Fast 0.004 % or 0.18 0.006 % or 0.18 0.0001 % or 0.008 PRT/RTD temperature accuracy Accuracy is for 4-wire 100 nominal PRT/RTD. When using 3-wire PRT/RTD add 0.039 C to the accuracy specification for internal resistance mismatch and voltage offset if using Channel 1, or add 0.15 C if using channels x01 through x20. If the environment temperature is outside the specified range, multiply the temperature coefficient number by the temperature deviation and add to the accuracy specification. Linear interpolation may be used between points in the table. Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor and characterization. Sample Rate Temperature DAQ-STAQ Multiplexer and Channel 1 High-Capacity Module T.C./ C Outside 18 C to 28 C Slow -200 C 0 C 300 C 600 C 0.002 C 0.005 C 0.012 C 0.02 C 0.008 C 0.008 C 0.018 C 0.03 C 0.002 C 0.003 C 0.006 C 0.01 C Medium -200 C 0 C 300 C 600 C 0.005 C 0.005 C 0.012 C 0.02 C 0.008 C 0.008 C 0.018 C 0.03 C 0.002 C 0.003 C 0.006 C 0.01 C Fast -200 C 0 C 300 C 600 C 0.013 C 0.013 C 0.014 C 0.02 C 0.015 C 0.015 C 0.018 C 0.03 C 0.002 C 0.003 C 0.006 C 0.01 C PRT/RTD measurement characteristics Range Temperature Display Resolution Fast Sample Rate 0 to 400 0.001 C 0.01 C 1 mA 400 to 4 k 0.001 C 0.01 C 0.1 mA Thermistor 6 Source Current Slow / Medium Sample Rate Temperature range -200 C to 400 C (depending on the sensor) Resistance range 0 to 1 M Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner Thermistor resistance accuracy Accuracy is given as (% of measurement + ). The basic accuracy specification is for 4-wire thermistor, slow sample rate. When using medium or fast sample rate, add the number given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. For 2-wire thermistor add 0.02 internal resistance if using Channel 1 or 1.5 if using channels x01 through x20, and add external lead wire resistance. Range Slow Sample Rate Medium Sample Rate Fast Sample Rate T.C./ C Outside 18 C to 28 C 0 to 2.2 k 0.004 % + 0.2 add 0.3 add 1 0.0005 % + 0.05 2.1 k to 98 k 0.004 % + 0.5 add 0.5 add 1.3 0.0005 % + 0.1 95 k to 1 M 0.015 % + 5 add 5 add 13 0.001 % + 2 Thermistor temperature accuracy Accuracy specifications are for 4-wire thermistor. When using 2-wire thermistor, add the number given in the table to the specification for internal resistance. If the environment temperature is outside the specified range, increase the accuracy specification by 25 % for every 1 C outside the specified environment temperature range. Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor. Range -40 C Accuracy 2.2 k Thermistor Slow Sample Rate Medium Sample Rate Fast Sample Rate 2-wire 0.001 C 0.001 C 0.01 C add 0.001 C 0 C 0.003 C 0.004 C 0.01 C add 0.004 C 25 C 0.006 C 0.011 C 0.02 C add 0.016 C 50 C 0.008 C 0.018 C 0.04 C add 0.05 C 100 C 0.047 C 0.114 C 0.28 C add 0.34 C 150 C 0.23 C 0.56 C 1.34 C add 1.7 C Range Accuracy 5 k Thermistor Slow Sample Rate Medium Sample Rate Fast Sample Rate 2-wire -40 C 0.003 C 0.004 C 0.01 C add 0.001 C 0 C 0.002 C 0.002 C 0.01 C add 0.002 C 25 C 0.004 C 0.006 C 0.01 C add 0.007 C 50 C 0.005 C 0.009 C 0.02 C add 0.022 C 100 C 0.022 C 0.052 C 0.13 C add 0.16 C 150 C 0.096 C 0.24 C 0.57 C add 0.7 C Range Accuracy 10 