Agilent 6650A Series Single-Output,
500 W DC Power Supplies, GPIB
Data Sheet
Speed and accuracy for test optimization
This series of 500 W linear-regulated
DC power supplies is designed to max-
imize the throughput of DUTs through
the manufacturing test process with
fast up and down programming time.
Valuable assemblies can be destroyed
by a minor component failure that
causes a surge of current to flow
into the DUT. Fast protection fea-
tures, including fast crowbar, mode
crossover protection, and the ability
to connect the protection circuitry of
multiple power supplies can increase
production yield.
Programming of the DC output and
the protection features can be done
either from the front panel or using
industry standard SCPI commands,
via the GPIB. Using the serial link,
up to 16 power supplies can be con-
nected through one GPIB address.
Test system integration can be further
simplified be using the VXIplug&play
drivers. The output voltage and current
can also be controlled with analog sig-
nals. This is helpful for certain types of
noisy environments, and also immedi-
ate reactions to process changes.
Lab bench use is enhanced by the fan
speed control, which helps to minimize
acoustic noise.
Fast, low-noise outputs
Analog control of output voltage and current
Fan-speed control to minimize acoustic noise
Built-in measurements and advanced programmable features
Protection features to ensure DUT safety
2
Specifi cations
Specifications
(at 0 ° to 55 °C unless otherwise
specified)
6651A 6652A 6653A 6654A 6655A
6651A-J05
Special order
option
Number of outputs 1 1 1 1 1 1
GPIB Yes Yes Yes Yes Yes Yes
Output ratings
Output voltage 0 to 8 V 0 to 20 V 0 to 35 V 0 to 60 V 0 to 120 V 10 V
Output current (40 °C) 0 to 50 A 0 to 25 A 0 to 15 A 0 to 9 A 0 to 4 A 50 A
Maximum current (50 °C/55 °C) 45 A/42.5 A 22.5 A/21.3 A 13.5 A/12.8 A 8.1 A/7.7 A 3.6 A/3.4 A 45 A/42.5 A
Programming accuracy (at 25 °C ± 5 °C)
Voltage 0.06% + 5 mV 10 mV 15 mV 26 mV 51 mV 6 mV
Current 0.15% + 60 mA 25 mA 13 mA 8 mA 4 mA 60 mA
Ripple and noise from 20 Hz to 20 MHz
Voltage rms 300 μV 300 μV 400 μV 500 μV 700 μV 300 μV
peak-to-peak 3 mV 3 mV 4 mV 5 mV 7 mV 3 mV
Current rms 25 mA 10 mA 5 mA 3 mA 2 mA 25 mA
Readback accuracy at 25 °C ± 5 °C
(percent of reading plus fi xed)
Voltage 0.07% + 6 mV 15 mV 25 mV 40 mV 80 mV 7.5 mV
+Current 0.15% + 67 mA 26 mA 15 mA 7 mA 3 mA 67 mA
–Current 0.35% + 100 mA 44 mA 24 mA 15 mA 7 mA 100 mA
Load regulation
Voltage 1 mV 2 mV 3 mV 4 mV 5 mV 1 mV
Current 2 mA 1 mA 0.5 mA 0.5 mA 0.5 mA 2 mA
Line regulation
Voltage 0.5 mV 0.5 mV 1 mV 1 mV 2 mV 0.5 mV
Current 2 mA 1 mA 0.75 mA 0.5 mA 0.5 mA 2 mA
Transient response time Less than 100 μs for the output voltage to recover to its previous level (within 0.1% of the
voltage rating of the supply or 20 mV, whichever is greater) following any step change in
load current of up to 50% of the output current rating of the supply
Supplemental characteristics
(Non-warranted characteristics
determined by design and useful in
applying the product)
6651A 6652A 6653A 6654A 6655A
6651A-J05
Special order
option
Average programming resolution
Voltage 2 mV 5 mV 10 mV 15 mV 30 mV 2.5 mV
Current 15 mA 7 mA 4 mA 2.5 mA 1.25 mA 15 mA
OVP 12 mV 30 mV 54 mV 93 mV 190 mV 16 mV
OVP accuracy 160 mV 400 mV 700 mV 1.2 V 2.4 V 200 mV
3
Specifi cations, continued
Specifications
(at 0 ° to 55 °C unless otherwise specified)
6651A-J03
Special order
option
6651A-J09
Special order
option
6652A-J03
Special order
option
6653A-J04
Special order
option
6653A-J17
Special order
option
Number of outputs 1 1 1 1 1
GPIB Yes Yes Yes Yes Yes
Output ratings
Output voltage 6 V 17 V/20 V 27 V 40 V 30 V
Output current (40 °C) 60 A 30 A/15 A 18.5 A 12.5 A 17.5 A
Maximum current (50 °C/55 °C) 54 A/51 A 27 A/25.5 A
13.5 A/12.75 A
16.65 A/15.72 A 11.25 A/10.6 A 15.75 A/14.87 A
Programming accuracy (at 25 °C ± 5 °C)
