PERFORMANCE
OUTPUT SIGNAL 0.5 - 5.0 VDC, 0 - 5 VDC, 0.5 - 5.5 VDC,
1 - 5 VDC, 1 - 6 VDC, 0 - 10 VDC, 1 - 11 VDC
ZERO & SPAN OFFSETS ±0.25% FSO @ 21°C
ACCURACY L & R ±0.25% of FSO @ 21°C (BFSL)
LONG-TERM STABILITY ±0.25% FSO/Year
COMPENSATED TEMP. -10° to +85°C (-22° to +185°F)
TEMPERATURE EFFECT ±1.0% FSO Over the Compensated
ON ZERO & SPAN Temperature Range
OPERATING
INPUT POWER 10 - 32 VDC (options 0.5 - 5 VDC, 0 - 5 VDC,
0.5 - 5.5 VDC, 1 - 5 VDC)
16 - 32 VDC (options 1 - 10 VDC, 1 - 11 VDC)
11 - 32 VDC (option 1 - 6 VDC)
INSULATION 100 MEGOHMS @ 50 VDC
RESISTANCE
REVERSE POLARITY Protected
PROOF PRESSURE 150% of Rated Range
BURST PRESSURE 200% of Rated Range or 30,000 PSI,
Whichever is Less
WEIGHT 0.25 Kilos (0.5 lbs)
ENVIRONMENTAL
MAXIMUM OPERATING -40° to +120°C (-40° to +250°F)
TEMPERATURE
EMI/RFI 10 Volts/Meter up to 1.0 Ghz
WETTED PARTS 316 Cres
ENCLOSURE NEMA 4 & 4X
VIBRATION 15 g’s 10-2000 Hz
SHOCK 100 g’s for 11 Milliseconds, Half-sine
CERTIFICATIONS
CE - IEC 801-2 Level 3, ESD
IEC 801-3 Level 3, Radiation RF
IEC 801-4 Electrical Fast Transient / Burst
IEC 801-5 Level 4, Surge to 4 kV
IEC 801-6 Level 3, Conducted Susceptibility
13100 Telfair Avenue Sylmar, CA 91342-3573
(818) 362-7151 Fax: (818) 362-0458
E-mail: sales@beipressure.com
A Subsidiary of BEI TECHNOLOGIES, INC.
Product Specifications
IAW ANSI/ASA S37.1-75
DOC 128618 REV A (05/99)
Model 6-10 General Purpose
Pressure Transducer
Ordering Information
6 - 10 X - X X X X - X X
INSTRUMENT
FAMILY
6 = Micro-machined
Silicon
SENSOR TYPE
10 = Pressure Transducer
0.5-5.0 VDC
REFERENCE PRESSURE
0 = Absolute
1 = Vented Gage
2 = Sealed Gage
PRESSURE RANGES**
0 = 3 PSI#† 12 = 300 PSI#
1 = 5 PSI#† 13 = 500 PSI*
2 = 15 PSI#14 = 1000 PSI*
3 = 20 PSI#15 = 1500 PSI*
4 = 25 PSI#16 = 2000 PSI*
5 = 30 PSI#17 = 3000 PSI*
6 = 50 PSI#18 = 5000 PSI*
7 = 60 PSI#19 = 6000 PSI*
8 = 100 PSI#20 = 7500 PSI*
9 = 150 PSI#21 = 10,000 PSI*
10 = 200 PSI#22 = 15,000 PSI*†
11 = 250 PSI#
PRESSURE FITTINGS
0 = MS33656E4 (SAE J514-4) 3 = 1/8NPT, Male
1 = SAE 7/16-20 Straight 4 = 1/4NPT, Female
2 = 1/4NPT, Male 5 = 1/2-14NPT, Male
ELECTRICAL CONNECTIONS
0 = PT1H-10-6P
1 = 1/2NPT 4-wire Cable Conduit
2 = PT1H-8-4P
3 = 5-Wire Shielded Cable
4 = DIN 43650 Form C
5 = Hirschmann ELST 412
CUSTOM CONFIGURATIONS
#
Available Absolute and Vented Gage
*
Available Sealed Gage Only
** Calibration in bars is also available. Contact factory for details.
†Consult Factory