k Thermistor Slow Sample Rate Medium Sample Rate Fast Sample Rate 2-wire -40 C 0.003 C 0.004 C 0.01 C add 0.001 C 0 C 0.002 C 0.002 C 0.01 C add 0.002 C 25 C 0.003 C 0.004 C 0.01 C add 0.004 C 50 C 0.005 C 0.009 C 0.02 C add 0.011 C 100 C 0.011 C 0.024 C 0.06 C add 0.067 C 150 C 0.04 C 0.098 C 0.24 C add 0.29 C Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 7 Thermistor measurement characteristics Range Temperature Display Resolution Source Current Slow / Medium Sample Rate Fast Sample Rate 0 to 2.2 k 0.0001 C 0.001 C 10 A 2.1 k to 98 k 0.0001 C 0.001 C 10 A 95 k to 1 M 0.0001 C 0.001 C 1 A Thermocouple Temperature range -200 C to 2315 C (depending on the sensor) Voltage range -15 mV to 100 mV Thermocouple voltage accuracy Accuracy is given as (|% of measurement| + V). Basic accuracy specification is for medium or slow sample rate. When using a fast sample rate add the number given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. Range Accuracy Channel 1 Ch. x01 - x20 Fast Sample Rate T.C./ C Outside 18 C to 28 C -15 mV to 100 mV 0.004 % + 4 V add 2 V add 1 V 0.0005 % + 0.0005 mV Thermocouple reference junction accuracy 8 Module CJC Accuracy T.C./ C Outside 18 C to 28 C DAQ-STAQ Multiplexer 0.25 C 0.02 C High-Capacity Module 0.6 C 0.05 C Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner Thermocouple temperature accuracy Accuracy specifications apply using medium or slow sample rate. When using fast sample rate, increase the accuracy specification by 25 %. If the environment temperature is outside the specified range, increase the accuracy specification by 12 % for every 1 C outside the specified environment temperature range. Accuracy with fixed/external CJC does not include the accuracy of the reference junction temperature. Linear interpolation may be used between points in the table. Specifications do not include sensor accuracy. The practical range of temperature measurement depends on the sensor. Type (Range) Temperature Accuracy Fixed / External CJC Internal CJC Channel 1 Ch. x01 - x20 DAQ-STAQ Multiplexer High-Capacity Module K -270 C to 1372 C -200 C 0 C 1000 C 0.28 C 0.10 C 0.14 C 0.41 C 0.15 C 0.20 C 0.76 C 0.29 C 0.32 C 1.60 C 0.62 C 0.64 C T -270 C to 400 C -200 C 0 C 200 C 400 C 0.27 C 0.10 C 0.08 C 0.08 C 0.40 C 0.15 C 0.12 C 0.11 C 0.76 C 0.30 C 0.23 C 0.20 C 1.60 C 0.65 C 0.47 C 0.41 C R -50 C to 1768 C 0 C 300 C 1200 C 1600 C 0.76 C 0.42 C 0.33 C 0.34 C 1.13 C 0.63 C 0.47 C 0.49 C 1.16 C 0.64 C 0.48 C 0.50 C 1.28 C 0.71 C 0.52 C 0.54 C S -50 C to 1768 C 0 C 300 C 1200 C 1600 C 0.74 C 0.45 C 0.37 C 0.39 C 1.11 C 0.67 C 0.54 C 0.56 C 1.14 C 0.68 C 0.55 C 0.57 C 1.26 C 0.76 C 0.60 C 0.63 C J -210 C to 1200 C -200 C 0 C 1000 C 0.20 C 0.08 C 0.11 C 0.29 C 0.12 C 0.14 C 0.65 C 0.28 C 0.25 C 1.41 C 0.61 C 0.53 C N -270 C to 1300 C -200 C 0 C 500 C 1000 C 0.42 C 0.15 C 0.12 C 0.14 C 0.62 C 0.23 C 0.17 C 0.19 C 0.90 C 0.34 C 0.24 C 0.26 C 1.69 C 0.64 C 0.44 C 0.45 C E -270 C to 1000 C -200 C 0 C 300 C 700 C 0.17 C 0.07 C 0.06 C 0.08 C 0.25 C 0.10 C 0.09 C 0.10 C 0.64 C 0.27 C 0.21 C 0.21 C 1.42 C 0.61 C 0.46 C 0.45 C B 100 C to 1820 C 300 C 600 C 1200 C 1600 C 1.32 C 0.68 C 0.41 C 0.38 C 1.97 C 1.02 C 0.60 C 0.55 C 1.97 C 1.02 C 0.60 