Voltage 0.06% + 5 mV 10 mV 13.5 mV 17.5 mV 15 mV
Current 0.15% + 75 mA 36 mA 25 mA 13 mA 16 mA
Ripple and noise from 20 Hz to 20 MHz
Voltage rms 300 μV 300 μV 450 μV 1.6 mV 400 μV
peak-to-peak 3 mV 4 mV 4.5 mV 5 mV 4 mV
Current rms 30 mA 13 mA 10 mA 5 mA 6 mA
Readback accuracy at 25 °C ± 5 °C
(percent of reading plus fi xed)
Voltage 0.07% + 6 mV 15 mV 20.5 mV 30 mV 25 mV
+Current 0.15% + 80 mA 40 mA 26 mA 15 mA 18 mA
–Current 0.35% + 150 mA 55 mA 44 mA 24 mA 28 mA
Load regulation
Voltage 1 mV 2 mV 2 mV 3.5 mV 3 mV
Current 6.5 mA 2 mA 1 mA 1 mA 0.5 mA
Line regulation
Voltage 0.5 mV 0.5 mV 0.5 mV 1 mV 1 mV
Current 2 mA 2 mA 2 mA 0.75 mA 0.75 mA
Transient response time Less than 100 μs for the output voltage to recover to its previous level (within 0.1%
of the voltage rating of the supply or 20 mV, whichever is greater) following any step
change in load current of up to 50% of the output current rating of the supply
Supplemental characteristics
(Non-warranted characteristics determined
by design and useful in applying the product)
6651A-J03
Special order
option
6651A-J09
Special order
option
6652A-J03
Special order
option
6653A-J04
Special order
option
6653A-J17
Special order
option
Average programming resolution
Voltage 2 mV 5 mV 6.75 mV 12 mV 10 mV
Current 18 mA 9 mA 7 mA 4 mA 5 mA
OVP 12 mV 30 mV 30 mV 65 mV 54 mV
OVP accuracy 160 mV 500 mV 400 mV 750 mV 700 mV
4
Specifications
(at 0 ° to 55 °C unless otherwise specified)
6654A-J04
Special order
option
6654A-J05
Special order
option
6654A-J12
Special order
option
6655A-J05
Special order
option
Special order
option
Number of outputs 1 1 1 1 1
GPIB Yes Yes Yes Yes Yes
Output ratings
Output voltage 70 V 50 V 80 V 150 V 156 V
Output current (40 °C) 7.5 V 10 A 6 A 3.2 A 3 A
Maximum current (50 °C/55 °C) 6.75 A/6.37 A 9 A/8.5 A 5.4 A/5.1 A 2.88 A/2.72 A 2.7 A/2.55 A
Programming accuracy (at 25 °C ± 5 °C)
Voltage 0.06% + 30 mV 26 mV 35 mV 64 mV 71 mV
Current 0.15% + 7 mA 9 mA 7 mA 3.5 mA 4 mA
Ripple and noise from 20 Hz to 20 MHz
Voltage rms
600 μV 500 μV 700 μV 800 μV 900 μV
peak-to-peak 6 mV 5 mV 7 mV 8 mV 8 mV
Current rms
5 mA 4 mA 3 mA 2 mA 3 mA
Readback accuracy at 25 °C ± 5 °C
(percent of reading plus fi xed)
Voltage 0.07% + 50 mV 40 mV 58 mV 100 mV 110 mV
+Current 0.15% + 6 mA 8 mA 6 mA 2.5 mA 3 mA
–Current 0.35% +
13 mA 17 mA 16 mA 6.5 mA 7.5 mA
Load regulation
Voltage 4 mV 4 mV 4 mV 6 mV 7 mV
Current 0.5 mA 0.5 mA 0.5 mA 0.5 mA 1 mA
Line regulation
Voltage 1 mV 1 mV 4.5 mV 2 mV 2 mV
Current 0.5 mA 0.5 mA 0.5 mA 0.5 mA 1 mA
Transient response time Less than 100 μs for the output voltage to recover to its previous level (within 0.1%
of the voltage rating of the supply or 20 mV, whichever is greater) following any step
change in load current of up to 50% of rated current
Supplemental characteristics
(Non-warranted characteristics determined
by design and useful in applying the product)
6654A-J04
Special order
option
6654A-J05
Special order
option
6654A-J12
Special order
option
6655A-J05
Special order
option
6655A-J10
Special order
option
Average programming resolution
Voltage 17.5 mV 15 mV 20 mV 37.5 mV 39.5 mV
Current 1.9 mA 2.75 mA 1.7 mA 8 mA 8 mA
OVP 110 mV 93 mV 130 mV 240 mV 250 mV
OVP accuracy 1.4 V 1.2 V 1.6 V 3 V 3.3 V
Specifi cations, continued
5
Agilent models: 6651A, 6652A, 6653A, 6654A, 6655A
425.5 mm
16.75"
497.8 mm
19.6"
518.2 mm
20.4"
128 mm
5.04"
12.7 mm
0.5"
132.6 mm
5.2"
145.1 mm
5.71"
Terminal Screw Size: 6-32 x 5/16 in
Metric Pan Head, M4 x 0.7 x 8 mm long
(center to center on these two screws is 20 mm, typ)
+
+
+– +–
Top
Side
Rear
Rear Output
IP
VDC Max VP IM IM+P S S
Supplemental characteristics
for all model numbers
DC floating voltage: Output terminals
can be floated up to ± 240 VDC from
chassis ground
Remote sensing: Up to half the rated
output voltage can be dropped in each
load lead. The drop in the load leads
subtracts from the voltage available
for the load.
