Functions Origin point setting (ABS length measurement
system)
Zero setting (INC length measurement system)
Hold
Function lock
Auto power ON/OFF
Data output *1
Error alarm
Degree of protection IP65 (IEC60529)*2
Measuring force 7-12N
Power supply Button type silver oxide battery (SR44), 1 piece
Position detection system Electromagnetic rotary sensor *3
Battery life Approx. 3 years under normal conditions
Standard accessories Reference bar, 1 piece (not for 0-25mm models)
Button type silver oxide battery (SR44, No.
938882), 1 piece
Spanner (No. 301336), 1 piece
n Common specifications
Origin point
setting (ABS length
measurement
system)
Pressing the ORIGIN button resets the ABS origin at the
current spindle position.
Zero setting
(INC length
measurement
system)
A brief press on the ZERO/ABS button sets display to
zero at the current spindle position and switches to
the incremental (INC) measuring mode. A longer press
resets to the ABS measuring mode.
Hold Pressing the HOLD button freezes the current value
in the display. This function is useful for preserving a
measurement in situations of poor visibility when the
instrument must be moved away from the workpiece
before the reading can be recorded. A second press
unfreezes the display ready for another measurement.
Function lock This function allows the ORIGIN (origin point setting)
function and the ZERO (zero setting) function to be
locked to prevent these points being reset accidentally.
Auto power
ON/OFF
The reading on the LCD disappears after this
instrument is idle for approx. 20 minutes, but the
origin point is retained. Turning the spindle causes the
reading on the LCD to reappear.
Data output Models equipped with this function have an output
port for transferring measurement data to a Statistical
Process Control (SPC) system.
Error alarm In case of an overflow on the LCD or a computing
error, an error message appears on the LCD, and the
measuring function stops. This prevents an instrument
from giving an erroneous reading. Also, when the
battery voltage drops to a certain level the low-
battery-voltage alarm annunciator appears well before
the micrometer becomes unusable.
n Functions
n Dimensions
n Selective specifications
Order No. Measuring
range
Resolution Instrumental
error*
Flatness of
measuring faces
Parallelism of
measuring faces
Weight
Data output function equipped 293-140 0-25mm
0.001mm ±1μm
0.3μm or less 1μm or less
265g
293-141 25-50mm 325g
Data output function not equipped 293-145 0-25mm 265g
293-146 25-50mm 325g
Data output function equipped 293-180 0-1”/0-25mm
.00005”/ 0.001mm ±.00005”/±1μm
265g
293-181 1-2”/ 25-50mm 325g
Data output function not equipped 293-185 0-1”/ 0-25mm 265g
293-186 1-2”/25-50mm 325g
L 37.548.5
19.3
6.35
b
27
a3
2.8
23.1
L a b
0-25mm (0-1”) 0 9 25
25-50mm (1-2”) 25 9.8 32
*1: Applicable only to 293-140/293-141/293-180/293-181
*2: This product is not waterproof. Rustproofing shall be applied after use.
*3: Patent pending (in Japan, USA, Europe, and China)
*Quantization error of ±1 count excluded
0-25mm (0-1”) range model
Unit: mm