C 0.55 C 1.97 C 1.02 C 0.60 C 0.55 C C 0 C to 2315 C 600 C 1200 C 2000 C 0.23 C 0.28 C 0.44 C 0.33 C 0.40 C 0.60 C 0.37 C 0.45 C 0.66 C 0.54 C 0.63 C 0.91 C D 0 C to 2315 C 600 C 1200 C 2000 C 0.22 C 0.26 C 0.39 C 0.32 C 0.36 C 0.53 C 0.34 C 0.39 C 0.56 C 0.44 C 0.49 C 0.69 C G 0 C to 2315 C 600 C 1200 C 2000 C 0.24 C 0.22 C 0.33 C 0.36 C 0.32 C 0.46 C 0.36 C 0.32 C 0.46 C 0.36 C 0.33 C 0.46 C L -200 C to 900 C -200 C 0 C 800 C 0.13 C 0.08 C 0.09 C 0.19 C 0.12 C 0.12 C 0.45 C 0.28 C 0.23 C 0.99 C 0.62 C 0.48 C M -50 C to 1410 C 0 C 500 C 1000 C 0.11 C 0.10 C 0.10 C 0.16 C 0.15 C 0.14 C 0.30 C 0.25 C 0.21 C 0.64 C 0.51 C 0.41 C U -200 C to 600 C -200 C 0 C 400 C 0.25 C 0.10 C 0.08 C 0.37 C 0.15 C 0.11 C 0.71 C 0.30 C 0.20 C 1.48 C 0.63 C 0.40 C W 0 C to 2315 C 600 C 1200 C 2000 C 0.24 C 0.22 C 0.33 C 0.36 C 0.32 C 0.46 C 0.36 C 0.32 C 0.46 C 0.36 C 0.33 C 0.46 C Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 9 Thermocouple measurement characteristics Range Temperature Display Resolution -270 C to 2315 C Slow / Medium Sample Rate Fast Sample Rate 0.01 C 0.1 C DC voltage Maximum input 50 V on any range Common mode rejection 140 dB at 50 Hz or 60 Hz (1 k unbalance in LOW lead) 50 V peak maximum Normal mode rejection 55 dB for power line frequency 0.1 %, 120 % of range peak maximum A/D linearity 2 ppm of measurement + 1 ppm of range Input bias current 30 pA at 25 C DC voltage accuracy Accuracy is given as (% measurement + % of range). Basic accuracy specification is for Channel 1, medium or slow sample rate. For channels x01 through x20 or when using Fast sample rate, add the numbers given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. Range Accuracy Channel 1 Ch. x01 - x20 Fast Sample Rate T.C./ C Outside 18 C to 28 C 100 mV 0.0037 % + 0.0035 % add 2 V add 0.0008 % of range 0.0005 % + 0.0005 % add 2 V 1 V 0.0025 % + 0.0007 % add 0.0008 % of range 0.0005 % + 0.0001 % 10 V 0.0024 % + 0.0005 % - add 0.0008 % of range 0.0005 % + 0.0001 % 50 V 0.0038 % + 0.0012 % - add 0.0008 % of range 0.0005 % + 0.0001 % DC voltage input characteristics Range 1 Resolution Input Impedance Slow / Medium Fast 100 mV 0.1 V 1 V 10 G 1 1 V 1 V 10 V 10 G 1 10 V 10 V 100 V 10 G 1 50 V 100 V 1 mV 10 M 1 % Input beyond 12 V is clamped. The clamp current is up to 3 mA. DC current Input protection 0.15 A resettable PTC DC current accuracy Accuracy is given as (% measurement + % of range). Basic accuracy specification is for medium or slow sample rate. When using a fast sample rate, add the number given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. 10 Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner Range Accuracy Fast Sample Rate T.C./ C Outside 18 C to 28 C 100 A 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 % 1 mA 0.015 % + 0.0011 % add 0.0008 % of range 0.002 % + 0.001 % 10 mA 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 % 100 mA 0.015 % + 0.0035 % add 0.0008 % of range 0.002 % + 0.001 % DC current input characteristics Range Resolution Burden Voltage Slow / Medium Fast 100 A 0.1 nA 1 nA <1 mV 1 mA 1 nA 10 nA <1 mV 10 mA 10 nA 100 nA <1 mV 100 mA 100 nA 1 A <1 mV Resistance Max. lead resistance (4-wire ohms) 10 per lead for 100 and 1 k ranges. 