Command processing time: Average
time required for the output voltage to
begin to change following receipt of
digital data is 20 ms for the power
supplies connected directly to the GPIB
Output programming response time:
The rise and fall time (10/90% and
90/10%) of the output voltage is less
than 15 ms. The output voltage change
settles within 1 LSB (0.025% x rated
voltage) of final value in less than 60 ms.
Down programming: An active down
programmer sinks approximately 20%
of the rated output current
Modulation: (Analog programming of
output voltage and current)
Input signal: 0 to –5 V
Input impedance: 10 kΩ nominal
AC input:
(AC input frequency 47 to 63 Hz)
Voltage 100 VAC 120 VAC 220 VAC 240 VAC
Current 12 A 10 A 5.7 A 5.3 A
Input power: 1,380 VA, 1,100 W at full
load; 120 W at no load
GPIB interface capabilities:
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1,
DT1, E1, and C0. IEEE-488.2 and SCPI-
compatible command set.
Software driver:
• IVI-COM
• VXIplug&play
Regulatory compliance: Listed to
UL 1244; conforms to IEC 61010-1.
Size:
425.5 mm W x 132.6 mm H x 497.8 mm D
(16.75 in x 5.22 in x 19.6 in)
Weight: Net, 25 kg (54 lb);
shipping, 28 kg (61 lb)
Warranty: One year
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Agilent Channel Partners
www.agilent.com/find/channelpartners
Get the best of both worlds: Agilent’s
measurement expertise and product
breadth, combined with channel
partner convenience.
For more information on Agilent
Technologies’ products, applications or
services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/fi nd/contactus
Americas
Canada (877) 894 4414
Brazil (11) 4197 3600
Mexico 01800 5064 800
United States (800) 829 4444
Asia Pacifi c
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 375 8100
Europe & Middle East
Belgium 32 (0) 2 404 93 40
Denmark 45 45 80 12 15
Finland 358 (0) 10 855 2100
France 0825 010 700*
*0.125 €/minute
Germany 49 (0) 7031 464 6333
Ireland 1890 924 204
Israel 972-3-9288-504/544
Italy 39 02 92 60 8484
Netherlands 31 (0) 20 547 2111
Spain 34 (91) 631 3300
Sweden 0200-88 22 55
United Kingdom 44 (0) 118 927 6201
For other unlisted countries:
www.agilent.com/fi nd/contactus
Revised: January 6, 2012
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2012
Published in USA, February 16, 2012
5990-9305EN
www.agilent.com
www.agilent.com/find/6650
Agilent Advantage Services is committed
to your success throughout your equip-
ment’s lifetime. To keep you competitive,
we continually invest in tools and
processes that speed up calibration and
repair and reduce your cost of ownership.
You can also use Infoline Web Services
to manage equipment and services more
effectively. By sharing our measurement
and service expertise, we help you create
the products that change our world.
www.agilent.com/quality
www.agilent.com/find/advantageservices
Quality Management SystemQuality Mana
g
ement Sy
s
ISO 9001:2008
Agilent Electronic Measurement Group
DEKRA Certified
Ordering information
Opt 100 87 to 106 VAC, 47 to 63 Hz
Opt 120 104 to 127 VAC, 47 to 63 Hz
Opt 220 191 to 233 VAC, 47 to 63 Hz
Opt 240 209 to 250 VAC, 47 to 63 Hz
Opt 0L1 Full documentation on
CD-ROM and printed user’s and pro-
gramming guides. CD-ROM includes
user’s guide, programming guide,
service manual, and quick start guide
Opt 0B3 Printed service manual
Accessories
1CM003A* Rack mount flange kit
132.6 mm H (3U) – two flange brackets
1CP002A* Rack mount flange and
handle kit 132.6 mm H (3U) – two
brackets and front handles
E3663AC Support rails for Agilent
rack cabinets
p/n 1494-0059 Accessory slide kit
p/n 1252-3698 7-pin analog plug
p/n 1252-1488 4-pin digital plug
p/n 5080-2148 Serial link cable
2 m (6.6 ft)
Application notes
10 Practical Tips You Need to Know
About Your Power Products,
5965-8239E
10 Hints for Using Your Power Supply
to Decrease Test Time, 5968-6359E
Understanding Linear Power Supply
Operation (AN1554), 5989-2291EN
Modern Connectivity–Using USB
and LAN I/O Converters (AN 1475-1)
5989-0123EN
Agilent DC Power Supplies for Base
Station Testing, 5988-2386EN
* Support rails required