1 k per lead on all other ranges Resistance accuracy Accuracy is given as (% measurement + % of range). Basic accuracy specification is for 4-wire resistance, medium or slow sample rate. For 2-wire resistance add 0.02 internal resistance if using Channel 1, or 1.5 if using channels x01 through x20, and add external lead wire resistance. When using fast sample rate, add the numbers given in the table to the accuracy specification. If the environment temperature is outside the specified range, multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification. Range Accuracy Fast Sample Rate T.C./ C Outside 18 C to 28 C 100 0.004 % + 0.0035 % add 0.001 % of range 0.0001 % + 0.0005 % 1 k 0.003 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 % 10 k 0.004 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 % 100 k 0.004 % + 0.001 % add 0.001 % of range 0.0001 % + 0.0001 % 1 M 0.006 % + 0.001 % add 0.002 % of reading plus 0.0008 % of range 0.0005 % + 0.0002 % 10 M 0.015 % + 0.001 % add 0.002 % of reading plus 0.0008 % of range 0.001 % + 0.0004 % 100 M 0.8 % + 0.01 % add 0.001 % of range 0.05 % + 0.002 % Resistance input characteristics Range Resolution Slow / Medium Fast Source Current (open-circuit voltage) 100 0.1 m 1 m 1 mA (4 V) 1 k 1 m 10 m 1 mA (4 V) 10 k 10 m 100 m 100 A (6 V) 100 k 100 m 1 100 A (12 V) 1 M 1 10 10 A (12 V) 10 M 10 100 1 A (12 V) 100 M 100 1 k 0.1 A (12 V) Fluke Calibration 1586A Super-DAQ Precision Temperature Scanner 11 Ordering information 1586A Super-DAQ Precision Temperature Scanner Super-DAQ and DAQ-STAQ Multiplexer 1586A/1DS 1586A/1DS/C Super-DAQ, 1 DAQ-STAQ Multiplexer Super-DAQ, 1 DAQ-STAQ Multiplexer, Accredited Calibration Super-DAQ, 2 DAQ-STAQ Multiplexers Super-DAQ, 2 DAQ-STAQ Multiplexers, Accredited Calibration 1586A/2DS 1586A/2DS/C Super-DAQ and High-Capacity Module 1586A/1HC 1586A/1HC/C Super-DAQ, 1 High-Capacity Module Super-DAQ, 1 High-Capacity Module, Accredited Calibration Super-DAQ, 2 High-Capacity Modules Super-DAQ, 2 High-Capacity Modules, Accredited Calibration 1586A/2HC 1586A/2HC/C Super-DAQ, High-Capacity Module, and DAQ-STAQ Multiplexer 1586A/DS-HC 1586A/DSHC/C Super-DAQ, 1 High-Capacity Module, 1 DAQ-STAQ Multiplexer Super-DAQ, 1 High-Capacity Module, 1 DAQ-STAQ Multiplexer, Accredited Calibration Accessories 1586-2586 1586-2586-KIT 1586-2588 1586-2588-KIT 1586-2588-CBL Y1586S Y1586D 1586-CASE 1586/DS-CASE High-Capacity Module without Relay Card High-Capacity Module with Relay Card DAQ-STAQ Multiplexer without Adapter Card DAQ-STAQ Multiplexer, Adapter Card, Interface Cable DAQ-STAQ Multiplexer Interface Cable Rack Mount Kit, Single (half rack) Rack Mount Kit, Dual (full rack) Super-DAQ Carrying Case (mainframe and internal modules) Super-DAQ/DAQ-STAQ Carrying Case (mainframe and external module) Fluke Calibration. Precision, performance, confidence.TM Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (877) 355-3225 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.flukecal.com (c)2013 Fluke Calibration. Specifications subject to change without notice. Printed in U.S.A. 9/2013 6000790B_EN Pub-ID 12118-eng Modification of this document is not permitted without written permission from Fluke Calibration.